← Product Code [MAX](/productcode/MAX) · K120401

# ALAMO T (K120401)

_Alliance Partners, LLC · MAX · Oct 5, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K120401

## Device Facts

- **Applicant:** Alliance Partners, LLC
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Oct 5, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Alamo T is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment prior to treatment with an intervertebral cage. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. The device system must be used with supplemental fixation and autograft to facilitate fusion and is to be implanted via a transforaminal approach.

## Device Story

Alamo T is an intervertebral body fusion device used in spinal fusion surgery to provide structural stability following discectomy. The device is implanted via a transforaminal (TLIF) approach by a surgeon. It features a hollow center to accommodate autograft bone to facilitate fusion and superior/inferior surface ridges to prevent migration. The device is used in conjunction with supplemental fixation. It provides mechanical support to the fusion site, potentially benefiting patients by stabilizing the spine and promoting bone integration.

## Clinical Evidence

Bench testing only. Performance was validated via mechanical testing per ASTM F2077 and F2267, including static compression, dynamic compression, subsidence, and expulsion tests.

## Technological Characteristics

Constructed from PEEK Optima LT1 (ASTM F2026) with tantalum markers (ASTM F560) for radiographic visualization. Features a hollow center for bone graft and pyramid-patterned ridges on superior/inferior surfaces for fixation. Dimensions range from 8mm-14mm in height. Mechanical performance tested per ASTM F2077 and F2267.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Globus Medical Signature TLIF Spacer ([K072970](/device/K072970.md))
- Genesys Spine Interbody Fusion System ([K103034](/device/K103034.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K|2α40|

ОСТ

5 2012

# 510(k) Summary

Summary of 510(k) Safety and Effectiveness

| Submitted By:               | Alliance Partners, LLC<br>121 Interpark Blvd, #601<br>San Antonio, TX 78216                       |  |
|-----------------------------|---------------------------------------------------------------------------------------------------|--|
| Date:                       | July 11, 2012                                                                                     |  |
| Contact Person:             | Jennifer Palinchik<br>Development and Regulatory Consultant                                       |  |
| Contact Telephone:          | (440) 933-8850                                                                                    |  |
| Device Trade Name:          | Alamo T                                                                                           |  |
| Device Classification Name: | Intervertebral Body Fusion Device with Bone Graft,<br>Lumbar                                      |  |
| Device Classification:      | Class II                                                                                          |  |
| Reviewing Panel:            | Orthopedic                                                                                        |  |
| Regulation Number:          | 888.3080                                                                                          |  |
| Product Code:               | MAX                                                                                               |  |
| Predicate Device:           | Globus Medical Signature TLIF Spacer (K072970)<br>Genesys Spine Interbody Fusion System (K103034) |  |

#### Device Description:

The Alamo T is used for spinal fusion surgery to provide support and structural stability at the fusion site following discectomy. The device is manufactured from PEEK Optima® LT1 per ASTM F2026 and includes tantalum markers per ASTM F560 for radiographic visualization.

The device footprint has a hollow centre to accommodate bone graft to facilitate bone integration and fusion between the end plates from a transforaminal (TLIF) surgical approach. The device is available in various heights to accommodate variability among patients and the inferior and superior surfaces are designed with ridges to improve fixation and stability and prevent back out and migration.

#### Intended Use:

The Alamo T is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment prior to treatment with an intervertebral cage. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. The device system must be used with supplemental fixation and autograft to facilitate fusion and is to be implanted via a transforaminal approach.

page 1 of 2

{1}------------------------------------------------

### Substantial Equivalence Information:

The design features, material, and indications for use of the Alamo T device are substantially equivalent to the predicate devices listed above. The safety and effectiveness is adequately supported by the substantial equivalence, material information, and analysis data provided within this Premarket Notification.

| Item                          | Alamo T                                                                                                                                                           | Globus TLIF Spacer                                                                                                                                                                                         | Genesys TLIF System                                                                                                                                                  |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product Code                  | MAX                                                                                                                                                               | MAX                                                                                                                                                                                                        | MAX, ODP, MQP                                                                                                                                                        |
| Classification Name           | Intervertebral Body<br>Fusion Device                                                                                                                              | Same                                                                                                                                                                                                       | Same                                                                                                                                                                 |
| Intended Use                  | degenerative disc disease<br>(DDD) at one or two<br>contiguous levels of the<br>lumbosacral spine (L2-<br>S1)                                                     | Same                                                                                                                                                                                                       | Same. Also intended<br>for vertebral body<br>replacement                                                                                                             |
| Footprint                     | Blocks have one length<br>and width, and a variety<br>of heights with axial and<br>lateral openings for bone<br>graft and rows of teeth.<br>No pivoting mechanism | Blocks in a variety of<br>lengths, widths, and<br>heights with axial and<br>lateral openings for bone<br>graft and rows of teeth.<br>Pivoting mechanism for<br>controlled articulation<br>during insertion | Blocks have one length<br>and width, and a<br>variety of heights with<br>axial and lateral<br>openings for bone graft<br>and rows of teeth. No<br>pivoting mechanism |
| Graft Opening                 | Large axial graft window                                                                                                                                          | Same plus anterior graft<br>windows                                                                                                                                                                        | Same                                                                                                                                                                 |
| Teeth to prevent<br>migration | Located on superior and<br>inferior surfaces.<br>Pyramid pattern                                                                                                  | Same                                                                                                                                                                                                       | Same, but linear pattern                                                                                                                                             |
| Radiographic<br>markers       | Yes                                                                                                                                                               | Yes                                                                                                                                                                                                        | Yes                                                                                                                                                                  |
| Axial Footprint<br>dimensions | 10 x 28mm                                                                                                                                                         | 10 x 28mm and 11 x<br>33mm                                                                                                                                                                                 | 14 x 35mm                                                                                                                                                            |
| Device Height                 | 8mm-14mm (1mm<br>increments)                                                                                                                                      | 7mm-17mm (1mm<br>increments), excluding<br>14mm                                                                                                                                                            | 6mm-15mm (1mm<br>increments)                                                                                                                                         |
| Material                      | PEEK Optima LT1 and<br>Tantalum (markers)                                                                                                                         | Same                                                                                                                                                                                                       | Same                                                                                                                                                                 |

### Mechanical Testing:

Performance testing was conducted via the following mechanical tests per ASTM F2077 and F2267 using the worst case device: Static Compression, Dynamic Compression, Subsidence, and Expulsion. The device functioned as intended and the performance results show that the Alamo T is substantially equivalent to the predicate devices.

> K120401 Page 2 of 2

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle with three wing-like shapes, symbolizing health and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle symbol. The logo is in black and white.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Alliance Partners, LLC % RP+M, LLC Ms. Jennifer Palinchik Development and Regulatory Consultant 33490 Pin Oak Parkway Avon Lake, Ohio 44012

Re: K120401

Trade/Device Name: Alamo T Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: September 26, 2012 Received: September 27, 2012

Dear Ms. Palinchik:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

OCT
5 2012

{3}------------------------------------------------

## Page 2 - Ms. Jennifer Palinchik

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

# Indications for Use

510(k) Number (if known): K12 0401

Device Name: Alamo T

#### Indications for Use:

The Alamo T is intended for spinal fusion procedures in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as back pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment prior to treatment with an intervertebral cage. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved levels. The device system must be used with supplemental fixation and autograft to facilitate fusion and is to be implanted via a transforaminal approach.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
| (Division Sign-Off)                                    |
| Division of Surgical, Orthopedic,                      |
| and Restorative Devices                                |

510(k) Number K1206401

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K120401](https://fda.innolitics.com/device/K120401)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).