Transtek Blood Pressure Monitor, Models LS-802, TMB-1018-A, TMB-1018-A, TMB-1112 and TMB-1112-A are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 22 cm (about 9 - 17 inches). These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings. · The Blood Pressure Monitor compares average blood pressure tesults to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHg. Transtek Blood Pressure Monitor, LS-802, TMB-1018-A, TMB-1112 and TMB-1112-A models are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
Device Story
Digital blood pressure monitor; uses oscillometric method to measure systolic/diastolic pressure and heart rate. Input: pressure sensor detects cuff oscillations; microprocessor processes signals to calculate values. Output: LCD displays blood pressure, heart rate, and irregular heartbeat warning. Used in home or clinical settings by patients or clinicians. Device compares results to AHA 135/85 mmHg guideline. Benefits: provides non-invasive, automated monitoring of blood pressure and heart rate; alerts users to potential irregular heartbeats for physician consultation.
Clinical Evidence
Bench testing only. Device performance validated against oscillometric measurement standards; accuracy verified within ±3mmHg for pressure and ±5% for heart rate.
Technological Characteristics
Oscillometric measurement; electronic pressure sensor; microprocessor-based; ABS housing; nylon cuff; powered by 4x AAA/AA batteries or 6V DC adapter; digital LCD display; irregular heartbeat detection algorithm.
Indications for Use
Indicated for adult patients with arm circumference 22-42 cm to measure blood pressure and heartbeat rate. Detects irregular heartbeats. Not for diagnostic use.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
TRANSTEK Blood Pressure Monitor, Model TMB-986 (K101681)
Submission Summary (Full Text)
{1}------------------------------------------------
# ittk
# Section 5 - 510(k) Summary
Date of Summary Preparation: 12/21/2011
## 1. Submitter's Identifications
Submitter's Name: ZHONGSHAN TRANSTEK ELECTRONICS CO., LTD. Address: Jin'an Road, Minzhong, Zhongshan City, Guangdong, China
Contact Person: Lisa Li
Contact Email Address: lishal@transtek.cn
Telephone: 086-760-88282982 ext. 876
Fax: 086-760-85339231
## 2. Correspondent's Identifications
Correspondent's Name: A03 Lab of BTS
Address: No. I Fanghua Street, Hi-tech Zone, Chengdu City, Sichuan, China
Contact Person: Leo Wang
Contact Email Address: leo.w@hibts.com
Telephone: 086-028-86083300
Fax: 086-4006070050 ext. 8
## 3. Name of the Device
Device Classification Name: System, Measurement, Blood-Pressure, Non-invasive
Product Name: TRANSTEK Blood Pressure Monitor
Trade Name: TRANSTEK
Models: LS-802, TMB-1018, TMB-1018-A, TMB-1112, TMB-1112-A
Classification Panel: Cardiovascular
Common/Usual Name: Automatic Blood Pressure Monitor
Product Code: DXN
Device Classification: Class II
Contraindications: Do not use the Analyzer if you have a pacemaker or other internal medical device.
### 4. The Predicate Devices
TRANSTEK, Blood Pressure Monitor, Model TMB-986, K101681
Section 5 - 510(k) Summary
{2}------------------------------------------------
### 5. Device Description
Transtek Blood Pressure Monitor, LS-802, TMB-1018, TMB-1018-A, TMB-1112 and TMB-1112-A are designed to measure the systolic and diastolic blood pressure and heartbeat rate of an individual by using a non-invasive technique in which an inflatable cuff is wrapped around the arm.
Our method to define systolic and diastolic pressure is similar to the auscultatory method but uses an electronic pressure sensor rather than a stethoscope and mercury manometer. The sensor converts tiny alterations in cuff pressure to electrical signals, by analyzing those signals to define the systolic and diastolic blood pressure and calculating heartbeat rate, which is a well-known technique in the market called the "Oscillometric method".
Transtek Blood Pressure Monitor is single-mounted devices of the main unit and cuff unit. ABS is used to outer housing of the main unit. The preformed cuff unit, which is applicable to arm circumference approximately between 22 and 42 cm, includes the inflatable bladder and nylon shell. All models of the arm blood pressure monitor can use the same two size of cuff (22 - 32cm, or 22 -42cm). Product package will contains only one cuff and which size is decide by business requirement. The device consists of the microprocessor, the pressure sensor, the operation keys, the pump, the electromagnetic deflation control valve and the LCD. The subject devices are powered by four AAA or AA alkaline batteries or by a DC 6V 400mA adapter.
