← Product Code [JOJ](/productcode/JOJ) · K120435

# BROWNE ALDAHOL GLUTARALDEHYDE INDICATOR (K120435)

_STERIS Corporation · JOJ · Jun 4, 2012 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K120435

## Device Facts

- **Applicant:** STERIS Corporation
- **Product Code:** [JOJ](/productcode/JOJ.md)
- **Decision Date:** Jun 4, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.2800
- **Device Class:** Class 2
- **Review Panel:** General Hospital

## Indications for Use

The Browne Aldahol Glutaraldehyde Indicator is a glutaraldehyde concentration monitor dedicated for use with Aldahol V High-Level Disinfectant solution cleared under K113015. The purpose of the Browne Aldahol Glutaraldehyde Indicator is to determine whether the glutaraldehyde concentration of an Aldahol V High-Level Disinfectant solution is above the minimum recommended concentration, allowing the solution to be re-used for reprocessing temperature-sensitive (and other) instruments if the glutaraldehyde concentration is found to be greater than 1.8%.

## Device Story

Chemical indicator strip for monitoring glutaraldehyde concentration in Aldahol V High-Level Disinfectant; consists of absorbent paper pad impregnated with reactive chemicals bonded to polymer film; used by healthcare personnel to verify solution concentration remains above 1.8% minimum recommended concentration (MRC) before reusing disinfectant for instrument reprocessing; provides visual indication of concentration status; ensures disinfectant efficacy for patient safety.

## Clinical Evidence

Bench testing only. Performance verified across 6 lots including: performance testing, blind study, simulated use with contaminants, strip life, chemical stability, specificity for glutaraldehyde, and storage stability. All tests met acceptance criteria (100% pass at 1.8% failure threshold; ≥80% pass at 2.3% threshold).

## Technological Characteristics

Chemical indicator strip; absorbent paper pad with reactive chemicals; polymer film backing; single-use; visual colorimetric sensing principle.

## Regulatory Identification

Biological sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor adequacy of sterilization. The device consists of a known number of microorganisms, of known resistance to the mode of sterilization, in or on a carrier and enclosed in a protective package. Subsequent growth or failure of the microorganisms to grow under suitable conditions indicates the adequacy of sterilization. Physical/chemical sterilization process indicator: A device intended for use by a health care provider to accompany products being sterilized through a sterilization procedure and to monitor one or more parameters of the sterilization process. The adequacy of the sterilization conditions as measured by these parameters is indicated by a visible change in the device.

## Predicate Devices

- 3M™ ComplyTM Cold Sterilog™ Glutaraldehyde Monitor 3987MM ([K924681](/device/K924681.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{1}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the STERIS logo, along with the code K120436. The STERIS logo is a stylized wave pattern. The image also contains the date JUN - 4 2012.

# 510(k) Summary For Browne Aldahol Glutaraldehyde Indicator

Albert Browne Ltd., a subsidiary of STERIS Corporation Chancery House 190 Waterside Road Hamilton Industrial Park Leicester LES IQZ United Kingdom Phone: +44 116 276 8636 Fax No: +44 116 276 8639

Contact:

Bill Brodbeck Senior Manager Regulatory Affairs Telephone: (440) 392-7690 Fax No: (440) 357-9198

Summary Date:

February 10, 2012

STERIS Corporation ■ 5960 Heisley Road ■ Mentor, OH 44060-1834 USA ■ 440-354-2600

.

{2}------------------------------------------------

## 1. Device Name

| Trade Name:          | Browne Aldahol Glutaraldehyde Indicator                                                       |
|----------------------|-----------------------------------------------------------------------------------------------|
| Common/Usual Name:   | Browne Aldahol Test Strip                                                                     |
| Classification Name: | Physical/chemical sterilization process indicator (21<br>CFR 880.2800 (b), Product Code JOJ). |

## 2. Predicate Device

- K924681 3M™ ComplyTM Cold Sterilog™ Glutaraldehyde Monitor 3987MM .
## 3. Description of Device

The Browne Aldahol Glutaraldehyde Indicator is a chemical indicator strip consisting of an absorbent paper pad impregnated with the reactive chemicals, which is adhesively bonded to one end of a polymer film. The Browne Aldahol Glutaraldehyde Indicator has been developed to monitor the active glutaraldehyde concentration of Aldahol V High-Level Disinfectant (K113015) solution that has an MRC of 1.8% glutaraldehyde.

