K120244 · Nexera Medical, Inc. · ONT · Jul 5, 2012 · General, Plastic Surgery
Device Facts
Record ID
K120244
Device Name
SPECTRASHIELD 9500 SURGICAL N95 RESPIRATOR
Applicant
Nexera Medical, Inc.
Product Code
ONT · General, Plastic Surgery
Decision Date
Jul 5, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4040
Device Class
Class 2
Indications for Use
The SpectraShield 9500 Surgical N95 Respirator is a single use, disposable, surgical N95 respirator, tested for continuous use up to 8 hours, embedded with a zeolite carrier containing a silver-copper agent on the outer layer and is not an antimicrobial drug. SpectraShield 9500 kills 99.99% of test bacteria after one hour of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill on the surface of the outer layer of the respirator when tested in vitro against single isolates of the following test bacteria: Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae, under tested contact conditions. No clinical studies have been conducted comparing the ability of an untreated surgical N95 respirator and the SpectraShield model 9500 surgical N95 respirator to protect the wearer from infection, and the antibacterial treatment cannot effect pathogens that are inhaled around the edges of the respirator. The SpectraShield 9500 Surgical N95 respirator is a single use device, intended for occupational use to protect against microorganisms, body fluids, and particulate material.
Device Story
SpectraShield 9500 is a molded, 4-layer surgical N95 respirator. Construction includes a 2-ply meltblown polypropylene middle layer sandwiched between polyester nonwoven fabric layers. Outer layer contains a zeolite carrier with a silver-copper antibacterial agent. Device features two latex-free elastic straps and an aluminum nose strip. Used in occupational settings to protect wearers from microorganisms, body fluids, and particulates. Antibacterial agent provides surface-contact kill (99.99%) against S. pyogenes, MRSA, and H. influenzae. Device is single-use and disposable; tested for up to 8 hours of continuous wear. Does not protect against pathogens inhaled around mask edges.
Clinical Evidence
No clinical studies conducted. Evidence is based on bench testing, including fluid resistance (ASTM F1862), particulate filtration (ASTM F2299), bacterial filtration (ASTM F2101), and biocompatibility (ISO 10993). Bio-efficacy testing demonstrated >99.99% kill rate against S. pyogenes, MRSA, and H. influenzae after 1 hour of contact, including after 12-hour perspiration exposure simulations.
Indicated for occupational use as a single-use, disposable surgical N95 respirator to protect the wearer against microorganisms, body fluids, and particulate material. Suitable for continuous use up to 8 hours.
Regulatory Classification
Identification
Surgical apparel are devices that are intended to be worn by operating room personnel during surgical procedures to protect both the surgical patient and the operating room personnel from transfer of microorganisms, body fluids, and particulate material. Examples include surgical caps, hoods, masks, gowns, operating room shoes and shoe covers, and isolation masks and gowns. Surgical suits and dresses, commonly known as scrub suits, are excluded.
Special Controls
*Classification.* (1) Class II (special controls) for surgical gowns and surgical masks. A surgical N95 respirator or N95 filtering facepiece respirator is not exempt if it is intended to prevent specific diseases or infections, or it is labeled or otherwise represented as filtering surgical smoke or plumes, filtering specific amounts of viruses or bacteria, reducing the amount of and/or killing viruses, bacteria, or fungi, or affecting allergenicity, or it contains coating technologies unrelated to filtration (e.g., to reduce and or kill microorganisms). Surgical N95 respirators and N95 filtering facepiece respirators are exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9, and the following conditions for exemption:(i) The user contacting components of the device must be demonstrated to be biocompatible.
(ii) Analysis and nonclinical testing must:
(A) Characterize flammability and be demonstrated to be appropriate for the intended environment of use; and
(B) Demonstrate the ability of the device to resist penetration by fluids, such as blood and body fluids, at a velocity consistent with the intended use of the device.
(iii) NIOSH approved under its regulation.
(2) Class I (general controls) for surgical apparel other than surgical gowns and surgical masks. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 878.9.
{0}------------------------------------------------
# K120244
Image /page/0/Picture/1 description: The image shows the logo for "NEXERA MEDICAL". The logo is in black and white, with the word "NEXERA" in large, bold letters. Above the X is a diamond shape. Below the ERA is the word "MEDICAL" in smaller letters.
