← Product Code [CHH](/productcode/CHH) · K120202

# EON CHOLESTEROL REAGENT (K120202)

_Vital Diagnostics Pty., Ltd. · CHH · Jul 27, 2012 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K120202

## Device Facts

- **Applicant:** Vital Diagnostics Pty., Ltd.
- **Product Code:** [CHH](/productcode/CHH.md)
- **Decision Date:** Jul 27, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1175
- **Device Class:** Class 1
- **Review Panel:** Clinical Chemistry

## Indications for Use

Eon Cholesterol Reagent is an in vitro diagnostic device intended for the quantitative determination of total cholesterol in human plasma and serum using the Eon 100 Clinical Chemistry Analyzer. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

## Device Story

Eon Cholesterol Reagent is an in vitro diagnostic reagent used with the Eon 100 Clinical Chemistry Analyser. It measures total cholesterol levels in human plasma and serum samples. The device is intended for use by healthcare professionals in clinical laboratory settings to assist in diagnosing and managing lipid and lipoprotein metabolism disorders. The analyzer processes the sample with the reagent to provide a quantitative cholesterol concentration, which clinicians use to assess patient health and guide therapeutic decisions.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

In vitro diagnostic reagent for quantitative cholesterol determination. Designed for use with the Eon 100 Clinical Chemistry Analyser. Operates via standard clinical chemistry enzymatic assay principles.

## Regulatory Identification

A cholesterol (total) test system is a device intended to measure cholesterol in plasma and serum. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE

A. 510(k) Number:
k120202

B. Purpose for Submission:
New device

C. Measurand:
Total Cholesterol

D. Type of Test:
Quantitative, Enzymatic Esterase – Oxidase Cholesterol

E. Applicant:
Vital Diagnostics (Manufacturing) Pty Ltd

F. Proprietary and Established Names:
Eon Cholesterol Reagent

G. Regulatory Information:
|  Product Code | Classification | Regulation Section | Panel  |
| --- | --- | --- | --- |
|  CHH | Class I, (meets limitations of exemption per 21 CFR 862.9(c)(4)) | 862.1175 | Clinical Chemistry  |

H. Intended Use:

1. Intended use(s):
Refer to indications for use below

2. Indication(s) for use:
Eon Cholesterol Reagent is an in vitro diagnostic device intended for the quantitative

{1}

determination of total cholesterol in human plasma and serum using the Eon 100 Clinical Chemistry Analyzer. Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders.

3. Special conditions for use statement(s):

For in vitro diagnostic use

For prescription use

3. Special instrument requirements:

Eon 100 Clinical Chemistry Analyzer (previously cleared in k100060)

I. Device Description:

The Eon Cholesterol Reagent kit includes 4 x 23 mL bottles that contain the following:

- 0.25 mmol/L 4-aminoantipyrine
- 10.0 mmol/L Hydroxybenzoic acid (HBA)
- 200 U/L cholesterol oxidase (microbial)
- 500 U/L cholesterol esterase (microbial)
- 600 U/L peroxidase (horseradish)

As well as, buffer, surfactants, and other ingredients including sodium azide

Handle and dispose of all human source materials as though capable of transmitting infectious agents using the universal precautions recommended by the Centers for Disease Control and Prevention (CDC).

J. Substantial Equivalence Information:

1. Predicate device name(s):

Roche Cholesterol (CHOL) Reagent

2. Predicate 510(k) number(s):

k952127

{2}

3. Comparison with predicate:

|  Item | Eon Cholesterol Reagent (Candidate Device) | Roche Cholesterol (CHOL) Reagent (Predicate - k952127)  |
| --- | --- | --- |
|  Intended Use | The Cholesterol Reagent is an in vitro diagnostic device intended for the quantitative determination of total cholesterol in human plasma and serum | Same  |
|  Indications for Use | Cholesterol measurements are used in the diagnosis and treatment of disorders involving excess cholesterol in the blood and lipid and lipoprotein metabolism disorders. | Same  |
|  Methodology | Enzymatic (esterase/oxidase end-point) | Same  |
|  Patient Sample Type | Serum and Plasma | Same  |
|  Assay Range | 6 – 775 mg/dL | 3 – 800 mg/dL  |
|  On-board stability | 28 days | Same  |
|  Limit of Detection | 2 mg/dL | 3 mg/dL  |
|  Reagent configuration | Single part liquid | Same  |

