PRIMUS HEARING INSTRUMENT TEST UNIT

K113831 · Auditdata A/S · ETW · Apr 19, 2012 · Ear, Nose, Throat

Device Facts

Record IDK113831
Device NamePRIMUS HEARING INSTRUMENT TEST UNIT
ApplicantAuditdata A/S
Product CodeETW · Ear, Nose, Throat
Decision DateApr 19, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 874.3310
Device ClassClass 2
AttributesPediatric

Intended Use

The Primus Hearing Instrument Test Unit is intended for use by professionals such as an audiologist, hearing healthcare specialist, or trained technician. Audiometric testing should take place in an extremely quiet environment and care should be taken to ensure optimal test conditions and safety of the client during testing. The Primus Hearing Instrument Test Unit is intended to give an objective indication of the characteristics of a Hearing Aid, by visualizing a signal recorded in the test coupler with reference information such as target curves in order to make adjustments of the Hearing Instrument settings. The Primus Hearing Instrument Test Unit is indicated for technical quality inspection of hearing instruments with no clients involved. The Primus Fitting Unit is intended for use by professionals such as an audiologist, hearing healthcare specialist, or trained technician. Audiometric testing should take place in an extremely quiet environment and care should be taken to ensure optimal test conditions and safety of the client during testing. The Primus Fitting Unit is intended for hearing test. The Primus Fitting Unit with stated accessories is indicated for non-continuous, noninvasive air and optionally bone conduction and speech audiometric testing in quiet office and sound treated environments. The Primus Fitting Unit is indicated for non-continuous real-ear measurements at the ear drum by means of noninvasive external ear canal insertion of a probe tube in quiet office environments. Finally the system can be used to present hearing instrument related sound examples through headsets or loudspeakers. The Primus Fitting Unit is indicated for use with both pediatric and adult age groups. The Primus Fitting Unit is not indicated as a sole means of diagnostics.

Device Story

PC-operated system comprising Primus Hearing Instrument Test Unit (PHITU) and Primus Fitting Unit (PFU). PHITU uses microphones and couplers to record hearing aid signals for technical quality inspection; visualizes signal against target curves to guide adjustments. PFU performs pure tone, bone conduction, and speech audiometry; conducts real-ear measurements (REM) via probe tubes in ear canals; presents sound examples via headsets/loudspeakers. Operated by audiologists/technicians in quiet office environments. Software runs on Windows OS, integrates with NOAH framework or operates standalone. Data files store historical session info. Benefits include objective hearing aid verification and standardized audiometric testing to support clinical fitting decisions.

Clinical Evidence

Bench testing only. No clinical data. Performance verified through compliance with standards: EN/IEC 60601-1, EN/IEC 60601-1-2, ANSI S3.22, IEC 60118-7, EN60645-1/ANSI S3.6 (Tone Audiometry), EN60645-2/ANSI S3.6 (Speech Audiometry), EN 61669, and ANSI S3.46.

Technological Characteristics

PC-based system with USB-powered hardware modules. Components include electronic circuit boards, digital signal processors, molded plastic housings, and sheet metal chassis. Uses industry-standard headsets, microphones, and couplers. Connectivity via USB to PC. Software-controlled signal processing. Compliant with EN/IEC 60601-1, ANSI S3.6, S3.22, S3.46, and IEC 60118-7.

Indications for Use

Indicated for pediatric and adult patients requiring non-continuous, noninvasive air/bone conduction and speech audiometric testing, and real-ear measurements. Also indicated for technical quality inspection of hearing instruments (no patient contact). Not indicated as a sole means of diagnostics.

Regulatory Classification

Identification

A hearing aid calibrator and analysis system is an electronic reference device intended to calibrate and assess the electroacoustic frequency and sound intensity characteristics emanating from a hearing aid, master hearing aid, group hearing aid or group auditory trainer. The device consists of an acoustic complex of known cavity volume, a sound level meter, a microphone, oscillators, frequency counters, microphone amplifiers, a distoration analyzer, a chart recorder, and a hearing aid test box.