JUL 5 5 2012
### Section 5.0: 510(k) SUMMARY
510(k) Owner: NexEra Medical, Inc. 3343 West Commercial Blvd, Suite 103 Ft. Lauderdale, FL 33309
| Contact: | Paul Sallarulo, President CEO |
|----------|-------------------------------|
| Phone: | 954-495-2020, x 2031 |
| Fax: | 954-491-7281 |
Establishment TBD Registration Number: Date Summarv July 2, 2012 Prepared: Device: Trade Name: SpectraShield model 9500 Surgical Mask Common /Classification Name: Surgical mask Classification Product Code: ONT Regulation Number: 21CFR 878.4040 Predicate K090414 SpectraShield 9500 Surgical N95 Respirator Device Information: The SpectraShield model 9500 Surgical Mask is a molded shape surgical mask composed Device of 4 layers of material, molded to form the mask. A 2-ply meitblown polypropylene Description: middle layer is sandwiched by inner and outer layers of 100% polyester nonwoven fabric. The inner and outside layers of polyester nonwoven fabric include fibers that have been embedded with an antibacterial agent to provide antibacterial performance. The mask has 2 latex-free non-allergenic elastic straps and an aluminum nose strip. Intended Use: The SpectraShield 9500 Surgical N95 Respirator is a single use, disposable surgical N95 respirator, tested for continuous use up to 8 hours, embedded with a zeolite carrier containing a silver-copper agent on the outer layer and is not an antimicrobial drug. SpectraShield 9500 kills 99.99% of test bacteria after one hour of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill on the 3343 West Commercial Blvd., Suite 103 - Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com
> Section 5.0: 510(K) Summary Page | 1
{1}------------------------------------------------
surface of the outer layer of the respirator when tested in vitro against single isolates of the following test bacteria: Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae under tested contact conditions.
No clinical studies have been conducted comparing the ability of the untreated surgical N95 respirator and the SpectraShield model 9500 surgical N95 respirator to protect the wearer from infection and the antibacterial treatment cannot effect pathogens that are inhaled around the edges of the respirator.
The SpectraShield 9500 Surgical N95 respirator is a single use device intended for occupational use to protect against microorganisms, body fluids and particulate material.
| | New Device | Predicate Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) # | To be determined | K090414 |
| Company | NexEra Medical, Inc. | NexEra Medical, Inc. |
| Name/Model | SpectraShield 9500 Surgical N95 Respirator *<br>(*with amended Intended use Statement) | SpectraShield 9500 Surgical N95 Respirator |
| Fabrics | Nonwoven polyester containing a silver-<br>copper zeolite (antibacterial agent) and a<br>meltblown polypropylene substrate. | Nonwoven polyester containing a silver-<br>copper zeolite (antibacterial agent) and a<br>meltblown polypropylene substrate. |
| Nosepiece | 100% Aluminum | 100% Aluminum |
| Straps | (2) Polyamide fiber and elastic straps, latex free | (2) Polyamide fiber and elastic straps, latex free |
| Mask Style | Molded shape | Molded shape |
| Fluid Resistance<br>ASTM F1862 | Pass: Fluid Resistant @ 160mm Hg | Pass: Fluid Resistant @ 160mm Hg |
| | New Device | Predicate Device |
| Particulate<br>Filtration Efficiency<br>ASTM F2299 | Pass: 99.87% at 0.1 microns | Pass: 99.87% at 0.1 microns |
| Differential<br>Pressure<br>Mil M36954C | Pass: 4.3mm H20/cm2 | Pass: 4.3mm H20/cm2 |
| Bacterial<br>Filtration<br>Efficiency<br>ASTM F2101 | Pass: 99.9% | Pass: 99.9% |
| Flammability<br>Class 16CFR 1610 | Class 1 | Class 1 |
| Cytotoxicity<br>10993-10 | Pass: USP reactivity score = < 2 | Pass: USP reactivity score = < 2 |
| Primary skin<br>irritation<br>ISO10993-10 | Pass: PSI Score = 0, Non-irritant | Pass: PSI Score = 0, Non-irritant |
| Repeated Patch<br>Dermal<br>Sensitization<br>ISO 10993-10 | Pass: 0% incidence sensitization response<br>"0" severity at each evaluated time point. | Pass: 0% incidence sensitization response<br>"0" severity at each evaluated time point. |
| Systemic Toxicity<br>ISO 10993-11 | Pass: No mortality or evidence of systemic<br>toxicity from the extracts was observed. | Pass: No mortality or evidence of systemic<br>toxicity from the extracts was observed. |
