← Product Code [MHX](/productcode/MHX) · K113623

# PATIENT MONITOR (K113623)

_Edan Instruments, Inc. · MHX · Feb 1, 2012 · Cardiovascular · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113623

## Device Facts

- **Applicant:** Edan Instruments, Inc.
- **Product Code:** [MHX](/productcode/MHX.md)
- **Decision Date:** Feb 1, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1025
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Pediatric

## Indications for Use

iM80 : The monitor monitors parameters such as ECG (3-lead, 5-lead, 12-lead selectable), Respiration (RESP), Functional arterial oxygen saturation (SpO2), Invasive or noninvasive blood pressure (2/4 channels IBP NIBP), Cardiac Output (CO), Temperature (dual-TEMP), Expired CO2 and Anesthetic gas (AG). The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both. iM50 : The monitor monitors parameters such as ECG (3-lead, 5-lead selectable), Respiration (RESP), Functional arterial oxygen saturation (SpO2), Invasive or noninvasive blood pressure (dual-IBP, NIBP), Temperature (dual-TEMP), Expired CO2 and Quick Temperature ( Quick TEMP. The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both. The arrhythmia detection and ST Segment analysis are not intended for neonatal patients.

## Device Story

iM50/iM80 patient monitors provide continuous monitoring of physiological parameters including ECG, SpO2, NIBP, IBP, RESP, TEMP, CO2, and AG (iM80 only). Devices process input signals from patient-connected sensors/modules; display real-time waveforms and numerical data on integrated touch screens (8.4-inch for iM50; 15-inch for iM80). Systems perform automated analysis of ECG for arrhythmia and ST segment changes; trigger visual/audible alarms for system faults or physiological limit violations. Used by clinicians in hospital environments to facilitate timely patient assessment and decision-making. iM80 supports networking and online software upgrades. Benefits include continuous oversight of patient status and early detection of abnormal physiological events.

## Clinical Evidence

No clinical data provided. Evidence consists of bench testing, including software validation, hardware testing, safety testing, environmental testing, and risk analysis.

## Technological Characteristics

Class II patient monitors. iM50 (8.4-inch) and iM80 (15-inch) touch screen displays. Modular design supporting SpO2, NIBP, ECG, RESP, TEMP, CO2, IBP, C.O., and AG. Connectivity includes networking and online software upgrades. Software-based arrhythmia and ST segment analysis. Sterilization not applicable (external use).

## Regulatory Identification

The arrhythmia detector and alarm device monitors an electrocardiogram and is designed to produce a visible or audible signal or alarm when atrial or ventricular arrhythmia, such as premature contraction or ventricular fibrillation, occurs.

## Special Controls

*Classification.* Class II (special controls). The guidance document entitled “Class II Special Controls Guidance Document: Arrhythmia Detector and Alarm” will serve as the special control. See § 870.1 for the availability of this guidance document.

## Predicate Devices

- MP5 ([K100939](/device/K100939.md))
- M3 and M3A ([K102825](/device/K102825.md))
- MP70 ([K100939](/device/K100939.md))
- M3B ([K083821](/device/K083821.md))
- ZOLL M series NIBP option ([K032363](/device/K032363.md))
- BX-10 ([K032857](/device/K032857.md))
- Carbon-dioxide analyzer ([K103604](/device/K103604.md))

## Reference Devices

- PC ECG ([K102854](/device/K102854.md))
- PC ECG ([K092010](/device/K092010.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K113623

Patient Monitor Traditional 510K Submission

Section 1

# 510(k) Summary of Safety and Effectiveness

This summary of 510k safety and effectiveness is being submitted in according with 21CFR part 807.92

| Submitter: | Edan Instruments, Inc<br>3/F - B, Nanshan Medical<br>Equipments Park, Nanhai Rd 1019#,<br>shekou, Nanshan Shenzhen,<br>518067 P.R. China<br>Tel: 86-755-26882220<br>Fax:86-755-26882223<br>Contact person: Randy Jiang |
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|

2011-12-02 Date of Preparation:

#### Patient Monitor (Models iM50 and iM80 ) Proprietary Name:

### Classification:

| Description                                                                        | Classification | Product<br>code |
|------------------------------------------------------------------------------------|----------------|-----------------|
| 21 CFR 870.2300 Cardiac monitor (including cardiotachoment and rate alarm)         | II             | DRT             |
| 21 CFR 870.1130 Non-Invasive blood pressure measurement System                     | II             | DXN             |
| 21 CFR 870.1110 Blood pressure computer                                            | II             | DSK             |
| 21 CFR 880.2910 Clinical Electronic Thermometers-Temperature Monitor with<br>Probe | II             | FLL             |
| 21 CFR 870.2700 Oximeter, Pulse                                                    | II             | DQA             |
| 21 CFR 870.1400 Carbon Dioxide Gas Analyzer                                        | II             | CCK             |
| 21 CFR 868.1500 Enflurane gas analyzer                                             | II             | CBQ             |
| 21 CFR 868.1620 Halothane gas analyzer                                             | II             | CBS             |
| 21 CFR 868.1700 Nitrous Oxide gas analyzer                                         | II             | CBR             |
| 21 CFR 868.1720 Oxygen gas analyzer                                                | II             | CCL             |
| 21 CFR 868.2900 cable, transducer and electrode, patient, (including connector)    | II             | DSA             |
| 21 CFR 870.2300 monitor, cardiac (incl. cardiotachometer & rate alarm)             | II             | DRT             |
| 21 CFR 870.1025 Detector and Alarm, Arrhythmia                                     | II             | DSI             |
| 21 CFR 870.1025 Monitor, ST Segment with Alarm                                     | II             | MLD             |

Regulatory Class:_ Class II

Page 1 OF 4

{1}------------------------------------------------

### Legally Marketed Predicate Devices::

| Manufacturer                   | Predicate Device             | 510(k) number | Cleared date   |
|--------------------------------|------------------------------|---------------|----------------|
| Philips Medical System         | MP5                          | K100939       | April.1.2010   |
| Edan Instruments, Inc          | M3 and M3A                   | K102825       | Dec.27.2010    |
| Philips Medical System         | MP70                         | K100939       | April.1.2010   |
| Edan Instruments, Inc          | M3B                          | K083821       | May.14,2009    |
| ZOLL Medical Corporation       | ZOLL M series NIBP<br>option | K032363       | July 30, 2003  |
| Colin Medical Instruments Corp | BX-10                        | K032857       | April.21,2003  |
| PHASEIN AB                     | Carbon-dioxide<br>analyzer   | K103604       | March.28, 2011 |

iM50 /iM80 Patient Monitor provides the following primary features: Device Description:

> iM50 /iM80 Patient Monitor can perform long-time continuous monitoring of multiple physiological parameters. Also, it is capable of storing, displaying, analyzing and controlling measurements, and it will indicate alarms in case of abnormity so that doctors and nurses can deal with them in time.

> iM50 Patient Monitor can monitor parameters including SpO2, NIBP, ECG, RESP, TEMP, CO2, IBP.

> iM80 Patient Monitor can monitor parameters including SpO2, NIBP, ECG, RESP, TEMP, CO2, IBP, C.O. and AG.

> The above is the maximum configuration for iM50 and iM80, the user may select different monitoring parameters in according with the requirement.

> iM50 is outfitted with a 8.4-inch display screen, iM80 is 15-inch, as well as an equal large touch screen, which enables the operation by touching the screen, thus offering convenience for doctors and nurses. Besides, iM80 supports software upgrade online and networking.

> iM50 Patient Monitor has parameter modules including SpO2 (pulse oxygen saturation, pulse rate and SpO2 plethysmogram) with EDAN SpO2 module or Nellcor SPO2 module, NIBP (systolic pressure, diastolic pressure, mean pressure and pulse rate), TEMP, ECG, RESP, CO2, IBP and Quick Temp.

> iM80 could be configured with three different NIBP modules, one is EDAN NIBP module, one is Omron NIBP module, the other is Suntech NIBP module; Omron NIBP module used in iM80 is the same to that used in BX-10, which has been cleared by K032857 in April.21.2003. Suntech NIBP module is cleared by FDA in K032363

> CO2 module and Nellcor SpO2 module used in iM50 and iM80 are the same to those used in M3B, which has been cleared by K083821 in

> > PAGE 2 OF 4

{2}------------------------------------------------

May 14, 2009.