The device also compares the longest and the shortest time intervals of detected pulse waves to mean time interval and displays a warning signal with the reading to indicate the detection of irregular heartbeat when the difference of the time intervals is over 25%.
#### 6. Intended Use of Device
Transtek Blood Pressure Monitor, Models LS-802, TMB-1018-A, TMB-1018-A, TMB-1112 and TMB-1112-A are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 22 cm (about 9 - 17 inches).
These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings.
· The Blood Pressure Monitor compares average blood pressure tesults to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHg.
Transtek Blood Pressure Monitor, LS-802, TMB-1018-A, TMB-1112 and TMB-1112-A models are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
{3}------------------------------------------------
.
.
・・・
. .
.
.
:
··
# RANSTEK
# 7. Summary of Substantial Equivalence
| Feature | LS-802 | TMB-1018 | TMB-1018-A | TMB-1112 | TMB-1112-A | Comparison | | |
|------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|------------------------------------------------------------|------------|--------------|------------------|
| Power supply | Battery powered mode | | | | | Similar | | |
| | 6VDC 4*AA<br>alkaline<br>batteries | 6VDC 4*AAA<br>alkaline<br>batteries | 6VDC 4*AAA<br>alkaline<br>batteries | 6VDC 4*AA<br>alkaline<br>batteries | 6VDC 4*AA<br>alkaline<br>batteries | | | |
| Power supply | AC adaptor powered mode | | | | | Similar | | |
| | AC 100~240V<br>50-60 Hz<br>400mA | AC 100~240V<br>50-60 Hz<br>400mA | AC 100~240V<br>50-60 Hz<br>400mA | AC 100~240V<br>50-60 Hz<br>400mA | AC 100~240V<br>50-60 Hz<br>400mA | | | |
| Display mode | Digital LCD<br>V.A. 78*92mm | Digital LCD<br>V.A. 140*36mm | Digital LCD<br>V.A. 140*36mm | Digital LCD<br>V.A. 61*93mm | Digital LCD<br>V.A. 41*60mm | Similar | | |
| Measurement<br>mode | Oscillographic<br>testing mode | Oscillographic<br>testing mode | Oscillographic<br>testing mode | Oscillographic<br>testing mode | Oscillographic<br>testing mode | Similar | | |
| Accuracy | Pressure:<br>15°C~25°C:<br>within ±3mmHg | Pressure:<br>15°C~25°C:<br>within ±3mmHg | Pressure:<br>15°C~25°C:<br>within ±3mmHg | Pressure:<br>15°C~25°C:<br>within ±3mmHg | Pressure:<br>15°C~25°C:<br>within ±3mmHg | Similar | | |
| | 10°C~40°C(out<br>of 15°C-25°C): | 10°C~40°C(out<br>of 15°C-25°C): | 10°C~40°C(out<br>of 15°C-25°C): | 10°C~40°C(out<br>of 15°C-25°C): | 10°C~40°C(out<br>of 15°C-25°C): | | | |
| | within ±5mmHg<br>Heartbeat:<br>within ±5% of<br>reading | within ±5mmHg<br>Heartbeat:<br>within ±5% of<br>reading | within ±5mmHg<br>Heartbeat:<br>within ±5% of<br>reading | within ±5mmHg<br>Heartbeat:<br>within ±5% of<br>reading | within ±5mmHg<br>Heartbeat:<br>within ±5% of<br>reading | | | |
| Measurement<br>range | Pressure: 0~<br>300mmHg<br>Heartbeat:<br>40~199<br>Pul/min | Pressure: 0~<br>300mmHg<br>Heartbeat:<br>40~199<br>Pul/min | Pressure: 0~<br>300mmHg<br>Heartbeat:<br>40~199<br>Pul/min | Pressure: 0~<br>300mmHg<br>Heartbeat:<br>40~199<br>Pul/min | Pressure: 0~<br>300mmHg<br>Heartbeat:<br>40~199<br>Pul/min | Similar | | |
