The ACE ALT Reagent is intended for the quantitative determination of alanine aminotransferase activity in serum using the ACE Axcel Clinical Chemistry System. Alanine aminotransferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE AST Reagent is intended for the quantitative determination of aspartate aminotransferase activity in serum using the ACE Axcel Clinical Chemistry System. Measurements of aspartate aminotransferase are used in the diagnosis and treatment of certain types of liver and heart disease. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE γ-GT Reagent is intended for the quantitative determination of gamma-glutamyltransferase activity in serum using the ACE γ-GT Reagent is intended for the quantitative determination of gamma-glutamyltransferase activity in serum using the ACE Axcel Clinical Chemistry System. Gamma-glutamyltransferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Device Story
Reagents for ACE Axcel Clinical Chemistry System; measure enzyme activity in serum samples. ALT assay: converts L-alanine/α-ketoglutarate to L-glutamate/pyruvate; LDH-catalyzed oxidation of reduced cofactor monitored via absorbance decrease at 340/647 nm. AST assay: converts L-aspartate/α-ketoglutarate to oxalacetate/L-glutamate; MDH-catalyzed reduction of oxalacetate with NADH oxidation monitored at 340/647 nm. γ-GT assay: transfers γ-glutamyl group from substrate to glycylglycine; product absorbance increase monitored at 408/486 nm. Used in clinical/physician office labs by trained personnel. Output is enzyme activity level (U/L) displayed on system interface; assists clinicians in diagnosing/monitoring liver and heart conditions.
Clinical Evidence
Bench testing only. Precision studies (22 days, 4 levels) and POL site studies (5 days) performed. Accuracy evaluated via correlation studies (n=102 for ALT, n=117 for AST, n=128 for γ-GT) comparing ACE Axcel to ACE system; correlation coefficients >0.999. Detection limits: 3.1 U/L (ALT), 1.5 U/L (AST), 2.7 U/L (γ-GT).
Technological Characteristics
Reagents consist of substrate and coenzyme/buffer bottles. ALT: L-alanine, α-ketoglutarate, NADH, LDH, Tris buffer. AST: L-aspartate, α-ketoglutarate, NADH, MDH, LDH, Tris buffer. γ-GT: glycylglycine, L-γ-glutamyl-3-carboxy-4-nitroanilide. Sensing principle: bichromatic spectrophotometry (absorbance monitoring). System: ACE Axcel Clinical Chemistry System.
Indications for Use
Indicated for quantitative determination of ALT, AST, and γ-GT activity in serum for diagnosis/treatment of liver and heart diseases. Intended for use in clinical laboratories or physician office laboratories (point-of-care).
Regulatory Classification
Identification
An alanine amino transferase (ALT/SGPT) test system is a device intended to measure the activity of the enzyme alanine amino transferase (ALT) (also known as a serum glutamic pyruvic transaminase or SGPT) in serum and plasma. Alanine amino transferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases.
Predicate Devices
Alfa Wassermann ACE plus ISE/Clinical Chemistry System; ACE Reagents (K931786)
Submission Summary (Full Text)
{0}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE
A. 510(k) Number:
k113382
B. Purpose for Submission:
New device
C. Measurand:
Alanine Aminotransferase activity
Aspartate Aminotransferase activity
Gamma-glutamyltransferase activity
D. Type of Test:
Quantitative, photometric assay
E. Applicant:
Alfa Wasserman Diagnostic Technologies, LLC
F. Proprietary and Established Names:
ACE ALT Reagent
ACE AST Reagent
ACE γ-GT Reagent
G. Regulatory Information:
1. Regulation section:
21 CFR § 862.1030, Alanine Aminotransferase (ALT/SGPT) test system
21 CFR § 862.1035, Aspartate Aminotransferase (AST/SGOT) test system
21 CFR § 862.1360, Gamma-glutamyltranspeptidase and isoenzymes test system
2. Classification:
Class I meets limitations of exemptions in 862.9(c)(9), for ALT and GGT
Class II for AST
{1}
3. Product code:
CKA
CIT
JPZ
4. Panel:
75 (Clinical Chemistry)
H. Intended Use:
1. Intended use(s):
See Indications for use below.
