AMEDITECH CHECKCUP MULTI DRUG TEST (FLAT/ROUND CUP)
K113501 · Ameditech, Inc. · DIO · Dec 5, 2011 · Clinical Toxicology
Device Facts
Record ID
K113501
Device Name
AMEDITECH CHECKCUP MULTI DRUG TEST (FLAT/ROUND CUP)
Applicant
Ameditech, Inc.
Product Code
DIO · Clinical Toxicology
Decision Date
Dec 5, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 862.3250
Device Class
Class 2
Intended Use
The CHECKCUP™ Multi-Drug Test is a one-step immunoassay for the qualitative detection of multiple drugs and drug metabolites in human urine at the following cutoff concentrations: [Table of 17 drugs/metabolites with cutoffs]. The CHECKCUP™ Multi-Drug Test consists of two test formats: flat cup and round cup. The configurations of the CHECKCUP™ Multi-Drug Test consist of any combination of the drugs listed above. The CHECKCUP™ is used to obtain a visual, qualitative result and is intended for professional use only. This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmation methods.
Device Story
Lateral flow immunoassay for qualitative detection of drugs/metabolites in human urine. Device consists of round or flat cup containing up to 17 test strips. Urine sample added to cup; results read visually by professional user. Presence of line in test region indicates negative; absence indicates positive. Control line confirms proper test performance. Results available in 1-10 minutes. Used in clinical/professional settings for preliminary screening; requires GC/MS or LC/MS confirmation. Benefits include rapid, multi-drug screening capability.
Clinical Evidence
No clinical studies were conducted. Performance characteristics were established via analytical performance verification and validation testing (bench testing only), leveraging data from the predicate device and K113046.
Technological Characteristics
Lateral flow immunoassay; visual readout. Consists of round or flat plastic cup housing multiple 1-drug or 2-drug test strips. No electronic components, energy source, or software. Connectivity: None. Sterilization: Not specified.
Indications for Use
Indicated for the qualitative detection of drugs of abuse in human urine specimens.
Regulatory Classification
Identification
A cocaine and cocaine metabolite test system is a device intended to measure cocaine and a cocaine metabolite (benzoylecgonine) in serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of cocaine use or overdose.
Special Controls
*Classification.* Class II (special controls). A cocaine and cocaine metabolite test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).
K113046 — AMEDITECH IMMUTEST MILTI-DRUG SCREEN · Ameditech, Inc. · Nov 10, 2011
K111322 — CHEMTRUE SINGLE/MULTI-PANEL DRUG SCREEN CASSETTE/DIPCARD · Chemtron Biotech, Inc. · Sep 14, 2011
K163704 — Healgen Multi-Drug Urine Test Cup; Healgen Multi-Drug Urine Test Dip Card · Healgen Scientific,, LLC · Aug 2, 2017
K173303 — INSTANT-VIEW plus Multi-Drug of Abuse Urine Test - Simple Cup (OTC Use), INSTANT-VIEW plus Multi-Drug Urine Test - Simple Cup (Prescription Use) · Alfa Scientific Designs, Inc. · Feb 23, 2018
Submission Summary (Full Text)
{0}------------------------------------------------
Ameditech, Inc.
DEC - 5 2011 K. 113
**CHECKCUP™ Multi-Drug Test**
#### 510(k) Summary
This 510(k) Summary is being submitted in accordance with the requirements of 21 CFR 807.92 and follows the Device Advice Guidance concerning "How To Prepare A Special 510(k)"
## A. Submitter:
| Submitter: | Ameditech, Inc.<br>10340 Camino Santa Fe.<br>San Diego, CA 92121 |
|-----------------|------------------------------------------------------------------------------------------------------------|
| Contact Person: | Brett Casey<br>General Manager<br>Phone: 858-805-3603<br>Fax: 858-535-1838<br>Email: brett.casey@alere.com |
| Date prepared: | 11/22/2011 |
Date prepared:
B. Device Name:
Ameditech CHECKCUP™
Predicates:
Ameditech DrugSmart Cup (K053175)
## C. Device Description:
The Ameditech CHECKCUP™ Multi-Drug Test is a lateral flow immunoassay intended for professional (prescription) use only. The device is used for the qualitative determination of parent compound and/or major metabolites of drugs in human urine specimens, and results are read visually. The device will be made available in two formats that use identical test strips: Round Cup and Flat Cup. The modified device contains up to seventeen (17) assays that include fourteen (14) different drugs at various cutoff concentrations. The assays detect different illicit drugs and medications that are commonly abused. Once the urine sample is administered, negative results can be read as soon as 1 minute and positive results should be read between 5 to 10 minutes. The presence of the line on the test region indicates a negative result for the drug and the absence of a line on the test region indicates a positive result for the drug. A visible line is also present at the control region of the test strip. This line should always appear, regardless of the presence of drugs or metabolites in the urine sample. This means that a negative urine sample will produce both a test line and control line, and a positive urine sample will generate only a control line. The presence of control line serves as a built-in control, which demonstrates that the test is performed properly.
