ACCULIF TL CAGE

K113465 · Coalign Innovations, Inc. · MAX · Dec 12, 2011 · Orthopedic

Device Facts

Record IDK113465
Device NameACCULIF TL CAGE
ApplicantCoalign Innovations, Inc.
Product CodeMAX · Orthopedic
Decision DateDec 12, 2011
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 888.3080
Device ClassClass 2
AttributesTherapeutic

Indications for Use

Intervertebral Body Fusion Device: The CoAlign Innovations AccuLIF TL Cage is indicated for intervertebral body fusion with autogenous bone graft material in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. The CoAlign Innovations AccuLIF TL Cage is always to be used with supplemental internal spinal fixation. Additionally, the CoAlign Innovations AccuLIF TL Cage is to be used with autogenous bone graft.

Device Story

AccuLIF TL Cage is an expandable spinal intervertebral body fusion device. Inserted in unexpanded state via articulating delivery handle; expanded in-situ to required height using hydraulic system (disposable flexible tubing and inflation syringe) via 2 hydraulic cylinder/piston arrangement. Device locks in 1mm increments. Features fixation ridges and graft opening window. Used by surgeons in clinical/OR settings to maintain intervertebral spacing and angulation following discectomy. Requires supplemental internal spinal fixation and autogenous bone graft. Benefits include restoration of disc height and stabilization of spinal segment.

Clinical Evidence

No clinical data. Bench testing only: dynamic/static compression and shear (ASTM F2077-03), static subsidence (ASTM F2267-04), and static expulsion (ASTM Draft Standard F-04.25.02.02). Surgical technique validated against predicate.

Technological Characteristics

Expandable spacer; Titanium-6AL-4V ELI (ASTM F136-08), Stainless Steel 316 LVM (ASTM F138-08), Silicone Rubber (MED-4870). Hydraulic expansion mechanism; 1mm locking increments. Dimensions: 11x34mm or 13x34mm footprint; 6-16mm height range. Fixation ridges; central graft window.

Indications for Use

Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two contiguous levels (L2-S1), including those with up to Grade I spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment.