2. Indication(s) for use:
The ACE ALT Reagent is intended for the quantitative determination of alanine aminotransferase activity in serum using the ACE Axcel Clinical Chemistry System. Alanine aminotransferase measurements are used in the diagnosis and treatment of certain liver diseases (e.g., viral hepatitis and cirrhosis) and heart diseases. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE AST Reagent is intended for the quantitative determination of aspartate aminotransferase activity in serum using the ACE Axcel Clinical Chemistry System. Measurements of aspartate aminotransferase are used in the diagnosis and treatment of certain types of liver and heart disease. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
The ACE γ-GT Reagent is intended for the quantitative determination of gamma-glutamyltransferase activity in serum using the ACE Axcel Clinical Chemistry System. Gamma-glutamyltransferase measurements are used in the diagnosis and treatment of liver diseases such as alcoholic cirrhosis and primary and secondary liver tumors. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
3. Special conditions for use statement(s):
For in vitro diagnostic use only.
For prescription use and Point-of-care settings
2
{2}
4. Special instrument requirements:
ACE Axcel Clinical Chemistry System
I. Device Description:
The ACE reagent kits, used with the ACE Axcel Clinical Chemistry System, consist of natural or brown plastic bottles containing liquid-stable reagents. The reagents are side-labeled to identify them to the user and have a dot code label applied to the bottom of the bottle to identify each bottle to the ACE Axcel via cameras, located in the bottom section of the reagent compartment. Reagent kits typically have either one reagent (R1) or sometimes two (R1, R2) or more different reagent bottles per kit. For example, the AST Reagent kit has both R1 (buffer) and R2 (substrate) bottles (three R1 and three R2 bottles) in each kit.
The ACE ALT Reagent consists of two reagent bottles (Substrate and Coenzyme). The combined AST Reagent consist of L-Alanine, $\alpha$-Ketoglutarate, Nicotinamide adenine dinucleotide, reduced (NADH), Lactate dehydrogenase (rabbit muscle), Tris, Preservative and Stabilizers
The ACE AST Reagent consists of two reagent bottles (Substrate and Coenzyme). The combined AST Reagent consists of L-Asprurate, $\alpha$-Ketoglutarate, Nicotinamide adenine dinucleotide, reduced (NADH), Malate dehydrogenase (microbial), Lactate dehydrogenase (Porcine), Tris, Preservative and Stabilizers.
The ACE $\gamma$-GT Reagent consists of two reagent bottles (Buffer and Substrate). The Buffer Reagent (R1) contains Glycylglycine, Preservative and stabilizers. The Substrate Reagent (R2) contains L-y-glutamyl-3-carboxy-4-nitroanilide, Buffer, Preservative and Stabilizers.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACE Clinical Chemistry System, ACE ALT Reagent
ACE Clinical Chemistry System, ACE AST Reagent
ACE Clinical Chemistry System, ACE $\gamma$-GT Reagent
2. Predicate 510(k) number(s):
k931786
3. Comparison with predicate:
{3}
| Item | ACE plus ISE/Clinical Chemistry System, ACE ALT Reagent (Candidate Device - k113382) | ACE plus ISE/Clinical Chemistry System, ACE ALT Reagent (Predicate Device - k931786) |
| --- | --- | --- |
| Intended Use/Indications for Use | It is intended for the quantitative determination of alanine aminotransferase activity in serum. | Same |
| Instrument | ACE Axcel Clinical Chemistry System | ACE and ACE Alera® Clinical Chemistry Systems |
| Sample type | Serum | Same |
| Reagent type | Two part liquid | Same |
| Reaction type | Kinetic | Same |
| Measuring range | 5 to 480 U/L | Same |
| On-Board Stability | 30 Days | Same |
| Shelf life | Stable until expiration date on box when stored at 2-8°C | Same |
| Item | ACE Axcel Clinical Chemistry System, ACE AST Reagent (Candidate Device - k113382) | ACE plus ISE/Clinical Chemistry System, ACE AST Reagent (Predicate Device - k931786) |
| --- | --- | --- |
| Intended Use/Indications for Use | It is intended for the quantitative determination of aspartate aminotransferase activity in serum. | Same |