## 4. Intended Use

The Browne Aldahol Glutaraldehyde Indicator is a glutaraldehyde concentration monitor dedicated for use with Aldahol V High-Level Disinfectant solution cleared under K113015. The purpose of the Browne Aldahol Glutaraldehyde Indicator is to determine whether the glutaraldehyde concentration of an Aldahol V High-Level Disinfectant solution is above the minimum recommended concentration, allowing the solution to be re-used for reprocessing temperature-sensitive (and other) instruments if the glutaraldehyde concentration is found to be greater than 1.8%.

## 5. Description of Safety and Substantial Equivalence

The proposed and predicate devices are all single use indicators used to monitor glutaraldehyde concentration in specific solutions. The differences between the proposed Browne Aldahol Glutaraldehyde Indicator and predicate device are limited to differences in the device design and materials. These differences do not raise any new issues of safety and efficacy.

{3}------------------------------------------------

A summary of the technological characteristics of the new device in comparison to those of the predicate device is provided in Section 12 of this premarket notification.

The following table summarizes the verification activities that were performed, with their respective acceptance criteria and results, to demonstrate that the Browne Aldahol Glutaraldehyde Indicator is safe and effective. These studies confirm that .the device's performance meets the requirements of its pre-defined acceptance criteria and intended uses.

| Test of 6 Lots                                            | Acceptance Criteria |                         | Study<br>Result |
|-----------------------------------------------------------|---------------------|-------------------------|-----------------|
|                                                           | FAIL @ 1.8%         | PASS @ 2.3%             |                 |
| Performance Testing                                       | 100%                | ≥ 80%                   | Pass            |
| Blind Study Testing                                       | 100%                | ≥ 80%                   | Pass            |
| Simulated Use<br>(Contaminants) Testing                   | 100%                | ≥ 80%                   | Pass            |
| Test Strip Life<br>Outside the Bottle                     | 100%                | ≥ 80%                   | Pass            |
| Aggressive Chemical<br>Stability Testing                  | 100%                | ≥ 80%                   | Pass            |
| Specificity Testing<br>for glutaraldehyde                 | 100%                | 100% FAIL for Tap Water | Pass            |
| Stability Testing in three<br>storage environments        | 100%                | ≥ 80%                   | Pass            |
| In-Use Stability Testing in<br>three storage environments | 100%                | ≥ 80%                   | Pass            |

The Browne Aldahol Glutaraldehyde Indicator is substantially equivalent to its predicate.

r

{4}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized depiction of a caduceus, a symbol often associated with medicine and healthcare. The caduceus features a staff with a snake winding around it, and a pair of wings at the top.

## Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Dr. Bill Brodbeck Senior Manager, Regulatory Affairs STERIS Corporation 5960 Heisley Road Mentor, Ohio 44060

JUN - 4 2012

Re: K120435

Trade/Device Name: Browne Aldahol Glutaraldehyde Indicator Regulation Number: 21 CFR 880.2800 Regulation Name: Sterilization Process Indicator Regulatory Class: II Product Code: JOJ Dated: May 10, 2012 Received: May 11, 2012

Dear Dr. Brodbeck:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

## Page 2 - Dr. Brodbeck

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting . (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

# Indications for Use

#### 510(k) Number (if known): K120435

Device Name:

## Browne Aldahol Glutaraldehyde Indicator

Indications for Use:

The Browne Aldahol Glutaraldehyde Indicator is a glutaraldehyde concentration monitor dedicated for use with Aldahol V High-Level Disinfectant. The purpose of the Browne Aldahol Glutaraldehyde Indicator is to monitor glutaraldehyde concentration greater than 1.8%.

Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use X (21 CFR 801 Subpart C)

# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Elizabeth F. Clannie-Willin.

(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices

K120435 510(k) Number:

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K120435](https://fda.innolitics.com/device/K120435)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