Regulatory Classification

Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval. (c) *Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K113465 DEC 1 2 2011 Page 1 of 3 #### 5. 510(k) Summary | Submitted By: | CoAlign Innovations<br>150 North Hill Drive, Suite 1<br>Brisbane, CA 94005 | |---------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------| | Establishment<br>Registration Number: | 10030843 | | Contact Person:<br>(Submission Prepared by) | Justin Eggleton<br>Musculoskeletal Clinical & Regulatory Advisers, LLC<br>1331 H Street NW, 12th Floor<br>Washington, DC 20005<br>202.552.5800 | | Date Prepared: | October 27, 2011 | | Device Trade Name: | AccuLIF® TL Cage | | Manufacturer: | CoAlign Innovations, Inc.<br>150 North Hill Drive, Suite 1<br>Brisbane, CA 94005 | | Common Name: | Spinal intervertebral body fixation orthosis | | Classification: | 21 CFR §888.3080 | | Class: | II | | Product Code: | MAX | # Device Description: The AccuLIF TL Cage acts as a spacer to maintain proper intervertebral spacing and angulation following discectomy. The AccuLIF TL Cage is manufactured from TI6Al4V ELI) as per ASTM F136-08, Stainless Steel (316 LVM) as per ASTM F138-08, and Silicone Rubber (MED-4870). The device is inserted in unexpanded state with an articulating delivery handle and expanded in-situ to the required height via 2 hydraulic cylinder and piston arrangement using a hydraulic system comprising disposable flexible expansion tubing set and inflation syringe. The device locks in 1mm increments as it expands. The Accul.IF Cage comes in 6 to 9mm, 8 to 12mm and 10 to 16mm sizes. Each size also comes in an 11mm x 34mm footprint and a 13mm x 34mm footprint. The device has fixation ridges on the top and bottom surface. It also has a graft opening window which extends from the bottom surface to the top surface. The device has a proximal {1}------------------------------------------------ boss which has a threaded connection port for connecting to the inserter and a fluid port for transporting the expansion fluid # Indications For Use: Intervertebral Body Fusion Device: The CoAlign Innovations AccuLIF TL Cage is indicated for intervertebral body fusion with autogenous bone graft material in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. The CoAlign Innovations AccuLIF TL Cage is always to be used with supplemental internal spinal fixation. Additionally, the CoAlign Innovations AccuLIF TL Cage is to be used with autogenous bone graft. ## Identification of Predicates - AccuLIF Cage (K110270) - - SEC IBF Cage (K093669) " - Globus Medical Caliber Spacer (K102293) i ## Summary of Technological Characteristics AccuLIF TL Cages are expandable spacers made from Titanium-6AL-4V ELI alloy that conforms to ASTM F136. The subject and predicate devices have nearly identical technological characteristics and the minor differences do not raise any new issues of safety and effectiveness. Specifically the following characteristics are identical between the subject and predicates: - Indications for Use - - Materials of manufacture - - Expansion mechanism - - Structural support mechanism - ### Discussion of Testing: The following non-clinical tests were conducted: - Dynamic compression testing, conducted in accordance with ASTM F2077-03 - Dynamic shear testing, conducted in accordance with -ASTM F2077-03 - Static compression testing, conducted in accordance with ASTM F2077-03 - Static shear testing, conducted in accordance with -: ASTM F2077-03 - Static Subsidence testing, conducted in accordance with -ASTM F2267-04 - Static Expulsion testing, conducted in accordance with -ASTM Draft Standard F-04.25.02.02 {2}------------------------------------------------ - Surgical technique validation, conducted in the same manner as the predicate surgical technique validation. ## Conclusions: The subject and predicate devices share the same indications for use, design, function, and materials of manufacture. The non-clinical test results demonstrate that any minor differences do not impact device performance as compared to the predicates. The AccuLIF TL Cage was shown to be substantially equivalent to the AccuLIF IBF Cage (K110270), SEC IBF Cage (K093669), and Globus Medical Caliber Spacer (K102293). - {3}------------------------------------------------ Image /page/3/Picture/1 description: The image is a black and white logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three wing segments, representing the department's mission to protect the health of all Americans and provide essential human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 DEC 1 2 2011 CoAlign Innovations, Inc. % Musculoskeletal Clinical Regulatory Advisers, LLC Mr. Justin Eggleton 1331 H Street NW, 12th Floor Washington, District of Columbia 20005 Re: K113465 Trade/Device Name: AccuLIF TL Cage Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: November 18, 2011 Received: November 22, 2011 Dear Mr. Eggleton: We have reviewed your Section-510(k)-premarket-notification of intent-to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must {4}------------------------------------------------ Page 2 - Mr. Justin Eggleton comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely vours. For D. Th. mait Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ #### Indications for Use 4. #### 510(k) Number (if known): K113465 Device Name: AccuLIF TL Cage Intervertebral Body Fusion Device: The CoAlign Innovations AccuLIF TL Cage is indicated for intervertebral body fusion with autogenous bone graft material in patients with degenerative disc disease (DDD) at one level or two contiguous levels from L2 to S1. DDD is defined as back pain of discogenic origin with degeneration of the disc confirmed by history and radiographic studies. These DDD patients may also have up to Grade I spondylolisthesis or retrolisthesis at the involved level(s). These patients should be skeletally mature and have completed six months of non-operative treatment. The CoAlign Innovations AccuLIF TL Cage is always to be used with supplemental internal spinal fixation. Additionally, the CoAlign Innovations AccuLIF TL Cage is to be used with autogenous bone graft. V Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-Off) Division of Surgical, Orth redic, and Restorative Devices 510(k) Number_________________________________________________________________________________________________________________________________________________________________
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...