| Physico-chemical<br>USP<br>Physico-chemical Test-<br>Plastics | Pass: Test results met the USP limits. | Pass: Test results met the USP limits. |
| Gas off Testing | Total antibacterial particles released from<br>the device were verified to be within safe<br>inhalation levels. | Total antibacterial particles released from<br>the device were verified to be within safe<br>inhalation levels. |
| Leach off testing | Total leachable antibacterial particles<br>released from the device were verified to be | Total leachable antibacterial particles<br>released from the device were verified to be |
| | New Device | Predicate Device |
| | within safe orally ingestible levels. | within safe orally ingestible levels. |
| BioEfficacy : | To Inoculums measured, >106<br>S.pyogenes: > 4.40log10reduction - 1 hour<br>H.influenzae: > 6.20log10reduction - 1 hour<br>MRSA: > 4.83log10reduction - 1 hour | To Inoculums measured, >106<br>S.pyogenes: > 4.40log10reduction - 1 hour<br>H.influenzae: > 6.20log10reduction - 1 hour<br>MRSA: > 4.83log10reduction - 1 hour |
| BioEfficacy :<br>after repeated<br>exposures to<br>perspiration over<br>12 hours | To Inoculums measured, >106<br>S.pyogenes: > 4.25log10reduction - 1 hour<br>H.influenzae: > 4.18log10reduction - 1 hour<br>MRSA: > 4.11log10reduction - 1 hour | To Inoculums measured, >106<br>S.pyogenes: > 4.25log10reduction - 1 hour<br>H.influenzae: > 4.18log10reduction - 1 hour<br>MRSA: > 4.11log10reduction - 1 hour |
| Intended Use<br>Statement | The SpectraShield 9500 Surgical N95<br>Respirator is a single use, disposable surgical<br>N95 respirator, tested for continuous use up<br>to 8 hours, embedded with a zeolite carrier<br>containing a silver-copper agent on the outer<br>layer and is not an antimicrobial drug.<br>SpectraShield 9500 kills 99.99% of test<br>bacteria after one hour of contact with the<br>surface of the respirator. In vitro (laboratory)<br>tests have demonstrated 99.99% kill on the<br>surface of the outer layer of the respirator<br>when tested in vitro against single isolates of<br>the following test bacteria: Streptococcus<br>pyogenes, MRSA (Methicillin Resistant<br>Staphylococcus aureus), and Haemophilus<br>influenzae under tested contact conditions.<br>No clinical studies have been conducted<br>comparing the ability of the untreated<br>surgical N95 respirator and the SpectraShield<br>model 9500 surgical N95 respirator to<br>protect the wearer from infection and the<br>antibacterial treatment cannot effect<br>pathogens that are inhaled around the edges<br>of the respirator. | The SpectraShield 9500 Surgical N95<br>Respirator is a single use, disposable surgical<br>N95 respirator, embedded with a zeolite<br>carrier containing a silver-copper agent on<br>the outer layer and is not an antimicrobial<br>drug. SpectraShield 9500 kills 99.99% of test<br>bacteria after one hour of contact with the<br>surface of the respirator. In vitro (laboratory)<br>tests have demonstrated 99.99% kill on the<br>surface of the outer layer of the respirator<br>when tested in vitro against single isolates of<br>the following test bacteria: Streptococcus<br>pyogenes, MRSA (Methicillin Resistant<br>Staphylococcus aureus), and Haemophilus<br>influenzae under tested contact conditions.<br>No clinical studies have been conducted<br>comparing the ability of the untreated<br>surgical N95 respirator and the SpectraShield<br>model 9500 surgical N95 respirator to<br>protect the wearer from infection and the<br>antibacterial treatment cannot effect<br>pathogens that are inhaled around the edges<br>of the respirator. |
| New Device | Predicate Device | |
| The SpectraShield 9500 Surgical N95<br>respirator is a single use device intended for<br>occupational use to protect against<br>microorganisms, body fluids and particulate<br>material. | respirator is a single use device intended for<br>occupational use to protect against<br>microorganisms, body fluids and particulate<br>material. | |
## 510(k) Summary Device Comparison Table
3343 West Commercial Blvd., Suite 103 - Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for "NEXERA MEDICAL". The logo is in black and features a stylized "X" with a diamond shape above it. The word "MEDICAL" is written in smaller letters below the "ERA" portion of "NEXERA".
