INVIA MOTION NEGATIVE PRESSURE WOUND THERAPY SYSTEM
K113678 · Medela AG · OMP · Oct 1, 2012 · General, Plastic Surgery
Device Facts
Record ID
K113678
Device Name
INVIA MOTION NEGATIVE PRESSURE WOUND THERAPY SYSTEM
Applicant
Medela AG
Product Code
OMP · General, Plastic Surgery
Decision Date
Oct 1, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4780
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The portable Medela® Invia Motion negative pressure wound therapy (NPWT) system is indicated to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. It is intended for the use in hospitals, clinics, Long Term Care (LTC) and Home Care (HC) settings on adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (such as diabetic, neuropathic, pressure or venous insufficiency), flaps and grafts.
Device Story
Invia Motion is a portable, single-patient-use negative pressure wound therapy (NPWT) suction pump; 60-day useful life. System includes pump, canister/tubing set, power supply, and carrying case. Operates in continuous or intermittent modes; powered by rechargeable battery or electrical supply. Used in hospitals, clinics, LTC, and home settings by medical personnel or trained lay users. Device monitors pressure; triggers acoustic and optical signals for faults or deviations from set values. Healthcare providers use these signals to troubleshoot and maintain therapy. Promotes wound healing by removing exudates/infectious material, reducing edema, and preparing wound bed for closure.
Clinical Evidence
No clinical testing performed. Evidence consists of a clinical evaluation of published literature and bench testing. Bench testing compared performance (pressure maintenance and fluid removal) against predicate devices. Usability study (n=30; 15 HCPs, 15 lay users) demonstrated 100% success in target tasks, including routine maintenance and alarm troubleshooting.
Technological Characteristics
Powered suction pump; rechargeable battery or AC power. Portable form factor. Single-patient use; 60-day life. Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-1-11 (home healthcare), and ISO 10993-1 (biocompatibility). Features optical and acoustic status display for pressure monitoring.
Indications for Use
Indicated for adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (diabetic, neuropathic, pressure, venous insufficiency), flaps, and grafts. Users must have normal visual acuity and hearing.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
KCI USA, Inc. V.A.C.Via Negative Pressure Wound Therapy System (K093526)
Submission Summary (Full Text)
{0}------------------------------------------------
K113678 Page 1/3
Medela AG Invia Motion Negative Pressure Wound Therapy System
# 510(k) Summary
1
20,
OCT
In accordance with 21 CFR 807.92 the following summary of information is provided:
DATE: August 30, 2012
# SUBMITTER:
Medela AG Lättichstrasse 4b 6341 Baar / Switzerland Phone +41 (0)41 769 52 47
#### PRIMARY CONTACT PERSON:
Adrienne Lenz, RAC Member Pathway Regulatory Consulting, LLC T 262-290-0023
#### SECONDARY CONTACT PERSON:
Markus Bütler Vice President Quality Management and Regulatory Affairs Medela AG
#### DEVICE:
TRADE NAME: Invia Motion
COMMON/USUAL NAME: Negative Pressure Wound Therapy System
CLASSIFICATION NAMES: 878.4780 Powered Suction Pump
PRODUCT CODE: OMP
PREDICATE DEVICE(S):
K080357 Medela Invia Wound Therapy
K083375 Smith and Nephew Renasys Go
K093526 KCI USA, Inc. V.A.C.Via Negative Pressure Wound Therapy System
{1}------------------------------------------------
K113678 Page 2/3
# Medela AG Invia Motion Negative Pressure Wound Therapy System
### DEVICE DESCRIPTION:
The Medela Invia Motion Negative Pressure Wound Therapy (NPWT) System is comprised of the Invia Motion NPWT Pump, canister/tubing set, power supply, carrying case, patient and user instructions, and Invia NPWT kits. The INVIA Motion is also compatible with Avance NPWT kits manufactured by Mölnlyke Healthcare.
Invia Motion NPWT pump is a suction pump for Negative Pressure Wound Therapy with an optical and acoustic status display. Invia Motion NPWT pump is a single patient use pump for continuous or intermittent operation and has a lifetime of 60 days.
Invia Motion NPWT pump is portable and can be operated independent of the electrical Invia Motion power supply due to a rechargeable battery. Acoustic and optical signals are triggered for variances from the set values as well as for faults.
Invia Motion NPWT system is intended for use in a home or other health care facility by medical personnel or trained lay users adhering to the instructions for use. The user may not be hard of hearing or deaf and must have normal visual acuity.