| | Measurement<br>perimeter of<br>upper arm | 22cm~42cm | 22cm~42cm | 22cm~42cm | 22cm~42cm | | 22cm~42cm | Similar |
| | Net weight | Approx. 385g | Approx. 300g | Approx. 270g | Approx. 305g | | Approx. 305g | Similar |
| Software | Ver. 1.0 | Ver. 1.0 | Ver. 1.0 | Ver. 1.0 | Ver. 1.0 | Similar | | |
| Feature | LS-802 | TMB-1018 | TMB-1018-A | TMB-1112 | TMB-1112-A | Comparison | | |
| Normal<br>working<br>condition | Temperature:<br>10℃~40℃ | Temperature:<br>10℃~40℃ | Temperature:<br>10℃~40℃ | Temperature:<br>10℃~40℃ | Temperature:<br>10℃~40℃ | Similar | | |
| | Relative<br>humidity:<br>15%~90% | Relative<br>humidity:<br>15%~90% | Relative<br>humidity:<br>15%~90% | Relative<br>humidity:<br>15%~90% | Relative<br>humidity:<br>15%~90% | | | |
| | Barometric<br>pressure:<br>105~80kPa | Barometric<br>pressure:<br>105~80kPa | Barometric<br>pressure:<br>105~80kPa | Barometric<br>pressure:<br>105~80kPa | Barometric<br>pressure:<br>105~80kPa | | | |
| Storage &<br>transportation<br>condition | Temperature:<br>-20~60℃ | Temperature:<br>-20~60℃ | Temperature:<br>-20~60℃ | Temperature:<br>-20~60℃ | Temperature:<br>-20~60℃ | Similar | | |
| | Relative<br>humidity:<br>≤90% | Relative<br>humidity:<br>≤90% | Relative<br>humidity:<br>≤90% | Relative<br>humidity:<br>≤90% | Relative<br>humidity:<br>≤90% | | | |
| | External<br>dimensions | 120*160*69mm | 180*100*40mm | 180*100*40mm | 120*140*70mm | | 120*140*70mm | Little different |
| Mode of<br>operation | Continuous<br>operation | Continuous<br>operation | Continuous<br>operation | Continuous<br>operation | Continuous<br>operation | Similar | | |
Table 1: The difference between subjects Transtek Blood Pressure Monitor :
Section 5 - 510(k) Summary
・
.
. .
・
Page 3 of 5 -
{4}------------------------------------------------
, '
.
:
:
+
. " :
# 网站: ਦੇ ਦੇ ਦਿ
.
Table 2: The difference between subject Transtek Blood Pressure Monitor and the predicate device,
Transtek Blood Pressure Monitor, TMB-986
| | Feature | LS-802, TMB-1018, TMB-1018-A,<br>TMB-1112, TMB-1112-A | Predicate: TMB-986 | Comparison |
|--|--------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------|------------|
| | Device name | Blood Pressure Monitor | Blood Pressure Monitor | Similar |
| | Indication for use | Measure the blood pressure and<br>heartbeat rate.<br>Irregular heartbeat detection.<br>These models are not intended to be a<br>diagnostic device. | Measure the blood pressure and<br>heartbeat rate.<br>Irregular heartbeat detection.<br>These models are not intended to be a<br>diagnostic device. | Similar |
| | Components | Main Unit, Cuff, Adapter, Instruction<br>Manual, 4*AAA/4*AA batteries,<br>Storage Case and Warranty Card | Main Unit, Cuff, Adapter, Instruction<br>Manual, 4*AAA batteries, Storage<br>Case and Warranty Card | Similar |
| | Energy used | Battery (4*AAA/4*AA) or AC<br>adapter (DC 6V Output) | Battery (4*AAA) or AC adapter (DC<br>6V Output) | Similar |
| | Display | Liquid crystal digital display | Liquid crystal digital display | Similar |
| | Software | Ver. 1.0 | Ver. 1.0 | Similar |
Section 5 -- 510(k) Summary
. . . . . .
・
・
.
: : : : :
:
.
