ACE CK REAGENT,ACE BUN/UREA REAGENT,ACE URIC ACID REAGENT,ACE CREATININE REAGENT
K113389 · Alfa Wassermann Diagnostic Technologies, Inc. · CDN · Aug 10, 2012 · Clinical Chemistry
Device Facts
Record ID
K113389
Device Name
ACE CK REAGENT,ACE BUN/UREA REAGENT,ACE URIC ACID REAGENT,ACE CREATININE REAGENT
Applicant
Alfa Wassermann Diagnostic Technologies, Inc.
Product Code
CDN · Clinical Chemistry
Decision Date
Aug 10, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.1770
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K113389 · Aug 10, 2012
ACE CK REAGENT,ACE BUN/UREA REAGENT,ACE URIC ACID REAGENT,ACE CREATININE REAGENT
Alfa Wassermann Diagnostic Technologies, Inc.
Physician Office Laboratory (POL) clinical sites
Performance data (precision and accuracy) were collected at three separate Physician Office Laboratory (POL) sites to demonstrate the device's performance in a routine clinical setting.
POL Precision and Correlation Studies; Multi-site clinical performance evaluation; Follow-up/Duration: 5 days
Patient samples in a Physician Office Laboratory setting; Sample Size: Not specified (multiple samples per site); Number of Sites: 3
Alfa Wassermann ACE Clinical Chemistry System
Precision (within-run and total CV), Accuracy (correlation coefficient, slope, intercept)
Indications for Use
The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum using the ACE Axcel Clinical Chemistry System. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum using the ACE Axcel Clinical Chemistry System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum using the ACE Axcel Clinical Chemistry System. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum using the ACE Axcel Clinical Chemistry System. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Device Story
Reagents for quantitative photometric/colorimetric analysis of serum samples (BUN, creatinine, uric acid, CK) on ACE Axcel Clinical Chemistry System. System performs enzymatic/chemical reactions; monitors absorbance changes at specific wavelengths (340nm, 505nm, 573nm, 610nm). Used in clinical labs and physician office labs by trained personnel. Output provides analyte concentration/activity levels to assist clinicians in diagnosing/monitoring renal, metabolic, and muscle disorders. Benefits include rapid, automated diagnostic testing.
Clinical Evidence
Bench testing only. Precision, accuracy, and detection limit studies performed. Correlation studies compared ACE Axcel system to predicate ACE system using 106-136 patient samples per analyte. Correlation coefficients ranged from 0.9858 to 0.9998. Precision studies (within-run and total CVs) conducted over 22 days and at three physician office laboratory sites.
Technological Characteristics
Photometric/colorimetric enzymatic assays. Reagents include urease, GLDH, picric acid, uricase, peroxidase, hexokinase, and G-6-PDH. Analyte detection via absorbance change at 340nm, 505nm, 573nm, 610nm. System: ACE Axcel Clinical Chemistry System. Calibration/on-board stability varies by reagent. Traceability to NIST SRM 909 and 967.
Indications for Use
Indicated for quantitative determination of BUN, creatinine, uric acid, and creatine kinase in serum. Used for diagnosis/treatment of renal/metabolic diseases, myocardial infarction, and muscle diseases (e.g., muscular dystrophy). Intended for clinical or physician office laboratories.
Regulatory Classification
Identification
A urea nitrogen test system is a device intended to measure urea nitrogen (an end-product of nitrogen metabolism) in whole blood, serum, plasma, and urine. Measurements obtained by this device are used in the diagnosis and treatment of certain renal and metabolic diseases.
{1}
1
SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k113389
B. Purpose for Submission:
New device
C. Measurand:
Urea Nitrogen (BUN)
Creatinine
Uric acid
Creatinine Kinase
D. Type of Test:
Quantitative, photometric/ colorimetric detection
E. Applicant:
Alfa Wassermann Diagnostic Technologies, LLC
F. Proprietary and Established Names:
ACE BUN/ Urea Reagent
ACE Creatinine Reagent
ACE Uric Acid Reagent
ACE CK Reagent
G. Regulatory Information:
| Analyte | Regulation | Product Code | Classification | Panel |
| --- | --- | --- | --- | --- |
| BUN | 21 CFR 862.1770
Urea nitrogen test system | CDN | II | Clinical Chemistry (75) |
| Creatinine | 21 CFR 862.1225
Creatinine test system | CGX | II | Clinical Chemistry (75) |
| Uric Acid | 21 CFR 862.1775
Uric acid test system | KNK | I. reserved | Clinical Chemistry (75) |
{2}
2
| CK | 21 CFR 862.1215
Creatine phosphokinase /
creatine kinase or isoenzymes | CGS | II | Clinical Chemistry
(75) |
| --- | --- | --- | --- | --- |
## H. Intended Use:
1. Intended use(s):
See indication for use below
2. Indication(s) for use:
The ACE BUN/Urea Reagent is intended for the quantitative determination of blood urea nitrogen (BUN) concentration in serum using the ACE Axcel Clinical Chemistry System. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum using the ACE Axcel Clinical Chemistry System. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum using the ACE Axcel Clinical Chemistry System. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, leukemia, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum using the ACE Axcel Clinical Chemistry System. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial infarction and muscle diseases such as progressive, Duchenne-type muscular dystrophy. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
3. Special conditions for use statement(s):
For *in vitro* diagnostic use only
For prescription use and Point-of-Care settings
4. Special instrument requirements:
{3}
ACE Axcel Clinical Chemistry System
# I. Device Description:
The ACE BUN/Urea Reagent consists of a single reagent bottle containing $\alpha$-Ketoglutarate (4.0 mmol/L), Urease (Jack Bean) >15,000 U/L, Glutamate dehydrogenase (GLDH) (Beef Liver) >1667 U/L, Adenosine diphosphate (ADP) 2.0 mmol/L, Nicotinamide adenine dinucleotide, reduced (NADH) 0.28 mmol/L, buffer, preservative and stabilizer.
