← Product Code [NDN](/productcode/NDN) · K113477

# IVAS 20MM (10 GAUGE) BALLOON CATHETER (K113477)

_Stryker Instruments, Instruments Div. · NDN · Jan 6, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113477

## Device Facts

- **Applicant:** Stryker Instruments, Instruments Div.
- **Product Code:** [NDN](/productcode/NDN.md)
- **Decision Date:** Jan 6, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 888.3027
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

The Stryker® iVAS Inflatable Vertebral Augmentation System (system) is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. This includes use during percutaneous vertebral augmentation. The system is to be used with cleared spinal Polymethylmethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

## Device Story

Stryker iVAS is an inflatable bone tamp catheter used in percutaneous vertebral augmentation. The device features an inflatable balloon at the distal end; when inserted into the vertebral body, the balloon is inflated to create a void in cancellous bone or reduce fractures. The procedure is performed by a physician in a clinical setting. Following void creation, the balloon is deflated and removed, allowing for the subsequent injection of cleared spinal PMMA bone cement. The device facilitates the stabilization of vertebral fractures, potentially reducing pain and restoring vertebral height. It is a mechanical instrument and does not involve electronic processing or software.

## Clinical Evidence

Bench testing only. Testing included functional evaluations such as insertion and retraction force, force to puncture, burst testing, and simulated use. Biocompatibility testing was conducted in accordance with ISO 10993-1.

## Technological Characteristics

Inflatable bone tamp catheter with a distal balloon component. Materials are biocompatible per ISO 10993-1. Mechanical operation via inflation/deflation. No energy source, software, or connectivity.

## Regulatory Identification

Polymethylmethacrylate (PMMA) bone cement is a device intended to be implanted that is made from methylmethacrylate, polymethylmethacrylate, esters of methacrylic acid, or copolymers containing polymethylmethacrylate and polystyrene. The device is intended for use in arthroplastic procedures of the hip, knee, and other joints for the fixation of polymer or metallic prosthetic implants to living bone.

## Special Controls

*Classification.* Class II (special controls). The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Polymethylmethacrylate (PMMA) Bone Cement.”

## Predicate Devices

- Stryker® iVAS Balloon Catheter ([K103807](/device/K103807.md))
- Stryker® iVAS Balloon Catheter ([K093419](/device/K093419.md))
- Kyphx Xpander Inflatable Bone Tamps ([K041454](/device/K041454.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K113477

JAN - 6 2012

## stryker

-1.

