Anesthesia conduction needles are intended for the transient delivery of anesthetics to provide regional anesthesia or to facilitate placement of an epidural catheter.
Device Story
Anesthesia conduction needles (AN-SI, AN-SII, AN-E) are sterile, single-use devices used for epidural or spinal anesthesia. The device consists of a needle tube, hub, stylet, and jacket. During use, the cannula is stabilized during puncture; the stylet is removed after reaching the target site to allow delivery of local anesthetics into the epidural or spinal space. Operated by professional anesthesiologists in clinical settings, the needles facilitate pain relief. The device design is based on established needle geometries (Quincke, Pencil Point, Tuohy) to ensure proper placement and anesthetic delivery.
Clinical Evidence
Bench testing only. Performance testing verified compliance with ISO 9626 (needle tubing), ISO 7864 (sterile hypodermic needles), and ISO 594-2 (conical fittings). Biocompatibility testing conducted per ISO 10993-1 confirmed material safety. No clinical data provided.
Technological Characteristics
Stainless steel needle tubing (ISO 9626); plastic hub (ISO 594-1/2); sterile, single-use; manual operation; luer lock fitting; dimensions and design equivalent to predicate (K070354).
Indications for Use
Indicated for patients requiring regional anesthesia or epidural catheter placement via transient delivery of anesthetics.
Regulatory Classification
Identification
An anesthesia conduction needle is a device used to inject local anesthetics into a patient to provide regional anesthesia.
Predicate Devices
IMD's Tuohy needle; Quincke needle; Pencil Point needle (K070354)
Submission Summary (Full Text)
{0}------------------------------------------------
## KIZOY75
## DEC 2 1 2012
Premarket Notification Section 510(k) Submission Section 3 510K Summary
Section 3 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
Date of Submission:
Submitted by:
Shanghai SA Medical & Plastic Instruments Co., Ltd Shanghai SA Medical & Plostion & Plostic Development Zone, Minhang District, Shanghai, P.R.C Contact Person: Mr. Xianming Cao Tel: +86-21-54332290 Fax: +86-21-54332600 Email: caoxm@sa001.com
## Contact:
Mike Gu Regulatory Manager Guangzhou Osmunda Medical Device Consulting Co., Ltd Guangzhou Osthunda models Building, No.982 Congyun Rd, Baiyun District, Guangzhou, P.R.C Tel: +86-20-37612981 Fax: +86-20-37612253 Email: Mike.gu@osmundacn.com
#### Device Name:
AN-SI spinal needle,AN-SII spinal needle,AN-E epidural needle Trade Names: Common Names: Anesthesia conduction needles
Classification Name: Anesthesia conduction needles
Regulation Number: 21 CFR 868.5150
Product Class: Class II
Product Code: BSP
Preparation Date of Summary: Now 30, 2011
Predicate Device: IMD's Tuohy needle; Quincke needle; Pencil Point needle (K 070354)
Device Description:
Device Description: needles, including Epidural Needle for Single Use(AN-E), Spinal Needle for Single Use(AN-S), 1
{1}------------------------------------------------
AN-SII), consist of needle tube, hub, stylet and jacket The product is mainly used for epidural and/or spinal block (also called as epidural and/or spinal anesthesia) in human bodies.
The needles have been categorized as following in detail:
1. AN-SI Spinal Needles (same to IMD's Quincke needle)
2. AN-SII Spinal Needle (same to IMD's Pencil point needle)
3. AN-E Epidural Needles (same to IMD's Tuchy needle)
These needles are provided as sterile, single use devices. They may be packaged individually or included in regional anesthesia trays (kits). Anesthesia conduction needle. This is a simple hypodernic regional allestilesia trays (Mis). Anoonball concession of the aresthesia onduction needle. The later needie to hake the intral punction and election of local and the end anesthelics into the epidural space for longer pain relief.
The Shanghai SA's anesthesia conduction needles – AN-SI, AN-SII, AN-SII, AN-E – are single use, sterile madical devices for like The Shanghal SA s anesulesia conductions. The cannula is stabilized during puncture with use of an Inner
thanslent delively of ancollobo acting regioned its reached its analogical site for regional style. This stylet is within atter the ansociety (i.e., within minutes) by the professional ansether and the possible. The altestiesia. Then the arestheses our as opplied below the anesthesia conduction needle. The needle is Allondavely of dealer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
## Intended Use:
Intended USe:
Anesthesia conduction needles are intended for the transient delivery of anesthelics to provide regional anesthesia or to facilitate placement of an epidural catheter.
