BORRELIA BURGDORFERI IGM BLOT TEST

K113846 · Gold Standard Diagnostics · LSR · Jun 1, 2012 · Microbiology

Device Facts

Record IDK113846
Device NameBORRELIA BURGDORFERI IGM BLOT TEST
ApplicantGold Standard Diagnostics
Product CodeLSR · Microbiology
Decision DateJun 1, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.3830
Device ClassClass 2
AttributesReal-World Evidence

Real-World Evidence

SubmissionDeviceSponsorRWD SourcesRWE Use SummaryKey Tags
K113846 · Jun 1, 2012BORRELIA BURGDORFERI IGM BLOT TESTGold Standard DiagnosticsClinically characterized patient serum samples from Massachusetts General HospitalRetrospective clinical samples were used to evaluate the sensitivity of the device across early, disseminated, and late stages of Lyme disease.Sensitivity study; Retrospective clinical samples; Lyme disease

Clinical Evidence

Study DesignPopulationComparatorKey Endpoints
Retrospective clinical evaluationPatients with early, disseminated, and late stages of Lyme disease; Sample Size: 100; Number of Sites: 1Commercially available deviceSensitivity

Indications for Use

The Gold Standard Diagnostics Borrelia burgdorferi B31 IgM Line Blot Test Kit is intended for the qualitative detection of IgM antibodies to B. burgdorferi sensu stricto (B31) in human serum. This test is intended for use in testing human serum samples which have been found positive or equivocal using an ELISA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi.

Device Story

The device is an in vitro diagnostic line blot assay for detecting IgM antibodies to B. burgdorferi. It utilizes nitrocellulose strips coated with purified B. burgdorferi antigens. Human serum samples are incubated on the strips; if specific IgM antibodies are present, they bind to the antigens. After washing, an enzyme-conjugated anti-human IgM is added, followed by a substrate (BCIP/NBT) that forms blue-violet precipitates at antigen-antibody binding sites. The resulting band pattern is visually interpreted by laboratory personnel to determine the presence or absence of specific IgM antibodies. The test is performed in a clinical laboratory setting as a secondary, supportive diagnostic step following initial screening by ELISA or IFA. The output provides qualitative evidence of infection, aiding clinicians in the diagnosis of Lyme disease.

Clinical Evidence

Prospective clinical study (n=310) compared subject device to predicate; 99.4% positive and negative percent agreement. Sensitivity evaluated on 100 clinically characterized samples (early, disseminated, late stages); subject device showed higher sensitivity than predicate across stages (e.g., 87.5% vs 77.5% in early stage). CDC reference panel testing showed high agreement with predicate. Analytical specificity 100% in 234 asymptomatic samples. Cross-reactivity tested against 215 specimens; minor cross-reactivity observed in Babesiosis, Ehrlichiosis, and Rickettsial disease samples, consistent with predicate performance.

Technological Characteristics

Line blot assay using purified or cloned B. burgdorferi proteins (23, 39, 41 kDa) sprayed onto nitrocellulose membranes. Detection via anti-human IgM antibody-enzyme conjugate and substrate-induced blue-violet precipitate. Manual interpretation of band patterns. No specialized instrumentation required.

Indications for Use

Indicated for qualitative detection of IgM antibodies to B. burgdorferi sensu stricto (B31) in human serum samples previously testing positive or equivocal by ELISA or IFA, to provide supportive evidence of Lyme disease infection.

Regulatory Classification

Identification

Treponema pallidum treponemal test reagents are devices that consist of the antigens, antisera and all control reagents (standardized reagents with which test results are compared) which are derived from treponemal sources and that are used in the fluorescent treponemal antibody absorption test (FTA-ABS), the Treponema pallidum immobilization test (T.P.I.), and other treponemal tests used to identify antibodies to Treponema pallidum directly from infecting treponemal organisms in serum. The identification aids in the diagnosis of syphilis caused by bacteria belonging to the genus Treponema and provides epidemiological information on syphilis.

