Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
Device Story
Multi-Modality Image Fusion (MMIF) is a PC-based software application for clinical/hospital use by physicians. It consists of two modules: offline and online. Offline module: used days/hours pre-procedure to delineate gland/lesion boundaries on DICOM images; saves data as surface meshes (VTK/STL) or binary volumes. Online module: performs rigid/elastic registration of pre-prepared DICOM data with incoming real-time TRUS images; enables side-by-side locked viewing (rotation, slicing, panning, zooming); allows target planning within suspected lesion boundaries. Deformation fields estimated during registration map planned targets back to the original ultrasound volume. Output assists physicians in procedure planning and intra-operative guidance. Benefits include improved visualization of anatomical structures and lesion targeting across modalities.
Clinical Evidence
Bench testing only. Verification and validation included hardware/software testing, phantom volume measurements, and registration accuracy tests using simulated prostate phantoms. Testing confirmed static and dynamic system performance, usability, and reliability under simulated clinical conditions.
Technological Characteristics
PC-based software; DICOM standard compliance (NEMA PS3.1-3.18); supports CT, MRI, and Ultrasound modalities. Features include rigid and elastic image registration, 3D/surface rendering, segmentation, ROI delineation, and volumetric measurement. Connectivity via standard DICOM/JPEG formats. ISO 14971 risk management and ISO 13485 quality management standards applied.
Indications for Use
Indicated for use by physicians in clinical or hospital settings for 2-D and 3-D visualization, registration, and fusion of MRI, CT, and Ultrasound images to map planning information across modalities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
# 510(k) Summary
# APR - 6 2012
# 510(k) Summary
Submitter's Name: Submitter's Address: Submitter's Telephone: Contact Name:
Date Summary was Prepared: Trade or Proprietary Name: Common or Usual Name: Classification Name:
Predicate Devices:
IGT, LLC dba Eigen 13366 Grass Valley Avenue, Grass Valley, CA 95945 530-274-1240 . William Mandel 21 Oct 2011
Multi-Modality Image Fusion Multi-Modality Image Fusion
System, Image Processing, Radiological, LLZ
Picture Archiving and Communications, 21CFR 892.2050
| Device Name | 510(k) Number |
|------------------|---------------|
| Syntegra | K041182 |
| UniSyn | K081987 |
| Aegis Navigation | K093672 |
| Abaris | K053610 |
### Intended Use
Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
### Description of the Device and Summary of the Technological Characteristics:
Multi-Modality Image Fusion (MMIF) is software, which comprises of two software components, which is referred to as offline and online. The offline component perfains to the preparation of gland and suspected lesion boundaries on a DICOM image file days or hours prior to the biopsy procedure. The online component fuses the DICOM image files, which were, prepared on the offline component, with a snap shot incoming TRUS image. Each of the two software components can work together or independently.
### Offline Software
The offline software can delineate suspected areas of interest using visualization and save this data in a standard format, e.g. as a surface mesh in VTK or STL format. Furthermore, the
#### EIGEN
PREMARKET NOTIFICATION - MULTI-MODALITY IMAGE FUSION
{2}------------------------------------------------
offline software can ontionally delineate the gland boundary from a DICOM image, or other volumes where the boundary can be easily observed. This addition data is also saved in a standard format. e.g. surface mesh in VTK or STL or as a binary volume. This software would be typically executed by Radiologists.
# Online Software
The online unit software is designed to fuse two images by entering a fusion mode and load the DICOM image and the boundary data provided by the offline unit. The online unit performs rigid alignment, either selecting globally align that orients the two DICOM volumes based on the known relative orientation difference between the two or specify landmarks to register images rigidly. The online unit then elastically warps the image to register on the incoming image from the offline unit.
The registered image volume is displayed on the incoming offline image frame of reference. Also overlaid on the incoming offline and registered image volumes annotating the suspected lesion boundaries. The two volumes are displayed side-by-side in a locked view. Rotation, slicing, panning, or zooming one of the volumes has the same effect on the other enabling visualization of similar features in both volumes simultaneously.
Plan targets on the volumes by selecting suitable locations within the suspected areas of interest boundary. When sufficient targets have been added, exiting fusion mode imports all targets planned in the offline frame of reference back to the original reconstructed ultrasound volume. This is done based on the deformation field that was estimated from registration (rigid followed by elastic registration).
