K120168 · Wipro GE Healthcare Private, Ltd. · LBI · May 2, 2012 · General Hospital
Device Facts
Record ID
K120168
Device Name
LULLABY LED PHOTTHERAPY SYSTEM
Applicant
Wipro GE Healthcare Private, Ltd.
Product Code
LBI · General Hospital
Decision Date
May 2, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5700
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The Lullaby™ LED Phototherapy System is used for the treatment of indirect hyperbilirubinemia in term and pre-term infants, in a hospital environment - NICUs, PICUs and Well-baby Nurseries - administered by trained, professional medical staff, on the order of a licensed medical practitioner. The Lullaby™ LED Phototherapy System is intended for use under the direct supervision of a licensed healthcare practitioner. The Lullaby™ LED Phototherapy System device is not intended to be operated in mobile vehicles including ambulances or other vehicles associated with health care facilities.
Device Story
Lullaby™ LED Phototherapy System treats neonatal hyperbilirubinemia (jaundice) in hospital settings. Device consists of lamp unit with 10 blue LEDs mounted on adjustable roll stand; base slides under bassinets, incubators, or radiant warmers. Emits blue light (400-550nm) to degrade bilirubin; lacks significant UV/IR emissions. Operated by trained medical staff. Provides high/low irradiance modes; includes overheat protection (power cutoff at ≥ 85°C). Benefits patient by reducing bilirubin levels without excessive warming or UV exposure. Requires protective eyeshades for infants.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including electrical, mechanical, safety, and reliability verification protocols. Performance testing conducted at unit and system levels.
Technological Characteristics
Phototherapy unit using 10 blue LEDs (400-550nm). Adjustable height/tilt plastic light enclosure. Power consumption: 20W. Lamp life: 50,000 hours. Overheat protection: power cutoff at ≥ 85°C. Standalone device on roll stand. No software-based algorithm described.
Indications for Use
Indicated for treatment of indirect hyperbilirubinemia in term and pre-term infants in hospital settings (NICU, PICU, Well-baby Nurseries). Administered by trained medical staff under licensed practitioner supervision. Not for use in mobile vehicles/ambulances.
Regulatory Classification
Identification
A neonatal phototherapy unit is a device used to treat or prevent hyperbilirubinemia (elevated serum bilirubin level). The device consists of one or more lamps that emit a specific spectral band of light, under which an infant is placed for therapy. This generic type of device may include supports for the patient and equipment and component parts.
{0}------------------------------------------------
K-120168
MAY - 2 2012
# Premarket Notification 510(k) Summary As required by section 807.92
# LullabyTM LED Phototherapy System
# GENERAL COMPANY INFORMATION as required by 807.92(a){1)
| Date:<br>Submitter: | 10th January 2012<br>Wipro GE Healthcare Private Ltd.<br>4, kadugodi industrial area<br>bangalore, INDIA 560067 |
|---------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Primary Contact Person: | Ms Agata Anthony<br>GE Healthcare,<br>Phone - +1410 456 0329<br>Fax - +1-410-888-0544 |
| Secondary Contact Person: | Sanjukta DebRoy<br>GE Healthcare,<br>Phone : +91 9632211022<br>Fax :+9180 28411645 |
| Device:<br>Trade Name: | Lullaby™ LED Phototherapy System |
| Common/Usual Name: | Phototherapy Device |
| Classification Names: | Neonatal Phototherapy Unit,<br>General Hospital, LBI, |
| Product Code: | |
| Regulation No: | |
| Predicate Device(s): | 21 CFR 880.5700 Unit, Neonatal Phototherapy<br>Lullaby™ Phototherapy System (K071828) |
| Device Description: | The Lullaby™ LED Phototherapy System is intended for the<br>treatment of neonatal hyperbilirubinemia, commonly known as<br>neonatal jaundice, in a hospital. The system can be used for<br>infants in a bassinet, incubator, open bed or radiant warmer. The<br>lamp unit emits blue light, which falls within the phototherapy<br>therapeutic spectrum range. The Lullaby™ LED Phototherapy<br>System consists of a lamp unit with 10 blue LED lamps mounted<br>on a roll stand.<br>The lamp unit consists of a lightweight plastic light enclosure. It<br>can be adjusted vertically and tilted if required. The light<br>enclosure can be tilted up to approximately 90° from the<br>horizontal about it pivot axis. The height of the pedestal can be<br>adjusted to change the vertical position of the lamp unit. The base<br>of the Lullaby™ LED Phototherapy System is designed to slide<br>conveniently under a bassinet, incubator, open bed, or radiant<br>warmer. The lamp unit is designed as a table top style to place<br>directly on the incubator |
4
{1}------------------------------------------------