| Instrument | ACE Axcel Clinical Chemistry System | ACE and ACE Alera® Clinical Chemistry Systems |
| Sample type | Serum | Same |
| Reagent type | Two part liquid | Same |
| Reaction type | Kinetic | Same |
| Measuring range | 7 to 450 U/L | Same |
| On-Board Stability | 30 Days | Same |
| Shelf life | Stable until expiration date on box when stored at 2-8°C | Same |
| Item | ACE Axcel Clinical Chemistry System, ACE γ-GT Reagent (Candidate Device - k113382) | ACE plus ISE/Clinical Chemistry System, ACE γ-GT Reagent (Predicate Device - k931786) |
| --- | --- | --- |
| Intended Use/Indications for Use | It is intended for the quantitative determination of gamma-glutamyltransferase activity in serum | Same |
| Instrument | ACE Axcel Clinical Chemistry System | ACE and ACE Alera® Clinical Chemistry Systems |
| Sample type | Serum | Same |
| Reagent type | Two part liquid | Same |
| Reaction type | Kinetic | Same |
| Measuring range | 7 to 950 U/L | 3 to 950 U/L |
| On-Board Stability | 30 Days | Same |
| Shelf life | Stable until expiration date on box when stored at 2-8°C | Same |
{4}
K. Standard/Guidance Document Referenced (if applicable):
- CLSI EP6-A: Evaluation of Linearity of Quantitative Analytical Methods
- CLSI EP7-A2: Method Comparison and Bias Estimation Using Patient Samples
- CLSI EP5-A2: Evaluation of Precision Performance of Clinical Chemistry Devices
- CLSI EP10-A3: Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures
- CLSI EP-A2-IR: Method Comparison and Bias Estimation Using Patient Samples, CLSI EP17-A: Protocols for Determination of Limits of Detection
L. Test Principle:
ACE ALT Reagent
Alanine aminotransferase in serum converts the L-alanine and α-ketoglutarate substrates, contained in the reagent, to L-glutamate and pyruvate. The pyruvate formed reacts with reduced nicotinamide adenine dinucleotide (NADH) in the presence of lactate dehydrogenase to form lactic acid and oxidized nicotinamide adenine dinucleotide (NAD+). The rate of conversion of the reduced cofactor to the oxidized cofactor results in a decrease in absorbance at 340 nm.
ACE AST Reagent
Aspartate aminotransferase in serum converts the L-aspartate and α-ketoglutarate substrates, contained in the reagent, to L-glutamate and oxalacetate. The oxalacetate formed reacts with reduced nicotinamide adenine dinucleotide (NADH) in the presence of malate dehydrogenase to form malate and oxidized nicotinamide adenine dinucleotide (NAD+). The rate of conversion of the reduced cofactor to the oxidized cofactor results in a decrease in absorbance at 340 nm. This rate of conversion is a proportional to the activity of AST in the sample.
ACE γ-GT Reagent
γ-GT in serum catalyzes the transfer of the γ-glutamyl group from the L-γ-glutamyl-3-carboxy-4-nitroanilide substrate to the glycylglycine product in the reagent. The product, 5-amino-2-nitrobenzoate, absorbs strongly at 408 nm. The rate of increase in absorbance is directly proportional to the γ-GT activity in the sample.
M. Performance Characteristics (if/when applicable):
1. Analytical performance:
a. Precision/Reproducibility:
In house: ACE ALT Reagent
Within-Run and Total precision evaluations were determined following CLSI EP5-A2. Four levels, three serum based pool and one normal human serum sample, were tested on the ACE Axcel Clinical Chemistry System analyzer in
{5}
two runs per day, with two replicates of each level per run for 22 days.
| Sample 1
Mean 30.0 U/L ALT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 2.1 | 0.0 | 0.0 | 2.1 |
| Coefficient of Variation | 6.9% | 0.0% | 0.0% | 6.9% |
| Sample 2
Mean 265.3 U/L ALT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 5.5 | 0.0 | 1.2 | 5.7 |
| Coefficient of Variation | 2.1% | 0.0% | 0.4% | 2.1% |
| Sample 3
Mean 389.2 U/L ALT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 3.2 | 2.3 | 2.2 | 4.5 |
| Coefficient of Variation | 0.8% | 0.6% | 0.6% | 1.1% |
| Sample 4*
Mean 55.3 U/L ALT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 0.4 | 0.6 | 1.1 | 1.3 |
| Coefficient of Variation | 0.8% | 1.2% | 2.0% | 2.4% |
*Data collected for 21 days.