#### Substantial Equivalence:
| Manufactu<br>rer | Eigen | ADAC<br>Laboratories | Convergent<br>Imaging<br>Solutions | Sentinelle Medical<br>Inc. | Traxtal<br>Technologies lhc. |
|------------------|------------------------------------------------------------------------|----------------------------------------------------------------|-----------------------------------------------------|-----------------------------------------------------------------------|---------------------------------------------------------|
| Product<br>Name | Multi-Modality<br>Image Fusion | Syntegra | UniSyn | Aegis Navigation | Abaris |
| 510(k)<br>number | pending | K041182 | K081987 | K093672 | K053610 |
| Intended<br>Use | Multi-Modality<br>Image Fusion is a<br>software<br>application used by | Syntegra is a<br>software<br>application for<br>multi-modality | UniSyn is a<br>software<br>application for<br>image | This device<br>provides two and<br>three-dimensional<br>image review. | ABARIS is a<br>stereotaxic<br>accessory for<br>Computed |
EIGEN
PREMARKET NOTIFICATION - MULTI-MODALITY IMAGE FUSION
{3}------------------------------------------------
| physicians in the | image registration | registration and | manipulation, and | Tomography | | | | |
|---------------------------------------|---------------------|------------------|---------------------|---------------------|------------------------------------------------------------------|--------------------|------------------------------------------------------------------------|------------|
| clinic or hospital | and diagnostic | fusion display | analysis tools to | (CT), Magnetic | | | | |
| for 2-D and 3-D | fusion. Images are | of scanned | assist users in | Resonance, (MR), | | | | |
| visualization, | registered and | image data | screening, | Ultrasound (US), | | | | |
| multi-modality | displayed in a | from CT, PET, | diagnosis, planning | Positron Emission | | | | |
| image registration, | "fused" (overlaid | SPECT and | and performing | Tomography | | | | |
| and fusion of | in the same spatial | MR scanners. | image-guided | (PET), Single | | | | |
| medical images. | orientation) | UniSyn creates | interventional | Photon Emission | | | | |
| | | | | | | | | |
| Additional | format to provide | multi-planar | procedures. | Computed | | | | |
| software features<br>include database | combined | reformat and | Supported imaging | Tomography | | | | |
| | functional and | maximum | modalities include | (SPECT), | | | | |
| management, | anatomical data | intensity | Magnetic | Fluoroscopy, | | | | |
| communication. | providing | projection | Resonance (MR), | Endoscopy and | | | | |
| surface rendering, | different angular | displays of the | Ultrasound (US), | other imaging | | | | |
| segmentation, ROI | perspectives for | data and | Single Photon | systems. it | | | | |
| delineation, | interpretation by | provides | Emission | displays the | | | | |
| measurements, and | trained | measurements | Computed | simulated image | | | | |
| reporting. | professionals | such as area, | Tomography | of a tracked | | | | |
| | | volume and | (SPECT), | insertion tool | | | | |
| | | Standard | Computed | such as a biopsy | | | | |
| | | Uptake Values | Tomography (CT), | needle, guidewire | | | | |
| | | for user defined | Positron | or probe on a | | | | |
| | | regions on the | Emission | computer monitor | | | | |
| | | image | Tomography | screen that shows | | | | |
| | | | (PET), | images of the | | | | |
| | | | Fluoroscopy and | target organs and | | | | |
| | | | Endoscopy. Images | the current and the | | | | |
| | | | and data are | projected future | | | | |
| | | | received from | path of the | | | | |
| | | | various imaging | interventional | | | | |
| | | | systems and other | instrument taking | | | | |
| | | | sources such as | into account | | | | |
| | | | calibrated spatial | movements of the | | | | |
| | | | positioning | patient. This is | | | | |
| | | | devices. | intended for | | | | |
| | | | This device | treatment | | | | |
| | | | provides the | planning and | | | | |
| | | | capability to | intra-operative | | | | |
| | | | overlay annotations | guidance for | | | | |
| | | | on 2D or 3D | surgical | | | | |
| | | | medical image | procedures. | | | | |
| | | | displays. | The device also | | | | |
| | | | These annotations | supports an | | | | |
| | | | may represent the | image-free mode | | | | |
| | | | position of | in which the | | | | |
| | | | instruments | proximity of the | | | | |
| | | | including but not | interventional | | | | |
| | | | limited to biopsy | device is | | | | |
| | | | needles, guidance | displayed relative | | | | |
| | | | wires, imaging | to another device. | | | | |
| | | | probes or other | The device is | | | | |
| | | | tracked devices. | intended to be | | | | |
| | | | This device is | used in clinical | | | | |
| | | | intended to assist | interventions and | | | | |
| | | | skilled medical | for anatomical | | | | |
| | | | professionals in | structures where | | | | |
| | | | | | skilled medical<br>professionals in<br>clinical screening<br>and | clinical screening | for anatomical<br>structures where<br>imaging is<br>currently used for | imaging is |
.
