Automatically Retractable Safety Syringe (with fixed needle) (5cc/ml 22×1-1/4) is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient.
Device Story
Automatically Retractable Safety Syringe is a single-use, sterile, piston-type syringe with an integrated fixed needle. Device functions like a conventional syringe for fluid injection; features a user-activated spring-loaded mechanism that retracts the contaminated needle into the barrel immediately following injection. Designed to prevent accidental needle sticks during disposal. Used in hospitals, clinics, and laboratories by healthcare professionals. Needle retraction occurs when the plunger is fully depressed. Benefits include reduced risk of healthcare worker exposure to hazardous drugs and protection from sharps injuries.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified through mechanical testing (hub/needle bond strength, needle cover strength, delivery accuracy) and biocompatibility testing per ISO 10993-1.
Technological Characteristics
Piston syringe with integrated fixed needle. Materials: Polypropylene (barrel, plunger, hub), Stainless steel (spring, cannula), Isoprene rubber (piston), Silicon dioxide (O-ring). Components bonded with acrylic glue. Active safety feature: manual spring-activated needle retraction. Conforms to ISO 7886-1:1993 for scale and delivery accuracy. Sterilization: Gamma.
Indications for Use
Indicated for intramuscular and subcutaneous injection of medication into patients.
Regulatory Classification
Identification
A piston syringe is a device intended for medical purposes that consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into, or withdraw fluids from, the body.
{0}------------------------------------------------
K113587 page 1 of
..............................................................................................................................................................................
FEB 2 4 2012
8
# 510(k) Summary
.
| Submitter's References |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Shantou Wealy Medical Instrument Co., Ltd.<br>North Jinhuan Road (Near of Qishan Mid-School)<br>Shantou City, Guangdong Province, China<br>Zip: 515064<br>FDA Registration Number: 3005202235<br>Owner/Operator Number: 9074040<br>Tel:+86-754-88218123<br>Fax:+86-754-82121654 |
| Official Correspondent |
| IRC- Mr. Charles Mack<br>77325 Joyce Way<br>Echo, Oregon 97826<br>Ph. 931-625-4938<br>Fax: 541-376-5063<br>Email: charliemack@irc-us.com |
| Summary Contents |
| Submitted Device Device Description Statement of Intended Use Substantial Equivalence Safety Effectiveness Conclusions |
{1}------------------------------------------------
K113587 page 2 of 8
### 1. Submitted Device
| Trade Name | Automatically Retractable Safety<br>Syringe (with fixed needle)<br>5cc/ml 22×11/4 |
|---------------------|-----------------------------------------------------------------------------------|
| Common Name | Syringe, anti-stick |
| Classification Name | Piston Syringe |
#### 1.1. Predicate/Legally Marketed Device
| Model | Manufacturer | K Number | Submitted Device |
|------------------------------------------------------------------------------------------|--------------------------------------|----------|-------------------------------------------------------------------------------------------|
| Anti-Stick<br>Syringe With<br>Integral<br>Hypodermic<br>Needle<br>Vanishpoint<br>Syringe | Retractable<br>Technologies,<br>Inc. | K970803 | Automatically<br>Retractable<br>Syringe(with<br>needle)<br>5cc/ml 22×11/4<br>Safety fixed |
#### Proposed Indications 1.2.
The proposed indications for the predicate Anti-stick Vanishpoint and the submitted Automatically Retractable Syringe are the same. The design, intended use and construction are exactly the same.
Both the predicate and the submitted device are single-use, anti-stick piston type syringes. The function of the predicate and the submitted device is to provide a safe and reliable method of injecting medication into a patient. Both devices work like a simple piston syringe except for the syringe's ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user.
{2}------------------------------------------------
## 2. Device Description
## 2.1 Description :
The Automatically Retractable Safety Syringe (with fixed needle) 5cc/ml 22×11/4 is a piston syringe. The device is intended for medical purposes and consists of a calibrated hollow barrel and a movable plunger. At one end of the barrel there is a male connector (nozzle) for fitting the female connector (hub) of a hypodermic single lumen needle. The device is used to inject fluids into the body.