## Point of Care Laboratory: ACE ALT Reagent
Point of care precision evaluations were determined following CLSI EP10-A3. Three serum based pools were tested on the ACE Axcel Clinical Chemistry System analyzer in one run per day, with three replicates of each level per run for 5 days.
{6}
7
| ALT | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 28.9 | 2.3 | 7.9 | 2.3 | 7.9 |
| POL 2 | 1 | 29.8 | 2.6 | 8.7 | 2.6 | 8.7 |
| POL 3 | 1 | 30.9 | 1.7 | 5.4 | 1.7 | 5.4 |
| | | | | | | |
| POL 1 | 2 | 256.5 | 5.9 | 2.3 | 5.9 | 2.3 |
| POL 2 | 2 | 263.3 | 5.4 | 2.1 | 6.1 | 2.3 |
| POL 3 | 2 | 261.4 | 2.1 | 0.8 | 4.4 | 1.7 |
| | | | | | | |
| POL 1 | 3 | 474.9 | 7.4 | 1.6 | 7.4 | 1.6 |
| POL 2 | 3 | 383.9 | 3.7 | 1.0 | 4.9 | 1.3 |
| POL 3 | 3 | 380.1 | 3.8 | 1.0 | 5.5 | 1.4 |
## In house: ACE AST Reagent
Within-Run and Total precision evaluations were determined following CLSI EP5-A2. Four levels, three serum based pool and one normal human serum sample, were tested on the ACE Axcel Clinical Chemistry System analyzer in two runs per day, with two replicates of each level per run for 22 days.
| Sample 1
Mean 23.9 U/L AST | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 1.7 | 1.1 | 0.0 | 2.0 |
| Coefficient of Variation | 7.1% | 4.8% | 0.0% | 8.5% |
| Sample 2
Mean 219.6 U/L AST | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 6.0 | 0.0 | 0.0 | 6.0 |
| Coefficient of Variation | 2.7% | 0.0% | 0.0% | 2.7% |
| Sample 3
Mean 398.4 U/L AST | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 5.2 | 0.4 | 2.1 | 5.7 |
| Coefficient of Variation | 1.3% | 0.1% | 0.5% | 1.4% |
| Sample 4*
Mean 115.2 U/L AST | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 1.1 | 0.8 | 1.8 | 2.3 |
| Coefficient of Variation | 0.9% | 0.7% | 1.6% | 2.0% |
*Data collected for 21 days.
{7}
Point of Care Laboratory: ACE AST Reagent
Point of care precision evaluations were determined following CLSI EP10-A3. Three serum based pools were tested on the ACE Axcel Clinical Chemistry System analyzer in one run per day, with three replicates of each level per run for 5 days.
| AST | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 23.3 | 2.6 | 11.3 | 2.6 | 11.3 |
| POL 2 | 1 | 25.5 | 2.2 | 8.7 | 2.2 | 8.7 |
| POL 3 | 1 | 23.7 | 1.8 | 7.6 | 2.3 | 9.6 |
| | | | | | | |
| POL 1 | 2 | 211.0 | 3.4 | 1.6 | 3.4 | 1.6 |
| POL 2 | 2 | 218.7 | 5.2 | 2.4 | 5.2 | 2.4 |
| POL 3 | 2 | 220.3 | 4.4 | 2.0 | 4.6 | 2.1 |
| | | | | | | |
| POL 1 | 3 | 383.7 | 4.3 | 1.1 | 7.5 | 2.0 |
| POL 2 | 3 | 397.3 | 4.8 | 1.2 | 4.8 | 1.2 |
| POL 3 | 3 | 394.5 | 6.5 | 1.6 | 8.3 | 2.1 |
In house: ACE γ-GT Reagent
Within-Run and Total precision evaluations were determined following CLSI EP5-A2. Four levels, three serum based pool and one normal human serum sample, were tested on the ACE Axcel Clinical Chemistry System analyzer in two runs per day, with two replicates of each level per run for 22 days.
