K113753 · Zimmer Dental, Inc. · DZE · Aug 30, 2012 · Dental
Device Facts
Record ID
K113753
Device Name
ZIMMER DENTAL TAPERED SCREW-VENT X
Applicant
Zimmer Dental, Inc.
Product Code
DZE · Dental
Decision Date
Aug 30, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.3640
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K113753 · Aug 30, 2012
ZIMMER DENTAL TAPERED SCREW-VENT X
Zimmer Dental, Inc.
Longitudinal clinical data collection
A longitudinal study was conducted to evaluate the device's performance and safety in normal clinical practice, including patients with elevated risk factors excluded from the controlled prospective study.
Longitudinal study; Normal clinical conditions; Real-world population
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Longitudinal data collection study; Longitudinal observational study
Patients in normal clinical conditions, including those with elevated risk factors (alcoholics, substance abusers, mentally unstable, smokers, osteoporotics, uncontrolled diabetics, parafunctional habits, dental or oral infections, and bone graft patients)
Not applicable for this study
Performance and functioning of the device
Indications for Use
The Tapered Screw-Vent® X Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The 4.1mmD Tapered Screw Vent X Implants should be splinted to additional implants when used in the posterior region.
Device Story
Tapered Screw-Vent® X is an endosseous dental implant system; used by dentists/oral surgeons in clinical settings to replace missing teeth. Device consists of titanium alloy (6Al-4V) body with Trabecular Metal™ (tantalum) component; features tapered design with triple-lead threads and internal hexagon interface. Implant is surgically placed into maxilla or mandible osteotomy; self-tapping design facilitates insertion. Provides foundation for prosthetic restoration. Benefits include osseointegration support and versatility for immediate or delayed loading protocols. Clinical decision-making relies on surgeon assessment of primary stability and occlusal load requirements.
Clinical Evidence
Prospective clinical study evaluated performance in posterior region with two-week immediate load protocol (6-month endpoint). Longitudinal data collection study included normal patients and those with elevated risk factors (e.g., smokers, diabetics, bone graft patients). No device-related adverse events reported in either study.
Indicated for patients requiring replacement of one or more missing teeth in the maxilla or mandible. Suitable for immediate loading (with good primary stability and appropriate occlusal load) or conventional healing periods. 4.1mmD implants require splinting in the posterior region.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
Predicate Devices
Zimmer Dental Tapered Screw-Vent® P Implant (K101880)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Zimmer Dental. The logo consists of a circle with a stylized letter "Z" inside it. Below the circle, the word "zimmer" is written in a bold, sans-serif font, and below that, the word "dental" is written in a smaller, serif font.
Zimmer Dental
1900 Aston Avenue Carlsbad, CA 92008 760.929.4300 (ph) 760.431.7811 (fax)
**510k No.:** K113753
Page No .: . A5-1
## Traditional 510(k) PRE-MARKET NOTIFICATION 510(k)
AUG 30 2012
# 510(k) SUMMARY (21CFR807.92(a))
#### 1. Submitter's Information:
Zimmer Dental Inc. Name:
1900 Aston Ave. Address: Carlsbad, CA 92008
760-929-4300 Phone:
Jeremy Markovich Contact:
Date Prepared: December 19, 2011
- 2. Device Name: Tapered Screw-Vent®X Implant Trade Name: K112160 510(k) Number: 872.3640 Regulation Number: DZE Classification Code: Device Classification Name: Endosseous Dental Implant
#### 3. Predicate Device(s):
Predicate Device No. 1 Zimmer Dental Tapered Screw-Vent® P Implant Trade Name: 510(k) Number: K101880 872.3640 Requlation Number: Classification Code: DZE Endosseous Dental Implant Device Classification Name:
Predicate Device No. 2 Zimmer Dental Tapered Screw-Vent® Implant Trade Name: System . K011028 / K953101 / K013227 / K061410 / 510(k) Number: K072589 Regulation Number: 872.3640 DZE Classification Code: Endosseous Dental Implant Device Classification Name:
{2}------------------------------------------------
Page No.
#### 4. Device Description:
The Tapered Screw-Vent® X Implant is an endosseous dental implant. The implant is composed of titanium alloy and Trabecular metal. The implant section is designed for ease of implantation and with greater surface area for osseointegration. The implant section surface is treated to facilitate osseointegration. In addition, the implant section is tapered with triple-lead threads.
The Tapered Screw-Vent X implants will be offered in two different texturing configurations: full texture to the top of the implant and texture to 0.5mm from the top of the implant. In addition, both texturing configurations of the implant will have coronal grooves on the collar to within 0.64mm of the top of the implant similar to the predicate Tapered Screw-Vent P Dental implant. The implant/abutment interface platform diameter will be offered in sizes of 3.5mm, 4.5mm, or 5.7mm depending on the outside implant thread diameter. The new device will feature MTX surface equivalent to existing Zimmer Dental implants. The MTX surface is used on the titanium body and is exposed on surfaces apical and coronal to the Trabecular Metal.
#### 5. Indications for Use:
The Tapered Screw-Vent® X Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The 4.1mmD Tapered Screw Vent X Implants should be splinted to additional implants when used in the posterior region.