The ACE Creatinine Reagent consists of two reagent bottles, the Sodium Hydroxide Reagent and Picric Acid Reagent. The Sodium Hydroxide Reagent (R1) contains Sodium Hydroxide 0.25 mol/L and surfactants. The Picric Acid Reagent (R2) contains Picric Acid 20.5 mmol/L.
The ACE Uric Acid Reagent consists of a single reagent bottle containing 4-Aminoantipyrine (AAP) 0.5 mmol/L, Dichlorohydroxybenzene sulfonic acid (DHBS) 1.8 mmol/L, Peroxidase (Horseradish) >3500 U/L, Uricase (Bacillus) >200 U/L and stabilizers and preservatives.
The ACE CK Reagent consists of two reagent bottles, Buffer and Substrate. The Buffer Reagent (R1) contains Imidazole Buffer (pH 6.5 at $25^{\circ}\mathrm{C}$), Glucose 27 mmol/L, N-acetylcysteine 27 mmol/L, Magnesium acetate 14 mmol/L, EDTA 2 mmol/L, NADP 2.7 mmol/L, Hexokinase (recombinant yeast, modified) >5 KU/L. The Substrate Reagent (R2) contains Creatine phosphate 160 mmol/L, EDTA 2 mmol/L, ADP 11 mmol/L, AMP 28 mmol/L, Diadenosine pentaphosphate 55 μmol/L and Glucose-6-phosphate dehydrogenase (E coli) >14 KU/L.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
- ACE BUN/Urea Reagent
- ACE Creatinine Reagent
- ACE Uric Acid Reagent
- ACE CK Reagent
2. Predicate 510(k) number(s):
- k930104
3. Comparison with predicate:
{4}
4
| | ACE BUN/Urea Reagent
(Candidate device) | ACE BUN/Urea Reagent
(Predicate device, k930104) |
| --- | --- | --- |
| Intended Use/Indications for use | The ACE BUN Reagent is intended for the quantitative determination of blood urea nitrogen concentration in serum. BUN measurements are used in the diagnosis and treatment of certain renal and metabolic diseases. | Same |
| Instrument/Platform | ACE Axcel Clinical Chemistry System | ACE Clinical Chemistry Systems |
| Calibration Stability | 7 days | Same |
| On- Board Stability | 30 days | Same |
| Basic Principle | Enzymatic method for urea nitrogen | Same |
| Sample Stability | 2-5 days at 2-8°C
up to several weeks at -20°C; and at -70°C or lower for longer periods | Same |
| Unopened Stability | until the expiration date shown on the box and bottle labels when stored in the refrigerator at 2-8°C | Same |
| | ACE Creatinine Reagent
(Candidate device) | Ace Creatinine Reagent
(Predicate, k930104) |
| --- | --- | --- |
| Intended Use/Indications for use | The ACE Creatinine Reagent is intended for the quantitative determination of creatinine concentration in serum. Creatinine measurements are used in the diagnosis and treatment of renal diseases, in monitoring renal dialysis, and as a calculation basis for measuring other urine analytes. | Same |
| Instrument/Platform | ACE Axcel Clinical Chemistry System | ACE Clinical Chemistry Systems |
{5}
| Calibration Stability | 2 days | Same |
| --- | --- | --- |
| On- Board Stability | 10 days | Same |
| Basic Principle | Alkaline picrate chemistry. | Same |
| Sample Stability | 7 days at 2-8°C; 3 month at -20°C | Same |
| Unopened Stability | until the expiration date shown on the box and bottle labels when stored in the refrigerator at 2-8°C | Same |
| | ACE Uric Acid Reagent (Candidate device) | ACE Uric Acid Reagent (Predicate, k930104) |
| --- | --- | --- |
| Intended Use/Indications for use | The ACE Uric Acid Reagent is intended for the quantitative determination of uric acid concentration in serum. Uric acid measurements are used in the diagnosis and treatment of numerous renal and metabolic disorders, including renal failure, gout, psoriasis, starvation or other wasting conditions and of patients receiving cytotoxic drugs. | Same |
| Instrument/ Platform | ACE Axcel Clinical Chemistry System, | ACE Clinical Chemistry System |
| Calibration & On- Board Stability | 30 days | Same |
| Basic Principle | Coupled Enzymatic method | Same |
| Sample Stability | 3-5 days at 2-8°C
6 months below -20°C. | Same |
| Unopened Stability | Until the expiration date shown on the box and bottle labels when stored in the refrigerator at 2-8°C | Same |
| | ACE CK Reagent (Candidate device) | ACE CK Reagent (Predicate, k930104) |
| --- | --- | --- |
{6}
| Intended Use/Indications for use | The ACE CK Reagent is intended for the quantitative determination of creatine kinase activity in serum. Measurement of creatine kinase is used in the diagnosis and treatment of myocardial and muscle diseases such as progressive, Duchenne-type muscular dystrophy. | Same |
| --- | --- | --- |