Predicate Devices

Submission Summary (Full Text)

{1} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: K113846 B. Purpose for Submission: New device C. Measurand: IgM antibodies to *Borrelia burgdorferi* proteins D. Type of Test: Western blot immunoassay E. Applicant: Gold Standard Diagnostics F. Proprietary and Established Names: *Borrelia burgdorferi* B31 IgM Line Blot Test Kit G. Regulatory Information: 1. Regulation section: 21 CFR 866.3830, Treponema pallidum treponemal test reagents 2. Classification: Class: II 3. Product code: LSR; Reagent, Borrelia Serological Reagent 4. Panel: 83 Microbiology {2} H. Intended Use: 1. Intended use(s): The Gold Standard Diagnostics Borrelia burgdorferi B31 IgM Line Blot Test Kit is intended for the qualitative detection of IgM antibodies to B. burgdorferi sensu stricto (B31) in human serum. This test is intended for use in testing human serum samples which have been found positive or equivocal using an ELISA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi. 2. Indication(s) for use: The Gold Standard Diagnostics Borrelia burgdorferi B31 IgM Line Blot Test Kit is intended for the qualitative detection of IgM antibodies to B. burgdorferi sensu stricto (B31) in human serum. This test is intended for use in testing human serum samples which have been found positive or equivocal using an ELISA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi. 3. Special conditions for use statement(s): For prescription use 4. Special instrument requirements: None I. Device Description: The Gold Standard Diagnostics Borrelia burgdorferi B31 IgM test is a line blot assay. The 3 antigenic proteins specific for B. burgdorferi (sensu stricto) are either purified or cloned and expressed in the host E.coli. The purified individual proteins are transferred to a nitrocellulose membrane using a spraying micro-dispensing method. Positions of the lines are defined on the filter and the antigen bands are assigned in the following order: 23, 39, and 41 kDa. During the test procedure, antibodies to Borrelia burgdorferi B31 (sensu stricto) present in the human serum sample will bind to the antigens coated onto the nitrocellulose strips. After removing serum and unbound antibodies by washing, the nitrocellulose strip is incubated with an antihuman IgM antibody-enzyme conjugate. After the unbound conjugate has been removed by a washing step, visualization of the antigen-antibody-antibody complex is accomplished by the addition of a substrate which forms a blue-violet precipitate at each site (antigen bands). The reaction is stopped by washing the nitrocellulose strip with distilled or deionized water. Depending on the observed band pattern one can interpret the presence or absence of specific IgM antibodies to B. burgdorferi infection. J. Substantial Equivalence Information: 1. Predicate device name(s): B. burgdorferi (IgM) Marblot Strip Test System from Trinity Biotech {3} 2. Predicate 510(k) number(s): K951709 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | Intended use | The Gold Standard Diagnostics Borrelia burgdorferi B31 IgM Line Blot Test Kit is intended for the qualitative detection of IgM antibodies to B. burgdorferi sensu stricto (B31) in human serum. This test is intended for use in testing human serum samples which have been found positive or equivocal using an ELISA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi. | MarDx B. burgdorferi (IgM) Marblot Strip Test System is a Western blot assay for the qualitative in vitro detection of human IgM antibody to individual proteins of B. burgdorferi in human serum. The MarDx B. burgdorferi (IgM) Marblot Strip Test System is intended for use in testing human samples which have been found positive or equivocal using an EIA or IFA test procedure to provide supportive evidence of infection with B. burgdorferi. | | Specimen type | Serum | Serum | | Method | Qualitative | Qualitative | | Assay | Immunoblot | Immunoblot | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Antigens | Individually purified or cloned proteins of B. burgdorferi | Whole cell extract of B. burgdorferi antigens | | Assay | Line blot | Western blot | {4} K. Standard/Guidance Document Referenced (if applicable): Not applicable (N/A) L. Test Principle: Enzyme-immunoassay on a blot