## Technology Characteristics:
Technology Characterisuics:
The Shanghai SA anesthesia conduction needles- AN-E,AN-SI,AN-SIAN CLAN,GLAN,SLL on onivalent I he Shangha SA anesthesia Conduction needies AN-Spart of Child Sh-Sl.AN-SI-AN-SIL-an-SIL-an-SIL-an-SI-an-SI-AN-SI-AN-SI-AN-SI-AN-SI-AN-SIpredicale devices dentilies above. The Sharghal of Childer.
in design physical dimensions, luer hub, metal and plastic materials, and packaging to the IMD anesthesia needle c under 510(K) number K070354
The Anesthesia Conduction Needles general design characteristics and functionality are similar in that they meet performance standards where applicable for:
Stainless Steel components: ISO 9626 Hub: ISO 594-1 and ISO 594-2 Hub to Needle Bond Strength: ISO 7864
All statements and representations set forth herein regarding or related to "substantially equivalential to the Endoral Frod equivalence" are in the limited context of the definition and purpose of substantial equivalence in the Federal Frodd in the and Cosmetic Act, as amended, and applicable regulations of the Food and Drug Administration as requiring including and Cosmetic Act, as almented, and application against interest under, any other laws or regulations, including
2
{2}------------------------------------------------
Premarket Notification Section 510(k) Submission Section 3 510K Summary
patent laws (whether in the context of patent infringement or otherwise).
## Nonclinical testing Conclusion:
Nonclinical testing Conclusion.
Performance testing was conducted to verify that the proposed device met all design specifications and was substantially equivalent to the predicate device:
substantally equivalent to the predicate down.
ISO 9626:-1991/Amd. 1:2001(E) "Stainless steel needle tubing for the manufacture of medical devices. "
ISO 7864:1993(E) "Sterile hypodermic needles for single use."
(SO 7864:1999(E) Stellie hypodernio no any of any of the synnges, needles and certain other medical equipment – Part 1:General Requirements."
1:General Requirements.
ISO 594-2:1998 "Conical fittings with 6% (Liter) taper for syringes, needles and certain other medical equipment- Pant 2: Lock fittings. "
Lock fittings. "
Results of performance testing indicate that the needles meet applicable sections of the standards referenced and are safe and effective for their intended use.
Biocompaibility testing based on the nature and duration of patient outlined in ISO 10993-12009 "Biological Biocompability testing based on the nature and testing within a risk management process" demonstrates that the materials used in the construction of the proposed needles are safe for their intended use.
Sterlization is equivalent to predicate device since the materials, packaging, and sterilization processed are the same.
## SE Conclusion:
SE Conclusion:
The companson between the predicate devices and the proposed devices . Anathering peoples can be The companson between the predicate Genes and the predicate devices. Anesthesia conduction needles can be
safe and effective, as well as substantially equivale prodicets Devi safe and effective, as well as substant to one predicate device IMD's Tuchy needle; Quincke needle; Pencil Point needte K 070354.
ಆ
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter of the circle. In the center of the circle is a stylized image of an eagle with its wings spread. The eagle is facing to the right and has three stripes across its body.
## December 21, 2012
Shanghai SA Medical & Plastic Instruments Company, Limited C/O Mike Gu Regulatory Manager Guangzhou Osmunda Medical Device Consulting Company, Limited 7th Floor, Jingui Business Building, No. 982 Congyun Road Baiyun District, Guangzhou China 510420
Re: K120475
Trade/Device Name: Anesthesia Conduction Needles Regulation Number: 21 CFR 868.5150 Regulation Name: Anesthesia Conduction Needle Regulatory Class: II Product Code: BSP Dated: December 12, 2012 Received: December 12, 2012
Dear Mr. Gu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
{4}------------------------------------------------
Page 2 - Mr. Gu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Anthony Lo Russo
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Premarket Notification Section 510(k) Submission Section 2 Indication for Use Statement Ref No.:
Section 2 Indication for Use Statement
510(k) Number: K120475
Device Name: Anesthesia conduction needles
Indications for Use:
Anesthesia conduction needles are intended for the transient delivery of anesthetics to provide regional anesthesia or to facilitate placement of an epidural catheter.
Prescription Use ___ X (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page_1_of_1
# Neel JJ Ratel
(Division Sign-Off) (Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K120475
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.