3343 West Commercial Blvd., Suite 103 – Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for "NEXERA MEDICAL". The logo is in black and white. The word "NEXERA" is in large, bold letters, with a diamond shape above the "X". Below "NEXERA" is the word "MEDICAL" in smaller letters.
3343 West Commercial Blvd., Suite 103 – Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com
{4}------------------------------------------------
The subject device (SpectraShield model 9500 Surgical mask with the revised IFU Conclusion: referencing "tested for continuous use up to 8 hours"), and the predicate device (K090414) are the same device. The intention of this 510k submittal is to change the IFU to include the statement "tested for continuous use up to 8 hours".
> The predicate device (K090414) was tested for bio-efficacy after repeated exposures to perspiration over a 12 hour period (see K090414 Repeat Challenge Protocol and Testing). The repeat challenge testing required the predicate device be repeatedly exposed to perspiration over a 12 hour period. Following the 12 hour exposure the predicate device was tested and demonstrated 99.99% kill on the surface of the outer layer of the respirator when tested in vitro against Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae under tested contact conditions. The intention of the repeated challenge and sustained exposure was to demonstrate that the device would still function as intended (99.99% kill) after wearing the device for 12 hours.
The IFU for the predicate device references "single use, disposable device". The proposed change to the IFU would read "single use, disposable device, tested for continuous use up to 8 hours."
It is our conclusion that the proposed change to the IFU does not change the intended use of the device, and we believe the change to the IFU further clarifies the intended use of the device. Additionally, we note the proposed change to the IFU demonstrate the device is as safe and as effective as the predicate device and performs equally as well.
3343 West Commercial Blvd., Suite 103 - Ft. Lauderdale, FL 33309 USA Telephone: 954-495-2020 ~ Fax: 954-495-7281 ~ www.nexeramed.com
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of an eagle with its wings spread, which is a common symbol of the United States. The logo is black and white.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
## Mr. Paul Sallarulo Nexera Medical Incorporated 3343 West Commercial Boulevard Suite 103 Fort Lauderdale, Florida 33309
5 2012 JUL
Re: K120244 Trade/Device Name: SpectraShield Model 9500 Surgical Respirator Regulation Number: 21 CFR 878.4040 Regulation Name: Surgical Apparel Regulatory Class: II Product Code: ONT Dated: June 21, 2012 Received: June 25, 2012
Dear Mr. Sallarulo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading:
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
### Page 2- Mr. Sallarulo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
For
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
### Indications for Use Statement Section 4.0:
510(k) Number: 510(k) submission K12 0244
SpectraShield model 9500 Surgical Respirator Device Name:
Indications for Use:
The SpectraShield 9500 Surgical N95 Respirator is a single use, disposable, surgical N95 respirator, tested for continuous use up to 8 hours, embedded with a zeolite carrier containing a silver-copper agent on the outer layer and is not an antimicrobial drug. SpectraShield 9500 kills 99.99% of test bacteria after one hour of contact with the surface of the respirator. In vitro (laboratory) tests have demonstrated 99.99% kill on the surface of the outer layer of the respirator when tested in vitro against single isolates of the following test bacteria: Streptococcus pyogenes, MRSA (Methicillin Resistant Staphylococcus aureus), and Haemophilus influenzae, under tested contact conditions.
No clinical studies have been conducted comparing the ability of an untreated surgical N95 respirator and the SpectraShield model 9500 surgical N95 respirator to protect the wearer from infection, and the antibacterial treatment cannot effect pathogens that are inhaled around the edges of the respirator.
The SpectraShield 9500 Surgical N95 respirator is a single use device, intended for occupational use to protect against microorganisms, body fluids, and particulate material.
Prescription Use
AND/OR
Over -the-counter Use X (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation (ODE)
Nalip. hanyhim. 1
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.