PENTARAY NAV HIGH-DENSITY MAPPING CATHETER

K120425 · Biosense Webster, Inc. · MTD · May 24, 2012 · Cardiovascular

Device Facts

Record IDK120425
Device NamePENTARAY NAV HIGH-DENSITY MAPPING CATHETER
ApplicantBiosense Webster, Inc.
Product CodeMTD · Cardiovascular
Decision DateMay 24, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 870.1220
Device ClassClass 2

Indications for Use

The Biosense Webster PENTARAY® NAV High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY® NAV High-Density Mapping Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.0.)

Device Story

PENTARAY® NAV is a 7 Fr deflectable electrophysiology catheter featuring a five-spine array with 20 Platinum/Iridium electrodes and an electromagnetic location sensor. Used in clinical electrophysiology labs by physicians to map cardiac structures; catheter is inserted into the heart to monitor electrical activity. Input consists of cardiac electrical signals; output is electrogram data transmitted to the CARTO® 3 EP Navigation System and recording equipment. The device provides real-time location information when integrated with the CARTO® 3 V3.0 system, allowing physicians to visualize catheter position during mapping. This aids in the diagnosis of heart arrhythmias by providing high-density electrical data from the endocardium. The catheter is maneuvered via a handle that controls deflection and torque, optimizing contact with the heart wall.

Clinical Evidence

Bench testing only; no clinical data provided. Testing included simulated use conditions to verify performance and safety.

Technological Characteristics

7 Fr deflectable catheter; 5-spine array with 20 Platinum/Iridium ring electrodes; electromagnetic location sensor in soft tip; 2 additional shaft ring electrodes. Connectivity via cables to CARTO® 3 EP Navigation System and recording equipment. Deflection capability minimum 160 degrees.

Indications for Use

Indicated for electrophysiological mapping (recording/stimulation) of atrial and ventricular cardiac structures. Contraindicated for RF ablation; patients with prosthetic valves; active systemic infection; and transseptal approach in patients with intracardiac thrombus, myxoma, or interatrial baffle/patch.

