← Product Code [HRS](/productcode/HRS) · K113733

# SMALL AND LARGE FRAGMENTS OSTEOSYNTHESIS SYSTEM NEOFIX (K113733)

_Neoortho Produtos Ortopedicos S/A · HRS · Mar 23, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113733

## Device Facts

- **Applicant:** Neoortho Produtos Ortopedicos S/A
- **Product Code:** [HRS](/productcode/HRS.md)
- **Decision Date:** Mar 23, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3030
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

Small and Large Fragments Osteosynthesis System NEOFIX is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, carpais, metacarpals, distal femur, proximal tibia, tibial pilon, fibula, calcaneus, tarsals and metatarsals; small fragments of the hand and wrist; pelvis and acetabulum fractures; periprosthetic fractures; metatarsal and phalangeal osteotomies; and carpal, metacarpal, wrist and ankle arthrodesis. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.

## Device Story

NEOFIX is an orthopedic internal fixation system comprising metallic plates and screws used for bone fracture stabilization, arthrodesis, and osteotomy. The system is utilized by orthopedic surgeons in clinical or surgical settings. Plates are provided in various geometric configurations with screw holes for locking or non-locking screws; screws are available in cortical and cancellous thread designs. The device provides mechanical stability to bone fragments, facilitating healing. It is intended for use in both small and large bone applications, including trauma and reconstructive surgery. The system is not for spinal use.

## Clinical Evidence

Bench testing only. No clinical data provided. Performance was demonstrated through engineering analysis and mechanical testing conducted in accordance with ASTM F382 (Standard Specification and Test Method for Metallic Bone Plates) and ASTM F543 (Standard Specification and Test Methods for Metallic Medical Bone Screws).

## Technological Characteristics

Materials: Commercially pure titanium or Ti-6Al-4V alloy. Components: Plates (various geometric configurations) and screws (cortical/cancellous, locking/non-locking). Standards: ASTM F382, ASTM F543. Sterilization: Standard methods (not specified). Connectivity: None (mechanical device).

## Regulatory Identification

Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.

## Predicate Devices

- Congruent Bone Plate System ([K012655](/device/K012655.md))
- Acumed Congruent Bone Plate System ([K071715](/device/K071715.md))
- Synthes (USA) LCP® Proximal Humerus Plates, long ([K041860](/device/K041860.md))
- Synthes LCP Proximal Humerus Plates ([K011815](/device/K011815.md))
- DePuy Anatomic Locked Plating System Long Plate Line Extension ([K101421](/device/K101421.md))
- Synthes (USA) Small Fragment Dynamic Compression Locking (DCL) System ([K000684](/device/K000684.md))
- Synthes One-Third Tubular Plate ([K011335](/device/K011335.md))
- Synthes (USA) Wrist Fusion Plates (WFP) ([K000558](/device/K000558.md))
- Synthes Calcaneal Plate ([K020401](/device/K020401.md))
- Synthes Pilon Plate ([K020602](/device/K020602.md))
- Synthes (USA) 4.5 mm Titanium LCP Proximal Tibia Plating System ([K023802](/device/K023802.md))
- Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking Compression Plates (LCP) ([K092609](/device/K092609.md))
- Synthes (USA) Large Fragment Locking Compression Plate (LCP) System-T Plate ([K010766](/device/K010766.md))
- Synthes (USA) TomoFix™ Medial Distal Femur Plates ([K081353](/device/K081353.md))
- Synthes (USA) TomoFix™ Osteotomy System ([K023941](/device/K023941.md))
- Synthes 4.0 and 5.0 mm Locking Screws ([K000089](/device/K000089.md))
- Peri-Prosthetic Screws ([K041533](/device/K041533.md))
- Synthes (USA) 3.5 mm LCP Distal Humerus System ([K033995](/device/K033995.md))
- Synthes (USA) 3.5 mm Cortex Screws ([K043185](/device/K043185.md))
- Synthes Anatomical Locking Plate System ([K961413](/device/K961413.md))
- Synthes (USA) 6.5 mm Cancellous Screws ([K061621](/device/K061621.md))
- Synthes 3.5 mm Low Profile Pelvic Reconstruction Plate ([K031573](/device/K031573.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

(1) 3)

