GLUCOSE METER-CHECK SOLUTION FOR INFOPIA

K113764 · Bionostics, Inc. · JJX · Jan 30, 2012 · Clinical Chemistry

Device Facts

Record IDK113764
Device NameGLUCOSE METER-CHECK SOLUTION FOR INFOPIA
ApplicantBionostics, Inc.
Product CodeJJX · Clinical Chemistry
Decision DateJan 30, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1
AttributesPediatric

Intended Use

Glucose Meter-Check® Solution for Infopia is intended for use to verify the performance and correct operation of Infopia blood glucose monitoring test systems utilizing Infopia glucose test strips. Glucose Meter-Check Solution for Infopia is intended for use by healthcare professionals and people with diabetes mellitus at home.

Device Story

Glucose Meter-Check® Solution for Infopia is an aqueous buffer solution containing D-glucose, viscosity modifiers, and preservatives; non-biological origin. Used as a quality control material to verify performance of Infopia blood glucose monitoring systems. Operated by healthcare professionals or patients with diabetes at home. User applies solution to Infopia glucose test strips; system measures glucose concentration; user compares result against lot-specific expected range (e.g., 88–132 mg/dL) to confirm system accuracy and functionality.

Clinical Evidence

No clinical data. Bench testing only. Stability studies confirmed 24-month shelf life at 2-30°C and 3-month open-vial stability. Traceability established to SRM 917 via YSI 2300 reference method.

Technological Characteristics

Buffered aqueous solution containing D-glucose, viscosity modifiers, and preservatives. No biological components. Formulated for use with Infopia blood glucose test systems. Packaged in 6 mL LDPE vials. Performance verified against ISO 15197:2003 requirements for intermediate precision (96 to 144 mg/dL range).