Arrhythmia and ST segment Analysis used for ECG diagnoses in iM80 is the same to that in PC ECG which has been cleared by FDA under K102854 and K092010

#### Comparison with predicate device

The iM50 and iM80 Patient Monitors have the following similarities to that which previously received 510(k) concurrence:

- · have the same indications for use,
- · use the similar operating principle,
- · have the same or similar performance specifications

In summary, the iM50 and iM80 Patient Monitor described in this submission is, in our opinion, substantially equivalent to the predicate device

### Intended Use:

#### iM80 :

The monitor monitors parameters such as ECG (3-lead, 5-lead, 12-lead selectable), Respiration (RESP), Functional arterial oxygen saturation (SpO2), Invasive or noninvasive blood pressure (2/4 channels IBP NIBP), Cardiac Output (CO), Temperature (dual-TEMP), Expired CO2 and Anesthetic gas (AG). The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both.

iM50 :

The monitor monitors parameters such as ECG (3-lead, 5-lead selectable), Respiration (RESP), Functional arterial oxygen saturation (SpO2), Invasive or noninvasive blood pressure (dual-IBP, NIBP), Temperature (dual-TEMP), Expired CO2 and Quick Temperature

( Quick TEMP. The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both.

The arrhythmia detection and ST Segment analysis are not intended for neonatal patients.

Contraindications:

It is not intended for use in patient's home or residence, or when it has not been ordered by a physician.

Test Summary:

The following quality assurance measures were applied to the development of the Patient Monitor

- · Software testing
- · Hardware testing

PAGE 3 OF 4

{3}------------------------------------------------

- Safety testing
- Environment test
- Risk analysis
- Final validation

### Conclusion:

Verification and validation testing was done on the Patient Monitor. This premarket notification submission demonstrates that Patient Monitor is substantially equivalent to the predicate device.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

FEB - 1 2012

Edan Instruments, Inc. c/o Mr. Randy Jiang Certification Engineer 3/F-B, Nanshan Medical Equipment Park, Nanhai Rd., 1019 No. Shenzhen CHINA 518067

Re: K113623

Trade/Device Name: Patient Monitor, Models iM50 and iM80 Regulation Number: 21 CFR 870.1025 Regulation Name: Patient Physiological Monitor (with arrhythmia detection or alarms) Regulatory Class: Class II (two) Product Codes: MHX, DXN, DRT, DSK, FLL, DQA, CCK, CBQ, CBS, CBR, CCL, DSA, DRT, DSI and MLD Dated: November 2, 2011

Received: December 3, 2011

Déar Mr. Jiang:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

#### Page 2 - Mr. Randy Jiang

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

### Indication for Use

510(k) Number (if known):

## Device Name: Patient Monitor Models iM50 and iM80

iM80 :

The monitor monitors parameters such as ECG (3-lead, 12-lead selectable), Respiration (RESP), Functional arterial oxygen saturation (SpO2), Invasive or noninvasive blood pressure (2/4 channels IBP NIBP), Cardiac Output (CO), Temperature (dual-TEMP), Expired CO2 and Anesthetic gas (AG). The monitor is equipped with alarms that indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both.

iM50 :

The monitor monitors parameters such as ECG (3-lead, 5-lead selectable), Respiration (RESP), Functional arterial oxygen saturation (SpO2), Invasive or noninvasive blood pressure (dual-IBP, NIBP), Temperature (dual-TEMP), Expired CO2 and Quick Temperature indicate system faults (such as loose or defective electrodes), physiologic parameters that have exceeded the limits set by the operator, or both.

The arrhythmia detection and ST Segment analysis are not intended for neonatal patients.

× Prescription Use (21 CFR Part 801 Subpart D) And/Or Over the Counter Use (21 CFR Part 801 Subpart C)

# Concurrence of CDRH, Office of Device Evaluation (ODE)

sion of Cardiovascular Devices

510(k) Number K113623

---

**Source:** [https://fda.innolitics.com/device/K113623](https://fda.innolitics.com/device/K113623)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
