ACE ALBUMIN REAGENT ACE TOTAL PROTEIN REAGENT ACE CALCIUM-ARSENAZO REAGENT ACE PHOSPHORUS U.V. REAGENT

K113374 · Alfa Wassermann Diagnostic Technologies, Inc. · CIX · Jul 13, 2012 · Clinical Chemistry

Device Facts

Record IDK113374
Device NameACE ALBUMIN REAGENT ACE TOTAL PROTEIN REAGENT ACE CALCIUM-ARSENAZO REAGENT ACE PHOSPHORUS U.V. REAGENT
ApplicantAlfa Wassermann Diagnostic Technologies, Inc.
Product CodeCIX · Clinical Chemistry
Decision DateJul 13, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1035
Device ClassClass 2

Intended Use

The ACE Albumin Reagent is intended for the quantitative determination of albumin concentration in serum using the ACE Axcel Clinical Chemistry System. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum using the ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum using the ACE Axcel Clinical Chemistry System. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.

Device Story

Reagents for quantitative photometric analysis of serum samples; used on ACE Axcel Clinical Chemistry System. Reagents contain specific chemical components (e.g., Bromcresol Green for albumin; copper sulfate/tartrate for total protein; Arsenazo III for calcium; ammonium molybdate for phosphorus). System performs endpoint reactions at 37°C. Operated by laboratory personnel in clinical or physician office settings. Output is concentration value (g/dL or mg/dL) displayed to clinician; used to aid diagnosis/treatment of liver, kidney, bone, and metabolic diseases. Benefits include standardized, automated diagnostic testing.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (n=88 per level; POL sites n=15 per level). Linearity evaluated per CLSI EP6-A. Interference evaluated per CLSI EP7-A2. Method comparison against predicate (k930104) using Deming regression showed high correlation (r² > 0.98) across all analytes in both in-house and POL studies.

Technological Characteristics

Single-reagent bottle format. Albumin: Bromcresol green/acetate buffer. Total Protein: Copper sulfate, sodium potassium tartrate, potassium iodide, sodium hydroxide. Calcium: Arsenazo III. Phosphorus: Ammonium molybdate, sulfuric acid. Bichromatic spectrophotometric sensing. Standalone reagent kits for use on the ACE Axcel Clinical Chemistry System.

Indications for Use

Indicated for quantitative serum analysis of albumin, total protein, calcium, and inorganic phosphorus in patients with suspected liver, kidney, bone marrow, parathyroid, or metabolic/nutritional disorders. Intended for use in clinical or physician office laboratories.

