← Product Code [EBF](/productcode/EBF) · K113617

# SPEE-DEE PATCH (K113617)

_Pulpdent Corporation · EBF · Mar 2, 2012 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113617

## Device Facts

- **Applicant:** Pulpdent Corporation
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** Mar 2, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental

## Indications for Use

Spee-Dee Patch is a visible light cured, glass-filled resin that is intended as a vital tooth build-up material where restorations were removed and must be built up for crown preparation; as a non-vital tooth post and core; as a build up material; and as a repair material to repair fractured cusps, lost partial or whole fillings, chipped crowns, broken teeth, crowns, or bridges.

## Device Story

Spee-Dee Patch is a visible light-cured, glass-filled resin dental material; used by dentists in clinical settings for tooth build-up and repair; applied to tooth structures where restorations were removed or damage occurred; functions as a restorative material for crown preparation, post and core, and repair of fractured dental components; provides structural support for subsequent dental restorations; benefits patients by restoring tooth integrity and function; clinician applies material to site and cures with visible light; output is a hardened resin structure integrated into the tooth.

## Clinical Evidence

Bench testing only. Performance metrics include: light cure setting time < 20 seconds; working time > 5 minutes; clinical depth of cure > 4 mm; compressive strength 312.0 ± 24.0 MPa; flexural strength 95.0 ± 10.0 MPa; diametral tensile strength 52.0 ± 2.0 MPa; specific gravity 1.520; shelf-life 2 years.

## Technological Characteristics

Visible light-cured, resin-based flowable material. Composition: Embrace monomer mix, photo-initiator, barium glass, and submicron silica filler. Bisphenol A-free. Form factor: flowable resin. Energy source: visible light for polymerization. Sterilization: not specified.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- Pulpdent Embrace™ Dual Cure Composite Materials ([K071278](/device/K071278.md))
- Pulpdent Embrace™ Pit and Fissure Sealant ([K020287](/device/K020287.md))
- Kuraray Clearfil™ Photo Core ([K882006](/device/K882006.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K 113617

# PULPDENT CORPORATION

80 Qakland Street Watertown, MA 02472 USA

MAR - 2 2012

Tel: 617 926 6666 Fax: 617 926 6262 Email: Pulpdent@pulpdent.com

## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS DATA

Kenneth J. Berk 80 Oakland Street Watertown, MA 02472 USA Telephone: 617-926-6666 Fax: 617-926-6262 Email: ken@pulpdent.com

#### DEVICE:

Trade Name: Pulpdent Spee-Dee™ Patch Classification Name: Tooth shade resin material FDA Product Code: 76 EBF, 21 CFR Part 872,3690

#### PREDICATE DEVICES:

Pulpdent Embrace™Dual Cure Composite Materials Pulpdent Embrace™ Pit and Fissure Sealant Kuraray Clearfil™ Photo Core

#### DESCRIPTION AND INTENDED USE:

Spee-Dee Patch is a visible light cured, glass-filled resin that is intended as a vital tooth build-up material where restorations were removed and must be built up for crown preparation; as a non-vital tooth post and core; as a build up material; and as a repair material to repair fractured cusps, lost partial or whole fillings, chipped crowns, broken teeth, crowns, or bridges.

#### COMPARISON WITH PREDICATE PRODUCTS:

Spee-Dee Patch is substantially equivalent in design, composition, performance, intended use, safety and effectiveness to the predicate products listed above. Spee-Dee Patch contains the same Embrace technology and resin composition as Embrace Pit and Fissure Sealant and Embrace Dual Cure Composite Materials (marketed as Spee-Dee™ Build-Up), except that Spee-Dee has dual cure chemistry.

