MANTARAY CATHETER

K120129 · Bridgepoint Medical · DQY · Feb 3, 2012 · Cardiovascular

Device Facts

Record IDK120129
Device NameMANTARAY CATHETER
ApplicantBridgepoint Medical
Product CodeDQY · Cardiovascular
Decision DateFeb 3, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 870.1250
Device ClassClass 2

Indications for Use

The BridgePoint Medical Mantaray™ Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the peripheral vasculature.

Device Story

Mantaray™ Catheter is a single-use, over-the-wire, dual-lumen percutaneous catheter. It facilitates placement, support, and steering of a guidewire into peripheral vasculature. Device features a flexible shaft with a distal non-compliant balloon segment used for fluoroscopic orientation and two sideports with radiopaque markers. Sideports connect to the central guidewire lumen, allowing the guidewire to exit at an angle for steering. Used by physicians in clinical settings. Modification involves removing the balloon segment coating to eliminate the 10-minute pre-use soaking requirement. Output is physical guidewire positioning; healthcare providers use fluoroscopic visualization of radiopaque markers to navigate the vasculature. Benefits include improved guidewire control and access to discrete peripheral regions.

Clinical Evidence

Bench testing only. Evaluated tensile strength, burst, fatigue, inflation/deflation time, dimensions, hydration, guidewire interaction, flexibility, trackability, kink resistance, torque, and surface defects. No new clinical or animal studies were required as the modification (coating removal) did not alter the fundamental design or material safety profile established by the predicate.

Technological Characteristics

Single-use, over-the-wire, dual-lumen percutaneous catheter. Features a distal non-compliant balloon segment and radiopaque markers for fluoroscopic orientation. Materials are identical to the predicate device. No software or electronic components.

Indications for Use

Indicated for directing, steering, controlling, and supporting a guidewire to access discrete regions of the peripheral vasculature.

Regulatory Classification

Identification

A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## FEB - 3 2012 ## 9. 510(K) SUMMARY This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. | 510(k) Number: | K120129 | |----------------|---------| |----------------|---------| #### Applicant Information: | Date Prepared: | January 13, 2012 | |----------------|-------------------------------------------------------------------------------------------------| | Name: | BridgePoint Medical | | Address: | 13355 10th Ave. N., Ste. #110<br>Plymouth, MN 55441<br>Phone: 763-225-8500<br>Fax: 763-225-8718 | | Contact Person: | Jill Munsinger | |-----------------|-------------------------------------------| | Phone Number: | office: 763-225-8510 / cell: 651-270-0572 | | E-mail: | jmunsinger@bridgepointmedical.com | #### Device Information: | Classification: | Class II Percutaneous Catheter | |----------------------|--------------------------------| | Trade Name: | Mantaray™ Catheter | | Common Name: | Percutaneous Catheter | | Classification Name: | Percutaneous Catheter | #### Predicate Devices: The modified BridgePoint Medical Mantaray™ Catheter is substantially equivalent in intended use, method of operation and technical aspects to the following predicate device: K111963 - Mantaray™ Catheter #### Device Description: The modified Mantaray™ Catheter is a single use, over-the-wire, disposable, dual lumen percutaneous catheter that facilitates the placement, support and steering of a guidewire into discrete regions of the peripheral vasculature through the central guidewire lumen or through one of two sideports (identified by radiopaque markers). The sideports connect with the central guidewire lumen and facilitate guidewire steering (at an angle to the central lumen) by allowing the guidewire to exit the catheter. The catheter contains a small non-compliant balloon segment used for fluoroscopic orientation on the distal tip of the flexible shaft. {1}------------------------------------------------ #### Intended Use: The BridgePoint Medical Mantaray™ Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the peripheral vasculature. #### Comparison to Predicate Device(s): The modified Mantaray™ Catheters are substantially equivalent to the current Mantaray™ Catheters, K111963 in that they are both designed to direct, steer, control and support a guidewire in accessing discrete regions of the peripheral vasculature. The modified Mantaray™ Catheter is constructed of the same materials as the current Mantaray™ Catheter. Modifications were made to remove the coating on the balloon segment, which alleviates the need to soak the device for 10 minutes prior to use. The modified and current Mantaray™ Catheters are manufactured using similar processes and components and have similar physical attributes (balloon performance, trackability, tensile, radiopacity, and torque, etc.). Both devices include radiopaque markers located within the balloon segment to indicate the location of the guidewire lumen ports. #### Performance Data: The modified Mantaray™ Catheters have been evaluated using the following in vitro bench testing to confirm the performance characteristics as compared to the predicate device: - Tensile - . Burst - . Fatigue - . Inflation & Deflation Time - Dimensional . - . Hydration - . Guidewire Insert & Withdrawal - . Flexibility - . Trackability - . Guidewire Re-Direction - Markerband Movement & Removal ● - Markerband & Guidewire Interaction ◆ - Kink Resistance . - Coating . - . Torque - . Surface Defects - . Balloon Protector Removal, and In vivo testing was not deemed necessary based on the significance of the proposed modification to the baseline device. Removal of the coating from the distal 6 cm of the catheter results in a device that meets the original design requirements of the currently {2}------------------------------------------------ marketed Mantaray™ Catheter as demonstrated in the bench tests above. Animal studies were successfully completed with the currently marketed Mantaray™. Likewise, the removal of coating from the distal 6 cm of the device did not warrant additional biocompatibility testing as all materials are included in the currently marketed Mantaray™ Catheter. ### Summary: Based upon the intended use and descriptive information provided in this pre-market notification, the modified BridgePoint Mantaray™ Catheters have been shown to be substantially equivalent to the currently marketed predicate device. {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or other bird-like figure. The image is black and white. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 FEB - 3 2012 BridgePoint Medical, Inc. c/o Jill Munsinger 13355 10th Ave N Plymouth, MN 55441 Re: K120129 Trade/Device Name: Mantaray™ Catheter Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: DQY Dated: January 13, 2012 Received: January 17, 2012 Dear Ms. Munsinger: We have reviewed your Section 510(k) premarket notification of intent to market the device we nave to viewed your becamed the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the encreated of the enactment date of the Medical Device Amendments, or to conninered pror to May 2014 sified in accordance with the provisions of the Federal Food. Drug, devices mut have been roomsomed require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The r ou may, merelore, mains of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adultier. Trease noter CDFL wire, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it If your device is classified (boo as or a) Existing major regulations affecting your device can be may be subject to additions, Title 21, Parts 800 to 898. In addition, FDA may round in wrther announcements concerning your device in the Federal Register. {4}------------------------------------------------ ## Page 2 - Ms. Jill Munsinger Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, M.J. Hillebrand Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # 8. INDICATIONS FOR USE STATEMENT 510(k) Number: (TBA) BridgePoint Medical Mantaray™ Catheter Device Name: ### Indications For Use: The BridgePoint Medical Mantaray™ Catheter is indicated for directing, steering, controlling, and supporting a guidewire in order to access discrete regions of the peripheral vasculature. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page I of 1 M.A.Kildenne (Division Sign-Off) Division of Cardiovascular Devices 510(k) Number K120129 Page 19 of 82
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...