| Instrument/Platform | ACE Axcel Clinical Chemistry System | ACE Clinical Chemistry System |
| Basic Principle | Conversion of creatine phosphate to creatine and ATP enzymatically coupled to the formation of NADPH | Same |
| Sample Stability | 4 hours at room temperature
48 hours at -8 °C
20 days below -20°C. | Same |
| Unopened Stability | until the expiration date shown on the box and bottle labels when stored in the refrigerator at 2-8 °C | Same |
# K. Standard/Guidance Document Referenced :
CLSI EP6-A Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach (2003)
CLSI EP5-A2 Evaluation of Precision Performance of Quantitative Measurement Methods; Second Edition (2004)
CLSI EP7-A2 Interference Testing in Clinical Chemistry; Approved Guideline; Second Edition (2005)
CLSI EP10-A3 Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures - Third Edition (2006)
CLSI EP9-A2-IR Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline- Second Edition (2002)
CLSI EPI 17-A Protocol for Determination of Limits of Detection; Second Edition (2004)
# L. Test Principle:
The ACE BUN/Urea Reagent is a photometric assay in which urea in serum is hydrolyzed by urease to yield ammonia and carbon dioxide, the ammonia formed subsequently reacts with 2-oxoglutarate and NADH in the presence of glutamate dehydrogenase (GLDH) to yield glutamate and NAD. NADH absorbs strongly at $340\mathrm{nm}$ , whereas $\mathrm{NAD^{+}}$ does not. The decrease in absorbance of NADH at $340\mathrm{nm}$ is proportional to the urea concentration in the sample.
{7}
The ACE Creatinine Reagent is a photometric assay based on Jaffe reaction, in which serum creatinine reacts with picric acid in an alkaline medium to form a red-orange colored complex, which absorbs strongly at 505 nm. The rate of complex formation is determined by the increase of absorbance at 505 nrn/573 nm during a fixed time interval; this rate is directly proportional to the creatinine concentration in the sample.
The ACE Uric Acid Reagent is based on a photometric assay in which serum uric acid is oxidized by uricase to allantoin and hydrogen peroxide, which subsequently reacts to coupled chromogen dichlorohydroxybenzene sulfonic acid (DHBS) and 4-aminoantipyrine (AAP) in a reaction catalyzed by peroxidase and produces a red colored quinoneimine complex; which are measured by the increase in absorbance at 505 nm/610 nm.
The ACE CK Reagent is a coupled enzymatic reaction in which serum creatine kinase (CK) first catalyzes the conversion of creatine phosphate and adenosine diphosphate (ADP) to creatine and ATP. Subsequently, hexokinase (HK) uses the ATP produced to phosphorylate D-glucose to form D-glucose-6-phosphate and ADP. In the final reaction, the enzyme glucose-6-phosphate dehydrogenase (G-6-PDH) uses D-glucose-6-phosphate to convert nicotinamide adenine dinucleotide phosphate (NADP+) into NADPH, absorbs strongly at 340 nm. The rate of conversion of NADP+ to NADPH, which is a function of the activity of CK in the sample, is determined by monitoring the increase in absorbance at 340 nm/378 nm.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
#### a. Precision/Reproducibility:
**In-House: ACE BUN/Urea Reagent**
Four serum samples (three serum based pool and one normal human serum sample), were tested on one ACE Axcel Clinical Chemistry System two times per run, two runs per day, for a total of 20 or more days. The results are summarized in the table below:
| Sample 1
Mean 15.9 mg/dL
BUN | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.5 | 0.0 | 0.6 | 0.7 |
| Coefficient of
Variation | 2.9% | 0.0% | 3.5% | 4.6% |
{8}
| Sample 2
Mean 52.7 mg/dL
BUN | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.7 | 0.4 | 1.7 | 1.9 |
| Coefficient of
Variation | 1.3% | 0.7% | 3.3% | 3.6% |
| Sample 3
Mean 87.6 mg/dL
BUN | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.9 | 1.0 | 2.7 | 3.0 |
| Coefficient of
Variation | 1.1% | 1.2% | 3.1% | 3.5% |
| Sample 4
Mean 8.0 mg/dL
BUN | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.3 | 0.0 | 0.2 | 0.3 |
| Coefficient of
Variation | 3.3% | 0.0% | 2.7% | 4.3% |
Point of Care Laboratory: ACE BUN/Urea Reagent
Three serum based samples were tested on three ACE Axcel Clinical Chemistry Systems (one at each POL site) at least three times per run, one run per day, for a total of 5 days. The mean, standard deviations and % coefficients of variation (CV) were calculated for each sample.