Automatically Retractable Safety Syringe (with fixed needle) (5cc/ml 22×11/4") works like a conventional hypodermic syringe except for its ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user. The needle retracting mechanism is activated by a spring action mechanism after injection is completed, while the exposed needle remains safety inside the empty syringe barrel for disposal.
{3}------------------------------------------------
K113587 page 4 of 8
Image /page/3/Figure/1 description: The image shows a diagram of a syringe and a corresponding table that labels the different parts of the syringe. The diagram labels 18 different parts of the syringe with numbers ranging from 1 to 18. The table lists the name of each part, such as 'Leak-proof ring', 'Zero line', 'Capacity graduation', 'Needle', and 'Needle covers', corresponding to the numbers in the diagram.
The plunger rod, barrel, needle hub, needle protector and slider are made from polypropylene. The spring and cannula are made from Stainless steel. The piston is made from isoprene rubber and the piston O-ring is made from silicon dioxide. The components are glued together with acrylic glue. All of these components have been tested to verify biocompatibility.
\$\gamma\$
20
{4}------------------------------------------------
K113587 page 5 of 8
#### Statement of Intended Use 3.
Automatically Retractable Safety Syringe (with fixed needle) (5cc/ml 22×1-1/4) is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient.
#### Substantial Equivalence 4.
The submitted Automatically Retractable Safety Syringe is constructed to perform like the predicate Vanishpoint Syringe. Both devices are simple piston syringes with ability to retract the used needle into the body after use. Both devices are single use devices and constructed in the same manner. The submitted device and predicate have undergone similar safety and performance testing and the submitted device has passed these tests.
#### 5. Differences
Please refer to the tables on the following pages for noted differences.
{5}------------------------------------------------
| Element of comparison | Company | Subject Device | Claimed SE Device |
|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | DA510(K) Number | Shantou Wealy Medical Instrument Co.,Ltd. | Retractable Technologies, Inc. |
| | | N/A | K970803 |
| Device Name | | Automatically Retractable Safety Syringe with Fixed Needle | VanishPoint™ Syringe |
| Volume | | 5ml | 5 ml and 10ml |
| Needle length | | 1-1/4 Inch | 1-1/2 Inch |
| Available needle gauge sizes | | 22G | 20/21/22/25G |
| Intended Use | | Automatically Retractable Safety Syringe (with fixed needle) (5ml/CC 22G×1 1/4") is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient. | The function of the VanishPoint™ Syringe is to provide a safe and reliable method of injecting medication into a patient or withdrawing blood from a patient. The VanishPoint™ Syringe works like a conventional hypodermic syringe except for its ability to retract the contaminated needle inside of the syringe immediately after patient injection. Needle retraction is activated by the syringe user. Because the contaminated needle is automatically withdrawn into the syringe barrel, The syringe user is protected from accidental needle sticks. These accidental needle sticks would occur between removing the needle from the patient and disposing of the syringe in a sharps disposable container. |
| Principle of operation | | The needle is automatically retracted from the patient into the barrel of the syringe when the plunger is fully depressed. Automatically Retractable Safety Syringe reduces the risk of healthcare worker exposure to hazardous drugs during reconstitution, preparation, transport, and disposal. | The needle is automatically retracted from the patient into the barrel of the syringe when the plunger is fully depressed. VanishPoint™ Syringe reduces the risk of healthcare worker exposure to hazardous drugs during reconstitution, preparation, transport, and disposal. |
| Syringe type | | Plunger, anti-stick with hypodermic needle | Plunger, anti-stick with hypodermic needle |
| Sites of use | | Hospitals, clinics, laboratories | Hospitals, clinics, laboratories |
| Safety Features | | Active safety feature, manually activated by users | Active safety feature, manually activated by users |