M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: The reproducibility of the assay was done by testing a negative sample, a high negative sample, a low positive sample, and a moderate positive sample in triplicate for five days, twice a day, at three sites with two technicians per site giving a total of 90 data points per sample. Results of band reproducibility and sample reproducibility are shown below: Band Reproducibility: | Sample/kDa | 41 | 39 | 23 | Number of Bands | | --- | --- | --- | --- | --- | | Negative | 0 | 0 | 0 | 0 significant bands | | High Negative | 0 | 0 | 79 | ≤1 significant bands | | Low Positive | 86 | 0 | 90 | 2 significant bands | | Moderate Positive | 90 | 90 | 90 | 3 significant bands | Sample Reproducibility: | | Band Reproducibility | Final Interpretation | | | --- | --- | --- | --- | | | | Positive | Negative | | Sample | | | | | Negative | 100% (0/0) | | 100% (90/90) | | High Negative | 87.8% (79/90) | | 100% (90/90) | | Low Positive | 97.8% (176/180) | 95.6% (86/90) | | | Moderate Positive | 100% (270/270) | 100% (90/90) | | *A low positive sample is expected to yield a positivity of 95%. {5} b. Linearity/assay reportable range: N/A c. Traceability, Stability, Expected values (controls, calibrators, or methods): N/A d. Detection limit: N/A e. Analytical specificity: Analytical Specificity Study: For the determination of analytical specificity for the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot Test, testing of 234 asymptomatic samples (blood donors) from both endemic and non-endemic regions was performed. The results are summarized in the following table: | Region | Number of Samples | Number Positive | Analytical Specificity | | --- | --- | --- | --- | | Endemic | 115 | 0 | 100% | | Non-endemic | 119 | 0 | 100% | Cross Reactivity: A cross reactivity study was performed on 215 specimens known to contain potentially cross reactive antibodies to B.burgdorferi. Sera from patients with infections and sera from patients with diagnoses that can be confused with the late manifestations of Lyme disease were tested. The results are summarized in the following table: | Infection / Diagnosis | Number of Sera Tested | # Positive / (%) | | --- | --- | --- | | Tick-borne Relapsing Fever / Rickettsial Diseases | 23 | 1 / (4%) | | Treponemal Infections | 12 | 0 / (0%) | | Ehrlichiosis | 20 | 1 / (5%) | | Babesiosis | 20 | 2 / (10%) | | Leptospirosis | 1 | 0 / (0%) | | Parvovirus B19 | 9 | 0 / (0%) | | Epstein-Barr Virus | 11 | 0 / (0%) | {6} 6 | Cytomegalovirus | 32 | 0 / (0%) | | --- | --- | --- | | H. pylori | 12 | 0 / (0%) | | Fibromyalgia | 10 | 0 / (0%) | | Rheumatoid Arthritis | 12 | 0 / (0%) | | Herpes Simplex Virus | 16 | 0 / (0%) | | Varicella Zoster Virus | 12 | 0 / (0%) | | Autoimmune Disease | 25 | 0 / (0%) | Two of the 20 Babesiosis samples and one each of the 23 Tick-borne relapsing fever / Rickettsial Disease and 20 Ehrlichiosis specimens were positive on the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot Test. These samples when tested on the predicate device also gave positive results. Interfering Substances: The effect of potential interfering substances on samples using the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot Test was evaluated. High levels of hemoglobin, bilirubin, cholesterol and intralipids in serum samples were tested on six sera (two positives, one low positive, one high negative, and two negatives). The recommended concentrations in the guideline "Interference Testing in Clinical Chemistry" from the Clinical and Laboratory Standards Institute were used. The interferents at the concentrations tested did not have any influence on the band pattern. A small variability in band intensity was seen that is in the normal range of deviation and did not change the final interpretation. The results are summarized in the following table: | Substance | Concentration | Interference | | --- | --- | --- | | Hemoglobin | 2 g/l | None detected | | Bilirubin | 342 μmol/l | None detected | | Cholesterol | 13 mmol/l | None detected | | Intralipids | 37 mmol/l | None detected | f. Assay cut-off: Determination of Cutoff: The cutoff was determined by testing a total of 365 sera of known anti-B. burgdorferi IgM positives and negatives. The cutoff concentration was determined by testing