4100 E. Milham Avenue Kalamazoo. MI 49001
t: 269 323 7700 f: 269 324 5412 www.stryker.com

ﺗ

## Instruments

## 510(k) Summary

| 510(k) Owner:                             | Stryker Instruments<br>4100 E. Milham Avenue<br>Kalamazoo, MI  49001<br>(p) 269-323-7700<br>(f) 269-324-5412                                                                                                                                                                                                                                                                                                                                            |
|-------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact Person:                           | Christina McKee                                                                                                                                                                                                                                                                                                                                                                                                                                         |
| Registration No .:                        | 1811755                                                                                                                                                                                                                                                                                                                                                                                                                                                 |
| Trade Name:                               | Stryker® iVAS Balloon Catheter                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Common Name:                              | Inflatable Bone Tamp                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Classification Name:                      | Arthroscope<br>Cement, Bone, Vertebroplasty                                                                                                                                                                                                                                                                                                                                                                                                             |
| Regulation Number:                        | \$888.1100<br>\$888.3027                                                                                                                                                                                                                                                                                                                                                                                                                                |
| Product Code:                             | HRX<br>NDN                                                                                                                                                                                                                                                                                                                                                                                                                                              |
| Predicate Device:                         | Stryker® iVAS Balloon Catheter (K103807)<br>Stryker® iVAS Balloon Catheter (K093419)<br>Kyphx Xpander Inflatable Bone Tamps (K041454)                                                                                                                                                                                                                                                                                                                   |
| Device Description:                       | The Stryker® iVAS balloon catheter is a bone tamp with an inflatable component<br>(balloon) at the distal end. The balloon is inflated to create a void within the<br>vertebral body.                                                                                                                                                                                                                                                                   |
| Indications for Use:                      | The Stryker® iVAS Inflatable Vertebral Augmentation System (system) is<br>intended to be used for the reduction of fractures and/or creation of a void in<br>cancellous bone in the spine. This includes use during percutaneous vertebral<br>augmentation. The system is to be used with cleared spinal<br>Polymethylmethacrylate (PMMA) bone cements indicated for use during<br>percutaneous vertebral augmentation procedures, such as kyphoplasty. |
| Testing:                                  | The Stryker® iVAS balloon catheter meets the specification and performance<br>characteristics and are substantially equivalent to the predicate devices. The<br>testing which was conducted included functional testing, such as insertion and<br>retraction force, force to puncture, burst and simulated use.                                                                                                                                         |
| Biocompatibility:                         | Biocompatibility testing of the Stryker® iVAS balloon catheter confirmed that the<br>device meets the applicable requirements of the FDA Blue Book Memorandum<br>G95-1 entitled Use of International Standards ISO-10993 Biological Evaluation of<br>Medical Devices Part -1: Evaluation and Testing and are biocompatible.                                                                                                                             |
| Substantial Equivalence<br>(SE) Rational: | The Stryker® iVAS balloon catheter is substantially equivalent in intended use,<br>technological characteristics, safety, and effectiveness to the Stryker® iVAS<br>Balloon Catheter and the Kyphx Xpander Inflatable Bone Tamp. The products<br>have the same fundamental scientific technology, basic design, functional<br>characteristics and the same clinical applications.                                                                       |
| Safety and Effectiveness:                 | The Stryker® iVAS balloon catheter does not raise any new safety and efficacy<br>concerns when compared to a similar device already legally marketed.<br>Therefore, the Stryker® iVAS balloon catheter is equivalent to the existing<br>predicate devices.                                                                                                                                                                                              |
| Submitted by:                             | Christina McKee                                                                                                                                                                                                                                                                                                                                                                                                                                         |
|                                           | Regulatory Affairs Associate Analyst                                                                                                                                                                                                                                                                                                                                                                                                                    |
|                                           | Signature                                                                                                                                                                                                                                                                                                                                                                                                                                               |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "stryker" in a bold, sans-serif font. The word is all lowercase, and there is a registered trademark symbol in the upper right corner. The text is black and the background is white.

Instruments

.

.

Date Submitted:

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. The eagle is black, and the text is also black.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN - 6 2012

Stryker Instruments % Ms. Christina McKee Regulatory Affairs Supervisor 4100 E. Milham Ave Kalamazoo, Michigan 49001

Re: K113477

Trade Name: iVAS 20mm (10 Gauge) Balloon Catheter Regulation Number: 21 CFR 888.3027 Regulation Name: Polymethylmethacrylate (PMMA) bone cement Regulatory Class: Class II Product Code: NDN, HRX Dated: November 22, 2011 Received: November 23, 2011

Dear Ms. McKee:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act 

{3}------------------------------------------------

Page 2 - Ms. Christina McKee

or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

510(K) Number (if known):

Device Name: Stryker Inflatable Vertebral Augmentation System (iVAS)

## Indications for Use

The Stryker® iVAS Inflatable Vertebral Augmentation System) is intended to be used for the reduction of fractures and/or creation of a void in cancellous bone in the spine. assed for the rouation of necessor overtebral augmentation. The system is to be used with cleared spinal Polymethacrylate (PMMA) bone cements indicated for use during percutaneous vertebral augmentation procedures, such as kyphoplasty.

Prescription Use _____________________________________________________________________________________________________________________________________________________________

and/or

.

Over-The-Counter Use

(Part 21 CFR 801 Subpart D)

(21 CFR 801 Subpart C)

(Please Do NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) .

Concurrence of CDRH, Office of Device Evaluation (ODE)

Division Sign-Off)

(DK43 Division of Surgical, Orthopedic, and Restorative Devices

510(k) Number K113477

---

**Source:** [https://fda.innolitics.com/device/K113477](https://fda.innolitics.com/device/K113477)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