Indications for Use

Indicated for healthcare professionals and people with diabetes mellitus to verify the performance and correct operation of Infopia blood glucose monitoring test systems using Element family test strips.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113764 JAN 3 0 2012 # 510(k) Summary1 Quality Solu (a) (1) Submitter's name, address Bionostics, Inc. 7 Jackson Road Devens, MA 01434 Contact Person Randy Byrd VP, Chief Technical Officer (978) 772-7070 x 272 Date of preparation of this summary: 20 December 2011 - (2) Device trade or proprietary name: Glucose Meter-Check Solution for Infopia ### Device common or usual name or classification name: Single (Specified) Analyte Control, All Types, Assayed and Unassayed אור | a prove and a program a comments of the many of the summer<br>REGULATION MEDICAL<br>OFFIAITY<br>And In Canada State 2 | and and the former of the many of the status and and any the more<br>REGULATION<br>"NUMBER | 2000 - 10 15 15 - 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 11 | OFGUIA I<br>SCRIPTION<br>11.07 | |-----------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------| | Chemistry | A /<br>1660 | | Glucose Control | #### l. Substantial Equivalence Glucose Meter-Check Control Solution for Infopia is substantially equivalent in function, safety and efficacy to currently marketed devices for the same intended use as shown in the following tables: | Characteristic | Predicate Device | Modified Device | |-------------------|----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------| | Name: | Infopia Control Solution | Glucose Meter-Check Control<br>Solution for Infopia | | 510(k), Date: | K0572369, Mar 21 2008 | K0572369, Mar 21 2008 | | Number of levels: | 1, typical fasting glucose | 1, typical fasting glucose | | Target ranges: | 94 – 140 mg/dL | 88 – 132 mg/dL | | Analytes: | glucose | glucose | | Container: | 6 mL LDPE vial with dispensing tip<br>and cap | 6 mL LDPE vial with dispensing tip<br>and cap | | Fill volume: | 4 mL | 4 mL | | Color: | red | red | | Matrix: | Buffered, aqueous solution of D-<br>Glucose, viscosity modifier,<br>preservatives and other, non-<br>reactive ingredients. | Buffered, aqueous solution of D-<br>Glucose, viscosity modifier,<br>preservatives and other, non-<br>reactive ingredients. | | Brands: | Infopia | Glucose Meter-Check | <sup>1</sup> This summary of safety and effectiveness is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. {1}------------------------------------------------ #### Description of the new device II. Glucose Meter-Check Solution for Infopia is a buffered aqueous solution with glucose containing no ingredients of biological origin, or in concentrations qualifying as a controlled product under the Controlled Products Regulation. The Glucose Meter-Check Solution is formulated for optimal performance on Infopia glucose meters using Infopia test strips. #### (a) (1) Intended use of the device Glucose Meter-Check Solution for Infopia is intended for use to verify the performance and correct operation of Infopia blood glucose monitoring test systems utilizing the Element family of blood glucose test strips. Glucose Meter-Check Solution for Infopia is intended for use by healthcare professionals and people with diabetes mellitus at home. #### Technological characteristics of the device. (a) (2) This material consists of viscosity-adjusted, aqueous glucose control solution prepared with a single concentration of D-glucose with recovery on the test systems in the range corresponding to Interval 2 of the glucose contration intervals for evalution of intermediate precision in ISO 151972 (96 to 144 mg/dL). This solution has been optimized to simulate the response of whole blood on the blood glucose test systems manufactured by Infopia using Infopia blood glucose test strips. The solution contains no hazardous, human or animal derived components. #### Summary of non-clinical tests submitted with the premarket notification for the device. (b) (1) Tests were conducted to verify specific performance requirements: - a) Closed bottle stability (Shelf-life) - b) Stability after opening (Use-life) - c) Transport Stability - d) Test response - Summary of clinical tests submitted with the premarket notification for the device. (b) (2) N/A #### Conclusions drawn from the clinical and non-clinical trials. (b) (3) Comparison of technological characteristics, formulation and intended use to predicate devices listed in this summary support the claim of substantial equivalence. <sup>2</sup> ISO 15197. In vitro diagnostic test systems – Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus. 155 edition, May 2003. {2}------------------------------------------------ ### TPLC Brand Name Detailed View | | | | | | Name | | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | | LANDLERS CREEN AR | |--------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------|--|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------| | | | <br>THE COLORAL COLORIAL COLLECTION COLLEGION COLLEGION COLLEGION COLLEGION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CON | <br> | <br>A production and annual a consideration and control and contractive de contractive de management de contrôleant de contrôleant de comments de combination of contrôleande de c | | | CARA A BALLERS A CONSULTION COLLEGION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONS | | CARACA LEASING CARACT CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTION CONSULTIO | | | ------------------------------------------------------------------------------------------------ | Казанские продавания по представленные продавания пристемента на сединание пристемента пристементальным полнение можение можение можение можение можение можение можение можен | | | | | | | | | | ### Registration and Listing ### Premarket Reviews Completed (None) ### Under Review, Withdrawn or Closed without Product Code (1) | K113764 | GLUCOSE METER-CHECK SOLUTION FOR INFOPIA | Review | Claudio Olga | |---------|-----------------------------------------------|---------|--------------| | Image | FDA.GOV JJK SINGLE ANALYTE CONTROL, ALL TYPES | K072369 | | | K113764 | GLUCOSE METER-CHECK SOLUTION FOR INFOPIA | Review | Claudio Olga | | Image | FDA.GOV JJK SINGLE ANALYTE CONTROL, ALL TYPES | K103021 | | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------- MDR Summary (None) MDR Analyst (None) MDR Distribution by Brand - Death or Injury (None) Patient Problems (None) Patient Outcomes (None) Device Problems (None) Manufacturer Evaluation Results (None) Manufacturer Evaluation Conclusions (None) Recalls (None) {3}------------------------------------------------ #### TPLC disclaimers We are continuing to improve and enhance the TPLC Universe and the TPLC sheets. When more data is available we will update this message. Please realize that in this initial release: - -Recall data is only available since October, 2009. - In the Premarket Under Review section and the Recall section, the premarket submission or recall is only included if the product code is specified in CTS. - . The MDR count is a count of reports and contains duplicate reports. - -MDR track action or additional information letters is not available yet. - EIR data related to inspections is not yet displayed. - - The TPLC name which consolidates variations on manufacturer names is not yet implemented. - " Publications are not yet linked in. We are working hard to address these issues and many can be addressed before the next release. {4}------------------------------------------------ Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized representation of a caduceus, a symbol often associated with medicine and healthcare. The caduceus consists of a staff with two snakes coiled around it, topped by a pair of wings. #### Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 JAN 3 0 2012 Bionostics. Inc. c/o Randy Byrd Chief Technical Officer 7 Jackson Road Devens, MA 01434 Re: k113764 Glucose Meter-Check® Solution for Infopia Trade/Device Name: Regulation Number: 21 CFR 862.1660 Regulation Name: Quality Control Material Regulatory Class: Class I, reserved . Product Code: JJX Dated: December 20, 2011 Received: December 21, 2011 Dear Mr. Byrd: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may. therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. {5}------------------------------------------------ ### Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ... You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm Sincerely yours, signature Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {6}------------------------------------------------ ## Indications for Use 510(k) Number: Device Name: Indications for Use: Glucose Meter-Check® Solution for Infopia Glucose Meter-Check Solution for Infopia is intended for use to verify the performance and correct operation of Infopia blood glucose monitoring test systems utilizing Infopia glucose test strips. Glucose Meter-Check Solution for Infopia is intended for use by healthcare professionals and people with diabetes mellitus at home. For In Vitro Diagnostic Use Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrent of CDRH, Office of Device Evaluation (OIVD) Division Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K113764 Page 1 of 1
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...