#### Device Comparison: હ.
The Tapered Screw-VentX implant is substantially equivalent to Zimmer's Tapered Screw-Vent implant. Both implants have the same friction fit internal hexagon connection and use the same prosthetic components to restore the implant. Both implants feature triple lead threads that taper inwardly from the coronal to the apical end of the implant (the overall taper for both of the implants are 2 degrees). Both implants have apical flutes that allow the implant to self tap into the osteotomy. Similar to the Tapered Screw-Vent®, the Tapered Screw-Vent® X is offered in diameters from 4.1, 4.7, and 6.0mmD (Tapered Screw-Vent® is also offered in 3.7mmD) and in lengths 8, 10, 11.5, and 13mmL (Tapered Screw-Vent® is also offered in 16mmL) The Tapered
.5
{3}------------------------------------------------
Tapered Screw-Vent® X Implant
Page No.
Screw-Vent® X and Tapered Screw-Vent® all use identical surgical instrumentation and an identical surgical protocol.
#### 7. Technological Characteristics
| Feature | Subject Device | Predicate Device 1 | Predicate Device 2 |
|----------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|------------------------------------------------------------|
| | Tapered Screw-Vent® X | Tapered Screw-Vent® P | Tapered Screw-Vent® |
| Implant Interface | Internal Hex | Internal Hex | Internal Hex |
| Implant Lengths | 8, 10, 11.5, 13mm | 10, 11.5, 13, 16mm | 8, 10, 11.5, 13, 16mm |
| Implant Diameters | 4.1, 4.7, 6.0mm | 4.7, 6.0mm | 3.7, 4.1, 4.7, 6.0mm |
| Material | Titanium 6Al-4V | Titanium 6Al-4V | Titanium 6Al-4V |
| Collars | Machined with grooves or<br>texture to top with<br>grooves | Machined with grooves or<br>texture to top with<br>grooves | Machined with grooves or<br>texture to top with<br>grooves |
| Thread Pattern | Triple lead threads,<br>pattern tightly spaced &<br>equal; partial cylinder<br>type body | Triple lead threads,<br>pattern tightly spaced &<br>equal; partial cylinder<br>type body | Triple lead threads,<br>pattern tightly spaced &<br>equal |
| Surface<br>Characteristics | MTX Surface<br>and Trabecular MetalTM<br>(tantalum) | MTX Surface<br>and CSTiTM coating | MTX Surface, or HA<br>Coated, or MP-1® HA<br>Coated |
#### 8. Non-clinical Testing
Testing was performed following "Guidance for Industry and FDA Staff -Class II Special Controls Guidance Document: Root-form Endosseous Dental Implants and Endosseous Dental Abutments." Performance testing demonstrated that the device performs appropriately for the proposed indications for use.
#### 9. Clinical Testing
A prospective clinical study was conducted to evaluate the performance of this device within a controlled population that included implantation within the posterior region only and a two week final restoration immediate load protocol with a 6-month end point.
A second, longitudinal data collection study was initiated to gather data on the routine placement and functioning within normal clinical conditions. This population included normal patients and subjects with elevated risk factors that were excluded from the controlled study, such as: alcoholics, substance abusers, mentally unstable, smokers,
{4}------------------------------------------------
Tapered Screw-Vent®X Implant
510(k) No. K113753 Page No. A5-4
osteoporotics, uncontrolled diabetics, parafunctional habits, dental or oral infections, and bone graft patients.
No device related adverse events were reported in either study to date.
#### 10. Conclusion
Based on our analysis, the device is substantially equivalent to the predicates and it is considered that the new device is as safe and effective for its indications for use.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three wing-like shapes extending to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
AUG 30 2012
Mr. Jeremv Markovich Seinor Regulatory Affairs Specialist Zimmer Dental, Incorporated 1900 Aston Ave. Carlsbad, California 92008
Re: K113753
Trade/Device Name: Tapered Screw-Vent® X Implant Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE Dated: August 24, 2012 Received: August 27, 2012
Dear Mr. Markovich:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing (reporting or increments as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to
http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{7}------------------------------------------------
Image /page/6/Picture/0 description: The image shows a logo with a stylized letter 'Z' inside a circle. Below the circle, the word 'zimmer' is written in a sans-serif font. Underneath 'zimmer', the word 'dental' is written in a smaller, slightly italicized font.
A4-1
Tapered Screw-Vent® X Implant
### Indications for Use
KII3753
510(k) Number (if known):
Device Name: Tapered Screw-Vent® X Implant
Indications For Use:
The Tapered Screw-Vent® X Implants are designed for use in the maxilla or mandible for immediate loading or for loading after a conventional healing period. Implants may be used to replace one or more missing teeth. Immediate loading is indicated when there is good primary stability and an appropriate occlusal load. The 4.1mmD Tapered Screw Vent X Implants should be splinted to additional implants when used in the posterior region.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Super Rese
(Division Sign-Off)
Division of Anesthesiology, General Hospital
Infection Control, Dental Devices
510(k) Number: K113753
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.