Regulatory Classification

Identification

An albumin test system is a device intended to measure the albumin concentration in serum and plasma. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K1B374 # 510(k) SUMMARY JUL 1 3 2012 | 510(k) Owner: | Alfa Wassermann Diagnostic Technologies, LLC<br>4 Henderson Drive<br>West Caldwell, NJ 07006 | | | |------------------------|----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|--| | Contact: | Hyman Katz, Ph.D.<br>Phone: 973-852-0158<br>Fax: 973-852-0237 | | | | Date Summary Prepared: | July 11, 2012 | | | | Device: | Trade/Device Name: | ACE Albumin Reagent | | | | Regulation Number: | 21 C.F.R. § 862.1035 | | | | Regulation Name: | Albumin Test System | | | | Regulatory Class: | Class 2 | | | | Product Code: | CIX | | | | Common/Classification Name: | Bromcresol Green Dye-Binding,<br>Albumin | | | | Trade/Device Name: | ACE Total Protein Reagent | | | | Regulation Number: | 21 C.F.R. § 862.1635 | | | | Regulation Name: | Total Protein Test System | | | | Regulatory Class: | Class 2, Exempt,<br>meets limits of exemptions per<br>21 CFR § 862.9 (c)(9) | | | | Product Code: | CEK | | | | Common/Classification Name: | Biuret (Colorimetric), Total Protein | | | | Trade/Device Name: | ACE Calcium-Arsenazo Reagent | | | | Regulation Number: | 21 C.F.R. § 862.1145 | | | | Regulation Name: | Calcium Test System | | | | Regulatory Class: | Class 2 | | | | Product Code: | CJY | | | | Common/Classification Name: | Azo Dye, Calcium | | {1}------------------------------------------------ | Trade/Device Name: | ACE Inorganic Phosphorus U.V.<br>Reagent | |-----------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Regulation Number: | 21 C.F.R. § 862.1580 | | Regulation Name: | Phosphorus (Inorganic) Test System | | Regulatory Class: | Class 1, Reserved | | Product Code: | CEO | | Common/Classification Name: | Phosphomolybdate (Colorimetric),<br>Inorganic Phosphorus | | Predicate Devices: | Manufacturer for analyzer/reagent system predicate:<br>Alfa Wassermann ACE plus ISE/Clinical Chemistry System<br>ACE Reagents (K931786) | | Device Descriptions: | In the ACE Albumin Reagent assay, Bromcresol green binds specifically to<br>albumin to form a green colored complex, which is measured<br>bichromatically at 629 nm/692 nm. The intensity of color produced is<br>directly proportional to the albumin concentration in the sample.<br><br>In the ACE Total Protein Reagent assay, cupric ions react with the peptide<br>bonds of proteins under alkaline conditions to form a violet colored<br>complex which is measured bichromatically at 544 nm/692 nm. The<br>intensity of color produced is directly proportional to the total protein<br>concentration in the sample.<br><br>In the ACE Calcium-Arsenazo Reagent assay, calcium reacts with Arsenazo<br>III in an acidic solution to form a blue-purple colored complex, which is<br>measured bichromatically at 647 nm/692 nm. The intensity of color<br>produced is directly proportional to the calcium concentration in the sample.<br><br>In the ACE Inorganic Phosphorus U.V. Reagent assay, under acidic<br>conditions, inorganic phosphorus in serum reacts with ammonium<br>molybdate to form an unreduced phosphomolybdate complex, which<br>absorbs strongly at 340 nm. The increase in absorbance, measured<br>bichromatically at 340 nm/378 nm, is directly proportional to the amount of<br>phosphorus in the sample. | | Intended Use: | Indications for Use:<br>The ACE Albumin Reagent is intended for the quantitative determination of<br>albumin concentration in serum using the ACE Axcel Clinical Chemistry<br>System. Albumin measurements are used in the diagnosis and treatment of<br>numerous diseases involving primarily the liver or kidneys. This test is<br>intended for use in clinical laboratories or physician office laboratories. For<br>in vitro diagnostic use only | . . ・ : {2}------------------------------------------------ The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum using the ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum using the ACE Axcel Clinical Chemistry System. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. {3}------------------------------------------------ | | Technological<br>Characteristics: | The ACE Albumin Reagent consists of a single reagent bottle. The reagent<br>contains Bromcresol green and acetate buffer. | |--|-----------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | | The ACE Total Protein Reagent consists of a single reagent bottle. The<br>reagent contains copper sulfate, sodium potassium tartrate, potassium iodide<br>and sodium hydroxide. | | | | The ACE Calcium-Arsenazo Reagent consists of a single reagent bottle.