Regulatory Classification

Identification

An electrode recording catheter or an electrode recording probe is a device used to detect an intracardiac electrocardiogram, or to detect cardiac output or left-to-right heart shunts. The device may be unipolar or multipolar for electrocardiogram detection, or may be a platinum-tipped catheter which senses the presence of a special indicator for cardiac output or left-to-right heart shunt determinations.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{1}------------------------------------------------ MAY 2 4 2012 # SECTION 2. # 510(K) SUMMARY K1204 ### 2. 510(K) SUMMARY This 510(k) Summary is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | APPLICANT | Biosense Webster, Inc.<br>3333 Diamond Canyon Road<br>Diamond Bar, CA 91765 | |---------------------------|-------------------------------------------------------------------------------------------------------------------------------------| | OFFICIAL<br>CORRESPONDENT | Wayne R. Hohman<br>Project Manager Regulatory Affairs<br>Telephone: 909-839-8597<br>Fax: 909-839-8804<br>Email: whohman@its.jnj.com | | TRADE NAME | PENTARAY® NAV High-Density Mapping Catheter | | COMMON NAME | Deflectable Tip Electrophysiology Catheter—Diagnostic | | CLASSIFICATION NAME | Electrode recording catheter or electrode recording<br>probe | | DEVICE CLASSIFICATION | Class II, 21 CFR §870.1220<br>Product Code: MTD | | PRODUCT CODES | D-1282-01-S, D-1282-02-S, D-1282-03-S<br>D-1282-04-S, D-1282-05-S, D-1282-06-S | | PREDICATE DEVICE | Biosense Webster Flower High-Density Mapping<br>Catheter (510(k) K050217), now named the PENTARAY®<br>High-Density Mapping Catheter | | REFERENCE DEVICE | Biosense Webster LASSO® 2515 NAV Variable Catheter<br>(510(k) K081258) | ### SUBSTANTIALLY EQUIVALENT TO: PENTARAY® NAV High-Density Mapping Catheter is substantially equivalent to Biosense Webster's Flower High-Density Mapping Catheter (510(k) K050217), now named the PENTARAY® High-Density Mapping Catheter, and a Reference Device, the Biosense Webster Lasso 2515 NAV Variable Catheter (510(k) K081258). Like the predicate device, the PENTARAY® NAV High-Density Mapping Catheter features a fivespine array of 20 electrodes at the distal tip of a 7 Fr diameter catheter, has the same patient contact materials, and has the same device geometry. The original intended use of the predicate device for diagnosis of heart arrhythmias remains the same in the proposed Biosense Webster, Inc. Pentaray NAV Catheters Traditional 510(k) Page 18 of 595 \$\frac{1}{3}\$ {2}------------------------------------------------ KI20425 device, however the proposed device will now have the added ability to provide location information when used with the CARTO® 3 EP Navigation System, Version 3.0. This latter capability is substantially equivalent to the Reference Device, the LASSO® 2515 NAV Variable Catheter. The LASSO® 2515 NAV Variable Catheter went through this same transition with FDA when FDA cleared the update of the non-navigational version to the navigational version. # DESCRIPTION OF THE DEVICE SUBJECT TO PREMARKET NOTIFICATION: The PENTARAY® NAV High-Density Mapping Catheter is a 7 Fr, multi-electrode, electrophysiological diagnostic mapping catheter that consists of an array of five flexible ences that support a total of 20 Platinum/Iridium ring mapping electrodes (four electrodes per spine). The five spines converge at a multi-lumen soft tip section of the catheter that is deflectable (a minimum of 160°) and maneuverable by the user. These features permit positioning the flexible electrode-bearing spines against the heart wall to monitor positioning the nemore is an electromagnetic location sensor location sensor located within the soft tip of the shaft. Two new ring electrodes are positioned on the exterior of the shaft in this same area. The electrical wires from the 22 electrodes, the sensor, the puller wires that effect deflection, and a plastic tube for heparinized saline run through the length of the catheter shaft to the handle. The handle is used to advance or retract the catheter, torque the catheter, and/or deflect the catheter to optimize positioning of the multi-spined tip. The catheter is connected via appropriate cables to the CARTO® 3 EP Navigation System and/or appropriate recording equipment. ### INDICATIONS FOR USE: The Biosense Webster PENTARAY® NAV High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the attial and ventricular regions of the heart. The PENTARAY® NAV High-Density Mapping Catheter provides location information when used with compatible CARTO® 3 V3.0 EP Navigation Systems. (This catheter is not compatible with CARTO®3 EP Navigation Systems prior to Version 3.0.) ### TECHNICAL CHARACTERISTICS: The PENTARAY® NAV High-Density Mapping Catheter is a typical electrophysiological catheter that is unique only in its geometrical arrangement of 20 ring electrodes on five individual spines at its distal tip. Otherwise, there are no special technical aspects of the ability of this catheter to detect electrical signals from heart endocardium and transmit this information to the CARTO® 3 EP Navigation System and/or recording equipment for display, analysis, and interpretation in detection of various heart arrhythmias. Biosense Webster, Inc. Pentaray NAV Catheters Traditional 510(k) Page 19 of 595 {3}------------------------------------------------ # SECTION 2. # 510(K) SUMMARY ### PERFORMANCE DATA: The PENTARAY® NAV High-Density Mapping Catheter underwent extensive Bench Testing that included simulated use conditions. After thorough investigations and mitigations where appropriate, the proposed catheter passed all intended criteria in accordance with appropriate test criteria and standards. ## BASIS FOR DETERMINATION OF SUBSTANTIAL EQUIVALENCE: A technological comparison and performance testing demonstrate that the PENTARAY® NAV High-Density Mapping Catheter is substantially equivalent to the predicate and reference devices. Biosense Webster, Inc. Pentaray NA V Catheters {4}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the department's name around the perimeter. In the center of the circle is an abstract symbol that resembles an eagle or bird-like figure. The symbol is composed of three curved lines that form the body and wings of the bird. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Biosense Webster, Inc. c/o Mr. Wayne Hohman Project Manager Regulatory Affairs 3333 Diamond Canyon Rd. Diamond Bar, CA 91765 MAY 2 4 2012 Re: K120425 Trade/Device Name: Pentaray Nav High-Density Mapping Catheter Regulation Number: 21 CFR §870.1220 Regulation Name: Catheter, Electrode Recording Regulatory Class: Class II (two) Product Code: MTD Dated: May 9, 2012 Received: May 10, 2012 Dear Mr. Hohman: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {5}------------------------------------------------ Page 2 - Mr. Wayne Hohman comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ### SECTION 1. INDICATIONS FOR USE STATEMENT ## 1. INDICATIONS FOR USE STATEMENT 510(k) Number (if known) Device Name: PENTARAY® NAV High-Density Mapping Catheter Indications for Use: The Biosense Webster PENTARAY® NAV High-Density Mapping Catheter is indicated for multiple electrode electrophysiological mapping of cardiac structures in the heart, i.e., recording or stimulation only. This catheter is intended to obtain electrograms in the atrial and ventricular regions of the heart. The PENTARAY® NAV High-Density Mapping Catheter provides location information when used with compatible CARTO® 3 EP Navigation Systems. (This catheter is not compatible with CARTO® 3 EP Navigation Systems prior to Version 3.0.) Contraindications - The Biosense Webster PENTARAY® NAV High-Density Mapping Catheter has not . been shown to be safe and effective for radio frequency (RF) ablation. - Use of this catheter may not be appropriate for use in patients with prosthetic valves. A ● relative contraindication for cardiac catheter procedures is active systemic infection. - The transseptal approach is contraindicated in patients with intracardiac thrombus or ● myxoma, or interatrial baffle or patch. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Cardiovascular Devices | 510(k) Number | K120425 | |---------------|---------| |---------------|---------| Page 1 of 1· Biosense Webster, Inc PentaRay NAV Catheters Traditional 510(k) (Replacement) Page 17 of 595
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...