### INTENDED USE:
The portable Medela® Invia Motion negative pressure wound therapy (NPWT) system is indicated to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. It is intended for the use in hospitals, clinics, Long Term Care (LTC) and Home Care (HC) settings on adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (such as diabetic, neuropathic, pressure or venous insufficiency), flaps and grafts.
#### DETERMINATION OF SUBSTANTIAL EQUIVALENCE:
SUMMARY OF TECHNOLOGICAL CHARACTERISTICS OF THE DEVICE COMPARED TO THE PREDICATE DEVICE
The Invia Motion Negative Pressure Wound Therapy System uses the same fundamental technology as the Invia Wound Therapy system for most features. The indications for use of Invia Motion are the same as KCI's V.A.C Via NPWT System. Invia Motion is similar to the predicate devices in its indications for use, contra-indications and patient population. The user interface is also similar to the predicated devices. The main difference is in the volume of the canister, which is reduced but still sufficient for the expected daily volumes. The smaller size improves mobility. The pump is also intended for a single patient and has a useful life of 60 days, which is sufficient for single patient use.
#### SUMMARY OF NON-CLINICAL TESTS:
The Invia Motion Negative Pressure Wound Therapy System complies with voluntary standards for electrical safety, electromagnetic compatibility, biocompatibility, sterilization and use in the
{2}------------------------------------------------
K 113678 Page 3/3
# Medela AG Invia Motion Negative Pressure Wound Therapy System
home healthcare environment. The following quality assurance measures were applied to the development of the system:
- Risk Analvsis .
- Software Validation .
- Electrical safety and electromagnetic compatibility testing per IEC 60601-1 and IEC . 60601-1-2 standards, respectively
- . Testing for use in the home healthcare environment per IEC 60601-1-11
- Biocompatibility testing per ISO 10993-1 standards .
- Sterilization validation .
- . Usability testing was conducted and demonstrated that the production versions of the labeling and device design provided first-time Invia® Motion™ NPWT System end-users with clear, concise instructions and the ability to avoid patterns of preventable use errors or difficulties when operating the device that could result in harms to the product end-users. Fifteen (15) Health Care Professionals (HCP) and 15 patient lay users (LU) were enrolled to participate in the study, which was conducted in the United States with English speakers. Representative product training for their user-population was provided and a training decay time was included prior to the evaluation. All 15 LU endusers (100%) successfully completed their target population use tasks by performing routine maintenance of the device and identifying and troubleshooting device acoustic and optical display signals to maintain safe and effective performance of the device. All close-calls and use-errors observed during the HCP use-scenarios and sub-tasks were resolved during the evaluation and did not affect patient safety.
- Bench Testing demonstrated that the design specifications were met, including device . performance and reliability. Comparison of performance of the Invia Motion with its compatible NPWT kits to the predicate device concluded that the Invia Motion is equivalent in maintaining set pressures across the specification range and in removing fluids in simulated wounds.
#### SUMMARY OF CLINICAL TESTS:
The Invia Motion Negative Pressure Wound Therapy System has not been the subject of clinical testing. A clinical evaluation of published literature has been conducted for Invia Motion.
#### CONCLUSION:
Medela AG considers the Invia Motion Negative Pressure Wound Therapy System to be as safe, as effective, and substantially equivalent to the predicate devices.
{3}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features a stylized caduceus-like symbol, which is a staff entwined with a serpent, representing medicine and health. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" are arranged in a circular pattern around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Medela AG % Pathway Regulatory Consulting, LLC Ms. Adrienne Lenz, RAC W324S3649 County Road E Dousman, Wisconsin 53118
OCT
1 2012
Re: K113678
Trade/Device Name: Invia Motion Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 · Regulation Name: Powered suction pump Regulatory Class: II Product Code: OMP Dated: August 30, 2012 Received: September 04, 2012
Dear Ms. Lenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
# Page 2 - Ms. Adrienne Lenz, RAC
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark A. Millerson
Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known): K113678
Invia Motion Negative Pressure Wound Therapy System Device Name:
Indications for Use:
Invia Motion NPWT System:
The portable Medela® Invia Motion negative pressure wound therapy (NPWT) system is indicated to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudates and infectious material. It is intended for the use in hospitals, clinics, Long Term Care (LTC) and Home Care (HC) settings on adult patients with chronic, acute, subacute, traumatic, dehisced wounds, partial-thickness burns, ulcers (such as diabetic, neuropathic, pressure or venous insufficiency), flaps and grafts.
Prescription Use X
AND/OR
Over-The-Counter Use__
(Part 21 CFR 801 Subpart D)
(Part 21 CFR 801 Subpart C) .
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Daniel Krone for MXM
(Division Sign Off)
Division of Surgical, Orthopedic, and Restorative Devices
510(k) Number K113678
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.