{1}------------------------------------------------
# D. Intended Use:
The CHECKCUP™ Multi-Drug Test is a one-step immunoassay for the qualitative detection of multiple drugs and drug metabolites in human urine at the following cutoff concentrations:
| Test | Calibrator | Cut-off (ng/ml) |
|---------|-----------------------------------|-----------------|
| AMP | d-Amphetamine | 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC 150 | Benzoylecgonine | 150 |
| COC | Benzoylecgonine | 300 |
| MDMA | 3,4-methylenedioxymethamphetamine | 500 |
| MET 500 | d-Methamphetamine | 500 |
| MET | d-Methamphetamine | 1000 |
| MTD | dl-Methadone | 300 |
| OPI 300 | Morphine | 300 |
| OPI | Morphine | 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-A9-THC-9-COOH | 50 |
The CHECKCUP™ Multi-Drug Test consists of two test formats: round cup and flat cup. The configurations of the CHECKCUP™ Multi-Drug Test consist of any combination of the drugs listed above. The CHECKCUP™ Multi-Drug Test is used to obtain a visual, qualitative result and is intended for professional use only.
This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmation methods.
{2}------------------------------------------------
## 510(k) Summary
#### CHECKCUP™ Multi-Drug Test
## Comparison with the Predicate
The modified device is identical to the cleared predicate referenced above (K053175). The drug strips cleared by FDA in submission K113046 will be placed in round cup that was cleared in K053175. In addition, the drug strips will be placed in a new format (flat cup).
The CHECKCUP™ Multi-Drug Test is a device modification of the previously 510(k) cleared Ameditech DrugSmart Cup (K053175). The CHECKCUP™ Multi-Drug Test expands the current drug test menu of the Ameditech DrugSmart Cup by adding the drug test strips cleared in K113046. The CHECKCUP™ Multi-Drug Test has two formats: round cup and flat cup. The CHECKCUP™ round cup is exactly the same cup as in the Drugsmart Cup. Both CHECKCUP™ round cup and DrugSmart Cup consist of multiple 1-drug strips. The plastic strip holder in the CHECKCUP™ flat and DrugSmart Cup are both flat except in the DrugSmart Cup, the plastic strip holder is a malleable piece of plastic that gets curved when inserted into the DrugSmart Cup. The CHECKCUP™ flat cup plastic strip holder stays flat when inserted into the CHECKCUP flat cup. The CHECKCUP™ flat cup consist of 1-drug/strip tests or 2-drug/strip tests or a mixture of both. Below is an illustration of DrugSmart Cup and CHECKCUP™ Multi-Drug Test (round and flat format). The modified and 510(k) cleared devices are identical in terms of intended use, principle, test strip manufacturing process and raw material used. The modified device detects 17 drug tests versus these drug tests are detected in separate devices K053175 and K113046. The purpose of this submission is to cover all the drug strips and their respective cutoffs with a single device. See Section 12.2 of this submission for a detailed comparison of the modified device with the predicates.
Image /page/2/Figure/6 description: The image shows three different drug testing cups. The first cup is labeled "DrugSmart Cup" and is described as "Unmodified". The text below the cup states that it detects up to 13 drugs with 1 strip per drug. The other two cups are labeled "CHECKCUP (round version)" and "CHECKCUP (flat version)" and are described as "Check Cup Multi Drug Test", detecting up to 17 drugs with 1 strip/drug, 2 strip/drug (flat cup) or both (flat cup).
