← Product Code [OVE](/productcode/OVE) · K113796

# SOLITAIRE-C CERVICAL SPACER SYSTEM (K113796)

_Ebi, LLC · OVE · Apr 26, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113796

## Device Facts

- **Applicant:** Ebi, LLC
- **Product Code:** [OVE](/productcode/OVE.md)
- **Decision Date:** Apr 26, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Solitaire -C Cervical Spacer System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2 to T1. Cervical degenerative disc disease is defined as intractable radior myclopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The Solitaire -C Cervical Spacer is to be used with autograft and implanted via an anterior approach. The Solitaire-C spacer must be implanted with the Solitaire-C titanium screws that are part of the system. This cervical device is to be used in patients who have had six weeks of non-operative treatment.

## Device Story

Solitaire-C Cervical Spacer System is an intervertebral body fusion device for stand-alone anterior cervical fusion. Device consists of PEEK spacer body, titanium faceplate/band, and tantalum markers; implanted with titanium bone screws. Used by surgeons in clinical settings to treat cervical degenerative disc disease. Spacer provides structural support for fusion; screws provide integrated fixation. Device implanted via anterior approach; requires autograft. Mechanical strength validated via static/dynamic compression, torsion, subsidence, and expulsion testing. Benefits include stabilization of cervical spine segments to alleviate nerve root or spinal cord compression.

## Clinical Evidence

Bench testing only. No clinical data presented. Mechanical testing performed per Class II Special Controls Guidance: Intervertebral Body Fusion Device, including static/dynamic axial compression, static/dynamic compression-shear, static/dynamic torsion, subsidence, and expulsion testing (ASTM F-2077, ASTM F-2267, ASTM Draft F-04.25.02.02). Additional testing included screw back-out, screw push-through, interconnection strength, and wear debris analysis.

## Technological Characteristics

PEEK-Optima LT1 (ASTM F-2026) main body; Ti-6Al-4V ELI (ASTM F-136) faceplate/band; unalloyed tantalum (ASTM F-560) markers. Titanium bone screws. Stand-alone interbody fusion design. Mechanical testing per ASTM F-2077, F-2267.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Solitaire PEEK Anterior Spinal System ([K081395](/device/K081395.md), [K093629](/device/K093629.md))
- Coalition Spacer ([K083389](/device/K083389.md))
- AVS-C Spacer ([K102606](/device/K102606.md))
- Synthes Zero-P Cervical Spacer ([K072981](/device/K072981.md), [K093629](/device/K093629.md))
- C-Thru Spacer System ([K092336](/device/K092336.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K113736
page 1 of 2

APR 2 6 2012

Image /page/0/Picture/2 description: The image shows the logo for Biomet Spine. The word "BIOMET" is in large, bold, sans-serif font. Below the word "BIOMET" is the word "SPINE" in a smaller, sans-serif font. The logo is black and white.

### 510(k) Summary

This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR \$ 807.92.

| Preparation Date:                      | April 25, 2012                                                           |
|----------------------------------------|--------------------------------------------------------------------------|
| Applicant/Sponsor:                     | Biomet Spine<br>100 Interpace Parkway<br>Parsippany, NJ 07054            |
| Contact Person:                        | Margaret F. Crowe<br>Phone: 973-299-9300, ext. 2260<br>Fax: 973-257-0232 |
| Trade name:                            | Solitaire®-C Cervical Spacer System                                      |
| Common Name:                           | Cervical interbody fusion device with integrated fixation                |
| Classification Name<br>(Product Code): | Intervertebral Body Fusion Device (OVE)                                  |
| Device Panel - Regulation No.:         | Orthopedics - 21 CFR 888.3080                                            |

#### Device Description:

The purpose of this submission is to gain market clearance for the Solitaire-C Cervical Spacer System. The Solitaire -C Cervical Spacer System consists of spacers and bone screws for stand-alone cervical intervertebral body fusion. The Solitaire -C spacer will be available in a variety of sizes, angles and footprints. This cervical spacer has a PEEK main body (PEEK-Optima LT1 per ASTM F-2026) with a titanium faccplate and band (Ti-6Al-4V ELI alloy per ASTM F-136), and tantalum markers (unalloyed tantalum per ASTM F-560). This device accepts titanium bone screws that are available in two diameters and multiple lengths.