{9}
9
| BUN/Urea | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (mg/dL) | SD | %CV | SD | %CV |
| POL 1 | 1 | 14.8 | 0.0 | 0.0% | 0.4 | 3.0% |
| POL 2 | 1 | 16.1 | 0.3 | 1.9% | 0.5 | 3.4% |
| POL 3 | 1 | 15.1 | 0.4 | 2.6% | 0.8 | 5.2% |
| | | | | | | |
| POL 1 | 2 | 48.9 | 0.3 | 0.6% | 1.7 | 3.5% |
| POL 2 | 2 | 52.9 | 0.4 | 0.8% | 1.7 | 3.2% |
| POL 3 | 2 | 49.2 | 0.6 | 1.2% | 1.5 | 3.1% |
| | | | | | | |
| POL 1 | 3 | 82.3 | 0.5 | 0.6% | 2.6 | 3.2% |
| POL 2 | 3 | 87.7 | 0.5 | 0.6% | 2.9 | 3.3% |
| POL 3 | 3 | 83.1 | 0.8 | 1.0% | 4.1 | 4.9% |
## In-House: ACE Creatinine Reagent
Four serum samples, three serum based pool and one normal human serum sample, were tested on one ACE Axcel Clinical Chemistry System two times per run, two runs per day, for a total of 22 days. The results are summarized in the table below:
| Sample 1
Mean 1.005 mg/dL
Creatinine | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.053 | 0.000 | 0.029 | 0.060 |
| Coefficient of
Variation | 5.3% | 0.0% | 2.8% | 6.0% |
| Sample 2
Mean 8.366 mg/dL
Creatinine | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.109 | 0.158 | 0.132 | 0.233 |
| Coefficient of
Variation | 1.3% | 1.9% | 1.6% | 2.8% |
| Sample 3
Mean 15.868 mg/dL
Creatinine | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
{10}
| Standard Deviation, mg/dL | 0.224 | 0.132 | 0.350 | 0.436 |
| --- | --- | --- | --- | --- |
| Coefficient of Variation | 1.4% | 0.8% | 2.2% | 2.7% |
| Sample 4
Mean 0.556 mg/dL
Creatinine | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, mg/dL | 0.054 | 0.009 | 0.000 | 0.054 |
| Coefficient of Variation | 9.6% | 1.6% | 0.0% | 9.8% |
## Point of Care Laboratory: ACE Creatinine Reagent
Four serum based samples, three serum-based pools and one normal human serum pool were tested on three ACE Axcel Clinical Chemistry Systems (one at each POL site) at least three times per run, one run per day, for a total of 5 days. The mean, standard deviations and % coefficients of variation (CV) were calculated for each sample.
