The MindChild Medical Meridian Fetal Heart Rate Monitor is an intrapartum fetal monitor that externally or internally measures and displays fetal heart rate (FHR). The MindChild Meridian acquires and displays the FHR tracing from abdominal surface electrodes that detect the fetal ECG signal (fECG). The MindChild Meridian may also be used to measure and display fetal heart rate using direct ECG (DECG) with a fetal scalp electrode. The MindChild Meridian is indicated for use on women who are at term (> 36 completed weeks), in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen. The MindChild Meridian is intended for use by healthcare professionals in a clinical setting.
Device Story
The MindChild Meridian Monitor is an AC-powered fetal heart rate (FHR) monitor housed in a roll-up cart. It acquires fetal ECG (fECG) signals via external abdominal surface electrodes or direct ECG (DECG) via a fetal scalp electrode. The device processes these analog signals using internal hardware and software algorithms to extract FHR and ECG morphology. Data is displayed in real-time on a 24-inch touch screen LCD and can be stored locally or printed offline. Used in clinical settings by trained healthcare professionals, the device provides continuous FHR monitoring to assist clinicians in intrapartum assessment. The system benefits patients by providing non-invasive (abdominal) or invasive (scalp) monitoring options to track fetal status during labor.
Clinical Evidence
Clinical validation compared the Meridian algorithm against the HP M1350A Fetal Monitor/FSE. Fetal ECG signals were recorded simultaneously. The study evaluated FHR accuracy and successful recording percentage. Results showed Meridian FHR calculations were within a clinically acceptable range of the comparator (mean RMS error of 5 BPM). The device achieved a successful signal reading on at least 94% of signals recorded with the comparator. No statistically significant differences were observed across BMI groups.
Technological Characteristics
Medical grade roll-up cart; 24" LCD touch screen (1900x1200); 85-264 VAC, 50/60 Hz power. Sensing: external abdominal electrodes or fetal scalp electrode (DECG). Signal processing: analog-to-digital conversion with proprietary software algorithms. Connectivity: USB, electrode lead connectors. Complies with FDA-recognized electrical safety and EMC standards.
Indications for Use
Indicated for women at term (>36 weeks), in labor, with singleton pregnancies. Used for intrapartum fetal heart rate monitoring via abdominal surface electrodes or fetal scalp electrode (DECG).
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
K120018
## 510(k) Summary MindChild Medical Meridian Monitor
SEP 1 9 2012
## Submitter:
(
, - 1
) ( ﻣﯿﮟ ﺍﯾﮏ ﺭﮨﺘﮯ ﮨﯿﮟ۔
ﺗ
ﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﺨﺪﻣﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
ﺎ
7
MindChild Medical 1600 Osgood Street, #2-17 North Andover, MA 01845 Phone: 978.975.1160 · Fax: 978.688.8875
Contact Person:
Leo Basta NorthStar Biomedical Associates 755 Westminster St., Unit 120 Providence, RI, 02903 Phone: 617.834.9866 lbasta@northstarbiomedical.com
| Date Prepared: | August 18, 2012 |
|------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade Name: | MindChild Medical Meridian Monitor |
| Regulation Name: | Perinatal monitoring system and<br>accessories |
| Classification Number: | 21 CFR 884.2740 |
| Product Code: | HGM |
| Predicate Devices: | Monica Healthcare AN24 (K101801)<br>HP M1350A Fetal Monitor (K900480) |
| Device Description: | The MindChild Medical Meridian Fetal Heart Rate Monitor<br>is an AC powered fetal heart rate (FHR) monitor. The<br>product consists of a signal processor containing hardware<br>circuitry and software which acquires and processes analog<br>signals using external, abdominal electrode sensors.<br>Software algorithms extract ECG morphology and monitor<br>fetal heart rate. Data is stored locally for retrieval in an off-<br>line mode. The patient data is displayed real time on a<br>touch screen LCD display monitor. Data may be printed<br>out offline for permanent hard copy records. The system<br>also possesses the ability to acquire and analyze signals |
84
{1}------------------------------------------------
from a fetal scalp electrode using the device's direct ECG (DECG) capability. The entire system is housed within a medical grade roll-up cart. The device is intended to be used in a hospital environment by trained medical staff.
The MindChild Medical Meridian Fetal Heart Rate Monitor is an intrapartum fetal monitor that externally or internally measures and displays fetal heart rate (FHR). The MindChild Meridian acquires and displays the FHR tracing from abdominal surface electrodes that detect the fetal ECG signal (fECG). The MindChild Meridian may also be used to measure and display fetal heart rate using direct ECG (DECG) with a fetal scalp electrode. The MindChild Meridian is indicated for use on women who are at term (> 36 completed weeks), in labor, with singleton pregnancies, using surface electrodes on the maternal abdomen. The MindChild Meridian is intended for use by healthcare professionals in a clinical setting.
Descriptive information, laboratory bench testing, electrical safety/EMC screening, software validation, and a biocompatibility assessment were provided to demonstrate the device meets its design specifications, performs as intended, and is safe for its intended use.