#### Indications for Use:

The Solitaire -C Cervical Spacer System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2 to T1. Cervical degenerative disc disease is defined as intractable radior myclopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The Solitaire -C Cervical Spacer is to be used with autograft and implanted via an anterior approach. The Solitaire-C spacer must be implanted with the Solitaire-C titanium screws that are part of the system. This cervical device is to be used in patients who have had six weeks of non-operative treatment.

#### Summary of Technologies:

The technological characteristics (material, design and sizing) of the Solitaire-C Cervical Spacer System is the same as, or similar to, the predicate devices. Examples of predicate devices include:

- Solitaire PEEK Anterior Spinal System (Biomet Spine K081395, K093629) .
- . Coalition Spacer (Globus Medical - K083389)
- A V S-C Spacer (Stryker Spine K 102606) .
- Synthes Zero-P Ccrvical Spacer (Synthes Spine K072981, K093762) .
- C-Thru Spacer System (Biomet Spine K092336) .

{1}------------------------------------------------

#### Performance Data

Mechanical testing recommended in the special controls guidance document entitled "Class II Special Controls Guidance Document: Intervertebral Body Fusion Device" was conducted. The testing conducted, along with the ASTM standard, are listed below:

- Static Axial Compression (ASTM F-2077) l)
- 2) Dynamic Axial Compression (ASTM F-2077)
- 3) Static Compression-Shear (ASTM F-2077)
- 4) Dynamic Compression-Shear (ASTM F-2077)
- 5) Static Torsion (ASTM F-2077)
- 6) Dynamic Torsion (ASTM F-2077)
- Subsidence (ASTM F-2267 and ASTM F-2077) 7)
- 8) Expulsion (ASTM Draft F-04.25.02.02)

Additional mechanical testing was conducted to evaluate screw back out, screw push through, and interconnection testing between the spacer body and the faceplate. Wear debris analysis was also presented.

Mechanical testing shows that the mechanical strength of the subject device is sufficient for its intended use.

#### Substantial Equivalence:

The Solitaire-C Cervical Spacer System is substantially equivalent to its predicate devices with respect to intended use and indications, technological characteristics, and principles of operation and do not present any new issues of safety or effectiveness. The predicates listed above are distributed for similar indications, and/or have similar design features.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or bird-like figure.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

APR 2 6 2012

Biomet Spine % Ms. Margaret F. Crowe 100 Interpace Parkway Parsippany, New Jersey 07054

Re: K113796

Trade/Device Name: Solitaire -C Cervical Spacer System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: II Product Code: OVE Dated: March 15, 2012 Received: March 16, 2012

Dear Ms. Crowe:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications) for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{3}------------------------------------------------

Page 2 - Ms. Margaret F. Crowe

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark M. Milhiven

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Indications for Use

# 510(k) Number (if known): ドリ3796

Device Name: Solitaire -C Cervical Spacer System

### Indications for Use:

The Solitaire -C Cervical Spacer System is indicated for stand-alone anterior cervical interbody fusion procedures in skeletally mature patients with cervical degenerative disc disease at one level from C2 to T1. Cervical degenerative disc disease is defined as intractable radioulopathy with herniated disc and/or osteophyte formation on posterior vertebral endplates producing symptomatic nerve root and/or spinal cord compression confirmed by radiographic studies. The Solitaire -C Cervical Spacer System is to be used with autograft and implanted via an anterior approach. The Solitaire-C spacer must be implanted with the Solitaire-C titanium screws that are part of the system. This cervical device is to be used in patients who have had six weeks of non-operative treatment.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

K113796 510(k) Number_

---

**Source:** [https://fda.innolitics.com/device/K113796](https://fda.innolitics.com/device/K113796)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