| Creatinine | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (mg/dL) | SD | %CV | SD | %CV |
| POL 1 | 1 | 1.000 | 0.038 | 3.8% | 0.040 | 4.0% |
| POL 2 | 1 | 0.941 | 0.048 | 5.1% | 0.058 | 6.1% |
| POL 3 | 1 | 0.967 | 0.032 | 3.3% | 0.051 | 5.3% |
| | | | | | | |
| POL 1 | 2 | 8.469 | 0.120 | 1.4% | 0.222 | 2.6% |
| POL 2 | 2 | 8.639 | 0.115 | 1.3% | 0.179 | 2.1% |
| POL 3 | 2 | 8.215 | 0.094 | 1.1% | 0.200 | 2.4% |
| | | | | | | |
| POL 1 | 3 | 16.262 | 0.254 | 1.6% | 0.486 | 3.0% |
| POL 2 | 3 | 16.485 | 0.175 | 1.1% | 0.367 | 2.2% |
| POL 3 | 3 | 15.773 | 0.144 | 0.9% | 0.334 | 2.1% |
## In-House: ACE Uric Acid Reagent
Four serum samples, (three serum based pool and one normal human serum sample), were tested on one ACE Axcel Clinical Chemistry System two times per run, two runs per day, for a total of 22 days. The results are summarized in the table below:
{11}
| Sample 1
Mean 2.90 mg/dL
Uric Acid | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.14 | 0.00 | 0.06 | 0.16 |
| Coefficient of
Variation | 4.9% | 0.0% | 2.1% | 5.4% |
| Sample 2
Mean 7.74 mg/dL
Uric Acid | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.18 | 0.00 | 0.08 | 0.20 |
| Coefficient of
Variation | 2.4% | 0.0% | 1.1% | 2.6% |
| Sample 3
Mean 12.34 mg/dL
Uric Acid | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.23 | 0.00 | 0.10 | 0.25 |
| Coefficient of
Variation | 1.8% | 0.0% | 0.8% | 2.0% |
| Sample 4
Mean 4.02 mg/dL
Uric Acid | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
mg/dL | 0.10 | 0.08 | 0.08 | 0.15 |
| Coefficient of
Variation | 2.6% | 2.0% | 2.0% | 3.8% |
**Point of Care Laboratory: ACE Uric Acid Reagent**
Four serum based samples, three serum-based pools and one normal human serum pool were tested on three ACE Axcel Clinical Chemistry Systems (one at each POL site) at least three times per run, one run per day, for a total of 5 days. The mean, standard deviations and % coefficients of variation (CV) were calculated for each sample.
{12}
12
| Uric Acid | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (mg/dL) | SD | %CV | SD | %CV |
| POL 1 | 1 | 2.74 | 0.12 | 4.4% | 0.14 | 5.2% |
| POL 2 | 1 | 2.95 | 0.09 | 3.0% | 0.09 | 3.0% |
| POL 3 | 1 | 2.84 | 0.10 | 3.6% | 0.10 | 3.6% |
| | | | | | | |
| POL 1 | 2 | 7.65 | 0.15 | 2.0% | 0.17 | 2.2% |
| POL 2 | 2 | 7.69 | 0.29 | 3.8% | 0.29 | 3.8% |
| POL 3 | 2 | 7.78 | 0.14 | 1.8% | 0.14 | 1.8% |
| | | | | | | |
| POL 1 | 3 | 12.45 | 0.19 | 1.5% | 0.29 | 2.3% |
| POL 2 | 3 | 12.53 | 0.26 | 2.1% | 0.26 | 2.1% |
| POL 3 | 3 | 12.55 | 0.23 | 1.8% | 0.23 | 1.8% |
## In-House: ACE CK Reagent
Four serum samples, three serum based pool and one normal human serum sample, were tested on one ACE Axcel Clinical Chemistry System two times per run, two runs per day, for a total of 22 days. The results are summarized in the table below:
| Sample 1
Mean 84.9 U/L
Creatine Kinase | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 2.5 | 0.0 | 0.7 | 2.6 |
| Coefficient of Variation | 2.9% | 0.0% | 0.8% | 3.0% |
| Sample 2
Mean 658.8 U/L
Creatine Kinase | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 11.5 | 0.0 | 6.1 | 13.0 |
| Coefficient of Variation | 1.7% | 0.0% | 0.9% | 2.0% |
{13}
13
| Sample 3
Mean 1182.9 U/L
Creatine Kinase | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
U/L | 13.4 | 10.3 | 9.1 | 19.2 |
| Coefficient of Variation | 1.1% | 0.9% | 0.8% | 1.6% |
| Sample 4
Mean 99.1 U/L
Creatine Kinase | Within
Run | Between
Run | Between
Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation,
U/L | 5.2 | 0.0 | 1.6 | 5.4 |
| Coefficient of Variation | 5.2% | 0.0% | 1.6% | 5.5% |
## Point of Care Laboratory: ACE CK Reagent
Four serum based samples, three serum-based pools and one normal human serum pool were tested on three ACE Axcel Clinical Chemistry Systems (one at each POL site) at least three times per run, one run per day, for a total of 5 days. The mean, standard deviations and % coefficients of variation (CV) were calculated for each sample.