<br>The Reagent contains Arsenazo III. | | | | The ACE Inorganic Phosphorus U.V. Reagent consists of a single reagent<br>bottle. The reagent contains ammonium molybdate and sulfuric acid. | | | Performance<br>Data: | Performance data for the Alfa Wassermann ACE Reagents run on the Alfa<br>Wassermann ACE Axcel Clinical Chemistry System included precision,<br>accuracy, and detection limit data. | | | | ACE Albumin Reagent | | | | Precision: In testing conducted at four albumin levels for 22 days, the<br>within-run CV ranged from 0.9 to 1.7%, and total CV ranged from 1.2 to<br>2.0%. In precision studies at three separate Physician Office Laboratory<br>(POL) sites over 5 days, the within-run CV ranged from 0.0 to 1.6% and<br>total CV ranged from 0.0 to 2.3%. | | | | Accuracy: In the correlation study, 118 samples with albumin values<br>ranging from 0.4 to 6.4 g/dL were assayed on the Alfa Wassermann ACE<br>Axcel Clinical Chemistry System (y) and the Alfa Wassermann ACE<br>Clinical Chemistry System (x). Least squares regression analysis yielded a<br>correlation coefficient of 0.9959, a standard error estimate of 0.08, a<br>confidence interval slope of 0.980 to 1.013, and a confidence interval<br>intercept of -0.04 to 0.10. In patient correlation studies at three separate<br>POL sites, using the Alfa Wassermann ACE Axcel Clinical Chemistry<br>System (y) and the Alfa Wassermann ACE Clinical Chemistry System (x),<br>least-squares regression analysis yielded correlation coefficients of 0.9894<br>to 0.9966, standard error estimates of 0.08 to 0.13, confidence interval<br>slopes of 0.946 to 1.037, and a confidence interval intercepts of -0.14 to<br>0.39.<br>Detection limit: The detection limit was 0.09 g/dL. | | | | | | | | ACE Total Protein Reagent | | | | Precision: In testing conducted at four total protein levels for 22 days, the<br>within-run CV ranged from 0.8 to 2.4%, and total CV ranged from 1.0 to<br>2.9%. In precision studies at three separate Physician Office Laboratory<br>(POL) sites over 5 days, the within-run CV ranged from 0.7 to 1.3% and<br>total CV ranged from 0.8 to 1.6%. | : 1999 - 1999 - 1999 {4}------------------------------------------------ Accuracy: In the correlation study, 121 samples with total protein values ranging from 0.4 to 13.5 g/dL were assayed on the Alfa Wassermann ACE Axcel Clinical Chemistry System (y) and the Alfa Wassermann ACE Clinical Chemistry System (x). Least squares regression analysis yielded a correlation coefficient of 0.9977, a standard error estimate of 0.12, a confidence interval slope of 0.978 to 1.002, and a confidence interval intercept of -0.12 to 0.06. In patient correlation studies at three separate POL sites, using the Alfa Wassermann ACE Axcel Clinical Chemistry System (y) and the Alfa Wassermann ACE Clinical Chemistry System (x), least-squares regression analysis vielded correlation coefficients of 0.9932 to 0.9987, standard error estimates of 0.09 to 0.24, confidence interval slopes of 0.973 to 1.047, and a confidence interval intercepts of -0.41 to 0.19. . Detection limit: The detection limit was 0.15 g/dL. ACE Calcium-Arsenazo Reagent Precision: In testing conducted at four calcium levels for 22 days, the within-run CV ranged from 1.3 to 2.3%, and total CV ranged from 1.4 to 2.3%. In precision studies at three separate Physician Office Laboratory (POL) sites over 5 days, the within-run CV ranged from 0.8 to 1.4% and total CV ranged from 1.1 to 2.9%. Accuracy: In the correlation study, 111 samples with calcium values ranging from 0.7 to 14.5 mg/dL were assayed on the Alfa Wassermann ACE Axcel Clinical Chemistry System (y) and the Alfa Wassermann ACE Clinical Chemistry System (x). Least squares regression analysis yielded a correlation coefficient of 0.9935, a standard error estimate of 0.22, a confidence interval slope of 0.998 to 1.042, and a confidence interval intercept of -0.50 to -0.08. In patient correlation studies at three separate POL sites, using the Alfa Wassermann ACE Axcel Clinical Chemistry System (v) and the Alfa Wassermann ACE Clinical Chemistry System (x). least-squares regression analysis yielded correlation coefficients of 0.9895 to 0.9977, standard error estimates of 0.16 to 0.23, confidence interval slopes of 0.969 to 1.075, and a confidence interval intercepts of -0.43 to 0.42. Detection limit: The detection limit was 0.11 mg/dL. ### ACE Inorganic Phosphorus U.V. Reagent Precision: In testing conducted at four phosphorus levels for 22 days, the within-run CV ranged from 1.4 to 1.9%, and total CV ranged from 1.5 to 2.5%. In precision studies at three separate Physician Office Laboratory (POL) sites over 5 days, the within-run CV ranged from 0.6 to 3.2% and total CV ranged from 1.0 to 3.9%. {5}------------------------------------------------ | | Accuracy: In the correlation study, 110 samples with phosphorus values<br>ranging from 0.6 to 19.6 mg/dL were assayed on the Alfa Wassermann<br>ACE Axcel Clinical Chemistry System (y) and the Alfa Wassermann ACE<br>Clinical Chemistry System (x). Least squares regression analysis yielded a<br>correlation coefficient of 0.9983, a standard error estimate of 0.16, a<br>confidence