| Sample 1*
Mean 23.0 U/L γ-GT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 0.7 | 0.9 | 0.9 | 1.4 |
| Coefficient of Variation | 3.0% | 3.7% | 3.8% | 6.1% |
*Data collected for 21 days
| Sample 2
Mean 401.9 U/L γ-GT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 5.0 | 4.4 | 0.0 | 6.7 |
| Coefficient of Variation | 1.2% | 1.1% | 0.0% | 1.7% |
| Sample 3
Mean 751.8 U/L γ-GT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
{8}
| Standard Deviation, U/L | 7.3 | 3.7 | 0.0 | 8.2 |
| --- | --- | --- | --- | --- |
| Coefficient of Variation | 1.0% | 0.5% | 0.0% | 1.1% |
| Sample 4* Mean 28.9 U/L γ-GT | Within Run | Between Run | Between Day | Total |
| --- | --- | --- | --- | --- |
| Standard Deviation, U/L | 0.7 | 1.0 | 0.8 | 1.5 |
| Coefficient of Variation | 2.3% | 3.6% | 2.9% | 5.2% |
*Data collected for 21 days
# Point of Care Laboratory: ACE $\gamma$ -GT Reagent
Point of care precision evaluations were determined following CLSI EP10-A3. Three serum based pools were tested on the ACE Axcel Clinical Chemistry System analyzer in one run per day, with three replicates of each level per run for 5 days.
| γ-GT | | | Within-Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 17.7 | 2.2 | 12.4 | 2.3 | 13.0 |
| POL 2 | 1 | 19.9 | 2.4 | 12.1 | 2.5 | 12.5 |
| POL 3 | 1 | 18.2 | 2.1 | 11.5 | 2.3 | 12.5 |
| | | | | | | |
| POL 1 | 2 | 415.5 | 4.8 | 1.2 | 8.9 | 2.1 |
| POL 2 | 2 | 412.7 | 4.5 | 1.1 | 14.7 | 3.6 |
| POL 3 | 2 | 421.3 | 4.5 | 1.1 | 10.0 | 2.4 |
| | | | | | | |
| POL 1 | 3 | 786.9 | 8.4 | 1.1 | 10.9 | 1.4 |
| POL 2 | 3 | 795.2 | 5.4 | 0.7 | 17.9 | 2.2 |
| POL 3 | 3 | 811.2 | 6.4 | 0.8 | 10.6 | 1.3 |
# b. Linearity/assay reportable range:
# ACE ALT Reagent
Linearity studies were carried out using dilutions of a spiked serum samples. Ten concentrations were prepared by mixing spiked serum samples in known portions of saline. All samples were measured in triplicate. The sample range tested was 4.7 to $480.3\mathrm{U / L}$ .
| Claimed Measuring Range | Intercept | Slope | r2 |
| --- | --- | --- | --- |
| 5 U/L - 480 U/L | 0.477 | 1.030 | 0.9988 |
Based on the linearity data, the measuring range claimed from 5 U/L - 480 U/L
{9}
was supported.
## ACE AST Reagent
Linearity studies were carried out using dilutions of a spiked serum samples. Ten concentrations were prepared by mixing spiked serum samples in known portions of saline. All samples were measured in triplicate. The sample range tested was 4.0 to 453.7 U/L.
| Claimed Measuring Range | Intercept | Slope | r² |
| --- | --- | --- | --- |
| 7 U/L - 450 U/L | 1.9 | 1.013 | 0.9990 |
Based on the linearity data, the measuring range claimed from 7 U/L - 450 U/L was supported.
## ACE g-GT Reagent
Linearity studies were carried out using dilutions of a spiked serum samples. Twelve concentrations were prepared by mixing spiked serum samples in known portions of saline. All samples were measured in triplicate. The sample range tested was 0.3 to 984.7 U/L.
| Claimed Measuring Range | Intercept | Slope | r² |
| --- | --- | --- | --- |
| 7 U/L - 950 U/L | 2.5 | 1.012 | 0.9991 |
Based on the linearity data, the measuring range claimed from 7 U/L - 950 U/L was supported.
c. Traceability, Stability, Expected values (controls, calibrators, or methods): Reagent Stability:
## ACE ALT Reagent
Traceable to another commercially available FDA cleared assay by method comparison.
## ACE AST Reagent
Traceable to another commercially available FDA cleared assay by method comparison.
## ACE γ-GT Reagent
Traceable to another commercially available FDA cleared assay by method comparison.
d. Detection limit:
Studies were carried out in accordance with CLSI Guidance Document EP17-A. Detection Limits (LoB and LoD) were performed using 60 blank and 60 low serum samples and LoQ were performed using 40 low serum samples according to the recommendations of CLSI EP 17A protocol. Sponsor defined
10
{10}
LoQ as the value with precision $< 20\%$ . LoB, LoD and LoQ were calculated to be: ALT: LoB = 2.9 U/L, LoD = 3.1 U/L, LoQ = 3.1 U/L; AST: LoB = 1.4 U/L, LoD = 1.5 U/L, LoQ = 4.0 U/L; $\gamma$ -GT: LoB = 2.4 U/L, LoD = 2.7 U/L, LoQ = 7.0 U/L.
