A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)
Device Story
Disposable vinyl patient examination gloves; worn on examiner's hands or fingers; intended to prevent cross-contamination between patient and examiner. Non-sterile; powder-free (residual powder < 2mg/glove). Used in clinical settings by healthcare providers. Device acts as a physical barrier. Benefits include reduction of contamination risk during medical examinations.
Clinical Evidence
Bench testing only. Testing included physical and dimension measurements (ASTM D5250-06), FDA 1000 ml water fill test for pinholes, and biocompatibility testing (primary skin irritation and skin sensitization).
Technological Characteristics
Vinyl material; powder-free (< 2mg/glove). Complies with ASTM D5250-06. Non-sterile. Physical barrier sensing/actuation. No electronic components or software.
Indications for Use
Indicated for use as a disposable medical glove worn on the examiner's hands or fingers to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
K120230 1 of 2
Shandong Taixiang Medical Products Co., Ltd.
No.15, East Road, Hongrun Industry Park Oingzhou, Shandong, China
JUL 2 7 2012
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# 510 (K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21CFR 807.92.
### Submitter's Identification: 1.
Shandong Taixiang Medical Products Co., Ltd. No.15, East Road, Hongrun Industry Park Oingzhou, Shandong, China
## Contact Person:
Yin, Xiaohong Product Manager Phone Number: 011-86-5365768606
Date summary prepared: January 10, 2012
### Name of the Device: 2.
Shandong Taixiang Medical Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves, Non-Sterile
### 3. Predicate Device Informaton:
Shijiazhuang Star Plastic Co., Ltd. Vinyl Examination Gloves - Powder Free (K100699)
Tangshan Zhonghong Pulin Food Products Co., Ltd. Vinyl Examination Gloves - Powder Free (K022091)
#### 4. Device Description:
A Powder-Free Vinyl Patient Examination Gloves is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880,6250). Our Powder-Free Vinyl Patient Examination Gloves do not contain any UPS powder according to the results of testing conducted on the gloves. The residual powder testing is conducted per standards of ASTM D-6124-06, which demonstrates that the powder free glove is less than 2mg/pc. Classified by FDA's General and Plastic Surgery Device panel as Class I, 21 CFR 880.6250. Powder-Free Vinyl Patient Examination Glove, 80LYZ, and meets all requirement of ASTM Standard D5250-06.
{1}------------------------------------------------
# Shandong Taixiang Medical Products Co., Ltd.
No.15, East Road, Hongrun Industry Park Oingzhou, Shandong, China
## ડ. Intended Use:
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)
## Comparison to Predicate Devices: 6.
Shandong Taixiang Medical Products Co., Ltd. Powder-Free Vinyl Patient Examination Gloves are substantially equivalent in safety and effectiveness to Tangshan Zhonghong Pulin Food Products Co., Ltd. (K022091) and Shijiazhuang Star Plastic., Co., Ltd. (K100699) Powder-Free Vinyl Patient Examination Gloves.
## 7. Discussion of Non-Clinical tests performed for Determination of Substantial Equivalence are as follows:
The standards used for Shandong Taixiang Medical Products Co., Ltd. glove production are based on ASTM-D-5250-06. All testing meets requirements for Physical and Dimensions Testing conducted on gloves, Inspection Level S-2, AOL 2.5.
The FDA 1000 ml. Water Fill Test was also conducted with samplings of AOL 2.5. Inspection Level I, meeting these requirements. Primary Skin irritation and Skin Sensitization (allergic contact dermatitis) testing was conducted with results showing no primary skin irritant or sensitization reactions.
There are no special labeling claims and we do not claim our gloves as hypoallergenic is conducted to insure that our gloves meet our "powder-free" claims (contain no more than 2 mg powder per glove).
## 8. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic claim.
## 9. Conclusions:
Shandong Taixiang Medical Products Co., Ltd. Powder-Free Viny] Patient Examination Gloves conforms fully to ASTM-D-5250-06 standard as well as applicable 21 CFR references, and, meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data in Section 7. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle with three tail feathers.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Shandong Taixiang Medical Products Company, Limited C/O Mr. Ray Zhou Basic Medical Industries, Incorporated 12390 East End Avenue Chino, California 91710
Re: K120230
Trade/Device Name: Powder-Free Vinyl Patient Examination Gloves, Non-Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYZ Dated: June 29, 2012 Received: July 3, 2012
Dear Mr. Zhou:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
JUL 2 7 2012
{3}------------------------------------------------
## Page 2 - Mr. Zhou
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements · of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free . number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Fo
Anthony D. Watson, B.S., M.S., M.B.A. Director
Division of Anesthesiology, General Hospital. Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
## Enclosure
{4}------------------------------------------------
# Shandong Taixiang Medical Products Co., Ltd.
No.15, East Road, Hongrun Industry Park Qingzhou, Shandong, China
# INDICATIONS FOR USE
Applicant: Shandong Taixiang Medical Products Co., Ltd.
K120230 510(k) Number:
Powder-Free Vinyl Patient Examination Gloves, Non-Sterile Device Name:
Indications of Use:
A patient examination glove is a disposable device intended for medical purposes that is worn upon the examiner's hands or finger to prevent contamination between patient and examiner (21CFR 880.6250)
Prescription Use
Over the Counter Use X
Factory Initials
Elizabet P. Clain-Weller
(Division Sian-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K120230
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.