Special Controls

*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # K113831 APR 1 9 2012 Image /page/0/Picture/2 description: The image shows the logo for Auditdata. The logo consists of the word "auditdata" in a bold, sans-serif font, with a stylized sound wave graphic to the right. Below the logo is the tagline "Your Partner in Audiology Solutions" in a smaller font. # Chapter 5 510(k) Summary #### 510(k) Information . The following 510(k) summary has been prepared pursuant to requirements specified in 21 CFR 807.92(a) ## Submitter/Owner Auditdata A/S Banestroeget 13 DK-2630 Taastrup Denmark Phone: +45 70 20 31 24 Establishment Registration No: 3008386587 Contact: Dan Haugbøl QA & Regulatory Manager dan@auditdata.com Summary date: March 08, 2012 # Device Identification Submission type: Traditional Reason for Submission: New Device Trade name: Primus Fitting System Common name: Primus Classification name: Hearing aid calibrator and analysis system Secondary class: Audiometer, 21 CFR 874.1050, EWO FDA: 21 CFR 874.3310 Device Product Code: ETW Classification Panel: Ear, Nose, and Throat Device Class: Class II © Auditdata 2012 {1}------------------------------------------------ Image /page/1/Picture/0 description: The image shows the logo for Auditdata. The logo features the word "auditdata" in a bold, sans-serif font, with a stylized graphic of sound waves emanating from the right side of the word. Below the logo is the tagline "Your Partner in Audiology Solutions" in a smaller font. # Predicate Marked Device Primus Fitting System has equivalence to: Product Name: Aurical Manufacturer: GN Otometrics 510(k) Number: K945199 # Device Description The Primus Fitting System is a PC operated system that contains Primus Hearing Instrument Test unit (PHITU) for Hearing Instrument testing (HIT), Primus Fitting Unit (PFU) for Audiometry (AUD) and Real Ear measurement (REM). The Primus Fitting System is operated from a dedicated software application installed on a standard Windows operating system. The Primus Fitting system can either be sold individually (PHITU or PFU) or together (PHITU and PFU). The Primus software solution operates within the NOAH framework, as well as stand-alone. On-top diagrams and control panels make it easy to use simultaneously with the proprietary hearing instrument fitting systems of your choice. The stand-alone client data files hold all historical session information, and makes fittings away from the office very easy. #### Audiometry (AUD) Primus AUD is a PC based audiometry module providing a wide range of possibilities within pure tone audiometry, bone conduction audiometry and speech audiometry including inserts probe and industry standard headsets with an option of high frequency testing. #### Real Ear Measurements (REM) Primus REM offers classic Real Ear Measurements as well as comprehensive Speech Mapping measurements, where all measurements include high frequency testing. The measurements are supported by a large sound library including technical sounds, daily life sounds, speech signals and dialogue. The Primus REM module contains 2 probes (one for each ear) each with 2 microphones used for measuring the sound level outside and inside the ear. The classic Real Ear Measurements, which can be performed by Primus REM, include: Unaided, Occluded and Aided Response as well as Insertion Gain. Toggling between SPL and Gain shows the response measurements in gain view. The PFU hardware is powered by a USB connection from the PC and a medical grade power supply (for extended sound pressure) and can be placed at the most convenient working place. #### Hearing Instrument Testing (HIT) Primus HIT offers full featured technical measurements for testing and troubleshooting hearing instruments. The test can be run as an automatic test sequence in accordance with European and American standards. The Primus HIT unit module contains of 2 microphones (one for measuring and one for reference) and various couplers for hearing aid testing. The PHITU hardware is powered by a USB connection from the PC and can be placed at the most convenient working place. {2}------------------------------------------------ Image /page/2/Picture/0 description: The image shows the logo for Auditdata. The logo consists of the word "auditdata" in a bold, sans-serif font, with a stylized sound wave graphic to the right of the word. Below the logo is the tagline "Your Partner in Audiology Solutions". # PHITU Indication for Use/Intended use #### Indications for use (PHITU): The Primus Hearing Instrument Test Unit is intended for use by professionals such as an audiologist, hearing healthcare specialist, or trained technician. Audiometric testing should take place in an extremely quiet environment and care should be taken to ensure optimal test conditions and safety of the client during testing. The Primus Hearing Instrument Test Unit is intended to give an objective indication of the characteristics of a Hearing Aid, by visualizing a signal recorded in the test coupler with reference information such as target curves in order to make adjustments of the Hearing Instrument settings. - The Primus Hearing Instrument Test Unit is indicated for technical quality inspection of . hearing instruments with no clients involved. #### Indications for use (PFU): The Primus Fitting Unit is intended for use by professionals such as an audiologist, hearing healthcare specialist, or trained technician. Audiometric testing should take place in an extremely quiet environment and care should be taken to ensure optimal test conditions and safety of the client during testing. - . The Primus Fitting Unit is intended for hearing test. - The Primus Fitting Unit with stated accessories is indicated for non-continuous, . noninvasive air and optionally bone conduction and speech audiometric testing in quiet ·office and sound treated environments. - . The Primus Fitting Unit is indicated for non-continuous real-ear measurements at the ear drum by means of noninvasive external ear canal insertion of a probe tube in quiet office environments. - . Finally the system can be used to present hearing instrument related sound examples through headsets or loudspeakers. - . The Primus Fitting Unit is Indicated for use with both pediatric and adult age groups. - The Primus Fitting Unit is not indicated as a sole means of diagnostics. . {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the word "auditdata" in a bold, sans-serif font. To the right of the word, there is a stylized graphic that resembles radio waves or sound waves emanating from a point. The overall impression is that of a company logo or brand name. Your Partner in Audiology Solutions # Technological Characteristics The PHITU and PFU unit uses the same overall technology as the predicate devices: | PFU/PHITU | Predicate devices | |-------------------------------------------------|-------------------| | Conventional electronic circuit<br>board design | Similar design | | | | | Industry standard Digital Signal<br>processor | Similar design | | | | | Conventional molded plastic<br>parts design | Similar design | | | | | Conventional sheet metal design | Similar design | | Industry standard headset | Same | | Mains operated energy source | Same | #### Comparison to Predicate Device Fitting systems like the Primus Fitting System have been produced for decades. The system most similar to the Primus Fitting System in terms of technical design and features is the Aurical from GN Otometrics. From a measurement perspective, the two systems utilize very similar signal processing and electrical design within the device, and both are controlled and operated by software running on the PC. The primary difference between the two devices is associated with the mechanical design, where Aurical is a combined AUD/REM/HIT unit and Primus Fitting System have independent HIT and REM/AUD unit operated by software running on the PC. #### Substantial Equivalence Performance testing Substantial equivalence to the Primus Fitting system is based a side-by-side design comparison (Aurical from GN Otometrics K945199) all verified by means of non-clinical performance testing of the Primus Fitting system as specified in ANSI S3.6, ANSI S3.22 and ANSI S3.46 This is further supported by the design implementations obtained as a result being compliant to the standards listed below. {4}------------------------------------------------ Image /page/4/Picture/0 description: The image shows the logo for Auditdata. The logo consists of the word "auditdata" in a bold, sans-serif font, with a stylized sound wave graphic to the right. Below the logo is the tagline "Your Partner in Audiology Solutions" in a smaller, sans-serif font. # Conclusion about Safety and Effectiveness Evaluating the results of testing to the following key standards has allowed us to make conclusions about the safety and effectiveness of the Primus Fitting System. | Standard | Description | |--------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | EN/IEC 60601-1 | Medical Electrical Equipment --<br>Part 1: General Requirements for<br>Safety. | | EN/IEC 60601-1-2 | Medical Electrical Equipment --<br>Part 1: General Requirements for<br>Safety; Electromagnetic<br>Compatibility -- Requirements and<br>Tests | | ANSI/ S3.22 | Specification of Hearing Aid<br>Characteristics | | IEC 60118-7 | Electroacoustics - Hearing aids -<br>Part 7: Measurement of the<br>performance characteristics of<br>hearing aids for production,<br>supply and delivery quality<br>assurance purposes | | EN60645-1/ANSI S3.6 Type 1 | Tone Audiometry | | EN60645-2/ANSI S3.6 Type A<br>or A-E | Speech Audiometry | | EN 61669 | Equipment for the measurement<br>of real-ear acoustical<br>characteristics of hearing aids. | | ANSI S3.46 | Methods of Measurement of<br>Real-Ear Performance<br> | The equivalent results from both the Primus fitting System and the Aurical System obtained when each device was tested to determine compliance with the standards clearly demonstrates the substantial equivalence between these two devices. Based on the results of testing to the applicable requirements of the aforementioned standards and achieving compliance to them, we hereby conclude that Primus Fitting System is both safe and effective. 5-5 {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs, enclosed in a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged around the upper portion of the circle. # DEPARTMENT OF HEALTH & HUMAN SERVICES Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 Auditdata A/S c/o Mr. Dan Haugbol Banestroeget 13 DK-2630 Taastrup Denmark APR 1 9 2012 Re: K113831 Trade/Device Name: Primus Fitting System Regulation Number: 21 CFR 874.3310 Regulation Name: Hearing aid calibrator and analysis system Regulatory Class: II Product Code: ETW, EWO Dated: March 8, 2012 Received: March 14, 2012 Dear Mr. Haugbol: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ # Page 2 - Mr. Dan Haugbol Please be advised that FDA's issuance of a substantial equivalence determination does not mean I loast ve devisou that I Driver and that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must of any I odolar statuation and regenanents, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours, Eda K. VMD Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ K113831 # Indications for Use Statement 510(k) Number: K113831 Device Name: Primus Fitting System #### Primus Hearing Instrument Test Unit (PHITU) The Primus Hearing Instrument Test Unit is intended for use by professionals such as an audiologist, hearing healthcare specialist, or trained technician. Audiometric testing should take place in an extremely quiet environment and care should be taken to ensure optimal test conditions and safety of the client during testing. The Primus Hearing Instrument Test Unit is intended to give an objective indication of the characteristics of a Hearing Aid, by visualizing a signal recorded in the test coupler with reference information such as target curves in order to make adjustments of the Hearing Instrument settings. - The Primus Hearing Instrument Test Unit is indicated for technical quality inspection of hearing instruments with no clients involved. #### Primus Fitting Unit (PFU): The Primus Fitting Unit is intended for use by professionals such as an audiologist, hearing healthcare specialist, or trained technician. Audiometric testing should take place in an extremely quiet environment and care should be taken to ensure optimal test conditions and safety of the client during testing. · The Primus Fitting Unit is intended for hearing test. - · The Primus Fitting Unit with stated accessories is indicated for non-continuous, noninvasive air and optionally bone conduction and speech audiometric testing in quiet office and sound treated environments. - The Primus Fitting Unit is indicated for non-continuous real-ear measurements at the ear drum by means of noninvasive external ear canal insertion of a probe tube in quiet office environments. - Finally the system can be used to present hearing instrument related sound examples through headsets or loudspeakers. - The Primus Fitting Unit is indicated for use with both pediatric and adult age groups. - · The Primus Fitting Unit is not indicated as a sole means of diagnostics. Prescription Use × AND/OR (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use X (Per 21 CFR 801.109) ch~gs~ (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices 510(k) Number. KIJ3831
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...