The claimed measuring range of ALT is 5 - 480 U/L, AST is 7 - 450 U/L, and $\gamma$ -GT is 7 - 950 U/L.
# e. Analytical specificity:
Studies were carried out in accordance with CLSI Guidance Document EP7-A2.
# ACE ALT Reagent
Interference studies were performed by using serum pools containing approximately $30~\mathrm{U / L}$ and $350~\mathrm{U / L}$ of ALT with individual interferents at a range of concentrations. The sera were assayed for ALT $(n = 3$ replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery changed by $> \pm 10\%$ . The results reported were obtained on ACE Axcel Clinical Chemistry System analyzer using fresh ACE ALT Reagent.
Interferents Claim
| Interferents | No Significant Interference At or Bellow |
| --- | --- |
| Unconjugated Bilirubin | 30 mg/dL |
| Hemolysis | 1000 mg/dL |
| Lipemia (Intralipid) | 502 mg/dL |
| Ascorbic Acid | 6 mg/dL |
# ACE AST Reagent
Interference studies were performed by using serum pools containing approximately $30~\mathrm{U / L}$ and $230~\mathrm{U / L}$ of AST with individual interferents at a range of concentrations. The sera were assayed for AST $(n = 3$ replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery changed by $> \pm 10\%$ . The results reported were obtained on ACE Axcel Clinical Chemistry System analyzer using fresh ACE AST Reagent.
Interferents Claim
| Interferents | No Significant Interference At or Bellow |
| --- | --- |
| Unconjugated Bilirubin | 58 mg/dL |
| Hemolysis | * |
| Lipemia (Intralipid) | 526 mg/dL |
| Ascorbic Acid | 6 mg/dL |
{11}
* Use clear, unhemolyzed serum. Because erythrocytes contain high levels of AST activity, hemolyzed samples should not be used
## ACE γ-GT Reagent
Interference studies were performed by using serum pools containing approximately 30 U/L and 400 U/L of γ-GT with individual interferents at a range of concentrations. The sera were assayed for γ-GT (n = 3 replicates) and the mean result calculated. Interference was considered to be significant by the sponsor if the analyte recovery changed by > ± 10%. The results reported were obtained on ACE Axcel Clinical Chemistry System analyzer using fresh ACE γ-GT Reagent.
### Interferents Claim
| Interferents | No Significant Interference At or Bellow |
| --- | --- |
| Unconjugated Bilirubin | 18 mg/dl |
| Hemolysis | 62.5 mg/dl* |
| Lipemia (Intralipid) | 1310 mg/dl |
| Acorbic Acid | 6 mg/dl |
Because hemolysis interferes with the GGT assay, the sponsor put the following limitations in the labeling for GGT:
“* Specimens showing any indication of hemolysis should not be analyzed.”
### f. Assay cut-off:
Not applicable
## 2. Comparison studies:
### a. Method comparison with predicate device:
Studies are carried out according to CLSI EP9-A2-IR.
## In house: ACE ALT Reagent
ACE ALT Reagent was used as the predicate method, using recommended applications and procedures on the ACE Clinical Chemistry System. One hundred two serum samples were assayed in parallel by both the test and predicate methods and the results compared by Deming regression. The range tested was 5 to 472 U/L. Seven samples were altered.
The comparison by Deming regression resulted in a slope of 1.041 (95%CI = 1.035 to 1.047), an intercept of 0.3 (95%CI = -0.3 to 0.9), correlation coefficient of R2 = 0.9996, and a std. error of 2.4.
## Point of Care Laboratory: ACE ALT Reagent
{12}
ACE Clinical Chemistry System, ACE ALT Reagent (k931786) was used as the predicate method, using recommended applications and procedures on the ACE Clinical Chemistry System analyzer and calibrating with ACE ALT Calibrators. Results compared by Deming regression. Eighteen samples were altered.
