K120477 · Konica Minolta Medical & Graphic, Inc. · MQB · Apr 9, 2012 · Radiology
Device Facts
Record ID
K120477
Device Name
AEROSYNC FOR AERODR SYSTEM
Applicant
Konica Minolta Medical & Graphic, Inc.
Product Code
MQB · Radiology
Decision Date
Apr 9, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1680
Device Class
Class 2
Attributes
Software as a Medical Device, Pediatric
Indications for Use
The AeroSync for AeroDR SYSTEM is a software device that is used in conjunction with AeroDR SYSTEM. This device is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures. This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
Device Story
AeroSync is a software enhancement for the AeroDR digital radiography system. It enables wireless X-ray detection by eliminating the physical cable connection between the detector and the X-ray generator. The software monitors for X-ray irradiation; upon detection, it triggers image acquisition. Used in clinical settings by radiology staff; it streamlines workflow by removing tethering requirements. The output is a digital radiographic image equivalent to the predicate system. Benefits include increased flexibility and ease of use in general diagnostic imaging.
Clinical Evidence
Bench testing only. Performance testing demonstrated that image quality is equivalent to the predicate device. No clinical data provided.
Technological Characteristics
Software-based X-ray detection system. Operates as an add-on to the AeroDR SYSTEM. Utilizes RF wireless technologies for communication. Complies with IEC 60601-1 (safety), IEC 60601-1-2 (EMC), and FCC Part 15 Subpart C/E (wireless).
Indications for Use
Indicated for generating radiographic images of human anatomy for general-purpose diagnostic procedures. Not for use in mammography, fluoroscopy, tomography, or angiography.
Regulatory Classification
Identification
A stationary x-ray system is a permanently installed diagnostic system intended to generate and control x-rays for examination of various anatomical regions. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Special Controls
*Classification.* Class II (special controls). A radiographic contrast tray or radiology diagnostic kit intended for use with a stationary x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
## 510(k) Summary
APR - 9 2012
as required by 807.92
## 1. Company Identification
Konica Minolta Medical & Graphic, Inc. No.1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Establishment Registration Number: 3004485675
## 2. Submitter's Name and Address
Shigeyuki Kojima
Manager
Regulations and Standards Section, Quality Assurance Center No. 1 Sakura-machi, Hino-shi, Tokyo 191-8511, Japan Telephone: 81-42-589-8429 Fax: 81-42-589-8053
## 3. Date of Submission
February 10, 2012
### 4. Device Trade Name
AeroSync for AeroDR SYSTEM
#### 5. Common Name
Digital Radiography
# 6. Classification, Product Code
Class II, 90MQB and 90LLZ
#### 7. Predicate Device
AeroDR SYSTEM, 510(k) number K102349
### 8. Indications for Use
The AeroSync for AeroDR SYSTEM is a software device that is used in conjunction with AeroDR SYSTEM. This device is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
{1}------------------------------------------------
## 9. Device Description
The AeroSync for AeroDR SYSTEM is a software device that is used in conjunction with our AeroDR SYSTEM, K102349. It adds minor change for our cleared AeroDR SYSTEM. It eliminates the need for an electrical connection between the AeroDR SYSTEM and X-ray generator. It can be detected X-ray irradiation without cables connected to X-ray generator.
## 10. Risk Analysis
The Risk Analysis for the AeroSync for AeroDR System (X-ray perception software) was conducted on the basis of ISO14971, "Medical devices – Application of risk management to medical devices". As a result of risk control measures, the risk associated with all of the identified hazards was reduced to an acceptable level or ALARP (As low as reasonably practicable).
Please refer to Section 13, Appendix-G
## 11. Compliance of Recognized Consensus Standard
In terms of Indications for use of this AeroSync for AeroDR SYSTEM, it has been tested and shown to meet the requirements of IEC 60601-1 and IEC 60601-1-2 as follows. Please refer to the Section 7, STANDARD DATA REPORTS FOR 510(k)s.
Safety standard: IEC60601-1 Ed.2 (1988) + A1 (1991)+A2(1995)
Electromagnetic Compatibility: IEC60601-1-2 Ed.3 (2007)
Besides, AeroSync for AeroDR SYSTEM includes RF wireless technologies (which customer can select), so that to ensure safety of device in case of wireless use, it also has been tested and shown to meet the requirements of FCC part15 Subpart C,E.
#### 12. Substantial Equivalence to Predicate Device
The predicate devices of AeroSync for AeroDR SYSTEM is our AeroDR SYSTEM, K102349.
The proposed device is a minor changed system of our cleared AeroDR SYSTEM. It eliminates the need for an electrical connection between the AeroDR SYSTEM and X-ray generator.
A comparison of the Indications for Use, Configuration, Specifications and Principals of operation by this proposed device and the predicate device has shown in Section 9, Substantial Equivalence Comparison Table.
The result of the Section 9, above mentioned the Risk Analysis and the
{2}------------------------------------------------
Compliance of Recognized Consensus Standard showed that there is no new safety and efficacy issue of the proposed device. Also, the results of performance testing shows that the image quality of proposed device is equivalent to the predicate device. Please refer to the section 14 and 15 Performance Testing.
The AeroSync for AeroDR SYSTEM is substantially equivalent to AeroDR SYSTEM, 510(k) Number, K102349.
### 9. Conclusion
Comprehensively, we conclude that the AeroSync for AeroDR System has the same technological characteristics as the predicate device. This 510(k) has demonstrated substantial equivalence as the predicate device.
5-3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The text is in all capital letters and is black.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
APR - 9 2012
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
Konica Minolta Medical & Graphics, Inc. % Mr. Ressell Munves Consultant Storch, Amini & Munves, P.C. 140 East 45th Street, 25th Floor Tow Grand Tower NEW YORK NY 10017
Re: K120477
Trade/Device Name: AeroSync for AeroDR SYSTEM Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: MQB Dated: February 14, 2012
Received: February 16, 2012
#### Dear Mr. Munves:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{4}------------------------------------------------
#### Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. .
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known)
AeroSync for AeroDR SYSTEM Device Name
Indications for Use:
The AeroSync for AeroDR SYSTEM is a software device that is used in conjunction with AeroDR SYSTEM. This device is indicated for use in generating radiographic images of human anatomy. It is intended to replace radiographic film/screen systems in general-purpose diagnostic procedures.
This device is not indicated for use in mammography, fluoroscopy, tomography and angiography applications.
Prescription Use X (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
AND/OR
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Rummellious
Jivision Sign-off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K.120477
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.