510(k) Summary K113733

## MAR 2 3 2012

# 510(k) Summary

## NEOORTHO Produtos Ortopédicos S/A

Small and Large Fragments Osteosynthesis System NEOFIX

#### K113733

February 22, 2012

#### ADMINISTRATIVE INFORMATION -

Manufacturer Name:

NEOORTHO Produtos Ortopédicos S/A Rua Ângelo Domingos Durigan 607, Cascatinha 82020-340 Curitiba - PR, Brazil +55 41 3535-1033 Telephone: Fax: +55 41 3535-1018

Official Contact:

Representative/Consultant:

Luciane Yumi Suzuki de Oliveira NEOORTHO Produtos Ortopédicos S/A

Kevin A. Thomas, PhD Linda K. Schulz, BSDH, RDH PaxMed International, LLC 11234 El Camino Real, Suite 200 San Diego, CA 92130 USA Telephone: +1 (858) 792-1235 Fax: +1 (858) 792-1236 kthomas@paxmed.com Email: lschulz@paxmed.com

### DEVICE NAME AND CLASSIFICATION

| Trade/Proprietary Name:     | Small and Large Fragments Osteosynthesis System NEOFIX                                                                               |                           |
|-----------------------------|--------------------------------------------------------------------------------------------------------------------------------------|---------------------------|
| Common Names:               | Plate, fixation, bone<br>Screw, fixation, bone                                                                                       |                           |
| Classification Names:       | Single/multiple component metallic bone fixation appliances and<br>accessories<br>Smooth or threaded metallic bone fixation fastener |                           |
| Classification Regulations: | 21 CFR 888.3030, Class II                                                                                                            | 21 CFR 888.3040, Class II |
| Product Codes:              | HRS                                                                                                                                  | HWC                       |
| Classification Panel:       | Orthopedic Products Panel                                                                                                            |                           |
| Reviewing Branch:           | Orthopedic Devices Branch                                                                                                            |                           |

{1}------------------------------------------------

2/3

#### INTENDED USE

Small and Large Fragments Osteosynthesis System NEOFIX is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, carpais, metacarpals, distal femur, proximal tibia, tibial pilon, fibula, calcaneus, tarsals and metatarsals; small fragments of the hand and wrist; pelvis and acetabulum fractures; periprosthetic fractures; metatarsal and phalangeal osteotomies; and carpal, metacarpal, wrist and ankle arthrodesis. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.

#### DEVICE DESCRIPTION

NEOFIX consists of plates and screws in a variety of designs and sizes and made from commercially pure titanium or Ti-6Al-4V alloy. Plates are provided in straight designs and in various geometric configurations that are commonly used in trauma and reconstructive surgery. Plates are provided with screw holes to accommodate non-locking and locking screws designs. Screws are provided in cortical (locking and non-locking) and cancellous thread designs in various diameters and lengths.

### EQUIVALENCE TO MARKETED DEVICE

NEOORTHO Produtos Ortopédicos S/A demonstrated that, for the purposes of FDA 's regulation of medical devices. Small and Large Fragments Osteosynthesis System NEOFIX is substantially equivalent in indications and design principles to the following predicate devices.

| Predicate<br>Number | K Number | Company Name            | Device Name                                                                           |
|---------------------|----------|-------------------------|---------------------------------------------------------------------------------------|
| 1                   | K012655  | Acumed, Inc.            | Congruent Bone Plate System                                                           |
| 2                   | K071715  | Acumed, Inc             | Acumed Congruent Bone Plate System                                                    |
| 3                   | K041860  | Synthes (USA)           | Synthes (USA) LCP® Proximal Humerus Plates, long                                      |
| 4                   | K011815  | Synthes (USA)           | Synthes LCP Proximal Humerus Plates                                                   |
| 5                   | K101421  | DePuy Orthopaedics, Inc | DePuy Anatomic Locked Plating System Long Plate Line Extension                        |
| 6                   | K000684  | Synthes (USA)           | Synthes (USA) Small Fragment Dynamic Compression Locking (DCL)<br>System              |
| 7                   | K011335  | Synthes (USA)           | Synthes One-Third Tubular Plate                                                       |
| 8                   | K000558  | Synthes (USA)           | Synthes (USA) Wrist Fusion Plates (WFP)                                               |
| 9                   | K020401  | Synthes (USA),          | Synthes Calcaneal Plate                                                               |
| 10                  | K020602  | Synthes (USA)           | Synthes Pilon Plate                                                                   |
| 11                  | K023802  | Synthes (USA)           | Synthes (USA) 4.5 mm Titanium LCP Proximal Tibia Plating System                       |
| 12                  | K092609  | Synthes (USA)           | Synthes 3.5 mm and 4.5 mm Curved Narrow and Broad Locking<br>Compression Plates (LCP) |
| 13                  | K010766  | Synthes (USA)           | Synthes (USA) Large Fragment Locking Compression Plate (LCP)<br>System-T Plate        |
| 14                  | K081353  | Synthes (USA)           | Synthes (USA) TomoFix™ Medial Distal Femur Plates                                     |