| Creatine Kinase | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean (U/L) | SD | %CV | SD | %CV |
| POL 1 | 1 | 82.1 | 2.2 | 2.7% | 2.2 | 2.7% |
| POL 2 | 1 | 86.8 | 2.4 | 2.8% | 3.9 | 4.5% |
| POL 3 | 1 | 83.6 | 1.4 | 1.7% | 1.9 | 2.2% |
| | | | | | | |
| POL 1 | 2 | 618.3 | 23.0 | 3.7% | 29.5 | 4.8% |
| POL 2 | 2 | 659.4 | 9.6 | 1.5% | 20.8 | 3.2% |
| POL 3 | 2 | 636.4 | 15.5 | 2.4% | 15.7 | 2.5% |
| | | | | | | |
| POL 1 | 3 | 1123.0 | 20.4 | 1.8% | 21.1 | 1.9% |
| POL 2 | 3 | 1170.5 | 26.7 | 2.3% | 3.6 | 3.1% |
| POL 3 | 3 | 1153.5 | 15.3 | 1.3% | 18.1 | 1.6% |
{14}
# b. Linearity/assay reportable range:
# ACE BUN/ Urea Reagent
Linearity studies were carried out using dilutions of a spiked serum samples. Twelve concentrations were prepared by mixing spiked serum samples in known proportion with saline. All samples were measured in triplicate. The sample range tested was 0.03 to $100.3\mathrm{mg / dL}$ . The linear regression between the expected values and the measured values yield the following correlations:
| Claimed Measuring Range | Intercept | Slope | r2 |
| --- | --- | --- | --- |
| 2 to 100 mg/dL | 1.4 | 0.994 | 0.9984 |
Based on the linearity data, the measuring range claimed from 2 to $100\mathrm{mg / dL}$ was supported.
# ACE Creatinine Reagent
Linearity studies were carried out using dilutions of a spiked serum samples. Fourteen concentrations were prepared by mixing spiked serum samples in known proportion with saline. All samples were measured in triplicate. The sample range tested was 0.12 to $24.75\mathrm{mg / dL}$ . The linear regression between the expected values and the measured values yield the following correlations:
| Claimed Measuring Range | Intercept | Slope | r2 |
| --- | --- | --- | --- |
| 0.23 to 25.0 mg/dL | 0.05 | 0.990 | 0.9996 |
Based on the linearity data, the measuring range claimed from 0.23 to $25.0\mathrm{mg / dL}$ was supported.
# ACE Uric Acid Reagent
Linearity studies were carried out using dilutions of a spiked serum samples. Fourteen concentrations were prepared by mixing spiked serum samples in known proportion with saline. All samples were measured in triplicate. The sample range tested was 0.43 to $17.43\mathrm{mg / dL}$ . The linear regression between the expected values and the measured values yield the following correlations:
| Claimed Measuring Range | Intercept | Slope | r2 |
| --- | --- | --- | --- |
| 1.2 to 16.0 mg/dL | -0.16 | 0.968 | 0.9964 |
Based on the linearity data, the measuring range claimed from 1.2 to $16.0\mathrm{mg / dL}$ was supported
# ACE CK Reagent
{15}
Linearity studies were carried out using dilutions of a spiked serum samples. Eight concentrations were prepared by mixing spiked serum samples in known proportion with saline. All samples were measured in triplicate. The sample range tested was 3.5 to 1547.5 U/L. The linear regression between the expected values and the measured values yield the following correlations:
| Claimed Measuring Range | Intercept | Slope | r² |
| --- | --- | --- | --- |
| 11 to 1350 U/L | -0.363 | 0.9987 | 0.9997 |
Based on the linearity data, the measuring range claimed from 11 to 1350 U/L was supported.
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
**Traceability:**
GEMCAL Reference Serum is traceable to NIST SRM 909 for BUN and uric acid and IDMS traceable to the NIST SRM 967 for creatinine.
For ACE CK reagent, enzyme activity (in IU/L) is determined directly by multiplying the absorbance change per minute of the unknown sample by a factor (10956) derived from the molar absorptivity of NADPH.
**On-Board stability:**
On-board stability studies were performed for all reagents by the sponsor. The stability protocols and acceptance criteria were reviewed and determined to be adequate. All ACE reagents in this submission demonstrated the on-board stability as stated in the table below when stored in the ACE Axcel reagent chamber (at 10-14°C) with Evap-Caps.
| Reagent | ACE Axcel On-Board |
| --- | --- |
| | Stability (Days) |
| BUN | 30 |
| Creatinine | 10 |
| Uric Acid | 30 |
| Creatinine Kinase | 25 |
d. Detection limit:
Detection limit studies (LoB, LoD, and LoQ) were performed in accordance with CLSI Guidance Document EP17-A. For LoB and LoD, testing was carried out using
{16}
blanks and low samples (total of 60 replicates each sample) over three days on two ACE Axcel Clinical Chemistry Systems. LoQ testing was performed using 5 low samples (total of 40 replicates for each sample). A curve was fit to estimate the relationship between the measured value and the $\% \mathrm{CV}$ . LoQ is defined as the lowest concentration for which the imprecision $\% \mathrm{CV}$ is $\leq 20\%$ .