The Lullaby™ LED Phototherapy System is used for the Indication for Use: treatment of indirect hyper-bilirubinemia in term and pre-term infants, in a hospital environment - NICUs, PICUs and Wellbaby Nurseries - administered by trained, professional medical staff, on the order of a licensed medical practitioner. The Lullaby "M LED Phototherapy System is intended for use under the direct supervision of a licensed healthcare practitioner. The Lullaby 110 LED Phototherapy System device is not intended to be operated in mobile vehicles including ambulances or other vehicles associated with health care facilities. Lullaby™ LED Phototherapy System is an intensive Technology: Phototherapy device with Blue LED light in the range of 400-550nm. This range corresponds to the spectral absorption of light by bilirubin and is thus considered to be the most effective for the degradation of bilirubin. Blue LEDs do not emit significant energy in the ultraviolet (UV) region of the spectrum, and so there is no unusual threat of exposure of the infant to UV radiation. In addition, blue LEDs do not emit significant energy in the infrared (IR) region of the spectrum, and so there is no unusual threat of exposure of the infant to IR radiation, or of excessive warming of the infant. As with all phototherapy lights, protective eyeshades must be used to protect the infant's eyes from excessive light exposure during treatment. The Lullaby™ LED Phototherapy System employs the same fundamental scientific technology as its predicate device Lullaby 110 Phototherapy System (K071828) by having the same mode of action for treating neonatal jaundice. But the source of light in Lullaby 110 LED Phototherapy System is from LED source as compared to CFL source in Lullaby TM Phototherapy System. Refer to the comparison table below for more details. Performance Factors LullabyTM Lullaby™ LED Phototherapy System Phototherapy System K071828 Proposed Predicate Light Intensity High Irradiance Mode: High irradiance mode: 30 µW.cm 2.nm 1 > 45 µW.cm -nm -Low Irradiance Mode: Low irradiance mode: 20 µW=cm " . nm " > 22 µW.cm 4.nm " 450-465 nm Light Spectrum 450-475 nm 60 cm X 30 cm 50 cm X 30 cm Surface Area Coverage Power Consumption 100W 20W (Low power consumption because of use of LED lights) Lamp Life 1000 hours 50000 hours Lamp Source CFL tubes LEDs Overheat protection Power cutoff for temp > Power cutoff for temperature ≥ 85° C 90° C
1
Confidential and Privileged. This document contains confidential and privileged trade secrets and other miermation of General Electrie Co. and as such may not be diselosed to others not employed by General Electric Co. All rights reserved
{2}------------------------------------------------
Determination of Substantial Equivalence:
い
Summary of Non-Clinical Tests:
Verification and Testing activities establish the performance, functionality, usability, safety, and reliability characteristics of Lullaby 11 LED Phototherapy System.
The Lullaby™ LED Phototherapy System comply with voluntary standards as detailed in Section 09, 17 and 18 of this premarket submission.
The following quality assurance measures were applied to the development of the system:
- Risk Analysis .
- . Requirements Reviews
- . Design Reviews
Summary of Simulated Use Setting:
The Design verification of Lullaby™ LED Phototherapy System has been divided into several protocols that include electrical, mechanical, safety Testing, reliability, and system design verification protocols.
The performance testing included testing on unit level, system level, as well as usability and safety parameters.
The results of the Design verification testing protocols have been documented in Section 18 of this 510(k) application.
The results demonstrate that the Lullaby TM LED Phototherapy System meets all design requirements and performance claims.
The subject of this premarket submission, LullabyTM LED Phototherapy System, did not require clinical studies to support substantial equivalence.
GE Healthcare considers the Lullaby TM LED Phototherapy Conclusion: Systems to be as safe and as effective as the predicate device, and the performance to be substantially equivalent to the predicate device.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features three wavy lines representing the branches of government. The caduceus is encircled by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" in a circular arrangement.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MAY - 2 2012
Wipro GE Healthcare Private Ltd. C/O Ms. Agata Anthony Regulatory Affairs Director Ohmeda Medical 8880 Gorman Road Laurel, Maryland 20723
Re: K120168
Trade/Device Name: Lullaby™ LED Phototherapy System Regulation Number: 21 CFR 880.5700 Regulation Name: Neonatal Phototherapy Unit Regulatory Class: II Product Code: LBI Dated: March 3, 2012 Received: March 7, 2012
Dear Ms. Anthony:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
## Page 2 - Ms. Anthony
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm 1 1 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Anthany D. awaton
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{5}------------------------------------------------
### Indications for Use
510(k) Number : K120168
20168
Device Name:
LullabyTM LED Phototherapy System
Indications for Use:
The Lullaby The LED Phototherapy System is used for the treatment of indirect hyperbilirubinemia in term and pre-term infants, in a hospital environment - NICUs, PICUs and Well-baby Nurseries - administered by trained, professional medical staff, on the order of a licensed medical practitioner.
The Lullaby 10 LED Phototherapy System is intended for use under the direct supervision of a licensed healthcare practitioner.
The Lullaby The LED Phototherapy System device is not intended to be operated in mobile vehicles including ambulances or other vehicles associated with health care facilities.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for RZe
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
8 Confidential and Privileged. This document contidential and privileged trade secrets and other milemation of General Electric Co, and as such may not be disclosed to others not employed by General Electric Co. All rights reserved
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.