← Product Code [LLZ](/productcode/LLZ) · K120161

# ALPHAPOINT IMAGING SOFTWARE (K120161)

_Radlogics, Inc. · LLZ · Apr 13, 2012 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K120161

## Device Facts

- **Applicant:** Radlogics, Inc.
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Apr 13, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

The AlphaPoint software is a device that allows review, analysis, and interchange of CT chest images. It is intended for use with CT Chest images to assist medical professionals in image analysis. It is not intended to be the primary interpretation. The software provides segmentation, Hounsfield numerical analysis, and substance indication. The user can review, verify and correct the results of the system and generate a report of the findings.

## Device Story

AlphaPoint Imaging Software is a DICOM-compliant application framework for CT chest image analysis. It retrieves imaging studies from PACS; performs automated segmentation and Hounsfield unit measurements to identify and quantify substances (e.g., air, lung, soft tissue, fat, water, blood, bone); and generates a Preliminary Findings Report. The system is operated by medical professionals in a clinical setting. Users review, verify, and correct the system-generated results before finalizing the report, which is then sent back to PACS. The device serves as an assistive tool for image analysis rather than a primary diagnostic interpretation, aiming to streamline workflow and provide quantitative data to support clinical decision-making.

## Clinical Evidence

No clinical data. Performance was established through bench testing, including verification of DICOM standard compliance and formal software validation. Validation included design reviews, code reviews, unit testing, and system testing traced to requirements. Testing was performed by independent engineers and documented in a Software Test Report (STR) confirming the software is ready for release.

## Technological Characteristics

Software-based imaging processing system. Written in C++, C#, and Matlab. Integrates with PACS via DICOM. Features include automated segmentation and Hounsfield numerical analysis. Operates as a standalone application framework.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Vitrea 2 software ([K060378](/device/K060378.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Kl2öi6i

# 510(k) Summary Prepared January 7, 2011

.

APR 1 3 2012

| Sponsor:                                                                          | RadLogics, Inc.                                                                  |
|-----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|
| Contact Person:                                                                   | Moshe Becker                                                                     |
| Telephone:                                                                        | 408 966 4874                                                                     |
| Fax:                                                                              | 408 966 4874                                                                     |
| Submission Date:                                                                  | December 6, 2011                                                                 |
| Device Name:                                                                      | AlphaPoint Imaging Software                                                      |
| Common Name:                                                                      | Imaging Software                                                                 |
| Classification:<br>Regulatory Class:<br>Review Category:<br>Classification Panel: | II<br>Class II<br>Radiology<br>System, Imaging Processing ; 21 CFR 892.2050; LLZ |

## A. Legally Marketed Predicate Devices

The modified software is substantially equivalent to the Vitrea 2 software manufactured by Vital Images, Inc. and cleared pursuant to K060378.

## B. Device Description:

The AlphaPoint system provides a full application framework with integration to PACS using DICOM. The system has the following functions:

- Communicates with PACS to get imaging studies for processing; .
- Activates one or more applications that process the imaging data . and use segmentation and Hounsfield measurements algorithms to find and measure various attributes in the images, and also identify particular slices as references images for the findings ;
- Formats the processing results for each study into a Preliminary . Findings Report
- Sends the results to PACS. .
- The software is written in C++, C# and Matlab. .

{1}------------------------------------------------

#### C. Intended Use

ﻨﺎ

The AlphaPoint software is a device that allows review, analysis, and interchange of CT chest images. It is intended for use with CT Chest images to assist medical professionals in image analysis. It is not intended to be the primary interpretation. The software provides segmentation, Hounsfield numerical analysis, and substance indication. The user can review, verify and correct the results of the system and generate a report of the findings.

## D. Substantial Equivalence

The submission device is substantially equivalent to the predicate software device with regard to both intended use and technological characteristics. Both devices retrieve and process chest images from Computed Tomography (CT), Both systems are DICOM compliant.

## E. Performance Data

The AlphaPoint software has been verified and validated according to the company's design control process. It was tested for compliance with the DICOM Standard and passed the six DICOM specific test cases provided in Section 4.4.2 of the Validation Test Report. The software development process complies with FDA Guidance documents related to software in Medical Devices and all of the documents specified have been submitted in the 510(k) Notification and can be summarized as follows

The software development life cycle includes various verification activities and the formal software validation. The software verification activities include design reviews, code reviews, unit tests and system testing.

The Software Test Description (STD) for the Alphapoint System describes the test cases for the device, along with its acceptance criteria, and the detailed test procedure. The test cases in the STD are traced to the requirements found in the SRS. The STD was developed using the template found in the IEEE Std 829-2008, IEEE Standard for Software Test Documentation and J-STD 016 (IEEE STD 1498), Software Development and Documentation.

The validation test runs were documented in the Software Test Report (STR) for the Alphapoint System. The STR contains the date, tester name, software versions, specific configurations used in the testing and the Pass/Fail grade for each test procedure step. The STR summarizes the validation assessment, which states that the software is ready for release. The STR was developed using the template found in the IEEE Std 829-2008, IEEE Standard for Software Test Documentation and J-STD 016 (IEEE STD 1498), Software Development and Documentation.

The validation was performed by a qualified independent software validation engineers who were not directly involved in the software design and implementation efforts. All anomalies have been resolved.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter of the circle.

# DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room - WO66-G609 Silver Spring, MD 20993-0002

RadLogics, Inc. Mr. Mark Job Accredited Third Party Reviewer % Regulatory Technology Services Incorporated 1394 25th Street, NW BUFFALO MN 55313

APR 1 3 2012

Re: K120161

Trade/Device Name: AlphaPoint Imaging Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: February 20, 2012 Received: February 21, 2012

#### Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into class II (Special Controls), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of

{3}------------------------------------------------

medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Parts 801 and 809), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely Yours,

Janine M. Morris

Acting Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

#### 510(k) Number (if known):

Device Name:

AlphaPoint Imaging Software

Indications for Use:

The AlphaPoint software is a device that allows review, analysis, and interchange of CT chest images. It is intended for use with CT Chest images to assist medical professionals in image analysis. It is not intended to be the primary interpretation. The software provides segmentation and Hounsfield numerical analysis values which are indicative of various substances (i.e., air, lung, soft tissue, fat, water, transudate, exudate, blood, muscle and bone). The user can review, verify and correct the results of the system and generate a report of the findings. .

Prescription Use · × (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

#### (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

(Division Sign-Off)
Division of Radiological Devices
Office of in Vitro Diagnostic Device Evaluation and Safety
510K
K120161

---

**Source:** [https://fda.innolitics.com/device/K120161](https://fda.innolitics.com/device/K120161)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