510(k) Summary
.
### EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
・
・
.
-
. . . . .
,
.
{4}------------------------------------------------
| | | | | 510(k) Summary | |
|----------------------------------------|--------------------------------------------------------|-----------------------------------------------------------|------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | | | | | |
| | | | | interventions, for<br>anatomical<br>structures where<br>imaging is<br>currently used for<br>visualizing such<br>structures,<br>including head and<br>neck, breast,<br>thoracic, and<br>abdominal<br>applications<br>(including | visualizing such<br>procedures. The<br>device is also<br>intended for use<br>in clinical<br>interventions to<br>determine the<br>proximity of<br>one device<br>relative to<br>another. |
| | | | | pelvis).<br>Lossy compressed<br>mammographic<br>images and<br>digitized film<br>screen images must<br>not be | |
| | | | | reviewed for<br>primary image<br>interpretations.<br>Mammographic<br>images may only<br>be interpreted<br>using | |
| | | | | a FDA approved<br>monitor that offers<br>at least 5 MPixel<br>resolution and<br>meets other<br>technical | |
| | | | | specifications<br>reviewed and<br>accepted by the<br>FDA. | |
| Where<br>Used | Office settings in<br>clinic or hospital | Office settings in<br>clinic or hospital | Office settings<br>in clinic or<br>hospital | Not specified | clinical<br>interventions |
| Software<br>Device | Yes | Yes | Yes | Yes | Yes |
| Image<br>Registratio<br>n | Multi-modality<br>image registration | Multi-modality<br>image registration | Multi-modality<br>image<br>registration | Multi-modality<br>image registration | Yes |
| Image<br>Fusion<br>/Overlay<br>Display | Fused overlay of<br>images from<br>multiple modalities | Fused overlay of<br>images from<br>multiple<br>modalities | Fusion overlay<br>of images from<br>multiple<br>modalities | overlay of images<br>from multiple<br>modalities | Yes |
| Opacity<br>Control | Yes | Yes | Yes | Not Specified | Not Specified |
| 3-D<br>Rendering | Yes | Yes | Yes | Yes | Yes |
| Surface<br>Rendering | Yes | Yes | Yes | Yes | Yes |
| Regions of<br>interest | Yes | Yes | Yes | Yes | Yes |
| Configura | Yes | Yes | Yes | Yes | Yes |
EIGEN
.
:
DIGE."
PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
·
{5}------------------------------------------------
510(k) Summary
| ble image layouts | | | | | |
|---------------------------------------------|------------------------|---------------------------------------------------------------------------------------|--------------------------------------|------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 3D<br>Contouring | Yes | Not Specified | Yes | Yes | Yes |
| Export of<br>3D contours<br>for<br>planning | Yes | Not Specified | Yes | Not Specified | Yes |
| Image<br>Storage<br>and<br>Communication | DICOM, JPEG | DICOM (check) | DICOM | DICOM | DICOM |
| Modalities | CT, MRI,<br>Ultrasound | Positron Emission<br>Tomography<br>(PET) and<br>X-ray Computed<br>Tomography<br>(CT). | CT, PET,<br>SPECT and<br>MR scanners | MRI, US, SPECT,<br>CT, PET,<br>Fluoroscopy,<br>Endoscopy and<br>others | Computed<br>Tomography<br>(CT), Magnetic<br>Resonance, (MR),<br>Ultrasound (US),<br>Positron Emission<br>Tomography<br>(PET), Single<br>Photon Emission<br>Computed<br>Tomography<br>(SPECT),<br>Fluoroscopy,<br>Endoscopy and<br>other imaging |
The product's technical features are substantially equivalent to Syntegra (K041182). UniSyn (K081987), Aegis Navigation (K093672) and Abaris (K053610). The Multi-Modality Image Fusion is a software product that runs on PC-based workstations. Image data is input to the devices and used to generate 3-D views and perform image processing. Like the four predicate devices, the software has image measurement, multi-planar reformatting, segmentation and image registration abilities, fusion of images from different modalities, image storage and retrieval, as well as patient information management functions.