several dilutions of each antigen and choosing the concentration that met the sample criteria best. 2. Comparison studies: a. Method comparison with predicate device: {7} Method Comparison Study: The performance of the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot test was determined by conducting a prospective clinical study at three different geographic sites (Pennsylvania, North Carolina, and California) in the U.S. The patient samples were tested by an anti-B. burgdorferi total antibody ELISA test first and the resulting equivocal and positive specimens were tested on the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot test and a commercially available B. burgdorferi IgM Blot test. The results are summarized in the following table: | | Predicate Device IgM Blot | | | | --- | --- | --- | --- | | | | Positive | Negative | | Gold Standard Diagnostics IgM Line Blot | Positive | 154 | 1 | | | Negative | 1 | 154 | Positive Percent Agreement = 99.4% (154/155) [95% CI: 96.5-100%] Negative Percent Agreement = 99.4% (154/155) [95% CI: 96.5-100%] The discrepant samples were tested on a second commercially available assay. On the one Line Blot negative sample that was positive by the predicate, the second assay gave a positive result. Of the one Line Blot positive and predicate negative samples, the second assay called the sample positive. The 310 samples were also tested by the Gold Standard B. burgdorferi IgG Line Blot test and 36.8% (114/310) were found to be positive and 63.2% (196/310) were found to be negative. Sensitivity Study: A sensitivity study for the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot Test was performed on 100 clinically characterized samples. The samples encompassed early, disseminated, and late stages of Lyme disease. The results are summarized in the following table: | Stage | Number of Samples | Line Blot Sensitivity with 95% CI | Commercially Available Device Sensitivity with 95% CI | | --- | --- | --- | --- | | Early | 40 | 87.5% (35/40) [73.2%-95.8%] | 77.5% (31/40) [61.6%-89.2%] | | Disseminated | 20 | 95.0% (19/20) [75.1%-99.9%] | 85% (17/20) [62.1%-96.8%] | | Late | 40 | 27.5% (11/40) [14.6%-43.9%] | 22.5% (9/40) [10.8%-38.5%] | CDC Reference Panel: A standard panel of positive and negative specimens provided by the Centers of Disease Control (CDC) for Lyme disease detection {8} was tested both on the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot Test and on the predicate device. The results are summarized in the following table: | Stage | Total | GSD Line Blot % Agreement | Commercially Available Device % Agreement | | --- | --- | --- | --- | | Healthy | 5 | 100% (5/5) | 100% (5/5) | | Early (0-2 months) | 15 | 86.7% (13/15) | 73.3% (11/15) | | Intermediate (3-12 months) | 13 | 92.3% (12/13) | 84.6% (11/13) | | Late (>1 year) | 7 | 57.1% (4/7) | 57.1% (4/7) | b. Matrix comparison: N/A 3. Clinical studies: a. Clinical Sensitivity: N/A b. Clinical specificity: N/A c. Other clinical supportive data (when a. and b. are not applicable): N/A 4. Clinical cut-off: N/A 5. Expected values/Reference range: The immune response to B. burgdorferi appears to follow a normal response pattern. IgM antibodies can be detected in some patients within days after onset, while IgG antibodies can be detected as early as two weeks after onset. IgM antibodies often decrease some weeks to months after convalescence, but it is also possible that IgM antibody titers remain constant up to some years. Band patterns will differ from sample to sample due to differences in patient immune responses and the stage to which the disease has progressed. The following table summarizes the frequency of bands observed with the Gold Standard Diagnostics B. burgdorferi B31 IgM Line Blot Test. | Stage / kDa | 41 | 39 | 23 | | --- | --- | --- | --- | | Early Lyme (n=55) | 81.8% | 25.5% | 89.0% | | Intermediate (3-12 months) | 13 | 92.3% | 84.6% | | Late (>1 year) | 7 | 57.1% | 57.1% | {9} | Disseminated (n=33) | 70.0% | 39.0% | 72.8% | | --- | --- | --- | --- | | Late Lyme (n=47) | 45.0% | 9.0% | 21.0% | | Prospective Samples (Not-staged) (n=310) | 53.5% | 24.5% | 60.6% | | Normal (n=115) (Non-Endemic) | 3.0% | 1.0% | 7.0% | | Normal (n=119) (Endemic) | 1.0% | 2.0% | 4.0% | # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...