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 40 | 6-457 | y = 1.014x + 1.5 | 0.9999 | 2.4 | 1.009 to 1.019 | 0.6 to 2.4 |
| 2 | 49 | 5-476 | y = 1.019x - 0.1 | 0.9997 | 2.9 | 1.012 to 1.027 | -1.1 to 0.9 |
| 3 | 40 | 6-453 | y = 1.026x + 1.1 | 0.9997 | 3.1 | 1.018 to 1.035 | 0.0 to 2.2 |
In house: ACE AST Reagent
ACE AST Reagent was used as the predicate method, using recommended applications and procedures on the ACE Clinical Chemistry System. One hundred seventeen serum samples were assayed in parallel by both the test and predicate methods and the results compared by Deming regression. The range tested was 8 to 440 U/L. Seven samples were altered.
The comparison by Deming regression resulted in a slope of 1.007 (95%CI = 1.002 to 1.012), an intercept of 2.3 (95%CI = 1.9 to 2.8), correlation coefficient of R2 = 0.9996, and a std. error of 2.2.
Point of Care Laboratory: ACE AST Reagent
ACE Clinical Chemistry System, ACE AST Reagent (k931786) was used as the predicate method, using recommended applications and procedures on the ACE Clinical Chemistry System analyzer and calibrating with ACE AST Calibrators. Results compared by Deming regression. Twenty one samples were altered.
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 51 | 7-409 | y = 1.010x - 0.6 | 0.9998 | 2.5 | 1.005 to 1.015 | -1.4 to 0.3 |
| 2 | 50 | 8-391 | y = 1.030x + 0.4 | 0.9996 | 2.8 | 1.021 to 1.038 | -0.6 to 1.3 |
| 3 | 42 | 7-383 | y = 1.030x + 1.1 | 0.9997 | 2.9 | 1.021 to 1.038 | 0.1 to 2.1 |
In house: ACE γ-GT Reagent
ACE γ-GT Reagent was used as the predicate method, using recommended applications and procedures on the ACE Clinical Chemistry System. One hundred twenty eight serum samples were assayed in parallel by both the test and predicate methods and the results compared by Deming regression. The range tested was 7 to 902 U/L. Four samples were altered.
{13}
The comparison by Deming regression resulted in a slope of 0.984 (95%CI = 0.981 to 0.988), an intercept of 0.1 (95%CI = -0.6 to 0.8), correlation coefficient of R2 = 0.9998, and a std. error of 3.4.
## Point of Care Laboratory: ACE γ-GT Reagent
ACE Clinical Chemistry System, ACE γ-GT Reagent (k931786) was used as the predicate method, using recommended applications and procedures on the ACE Clinical Chemistry System analyzer and calibrating with ACE γ-GT Calibrators. Results compared by Deming regression. Twenty five samples were altered.
| POL | n | Range | Regression Equation | Correlation Coefficient | Standard Error | Confidence Interval Slope | Confidence Interval Intercept |
| --- | --- | --- | --- | --- | --- | --- | --- |
| 1 | 62 | 7-814 | y = 0.970x - 0.6 | 0.9999 | 3.7 | 0.967 to 0.973 | -1.7 to 0.5 |
| 2 | 52 | 13-848 | y = 0.981x + 2.0 | 0.9992 | 8.8 | 0.970 to 0.992 | -0.8 to 5.0 |
| 3 | 49 | 8-912 | y = 1.046x - 0.5 | 0.9997 | 3.6 | 1.039 to 1.053 | -1.6 to 0.7 |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not Applicable
b. Clinical specificity:
Not Applicable
c. Other clinical supportive data (when a. and b. are not applicable):
4. Clinical cut-off:
Not Applicable
5. Expected values/Reference range:
ALT expected values are: 10-40 U/L (Males), 7-35 U/L (Females)¹
AST expected values are: 5-34 U/L²
{14}
$\gamma$ -GT expected values are: 13-68 U/L (Males), 11-48 U/L (Females) $^{3}$
1. Tietz, N. W. (Ed.), Clinical Guide to Laboratory Tests, 3rd Edition, W.B. Saunders Co., Philadelphia, PA (2005).
2. Kaplan L.A., Pesce, A.J., Clinical Chemistry, St. Louis, C.V. Mosby, 911-912 (1989).
3. Reference Intervals-UNC Hospitals, McLendon Clinical Laboratories http://unchealthcare.org/labtestinfo
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
15
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.