The sponsor's claimed detection limits for the four analytes are summarized below:
| Reagent | Limit of Blank (LoB) | Limit of Detection (LoD) | Limit of Quantification (LoQ) |
| --- | --- | --- | --- |
| BUN | 0.9 mg/dL | 1.1 mg/dL | 1.9 mg/dL |
| Creatinine | 0.14 mg/dL | 0.19 mg/dL | 0.23 mg/dL |
| Uric Acid | 0.97 mg/dL | 1.13 mg/dL | 1.2 mg/dL |
| CK | 3.2 U/L | 3.5 U/L | 10.6 U/L |
The sponsor's claimed measuring range for BUN is 2 to $100\mathrm{mg / dL}$ , for creatinine is 0.23 to $25.0\mathrm{mg / dL}$ , for uric acid is 1.2 to $16.0\mathrm{mg / dL}$ , and for CK is 11 to $1350\mathrm{U / L}$ .
# e. Analytical specificity:
# ACE BUN/ Urea Reagent
Interference studies were performed by using two serum pools containing $20\mathrm{mg / dL}$ and $50\mathrm{mg / dL}$ BUN with individual interferents at a range of concentrations. The sera were assayed for BUN $(n = 3$ replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery changed by $\pm 10\%$ . The results reported were obtained on the ACE Axcel Clinical Chemistry System analyzer.
| Interferent | No Significant Interference At or Below: |
| --- | --- |
| Unconjugated Bilirubin | 54 mg/dL |
| Hemolysis | 1000 mg/dL |
| Lipemia (Intralipid) | 1000 mg/dL |
| Ascorbic Acid | 6 mg/dL |
# ACE Creatinine Reagent
Interference studies were performed by using two serum pools containing $1\mathrm{mg / dL}$ and $9\mathrm{mg / dL}$ creatinine with individual interferents at a range of concentrations. The sera were assayed for creatinine $(n = 3$ replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery
{17}
changed by ± 10%. The results reported were obtained on the ACE Axcel Clinical Chemistry System analyzer.
| Interferent | No Significant Interference At or Below: |
| --- | --- |
| Unconjugated Bilirubin | 6.2 mg/dL |
| Hemolysis | 500 mg/dL |
| Lipemia (Intralipid) | 1000 mg/dL |
| Ascorbic Acid | 6 mg/dL |
## ACE Uric Acid Reagent
Interference studies were performed by using two serum pools containing 4 mg/dL and 11 mg/dL uric acid with individual interferents at a range of concentrations. The sera were assayed for uric acid (n = 3 replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery changed by ± 10%. The results reported were obtained on the ACE Axcel Clinical Chemistry System analyzer.
| Interferent | No Significant Interference At or Below: |
| --- | --- |
| Unconjugated Bilirubin | 32 mg/dL |
| Hemolysis | 62.5 mg/dL |
| Triglycerides | 325 mg/dL |
| Ascorbic Acid | 1.34 mg/dL |
The sponsor has the following limitation in the labeling due to hemoglobin interference, "Hemolyzed samples cannot be used. Use clear, unhemolyzed sample."
For ascorbic acid, the following limitation was noted in the labeling: "Ascorbic acid concentrations as low as 1.3 mg/dL have been shown to interfere with this assay. Non-fasting patients taking a high dose of vitamin C may cause low uric acid levels and results should be interpreted with caution."
## ACE CK Reagent
Interference studies were performed by using two serum pools containing 100 U/L and 600 U/L creatine kinase with individual interferents at a range of concentrations. The sera were assayed for creatine kinase (n = 3 replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery changed by ± 10%. The results reported were obtained on the ACE Axcel
{18}
Clinical Chemistry System analyzer.
| Interferent | No Significant Interference At or Below: |
| --- | --- |
| Unconjugated Bilirubin | 28 mg/dL |
| Hemolysis | 125 mg/dL |
| Lipemia (Intralipid) | 1000 mg/dL |
| Ascorbic Acid | 6 mg/dL |
The sponsor has the following limitation in the labeling due to hemoglobin interference, "Hemolyzed samples cannot be used. Use clear, unhemolyzed sample."
f. Assay cut-off:
Not applicable
2. Comparison studies:
Studies were carried out according to CLSI EP09-A2-IR.
In house: ACE BUN/ Urea Reagent
One hundred thirteen serum samples were assayed in parallel by both the candidate and predicate methods. The results were analyzed by using Deming regression. The range tested was 4 to 96 mg/dL. Altered samples were included in the study (no more than 10%).
The comparison by Deming regression resulted in a slope of 1.012 (95%CI = 0.995 to 1.028), an intercept of 0.2 (95%CI = -0.3 to 0.6), correlation coefficient (R²) = 0.9963, and a standard error of 1.5.
Point of Care Laboratory: ACE BUN/ Urea Reagent
Serum samples were assayed in parallel by both the candidate and predicate methods at three POC sites. The results were analyzed by using Deming regression. (Some altered samples were used among all three sites.)