interval slope of 0.994 to 1.017, and a confidence interval<br>intercept of -0.06 to 0.05. In patient correlation studies at three separate<br>POL sites, using the Alfa Wassermann ACE Axcel Clinical Chemistry<br>System (y) and the Alfa Wassermann ACE Clinical Chemistry System (x),<br>least-squares regression analysis yielded correlation coefficients of 0.9977<br>to 0.9996, standard error estimates of 0.11 to 0.19, confidence interval<br>slopes of 1.014 to 1.067, and a confidence interval intercepts of -0.33 to<br>0.09. | |--------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Detection limit: The detection limit was 0.07 mg/dL. | | Conclusions: | Based on the foregoing data, the device is safe and effective. These data<br>also indicate substantial equivalence to the predicate device. | and the control of the country of the country : Career States of Career . . . . . . . {6}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/6/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features the department's name encircling a stylized image of an eagle. The eagle is depicted with its wings spread, and its head turned to the right. The seal is black and white. 10903 New Hampshire Avenue Silver Spring, MD 20993 Alfa Wassermann Diagnostic Technologies, LLC c/o Hyman Katz, Ph. D. Vice President, Quality and Regulatory Affairs 4 Henderson Drive West Caldwell, NJ 07006 JUL 1 3 2012 Re: k113374 Trade/Device Name: ACE Albumin Reagent, ACE Total Protein Reagent, ACE Calcium-Arsenazo Reagent, ACE Inorganic Phosphorus U.V. Reagent Regulation Number: 21 CFR§ 862.1035 Regulation Name: Albumin test system Regulatory Class: Class II Product Code: CIX, CEK, CJY, CEO Dated: May 29, 2012 Received: May 31, 2012 Dear Dr. Katz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {7}------------------------------------------------ #### Page 2 If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (301) 796-5450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (301) 796-5760. For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/Medical Devices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance ... You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-5680 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm Sincerely yours, - Signature Counney H. Lias, Ph.D. Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## Indications for Use 8 113374 510(k) Number (if known): __ Device Name: ACE Albumin Reagent The ACE Albumin Reagent is intended for the quantitative Indications for Use: determination of albumin concentration in serum using the ACE Axcel Clinical Chemistry System. Albumin measurements are used in the diagnosis and treatment of numerous diseases involving primarily the liver or kidneys. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Device Name: ACE Total Protein Reagent Indications for Use: The ACE Total Protein Reagent is intended for the quantitative determination of total protein concentration in serum using the ACE Axcel Clinical Chemistry System. Total protein measurements are used in the diagnosis and treatment of a variety of diseases involving the liver, kidney, or bone marrow as well as other metabolic or nutritional disorders. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Prescription Use X (21 CFR Part 801 Subpart D) AND/OR Over-The-Counter Use. (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In vitro Diagnostic Devices (OIVD) Rute Chuler Division Sign-Off Office of In vitro Diagnostic Device Evaluation and Safety 510(k) 113374 Page 1 of 2 {9}------------------------------------------------ ## Indications for Use 510(k) Number (if known): K113374 Device Name: ACE Calcium-Arsenazo Reagent Indications for Use: The ACE Calcium-Arsenazo Reagent is intended for the quantitative determination of calcium concentration in serum using the ACE Axcel Clinical Chemistry System. Calcium measurements are used in the diagnosis and treatment of parathyroid disease, a variety of bone diseases, chronic renal disease and tetany (intermittent muscular contractions or spasms). This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Device Name: ACE Inorganic Phosphorus U.V. Reagent Indications for Use: The ACE Inorganic Phosphorus U.V. Reagent is intended for the quantitative determination of inorganic phosphorus concentration in serum using the ACE Axcel Clinical Chemistry System. Measurements of inorganic phosphorus are used in the diagnosis and treatment of various disorders, including parathyroid gland and kidney diseases and vitamin D imbalance. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Prescription Use X (21 CFR Part 801 Subpart D) AND/OR Over-The-Counter Use. (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In vitro Diagnostic Devices (OIVD) Rute C. hulen Division Sign-Off Office of In vitro Diagnostic Device Evaluation and Safety 510(k) 113374 Page 2 of 2
Innolitics
510(k) Summary
Decision Summary
Classification Order