In addition, a clinical validation of the Meridian algorithm responsible for calculating FHR from fECG signals was performed. For this validation of the algorithm, fetal ECG signals were acquired and recorded using the Meridian Monitor technology and HP M1350A Fetal Monitor/FSE simultaneously. The HP M1250A Fetal Monitor/FSE was the comparator for the clinical validation study. The Meridian Monitor signals were subjected to analysis by the Meridian algorithm to calculate fetal heart rate. The resulting heart rates were compared to heart rates calculated using HP M1350A Fetal Monitor/FSE. The analyses performed included FHR accuracy and percent successful recordings (i.e., # comparator signals/ # of Meridian signals x 100). This algorithm validation demonstrated that in either mode (abdominal electrode or DECG), the fetal heart rates calculated using the Meridian Monitor algorithm were close to the comparator's fetal heart rate calculations within a clinically acceptable number of beats per minute (BPM). There were no
## Intended Use:
### Functional Testing:
{2}------------------------------------------------
statistically significant differences across different BMI groups analyzed. In addition, for all signals recorded with the comparator FSE device, the Meridian technology provided a successful signal reading on at least 94% of the signals.
Lastly, a single Bland Altman (BA) difference plot was generated for the Meridian Monitor FHR calculations versus those calculated using the Gold Standard device for each patient and a root mean square (RMS) error was determined. A BA difference plot is a scatter plot of the difference between the device and Gold Standard measurements versus the Gold Standard measurement. RMS error is the square root of the mean of the squared differences. The mean RMS error of 5 BPM (Beats Per Minute) for the Meridian algorithm is similar to results observed in other FDA cleared fetal heart rate monitors.
જર્
{3}------------------------------------------------
| Non-Resident |
|--------------------------------------------------------------------------------------|
| |
| |
| and the may been and the been and the been and the been and the best of the been and |
| 10-40-40 10-40, 40- -- 4-4- 4-4 |
| |
| 1 |
| |
| 4.5 |
| |
| . FOR AND AND AND A |
| 1 |
| |
| |
| |
| |
| |
| 1 |
,
| Characteristic | Proposed MindChild Meridian Device - K120018 | Monica AN24, Monica Healthcare - K101801 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for<br>Use | The MindChild Medical Meridian Fetal Heart Rate Monitor is an intrapartum<br>fetal monitor that externally or internally measures and displays fetal heart rate<br>(FHR). The MindChild Meridian acquires and displays the FHR tracing from<br>abdominal surface electrodes that detect the fetal ECG signal (fECG). The<br>MindChild Meridian may also be used to measure and display fetal heart rate<br>using direct ECG (DECG) with a fetal scalp electrode. The MindChild<br>Meridian is indicated for use on women who are at term (> 36 completed<br>weeks), in labor, with singleton pregnancies, using surface electrodes on the<br>maternal abdomen. The MindChild Meridian is intended for use by healthcare<br>professionals in a clinical setting. | The Monica AN24 is an intrapartum maternal-fetal monitor that non-invasively<br>measures and displays fetal heart rate (FHR) and uterine activity (UA). The<br>AN24 acquires and displays the FHR tracing from abdominal surface electrodes<br>that pick up the fetal EGG (fECG) signal. Using the same surface electrodes, the<br>AN24 also acquires and displays the UA tracing from the uterine<br>electromyography (EMG) signal. The AN24 is indicated for use on women who<br>are at term (>36 completed weeks), in labor, with singleton pregnancies, using<br>surface electrodes on the maternal abdomen. The AN24 maternal-fetal monitor is<br>intended for use by healthcare professionals in a clinical setting. |
| Method | Acquisition of fetal ECG signals via external abdominal sensor electrodes.<br>From these signals the fetal heart rate is extracted. Also may use direct ECG<br>(DECG) acquisition using a fetal scalp electrode. | The electrical signals are passively<br>monitored on three channels using external sensor electrodes placed on the<br>pregnant abdomen. in specific locations. From these electrical signals the Fetal<br>Heart Rate (FHR) are continuously extracted and displayed. No DECG option. |
| Software | Proprietary algorithm to calculate heart rate. | Proprietary algorithm to calculate heart rate. |
| Signal<br>Processing | Analog to digital | Analog to digital |
| Display | 24" diagonal real time LCD display (1900 x 1200 pixels) that graphically<br>displays clinical parameters received from signal processors. Touch screen<br>provides user with interface to control operation of unit. | Ability to download data to PC or view via optional 15" cart mounted monitor or<br>10" mobile touch screen. |
| Controls | Power "On" button on control unit activates main operating screen on LCD<br>display. On-screen display buttons to adjust various parameters. | On/Off switch, "Event" button, battery status, signal quality, memory status. |
| Connections | Five connectors for electrode leads and optional fetal scalp electrode, standard<br>USB connector,<br>standard AC plug and power connection | Electrode leads connector attaches to the top of the unit. |
| Input Voltage<br>Range | 85 - 264 VAC | AN24 and optional mobile 10" viewing monitor (VS10) - Battery operated |
| Frequency | 50/60 Hz | Battery operated |
| Measurement<br>Range | 80 - 240 BPM; abdominal sensor mode, 30 - 50 BPM; DECG Mode | UNK |
| Safety | Complies with FDA recognized standards pertaining to electrical safety.<br>Complies with FDA recognized standards pertaining to electrical safety and<br>EMC. | Complies with FDA recognized standards pertaining to electrical safety.<br>Complies with FDA recognized standards pertaining to electrical safety and<br>EMC. |
MindChild 510(k)
Rev. B
87
{4}------------------------------------------------
## Summary of Substantial Equivalence:
The design, intended use, principles of operation, and technological characteristics of the MindChild Meridian device are substantially equivalent to those of the predicate devices cited above. Substantial equivalence is based upon descriptive characteristics of the various cited predicate devices and upon the testing conducted to demonstrate that the subject device performs as intended and is substantially equivalent to the predicate devices in terms of its ability to monitor fetal heart rate using its abdominally placed external electrode sensors. In addition, fetal heart rate monitoring through the device's DECG mode was demonstrated to be substantially equivalent to that of marketed devices.