| Pulpdent Spee-Dee<br>Patch                             | Visible light-cure, resin-based,<br>flowable material that<br>contains no Bisphenol A.    | Used to patch teeth or core<br>preparations and as a core<br>build-up material. | Embrace Monomer Mix<br>Photo-initiator<br>Filler: Barium glass and submicron silica                       |
|--------------------------------------------------------|-------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------|
| Pulpdent Spee-Dee<br>Build-Up Material<br>K 071278     | Dual-cure, resin-based<br>material that contains no<br>Bisphenol A.                       | Build-up and repair material.                                                   | Embrace Monomer Mix<br>Photo-initiator<br>Chemical initiator<br>Filler: Barium glass and submicron silica |
| Pulpdent Embrace™<br>Pit & Fissure Sealant<br>K 020287 | Hydrophilic, visible light-cure,<br>resin-based material that<br>contains no Bisphenol A. | Seals the pits and fissures in<br>teeth.                                        | Embrace Monomer Mix<br>Photo-initiator<br>Filler: Barium glass and submicron silica                       |
| Kuraray Clearfil™<br>Photo Core<br>K882006             | Light cured, resin-based<br>composite.                                                    | Build-up and repair material                                                    | Monomer Mix<br>Photo-initiator<br>Filler                                                                  |

{1}------------------------------------------------

## PULPDENT CORPORATION

0 Oakland Street Watertown, MA 02472 IISA

#### Summary of Performance Testing - Bench

The following test results demonstrate that Spee-Dee Patch performs as intended:

| Light Cure Setting Time       | < 20 seconds                            |
|-------------------------------|-----------------------------------------|
| Working time in ambient light | > 5 minutes                             |
| Clinical depth of cure        | > 4 mm                                  |
| Compressive Strength          | 45,240 ± 3480 p.s.i. / 312.0 ± 24.0 MPa |
| Specific gravity              | 1.520                                   |
| Flexural Strength             | 13,775 ± 1450 p.s.i. / 95.0 ± 10.0 MPa  |
| Diametral Tensile Strength    | 7540 ± 290 p.s.i. / 52.0 ± 2.0 MPa      |
| Shelf-life                    | Two years                               |

#### SAFETY AND EFFECTIVENESS:

From the above comparisons, the bench testing and the ten years organizational experience with Embrace resins, it can be concluded that Spee-Dee Patch is substantially equivalent in design, composition, performance, intended use, safety and effectiveness to the predicate products listed above.

The predicate products have been found substantially equivalent under the 510(k) Premarket Notification process as Class II Dental Devices under CFR 872.3690, 872.3275 and have been on the market for years with no reports of safety or effectiveness issues.

{2}------------------------------------------------

## DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like figure with three curved lines forming its body and wings. The logo is surrounded by text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" in a circular arrangement.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

Mr. Kenneth J. Berk Director of Research Pulpdent Corporation 80 Oakland Street Watertown, MA 02472

Re: K113617

Trade/Device Name: Spee-Dee™ Patch Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Codes: EBF Dated: December 1, 2011 Received: December 7, 2011

Dear Mr. Berk:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Eederal Register.

MAR - 2 2012

{3}------------------------------------------------

## Page 2 - Mr. Berk:

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Anthony D. Watson, B.S., M.S., M.B.A. Director

Division of Anesthesiology, General Hospital, Infection Control and Dental Devices

Office of Device Evaluation

Center for Devices and Radiological Health

### Enclosure

{4}------------------------------------------------

# INDICATIONS FOR USE

510 (k) Number:

## K113617

Device Name:

# PULPDENT SPEE-DEE™ PATCH

Indications for Use:

Spee-Dee Patch is a visible light cured, glass-filled resin that is intended as a vital tooth build-up material where restorations were removed and must be built up for crown preparation; as a non-vital tooth post and core; as a build up material; and as a repair material to repair fractured cusps, lost partial or whole fillings, chipped crowns, broken teeth, crowns, or bridges.

Prescription Use X or (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 807 Subpart C)

Please do not write below this line. Continue on another page if needed.

Concurrence of CDRH, Office of Device Evaluation (ODE)

Susan Curry

(Division Sign-Üff) Division of Anesthesiology, General Hospital Infection Control, Denial Devices

510(k) Number: K113617

---

**Source:** [https://fda.innolitics.com/device/K113617](https://fda.innolitics.com/device/K113617)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com