K120203 · Zimmer Trabecular Metal Technology, Inc. · MAX · Jun 15, 2012 · Orthopedic
Device Facts
Record ID
K120203
Device Name
TM-400 DEVICE
Applicant
Zimmer Trabecular Metal Technology, Inc.
Product Code
MAX · Orthopedic
Decision Date
Jun 15, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3080
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The TM-400 device is a Vertebral Body Replacement device intended for use in the thoracolumbar spine (T1 – L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The TM-400 device is intended for use with supplemental internal fixation systems, and may be used with autograft or allograft. The TM-400 device is also intended for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain with discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. The TM-400 device is intended for use with supplemental internal fixation and autogenous bone graft.
Device Story
TM-400 is an oval-shaped, porous tantalum intervertebral body fusion and vertebral body replacement device. Implanted surgically in the thoracolumbar spine to replace collapsed/damaged vertebrae or fuse adjacent bony surfaces in DDD patients. Features include patterned teeth on superior/inferior surfaces for initial stability and slots for insertion instrumentation. Used with supplemental internal fixation and bone graft (autograft/allograft). Provides structural support to the anterior column of the spine; facilitates fusion. Available in various heights, cross-sections, and lordotic angles (7 and 13 degrees) to match patient anatomy.
Clinical Evidence
Bench testing only. Mechanical performance evaluated per FDA Class II Special Controls Guidance Document: Intervertebral Fusion Device. Testing included static and dynamic axial compression (ASTM F2077-03), static and dynamic torsion (ASTM F2077-01), expulsion, and subsidence (ASTM F2267-04 & F2077). Results demonstrated the system meets predetermined mechanical performance requirements.
Technological Characteristics
Material: Trabecular Metal Porous Tantalum. Form factor: Oval-shaped implant with surface teeth and insertion slots. Lordotic angles: 7 and 13 degrees. Mechanical testing standards: ASTM F2077-03, F2077-01, F2267-04. No software or electronic components.
Indications for Use
Indicated for skeletally mature patients requiring vertebral body replacement (T1-L5) due to tumor or trauma (fracture), or intervertebral body fusion (L2-S1) for degenerative disc disease (DDD) with up to Grade I spondylolisthesis or retrolisthesis. DDD patients must have discogenic back pain, radiographic confirmation, and 6 months of failed non-operative treatment. Requires supplemental internal fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{1}------------------------------------------------
# JUN 1 5 2012
### 510(k) Summary
| Submitter: | Zimmer Trabecular Metal Technology, Inc.<br>10 Pomeroy Road<br>Parsippany, New Jersey 07054 |
|-----------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person: | Kathleen Rutherford<br>Associate Director, Regulatory Affairs<br>Telephone: (973) 576-0139<br>Fax: (973) 884-8792 |
| Date: | June 6, 2012 |
| Trade Name: | TM-400 Device |
| Common Name:<br>Replacement<br>Classification Name: | Intervertebral body fusion device & Spinal Vertebral Body<br>Device<br>Intervertebral body fusion device, 21 CFR § 888.3080,<br>Spinal intervertebral body fixation orthosis, 21 CFR § 888.3060 |
Device Panel/Product Code:Orthopedic MAX & MQP
### Device Description:
The TM-400 Device is an oval shaped ALIF device for interbody fusion of the anterior column of the spine. The TM-400 Device is currently cleared to accommodate the replacement of a vertebral body in the thoracic and lumbar region of the spine. Use of this device is expanded to include use as an interbody fusion device. TM-400 Device is designed for fusing the adjacent bony surfaces and may be used to replace a disc at one or two contiguous levels in L2-S1. The device is available in a variety of cross sections and heights to accommodate variations in the individual pathology and anatomic condition of the patient. TM-400 implants are also available in two lordotic angles 7 degree and 13 degree. The superior and inferior surfaces of the device contain a pattern of teeth to provide for initial stability. The device also has slots to mate with the insertion instrument. The TM-400 device is wholly comprised of Trabecular Metal Porous Tantalum. Surgical instrumentation for use with the proposed system are fabricated from surgical grade stainless steel and other applicable materials.
### Indications for Use:
The TM-400 device is a Vertebral Body Replacement device intended for use in the thoracolumbar spine (T1 – L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The TM-400 device is intended for use with supplemental internal fixation systems, and may be used with autograft or allograft.
The TM-400 device is also intended for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-S1. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain with discogenic origin with degeneration of the
{2}------------------------------------------------
disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. The TM-400 device is intended for use with supplemental internal fixation and autogenous bone graft.
### Device Technological Characteristics and Comparison to Predicate Device(s):
The TM-400 Device was shown to be substantially equivalent to legally marketed predicate devices. The predicate devices include this TM-400 Device as a VBR (K052950, K031823, K010378 and, for the instrumentation, K093127), the Zimmer Spine, Inc. Ardis® Interbody System cleared under K073202, the Biomet Enclave cleared under K081636 and the Orthofix Pillar AL cleared under K082235.
The TM-400 Device has the identical material as previously cleared devices. The intended use and indications for use of the subject device are similar to those of its predicate devices. The sizes, design features and overall geometry of the device in the current submission are similar to the cleared predicate devices.
There are no significant differences between the TM-400 Device and the predicate devices currently being marketed that would adversely affect the use of the product. Any differences in technological characteristics do not raise new issues of safety or efficacy. The subject system is similar to its predicate devices with respect to intended use/indications for use, material, technological characteristics and basic principles of operation.
### Performance Data:
Mechanical testing was performed on the TM-400 Device as recommended by the FDA Class II Special Controls Guidance Document: Intervertebral Fusion Device. The tests that were performed and the relevant ASTM Standards are as follows:
| Mechanical Tests | ASTM Standard |
|---------------------------|------------------|
| Static Axial Compression | F2077-03 |
| Dynamic Axial Compression | F2077-03 |
| Static Torsion | F2077-01 |
| Dynamic Torsion | F2077-01 |
| Expulsion | N/R |
| Subsidence | F2267-04 & F2077 |
The results of testing and analyses conducted demonstrate that the proposed system adequately meets the predetermined requirements established for its mechanical performance.
### Substantial Equivalence:
The TM-400 Device is substantially equivalent to its predicate devices with respect to intended use/indications for use, materials, technological characteristics and basic principles of operation as demonstrated by the supporting performance testing data.
{3}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JUN 1 5 2012
Zimmer Trabecular Metal Technology, Inc. % Ms. Kathleen Rutherford Associate Director, Regulatory Affairs 10 Pomeroy Road Parsippany, New Jersey 07054
Re: K120203
Trade/Device Name: TM-400 Device Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX, MQP Dated: May 21, 2012 Received: May 29, 2012
Dear Ms. Rutherford:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{4}------------------------------------------------
### Page 2 - Ms. Kathleen Rutherford
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark McMillan
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: TM-400 Device
Indications for Use:
The TM-400 device is a Vertebral Body Replacement device intended for use in the thoracolumbar spine (T1 - L5) to replace a collapsed, damaged or unstable vertebral body due to tumor or trauma (i.e., fracture). The TM-400 device is intended for use with supplemental internal fixation systems, and may be used with autograft or allograft.
The TM-400 device is also intended for use as an intervertebral body fusion device in skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels from L2-SI. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). DDD is defined as back pain with discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These patients should have had six months of non-operative treatment. The TM-400 device is intended for use with supplemental internal fixation and autogenous bone graft.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
:
(Division Sign-Off) (Division of Surgical, Orthopedic, and Restorative Devices
K120203 510(k) Number_
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.