、
{5}------------------------------------------------
| Feature | LS-802, TMB-1018, TMB-1018-A,<br>TMB-1112, TMB-1112-A | Predicate: TMB-986 | Comparison |
|---------------------------------------|----------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------|------------|
| External dimensions | LS-802: 120*160*69mm<br>TMB-1018 & TMB-1018-A:<br>180*100*40mm<br>TMB-1112 & TMB-1112-A:<br>120*140*70mm | 180*100*39mm | Similar |
| Applicable cuff | Wrap around cuff<br>(Model numbers, 22/32, 22/42) | Wrap around cuff<br>(Model numbers, 22/32, 22/42) | Similar |
| Accuracy of pressure<br>indicator | Within±3mmHg (15℃~25℃)<br>Within ±5mmHg (10℃~40℃ [out of<br>15℃~25℃]) | Within±3mmHg (15℃~25℃)<br>Within ±5mmHg (10℃~40℃ [out of<br>15℃~25℃]) | Similar |
| Accuracy of heartbeat<br>rate | Within ±5% of reading | Within ±5% of reading | Similar |
| Measurement range | Pressure: 0~300mmHg<br>Heartbeat: 40~199 Pul/min | Pressure: 0~300mmHg<br>Heartbeat: 40~199 Pul/min | Similar |
| Cult inflation | Automatic inflation with air pump | Automatic inflation with air pump | Similar |
| Dellation of pressure | Automatic air release | Automatic air release | Similar |
| Operating voltage | DC 6V | DC 6V | Similar |
| Measurement<br>perimeter of upper arm | 22cm~42cm | 22cm~42cm | Similar |
| Operating<br>environment | Temperature: 10℃~40℃<br>Relative humidity: 15%~90%<br>Barometric pressure: 105~80kPa | Temperature: 10℃~40℃<br>Relative humidity: 15%~90%<br>Barometric pressure: 105~80kPa | Similar |
| Transport and storage<br>environment | Temperature: -20~60℃<br>Relative humidity: ≤90% | Temperature: -20~60℃<br>Relative humidity: ≤90% | Similar |
# 8. Conclusions
The subject devices have all features of the predicate device TMB-986 except display and external dimensions. These differences do not affect the safety and effectiveness of the subject devices. Thus, the subject devices are substantially equivalent to the predicate device.
--- End of this section ---
Section 5 - 510(k) Summary
Page 5 ol 5
{6}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three curved lines representing its wings or feathers. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN 3 1 2012
Zhongshan Transtek Electronics Co., Ltd. c/o Mr. Leo Wang Consulting Manager A03 Lab of BTS No. 1 Fanghua Street Hi-Tech District Chengdu, CHINA 610041
K120058 Re:
K120036
Trade/Device Name: Transtek Blood Pressure Monitor Models: LS-802, TMB-1018, TMB-1018-A, TMB-1112, and TMB-1112-A
Regulatory Number: 21 CFR 870.1130 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: II (two) Product Code: 74 DXN Dated: January 7, 2012 Received: January 9, 2012
Dear Mr. Wang:
We have reviewed your Section 510(k) premarket notification of intent to market the device we nave reviewed your Section 910(x) premained is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in intersate for use stated in the encrosule) to regars mant date of the Medical Device Ameral Food Drug commerce prof to May 20, 1970, the enacified in accordance with the provisions of the Federation (DMA) devices that have been reclassified in accordance while as proval application (PMA).
and Cosmetic Act (Act) that do not require approval of a premarket approval organisms of and Cosment Act (Act) that do not require approval controls provisions of the Act. The Act. The Y ou may, therefore, market the devices, bayers to usements for annual registration, listing of the col general controls provisions of the Feet mends as a prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (Scc above) into existing major regulations affecting your device can be
{7}------------------------------------------------
Page 2 -- Mr. Leo Wang
found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean i lease be advised that I Drivedants over device complies with other requirements of the Act that I DA has made a acterimations administered by other Federal agencies. You must or any I edital statutes and regenments, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical CPN i art 607); adocing (21 CFR 803); good manufacturing practice requirements as set de rice-related daverse overse) (2 regulation (7) CFR Part 820); and if applicable, the electronic forth in the quality bytellio (Ques (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please 11 you desire specific acrisboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for go to intep://www.radiological Health's (CDRH's) Office of Compliance. Also, please the Center 101 Devices and note the regulation onlined, "Thisstanding of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the I ou may other built belief generational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours
Bram D Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Renstek
# Section 4 - Indications for Use
#### 510(k) Number (if known):
Device Name:
Transtek Blood Pressure Monitor Models: LS-802, TMB-1018, TMB-1018-A, TMB-1112, TMB-1112-A
Indications for Use:
This series of devices are digital monitors intended for use in measuring blood pressure and heartbeat rate in adult patient population with arm circumference ranging from 22 cm to 42 cm (about 9 - 17 inches).
These devices detect the appearance of irregular heartbeats during measurement and give a warning signal with readings. The Blood Pressure Monitor compares average blood pressure results to pre-established AHA (American Heart Association) hypertension guideline of 135/85 mmHg.
Transtek Blood Pressure Monitor, LS-802, TMB-1018, TMB-1018-A, TMB-1112, and TMB-1112-A models are not intended to be a diagnostic device. Contact your physician if hypertensive values are indicated.
Prescription Use
AND/OR
Over-The-Counter Use
(Part 21 CFR 801 Subpart D)
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) .. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . consider Devices 21200 S
Section 4 - Indications for Use
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.