| Element of<br>comparison | Subject Device | Claimed SE Device | |
| Specific drug use | Conventional drugs | Conventional drugs | |
| Needle tip configuration | Tri-Beveled Tip | Tri-Beveled Tip | |
| Nozzle type | Needle and hub are integral to the syringe, not separable. | Needle and hub are integral to the syringe, not separable. | |
| Barrel marking specs | Scale :conforms to ISO 7886-1:1993/ Corrigendum 1:1995 | Scale :conforms to ISO 7886-1:1993/ Corrigendum 1:1995 | |
| Gradations legibility | 0.2ml | 0.2ml | |
| Needle cover color | colorless | colorless | |
| Lubricant composition | Polydimethylsiloxane (PDMS) | Polydimethylsiloxane (PDMS) | |
| Lubricant amount/cm² | <0.25mg/cm² | <0.25mg/cm² | |
| Barrel transparency | Clear | Clear | |
| Delivery accuracy | conforms to ISO 7886-1:1993/ Corrigendum 1:1995 | conforms to ISO 7886-1:1993/ Corrigendum 1:1995 | |
| Needle cover strength | <15N | <15N | |
| Hub/needle bond<br>strength | 40N | 40N | |
| Biocompatibility | Conform to ISO 10993-1 | Conform to ISO 10993-1 | |
| omparisor<br>Element of | Subject Device | Claimed SE Device | |
| Materials | Plastic parts : Polypropylene<br>Piston : Isoprene rubber<br>Needle : Stainless steel<br>)-ring: Silicon dioxide | Plastic parts : Polypropylene<br>Piston : Isoprene rubber<br>Needle : Stainless steel<br>O-ring: Silicon dioxide | |
| ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Syringe type | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>Antistick syringe<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Antistick syringe<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | |
| Reuse | lon-reusable | Non-reusable | |
| Sterilization | ______________________________________________________________________________________________________________________________________________________________________________ | Gamma Sterilizatio | |
| | | | |
{6}------------------------------------------------
3-
1. f
page 7 of 8
{7}------------------------------------------------
A Production Production Comments of Comments of Comments of Comments of Comments of Comments of Comments of Comments of Comments of Comments of Comments of Comments of Commen
END
K113587 page 8 of 8
{8}------------------------------------------------
Image /page/8/Picture/0 description: The image shows the text "DEPARTMENT OF HEALTH & HUMAN SERVICES" in bold, black font. The text is horizontally oriented and appears to be a title or heading. The words are capitalized and evenly spaced.
Image /page/8/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the department's emblem in the center. The emblem is a stylized representation of a human figure, with three profiles overlapping each other. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room ~WO66-G609 Silver Spring, MD 20993-0002
Shantou Wealy Medical Instrument Company, Limited C/O Mr. Charlie Mack Principal Engineer IRC 77325 Joyce Way Echo, Oregon 97826
FEB 2 4 2012
Re: K113587
Trade/Device Name: Automatically Retractable Safety Syringe (with Fixed Needle) 5ml/CC 22G*1 1/4"
Regulation Number: 21 CFR 880.5860 Regulation Name: Piston Syringe Regulatory Class: II Product Code: MEG Dated: November 27, 2011 Received: December 7, 2011
Dear Mr. Mack:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{9}------------------------------------------------
### Page 2 - Mr. Mack
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
### Enclosure
{10}------------------------------------------------
## Indications for Use
510(k) Number (if known):
Device Name: Automatically Retractable Safety Syringe (with Fixed Needle) 5ml/CC 22G×1 1/4"
Indications For Use:
Automatically Retractable Safety Syringe (with fixed needle) (5cc/ml 22×1-1/4) is a sterile, single-use, disposable and non-reusable, retractable safety syringe which is intended to provide a safe and reliable method for intramuscular and subcutaneous injection of medication into a patient.
AND/OR Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
IPLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page 1 of _
11
\$\frac{l_{i} \mathbb{N}}{(\mathbb{D}i: \mathbb{C}i} for REC Feb 22, 2012
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number: K11358 +
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.