← Product Code [MAX](/productcode/MAX) · K113447

# RISE SPACER (K113447)

_Globus Medical, Inc. · MAX · Jan 26, 2012 · Orthopedic · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113447

## Device Facts

- **Applicant:** Globus Medical, Inc.
- **Product Code:** [MAX](/productcode/MAX.md)
- **Decision Date:** Jan 26, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 888.3080
- **Device Class:** Class 2
- **Review Panel:** Orthopedic
- **Attributes:** Therapeutic

## Indications for Use

RISE™ Spacers are lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s). RISE™ Spacers are to be filled with autogenous bone graft material. This device is intended to be used with supplemental fixation, such as the REVERE® or REVOLVE™ Stabilization Systems.

## Device Story

RISE™ Spacer is an expandable lumbar interbody fusion device designed to provide structural stability following discectomy. The device is implanted by a surgeon during spinal fusion procedures. It features an all-titanium exterior with superior/inferior surface protrusions to grip vertebral endplates and resist expulsion; an internal PEEK component facilitates the expansion mechanism. The device is filled with autogenous bone graft material to promote fusion and must be used with supplemental fixation systems (e.g., REVERE® or REVOLVE™). By expanding to a desired height, the device restores disc space and provides stability, potentially reducing pain associated with degenerative disc disease.

## Clinical Evidence

Bench testing only. Mechanical testing, including static and dynamic compression, static and dynamic compression-shear, and subsidence, was performed to demonstrate substantial equivalence to the predicate device.

## Technological Characteristics

Lumbar interbody fusion device; titanium alloy (ASTM F136, F1295) exterior; internal PEEK polymer (ASTM F2026) component; expandable mechanism; surface protrusions for endplate fixation; requires autogenous bone graft; used with supplemental fixation.

## Regulatory Identification

An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.

## Special Controls

*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- Globus Medical CALIBER™ Spacer ([K102293](/device/K102293.md))

## Reference Devices

- REVERE® Stabilization System
- REVOLVE™ Stabilization System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY: RISE™ Spacer

| Company:        | Globus Medical Inc.<br>2560 General Armistead Avenue.<br>Audubon, PA 19403<br>(610) 930-1800                                        |
|-----------------|-------------------------------------------------------------------------------------------------------------------------------------|
| Contact:        | Kelly J. Baker, Ph.D.<br>Vice President, Regulatory and Clinical Affairs                                                            |
| Date Prepared:  | November 18, 2011                                                                                                                   |
| Device Name:    | RISETM Spacer                                                                                                                       |
| Classification: | Per 21 CFR as follows:<br>§888.3080 Intervertebral Body Fusion device.<br>Product Code: MAX<br>Regulatory Class: II, Panel Code: 87 |
| Predicate(s):   | Globus Medical CALIBERTM Spacer (K102293)<br>SE date: January 5, 2011                                                               |

#### Purpose:

The purpose of this submission is to request clearance for RISE™ Spacers, which have an all titanium design, a smaller overall implant profile, and a modified expansion mechanism, compared to the predicate titanium/PEEK CALIBER device.

### Device Description:

RISE™ Spacers are lumbar interbody fusion devices used to provide structural stability in skeletally mature individuals following discectomy. RISE™ Spacers are provided in different shapes to accommodate various surgical approaches to the lumbar spine (posterior, transforaminal [posterolateral] or lateral) and can expand to the desired height. The implants are available in various heights and geometric options to fit the anatomical needs of a wide variety of patients. This device is to be filled with autogenous bone graft material. Protrusions on the superior and inferior surfaces of each device grip the endplates of the adjacent vertebrae to resist expulsion.

RISE™ Spacers are manufactured from titanium alloy, as specified in ASTM F136 and F1295. An internal component is manufactured from radiolucent PEEK polymer, as specified in ASTM F2026.

#### Indications for Use:

RISE™ Spacers are lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the 

{1}------------------------------------------------

lumbosacral spine (L2-S1). DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).

RISE™ Spacers are to be filled with autogenous bone graft material. This device is intended to be used with supplemental fixation, such as the REVERE® or REVOLVE™ Stabilization Systems.

#### Performance Data:

Mechanical testing (static and dynamic compression, static and dynamic compression-shear and subsidence) was conducted to demonstrate substantial equivalence to the predicate device.

#### Basis for Substantial Equivalence:

The RISE™ Spacer has been found to be substantially equivalent to the predicate with respect to technical characteristics, performance, design, materials, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate device.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, and a ribbon-like element flowing beneath it. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

JAN 26 2012

Globus Medical, Inc. % Kelly J. Baker, Ph.D. Vice President, Regulatory and Clinical Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403

Re: K113447

Trade/Device Name: RISE™ Spacer Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: December 29, 2011 Received: December 30, 2011

Dear Dr. Baker:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 - Kelly J. Baker, Ph.D.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part) 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,
Erin Keith

Mark N. Melkerson Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

K113447 510(k) Number:

Device Name: RISE™ Spacer

Indications:

RISE™ Spacers are lumbar interbody fusion devices intended for use in patients with degenerative disc disease (DDD) at one or two contiguous levels of the DDD is defined as discogenic back pain with lumbosacral spine (L2-S1). degeneration of the disc confirmed by history and radiographic studies. These patients should be skeletally mature and have had at least six (6) months of nonoperative treatment. In addition, these patients may have up to Grade 1 spondylolisthesis or retrolisthesis at the involved level(s).

RISE™ Spacers are to be filled with autogenous bone graft material. This device is intended to be used with supplemental fixation, such as the REVERE® or REVOLVE™ Stabilization Systems.

OR

Prescription Use × (Per 21 CFR §801.109) Over-The-Counter Use

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) ·

Concurrence of CDRH, Office of Device Evaluation (ODE)

(Division Sign-Off) Division of Surgica' Orthopedic, and Restorative Devices

KI 1 344 7 510(k) Number_

---

**Source:** [https://fda.innolitics.com/device/K113447](https://fda.innolitics.com/device/K113447)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