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
{19}
19
| 1 | 59 | 4-97 | y = 0.996x - 0.3 | 0.9988 | 1.0 | 0.983 to 1.009 | -0.7 to 0.1 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 2 | 50 | 2-100 | y = 1.022x + 0.6 | 0.9982 | 1.6 | 1.003 to 1.038 | -0.1 to 1.2 |
| 3 | 45 | 3-98 | y = 1.020x + 1.1 | 0.9987 | 1.1 | 1.004 to 1.036 | 0.5 to 1.6 |
## In house: ACE Creatinine Reagent
One hundred thirty six serum samples were assayed in parallel by both the candidate and predicate methods. The results were analyzed by using Deming regression. The range tested was 0.28 to 22.95 mg/dL. Altered samples were included in the study (no more than 10%).
The comparison by Deming regression resulted in a slope of 0.979 (95%CI = 0.975 to 0.983), an intercept of -0.006 (95%CI = -0.022 to 0.010), correlation coefficient (R²) = 0.9998, and a standard error of 0.082.
## Point of Care Laboratory: ACE Creatinine Reagent
Serum samples were assayed in parallel by both the candidate and predicate methods at three POC sites. The results were analyzed by using Deming regression. (Some altered samples were used among all three sites.)
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 62 | 0.34-24.29 | y = 1.022x - 0.036 | 0.9998 | 0.123 | 1.016 to 1.027 | -0.072 to 0.001 |
| 2 | 68 | 0.47-22.98 | y = 0.969x - 0.040 | 0.9995 | 0.129 | 0.961 to 0.976 | -0.077 to -0.004 |
| 3 | 52 | 0.65-22.28 | y = 1.006x - 0.073 | 0.9994 | 0.192 | 0.996 to 1.015 | -0.136 to -0.009 |
## In house: ACE Uric acid Reagent
One hundred and six serum samples were assayed in parallel by both the candidate and predicate methods. The results were analyzed by using Deming regression. The range tested was 1.7 to 15.9 mg/dL. Altered samples were included in the study (no more than 10%).
{20}
The comparison by Deming regression resulted in a slope of 1.042 (95%CI = 1.023 to 1.060), an intercept of - 0.06 (95%CI = -0.18 to 0.07), correlation coefficient (R²) = 0.9958, and a standard error of 0.23.
## Point of Care Laboratory: ACE Uric acid Reagent
Serum samples were assayed in parallel by both the candidate and predicate methods at three POC sites. The results were analyzed by using Deming regression. (Some altered samples were used among all three sites.)
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 52 | 1.4-15.6 | y = 1.024x - 0.11 | 0.9961 | 0.25 | 0.998 to 1.050 | -0.28 to 0.05 |
| 2 | 72 | 2.1-13.6 | y = 1.013x - 0.02 | 0.9858 | 0.44 | 0.972 to 1.053 | -0.31 to 0.28 |
| 3 | 45 | 2.4-15.8 | y = 1.026x - 0.08 | 0.9961 | 0.22 | 0.998 to 1.054 | -0.27 to 0.11 |
## In house: ACE CK Reagent
One hundred nineteen serum samples were assayed in parallel by both the candidate and predicate methods. The results were analyzed by using Deming regression. The range tested was 16 to 1345 U/L. Altered samples were included in the study (no more than 10%).
The comparison by Deming regression resulted in a slope of 1.016 (95%CI = 1.009 to 1.023), an intercept of - 0.1 (95%CI = -2.6 to 2.4), correlation coefficient (R²) = 0.9994, and a standard error of 10.6.
## Point of Care Laboratory: ACE CK Reagent
Serum samples were assayed in parallel by both the candidate and predicate methods at three POC sites. The results were analyzed by using Deming regression. (Some altered samples were used among all three sites.)
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
{21}
| 1 | 62 | 28-1325 | y = 1.048x - 2.9 | 0.9997 | 10.9 | 1.041 to 1.055 | -6.4 to 0.6 |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 2 | 46 | 21-1165 | y = 1.044x - 3.0 | 0.9997 | 5.9 | 1.036 to 1.052 | -5.2 to -0.8 |
| 3 | 48 | 15-1281 | y = 0.996x + 0.9 | 0.9997 | 7.2 | 0.988 to 1.003 | -1.8 to 3.5 |
3. Clinical studies:
Not applicable
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
BUN: 6-20 mg/dL
Creatinine:
Male: 0.9-1.3 mg/dL
Female: 0.6-1.1 mg/dL
Uric acid:
Male: 3.5 - 7.2 mg/dL
Female: 2.6 - 6.0 mg/dL
CK:
Male: 38 - 174 U/L
Female: 26 - 140 U/L
Reference: Tietz N.W..Clinical Guide to Laboratory Tests 3rd Ed (W.B. Saunders, 1995)
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.