The Syntegra and UniSyn are software products that accept multiple image data types including magnetic resonance, and computed tomography. Aegis Navigation and Abaris are systems that accept multiple modalities that include magnetic resonance, computed tomography and ultrasound. The Multi-Modality Image Fusion has been designed for processing medical images in standard DICOM format. The system is composed of two modules: an offline and an online module for image fusion for planning a procedure.
EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
{6}------------------------------------------------
# Testing and Performance Data:
All product and engineering specifications were verified and validated. Test phantoms incorporating simulated prostates were developed and were used to verify system performance.
The device has been designed and manufactured to conform to the following standards:
ACRINEMA PS3.1-3.18 Digital Imaging and Communications in Medicine (DICOM)
ISO 14971 Application of Risk Management to Medical Devices
ISO 13485 Medical Devices - Quality Management Systems - Requirements For Regulatory purposes
The following tests were performed:
MMIF Hardware verification Test Procedure
MMIF Online Verification Test Procedure
MMIF Offline Software Application Verification Test Procedure .
MMIF Risk Mitigation Verification Test Procedure
Registration Accuracy for Clinical Data (bench test)
Phantom Volume Measurements (bench test)
Phantom Registration Accuracy (bench test)
All the above mentioned tests passed. Theses testing activities were conducted to establish the performance, safety, effectiveness, functionality usability, and reliability characteristics of the new device.
The purpose of the performance testing (bench testing) was to ensure the performance of the new MMIF software by verifying the accuracy of the specifications and simulating real customer data collected from the hospital setting.
The conducted verification tests have confirmed static and dynamic performance of the complete system consisting of both the online and off line software installed on specified hardware. Furthermore these tests establish compliance to the specification under conditions simulation real environment of use.
# Conclusion:
The results of comparing the intended use, function, technological characteristics, mode of operation and specifications of the Multi-Modality Image Fusion with those of the two predicate devices demonstrate that the Multi-Modality Image Fusion is substantially
- EIGEN
PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
{7}------------------------------------------------
equivalent to existing products on the market today.
. . . .
.
ﺎ ﺳﮯ ﺍ
1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 - 1999 -
·
The verification and validation testing included bench performance tests, functionality tests, and system test. All tests were successfully completed.
. '·
-
'
.
: :
#### EIGEN PREMARKET NOTIFICATION – MULTI-MODALITY IMAGE FUSION
. .
{8}------------------------------------------------
Image /page/7/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three stripes forming its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002
APR - 6 2012
IGT LCC dba Eigen % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services 1394 25th Street NW BUFFALO MN. 55313
Re: K120187
Trade/Device Name: Multi-Modality Image Fusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications System Regulatory Class: II Product Code: LLZ Dated: March 1, 2012 Received: March 2 2012
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls: Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or uny 1 vict all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of
{9}------------------------------------------------
### Page 2
medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely Yours,
Janine M. Morris
Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{10}------------------------------------------------
# Indications for Use Form
510(k) Number (if known):
Device Name: Multi-Mogality Image Fusion
Indications for Use: Multi-Modality Image Fusion is a software application to be used by physicians in the clinic or hospital for 2-D and 3-D visualization, image registration, and fusion of MRI, CT and Ultrasound imaging modalities for mapping planning information across modalities. Additional software features include database management, data communication, surface rendering, segmentation, regions of interest (ROI) delineation, volumetric measurements, and data reporting.
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|----------------------------------------------|--------|
| | AND/OR |
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE OF ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k) K120187
1-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.