MindChild 510(k) Rev. B
88
{5}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image is a circular logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing segments, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circle's perimeter. The logo is black and white.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
MindChild Medical % Mr. Leo Basta Owner NorthStar Biomedical Associates 755 WestMinster Street, Unit 120 PROVIDENCE RI 02903
SEP 1. 9 2012
Re: K120018
Trade/Device Name: MindChild Meridian Fetal Heart Rate Monitor Regulation Number: 21 CFR§ 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: September 14, 2012 Received: September 17, 2012
Dear Mr. Basta:
We have reviewed your Section 510(k) premarket notification of intent to market the device indication We nave reviewed your Scellon >10(t) premierity is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate for use stated in the enclosule) to regarly manated producal Device Amendments, or to
commerce prior to May 28, 1976, the enactment date of the Medical Frederal Frederal Fred commerce prior to May 20, 1976, the excordance with the provisions of the Federal Food, Drug, devices that have been recalismed in accession in a premarket approval application (PMA).
and Cosmetic Act (Act) that do not require approval of a presisions of the Act . The and Costhere Act (Act) that do not require apple the general controls provisions of the Act. The Act. The Act. The Act. The You may, dielefone, market the device, basjever to the may begistration, listing of general condois provisions of the rice labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability adulteration. Flease note. ODIC access that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), If your device is classified (see above) and or regulations affecting your device can be
it may be subject to additional controls. Existing major regulations affecting your it may be subject to adultional controls. Entraling migrid to 898. In addition, FDA may
round in the Code of I cacial resgueening your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean Please be advised that FDA s issualled of a buce complies with other requirements of the Act
that FDA has made a determination that your device complies with other requiremen that FDA has made a decemination and regulations administered by other Federal and listing of any Federal Statutes and regulations annline, but not limited to: registration and listing of media comply with an the Act sTequilements, merading, outsist and the more of medical
{6}------------------------------------------------
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set device-leheu adverse events) (21 CFR 003); good many (21 CFR Part 820); and if applicable, the electronic forth in the quality Systems (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please IT you desire specific advice for your as nte coffices/CDRH/CDRHOffices/ucm115809.htm for go to mip.//www.ida.gov/rtoodid Drives/Health's (CDRH's) Office of Compliance. Also, please the Concer for Dorroos and reading by reference to premarket notification" (21CFR Part note the regulation emilied, "Misoranding of releases to presents under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the You may oodain outer general mormational and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address (000) 030-2011 01 (00:00)Devices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Benjamin R. Tucker
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## INDICATION FOR USE
K120018
Not Assigned
## 510(k) Number (if known):
Device Name:
MindChild Meridian Fetal Heart Rate Monitor
Indications for Use:
The MindChild Medical Meridian Fetal Heart Rate Monitor is an intrapartum fetal monitor than The MindChild Wedical Meridian Peter Fetal heart rate (FHR). The MindChild Meridian externally of medially measures and displays from abdominal surface electrodes that detal acquires and displays the Prix training nominal may also be used to measure and display fetal heart rate using direct ECG (DECG) with a fetal scalp electrode. The MindChild Meridian is indicated for use on women who are at term (> 36 completed weeks), in labor, with singleton in indicated for use on women who are at term ( - 30 - 8 - 6 - 6 - 6 - 6 - 1 - MindChild Meridian is intended for use by healthcare professionals in a clinical setting.
AND/OR X Prescription Use: _ (Per 21 CFR 801 Subpart D)....................................................................................................................................................
Over-The Counter Use: (Per 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of 1
Aogui M. Whim
(Division Sign-Off)
Division of Reproductive, Gastro-Renal, and
Urological Devices
510(k) Number K120018
MindChild 510(k) Rev. A
Confidential
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.