K. Standard/Guidance Document Referenced (if applicable):

CLSI EP06-A: Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline

CLSI EP05-A2: Evaluation of Precision Performance of Clinical Chemistry Devices; Approved Guideline – Second Edition

CLSI EP09-A2-IR: Method Comparison and Bias Estimation using Patient Samples; Approved Guideline – Second Edition (Interim Revision)

CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline

L. Test Principle:

Free cholesterol and cholesterol produced from the enzymatic hydrolysis of cholesterol esters are oxidized in the presence of cholesterol oxidase. The resulting hydrogen peroxide reacts with Hydroxybenzoic acid (HBA) and 4-aminoantipyrine to produce a red quinoneimine dye in the presence of peroxidase. The red quinoneimine dye absorbs at 505 nm. The final

{3}

absorbance at this wavelength is proportional to the concentration of total cholesterol in the sample.

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

Four levels of control serum were tested on the Eon 100 analyzer two times per run, two runs per day for 20 days (n = 80). The results are summarized in the tables below:

|  Sample 1
Mean 120.6 mg/dL
Cholesterol | Within
Run | Total  |
| --- | --- | --- |
|  Standard Deviation, mg/dL | 1.6 | 2.3  |
|  Coefficient of Variation, % | 1.3 | 1.9  |
|  Sample 2
Mean 185.6 mg/dL
Cholesterol | Within
Run | Total  |
| --- | --- | --- |
|  Standard Deviation, mg/dL | 2.6 | 3.4  |
|  Coefficient of Variation, % | 1.4 | 1.8  |
|  Sample 3
Mean 249.2 mg/dL
Cholesterol | Within
Run | Total  |
| --- | --- | --- |
|  Standard Deviation, mg/dL | 2.4 | 3.8  |
|  Coefficient of Variation, % | 1.0 | 1.5  |
|  Sample 4
Mean 413.2 mg/dL
Cholesterol | Within
Run | Total  |
| --- | --- | --- |
|  Standard Deviation, mg/dL | 4.2 | 6.2  |
|  Coefficient of Variation, % | 1.0 | 1.5  |

#### b. Linearity/assay reportable range:

Linearity studies were carried out using serial dilutions of a commercially available serum cholesterol concentrate. The range tested was 5.0 – 832.0 mg/dL. A polynomial fit analysis did not indicate statistically significant non linearity. The linear regression analysis is provided below:

{4}

|  Claimed Measuring Range | Intercept (95%CI) (mg/dL) | Slope (95% CI)  |
| --- | --- | --- |
|  6 – 775 mg/dL | -0.7 (-5.4 – 4.0) | 0.996 (0.989 – 1.006)  |

Based on the linearity data, the measuring range claim 6 – 775 mg/dL is supported.

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

Reagent Traceability

A Cholesterol Reference Method Laboratory Network (CRMLN) Certificate of Traceability for the Eon Cholesterol Reagent was provided.

Reagent Stability

Closed vial/Shelf life

Accelerated heat stress stability studies have concluded that the reagent is stable for 21 months when stored at 2 – 8 °C. Initial shelf-life claims will be for 18 months; however, real-time studies are on-going and the final claim will be reflected by real-time results. Closed vial/Shelf life protocols and acceptance criteria were reviewed and found to be acceptable.

Open vial/On-board Stability

Open vial/On-board stability protocols and acceptance criteria were reviewed and found to be acceptable. The Eon Cholesterol Reagent demonstrated on-board stability of 28 days.

Eon calibrators, as referenced in the package insert, were cleared in k110394 under the name ATAC Serum Calibrator (Direct Bilirubin and Iron)

Eon controls, as referenced in the package insert, were cleared in k111063 under the name Vital Serum Controls

d. Detection limit:

Protocols for the determination of the limit of blank (LoB) and the limit of detection (LoD) were performed in accordance with the recommendations in the CLSI Guideline EP17-A. Testing was carried out using true blanks and low level samples (total 60 each) on two Eon 100 analyzers. The calculated LoB of 0.70 mg/dL and LoD of 1.5 mg/dL support a LoB and LoD claim of 1 and 2 mg/dL, respectively. The measuring range of the assay is 6 to 775 mg/dL.