{2}------------------------------------------------

3 3)

| Predicate<br>Number | K Number | Company Name  | Device Name                                            |
|---------------------|----------|---------------|--------------------------------------------------------|
| 15                  | K023941  | Synthes (USA) | Synthes (USA) TomoFix™ Osteotomy System                |
| 16                  | K000089  | Synthes (USA) | Synthes 4.0 and 5.0 mm Locking Screws                  |
| 17                  | K041533  | Synthes (USA) | Peri-Prosthetic Screws                                 |
| 18                  | K033995  | Synthes (USA) | Synthes (USA) 3.5 mm LCP Distal Humerus System         |
| 19                  | K043185  | Synthes (USA) | Synthes (USA) 3.5 mm Cortex Screws                     |
| 20                  | K961413  | Synthes (USA) | Synthes Anatomical Locking Plate System                |
| 21                  | K061621  | Synthes (USA) | Synthes (USA) 6.5 mm Cancellous Screws                 |
| 22                  | K031573  | Synthes (USA) | Synthes 3.5 mm Low Profile Pelvic Reconstruction Plate |

The subject device and the predicate devices have the same intended use and have the same technological characteristics. The subject and predicate devices are all fabricated from the same or similar materials and share similar design characteristics. The subject and predicate devices encompass the same range of physical dimensions, are packaged using the same materials, and are to be sterilized by the same methods. Any differences in the technological characteristics do not raise new issues of safety or efficacy.

Performance data provided to demonstrate substantial equivalence included engineering analysis and mechanical testing according to ASTM F382 and ASTM F543.

Overall, Small and Large Fragments Osteosynthesis System NEOFIX has the following similarities to the predicate devices:

- has the same intended use, .
- . uses the same operating principle,
- incorporates the same basic design, .
- incorporates the same or very similar materials, and ●
- has similar packaging and is sterilized using the same materials and processes. .

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. The logo is black and white.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

NEOOTHO Produtos Orthopédicos S/A % PaxMed International, LLC Kevin A. Thomas, Ph.D. 11234 El Camino Real, Suite 200 San Diego, California 92130

MAR 2 3 2012

Re: K113733

Trade/Device Name: Internal Fixation System NEOFIX Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: March 9, 2012

Received: March 12, 2012

Dear Dr. Thomas:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{4}------------------------------------------------

#### Page 2 - Kevin A. Thomas, Ph.D.

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

510(k) Summary K113733

Small and Large Fragments Osteosynthesis System NEOFIX

### Indications for Use

K113733 510(k) Number:

Device Name:

Internal Fixation System NEOFIX

Indications for Use:

Small and Large Fragments Osteosynthesis System NEOFIX is intended for small and large bone fracture fixation, arthrodesis and osteotomy fixation. Examples include: fractures of the clavicle, scapula, humerus, olecranon, radius, ulna, carpals, metacarpals, distal femur, proximal tibia, tibial pilon, fibula, calcaneus, tarsals and metatarsals; small fragments of the hand and wrist; pelvis and acetabulum fractures; periprosthetic fractures; metatarsal and phalangeal osteotomies; and carpal, metacarpal, wrist and ankle arthrodesis. The use of locking plate/screw systems is suited for treatment of fractures in osteopenic bone. This system is not indicated for use in the spine.

Prescription Use X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

111

Duyal

(Division Sign-Off) Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K113733

---

**Source:** [https://fda.innolitics.com/device/K113733](https://fda.innolitics.com/device/K113733)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