{3}------------------------------------------------
## 510(k) Summary
# Nonclinical Data:
As the test strips in both the modified and predicate devices are identical in terms of intended use, test principle, manufacturing process, and raw materials; simplified performance studies were conducted to demonstrate that the sensitivity, precision, and specificity of the modified device was unchanged from the predicate devices. The study approach was based on previous similar submission (k113046). The analytical performance verification and validation testing confirms that the Ameditech CHECKCUP™ has equivalent performance compared with the predicate devices in terms of precision (reproducibility), analytical sensitivity, analytical specificity, and interfering substances. Refer to Appendix D- Testing Strategy. .. .
# E. Clinical Data:
Both the modified and predicate devices use identical test strips, therefore we did not conduct any clinical studies for this special 510(k) submission, because the performance characteristics that were validated in the predicate submissions are applicable to the modified device.
#### F. Conclusions Drawn from Testing Results:
Performance verification and validation studies have demonstrated that the device is as safe, as effective, and performs as well or better than the predicate device.
{4}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with a bird-like figure intertwined around a staff. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the caduceus.
10903 New Hampshire Avenue Silver Spring, MD 20993
AMEDITECH. INC C/O Brett Casey 10340 Camino Santa Fe. Suite F San Diego, CA 92121
DEC - 5 2011
Re: k113501
> Trade Name: Ameditech CHECKCUP™ Multi-Drug Test Regulation Number: 21 CFR §862.3100 Regulation Name: Amphetamine test system Regulatory Class: Class II Product Codes: DKZ, DIS, JXM, DIO, DJC, DJR, DJG, LCM, JXN. LFG Dated: November 25, 2011 Received: November 25, 2011
Dear Mr. Casey:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820).
{5}------------------------------------------------
#### Page 2
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical
Devices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ...
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm
Sincerely yours.
N
Countney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
Ameditech, Inc.
#### Indications for Use Statement
510(k) Number (if known): K || 350 |
Device Name: CHECKCUP™ Multi-Drug Test
Indications for Use:
The CHECKCUP™ Multi-Drug Test is a one-step immunoassay for the qualitative detection of multiple drugs and drug metabolites in human urine at the following cutoff concentrations:
| Test | Calibrator | Cut-off (ng/ml) |
|---------|-----------------------------------|-----------------|
| AMP | d-Amphetamine | 1000 |
| BAR | Secobarbital | 300 |
| BUP | Buprenorphine | 10 |
| BZO | Oxazepam | 300 |
| COC 150 | Benzoylecgonine | 150 |
| COC | Benzoylecgonine | 300 |
| MDMA | 3,4-methylenedioxymethamphetamine | 500 |
| MET 500 | d-Methamphetamine | 500 |
| MET | d-Methamphetamine | 1000 |
| MTD | dl-Methadone | 300 |
| OPI 300 | Morphine | 300 |
| OPI | Morphine | 2000 |
| OXY | Oxycodone | 100 |
| PCP | Phencyclidine | 25 |
| PPX | Propoxyphene | 300 |
| TCA | Nortriptyline | 1000 |
| THC | 11-nor-Δ9-THC-9-COOH | 50 |
The CHECKCUP™ Multi-Drug Test consists of two test formats: flat cup and round cup. The configurations of the CHECKCUP™ Multi-Drug Test consist of any combination of the drugs listed above. The CHECKCUP™ is used to obtain a visual, qualitative result and is intended for professional use only.
This assay provides only a preliminary result. Clinical consideration and professional judgment must be applied to any drug of abuse test result, particularly in evaluating a preliminary positive result. In order to obtain a confirmed analytical result, a more specific alternate chemical method is needed. Gas Chromatography/Mass Spectrometry (GC/MS) and Liquid Chromatography/Mass Spectrometry (LC/MS) are the preferred confirmation methods.
Prescription Use X (Part 21 CFR 801, Subpart D) AND/OR
Over-the-Counter Use (Part 21 CFR 807, Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
ARS
Division Sign-Off Office of In Vitro Diagnostic Devices Evaluation and Safety
510(k): K113501
Page 1 of 1
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.