{5}

# e. Analytical specificity:

Interference studies were performed by using serum pools containing cholesterol (2 levels) with individual interferents at a range of concentrations. The sera were assayed for cholesterol  $(n\geq 3$  replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery exceeded  $\pm$ $10\%$ . The results were obtained on the Eon 100 analyzer.

|   | Level 1 |   | Level 2 |   |   |
| --- | --- | --- | --- | --- | --- |
|  Interferent | Cholesterol Conc. (mg/dL) | No interference at this level of interferent (mg/dL) | Cholesterol Conc. (mg/dL) | No interference at this level of interferent (mg/dL) | Interference Claim (mg/dL)  |
|  Hemoglobin | 75 | 500 | 230 | 1000 | No interference up to 500  |
|  Lipemia | 66 | 1000 | 208 | 1000 | No interference up to 1000  |
|  Bilirubin, conjugated | 143 | 15 | 242 | 7.5 | No interference up to 7.5  |
|  Bilirubin, unconjugated | 140 | 7.5 | 235 | 7.5 | No interference up to 7.5  |
|  Ascorbic acid | 144 | 5 | 229 | 10 | No interference up to 5  |
|  Gentisic acid | 164 | 117 (μmol/L) | 254 | 117 (μmol/L) | No interference up to 117 (μmol/L)  |
|  Dopamine | 153 | 5.9 (μmol/L) | 251 | 5.9 (μmol/L) | No interference up to 5.9 (μmol/L)  |

# f. Assay cut-off:

Not applicable

# 2. Comparison studies:

a. Method comparison with predicate device:

{6}

Studies were carried out according to CLSI EP09-A2-IR. Serum samples (n=101) were assayed in parallel by both the test (Y) and predicate (X) methods. The results were analyzed by using both Least squares and Deming regression. The range tested 10 – 750 mg/dL. Altered samples were included in the study.

|  Linear Regression | Range (Predicate) | n | Slope | Intercept | R | Standard error est.  |
| --- | --- | --- | --- | --- | --- | --- |
|  Least squares | 10 – 750 (mg/dL) | 10
1 | 0.995
(0.979 – 1.011) | 3.422
(-0.630 – 7.474) | 0.9968 | 10.294  |
|  Deming |   |   | 0.998
(0.982 – 1.012) | 2.722
(-1.33 – 6.777) | 0.9968 | 10.303  |

b. Matrix comparison:

Parallel drawn (matched) samples from 68 individuals were collected as serum and anti-coagulated plasma (Lithium heparin. The paired samples for each individual were assayed in parallel with Eon cholesterol reagent on the Eon 100 analyzer. The results were analyzed by using both Least squares and Deming regression. The range tested 14 - 680 mg/dL. Altered samples were included in the study.

|  Linear Regression | Range (serum) | n | Slope | Intercept | R | Standard error est.  |
| --- | --- | --- | --- | --- | --- | --- |
|  Least squares | 14 - 680 (mg/dL) | 68 | 0.965
(0.943 – 0.987) | 5.389
(0.362 – 10.416) | 0.9956 | 9.926  |
|  Deming |   |   | 0.969
(0.947 – 0.992) | 4.576
(-0.456 – 9.608) | 0.9956 | 9.936  |

3. Clinical studies:

a. Clinical Sensitivity:

- Not applicable

b. Clinical specificity:

- Not applicable

c. Other clinical supportive data (when a. and b. are not applicable):

- Not applicable

4. Clinical cut-off:

- Not applicable

{7}

5. Expected values/Reference range:

For cholesterol, the National Cholesterol Education Program (NCEP) recommends classifying patient results as being desirable, borderline or high based on the following cut-off thresholds established by the Adult Treatment Panel (ATP) III:

Desirable: &lt; 200 mg/dL
Borderline: 200 – 239 mg/dL
High: ≥ 240 mg/dL

N. Proposed Labeling:

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

8

---

**Source:** [https://fda.innolitics.com/device/K120202](https://fda.innolitics.com/device/K120202)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
