K113268 · Applied Medical Resources Corp. · KGW · Feb 8, 2013 · General, Plastic Surgery
Device Facts
Record ID
K113268
Device Name
ALEXIS O WOUND PROTECTOR/RETRACTOR
Applicant
Applied Medical Resources Corp.
Product Code
KGW · General, Plastic Surgery
Decision Date
Feb 8, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4370
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Applied Medical Alexis wound protector/retractor models C8405 and C8406 have the same intended use as the predicate devices: accessing the abdominal cavity during surgery through an atraumatically retracted incision. The protector/retractors deliver maximum exposure of the abdominal cavity with minimum incision size and protect against wound contamination during laparoscopic and open surgery.
Device Story
Alexis O Wound Protector/Retractor is a single-use, disposable surgical device. It consists of a clear flexible membrane sleeve with two semi-rigid polymer rings. During surgery, one ring is inserted through an abdominal incision to anchor the device; the second ring remains outside the body. The surgeon rolls the external ring over itself to create tension, retracting the abdominal tissue into a circular opening. Used in ORs by surgeons to provide 360-degree circumferential retraction and wound margin protection. Benefits include reduced surgical site infection (SSI) risk and improved abdominal exposure. No software or electronic components.
Clinical Evidence
No clinical testing performed on subject devices. Supporting literature (Reid K et al., 2010) for the predicate device (n=130) showed a significant reduction in surgical site infection (SSI) rates: 4.7% with Alexis vs 22.7% with standard retraction (p-value not provided).
Technological Characteristics
Clear flexible membrane sleeve with two semi-rigid polymer rings. Mechanical retraction principle. Biocompatibility per ISO 10993. Microbial barrier performance per ANSI AAMI PN70: 2003/R 2009 and ASTM F1671 (Level 4 surgical gown equivalent). Single-use, disposable.
Indications for Use
Indicated for use in patients undergoing laparoscopic or open surgery to access the abdominal cavity through an atraumatically retracted incision, provide maximum exposure with minimum incision size, and protect against wound contamination.
Regulatory Classification
Identification
A surgical drape and drape accessories is a device made of natural or synthetic materials intended to be used as a protective patient covering, such as to isolate a site of surgical incision from microbial and other contamination. The device includes a plastic wound protector that may adhere to the skin around a surgical incision or be placed in a wound to cover its exposed edges, and a latex drape with a self-retaining finger cot that is intended to allow repeated insertion of the surgeon's finger into the rectum during performance of a transurethral prostatectomy.
Special Controls
*Classification.* Class II (special controls). The device, when it is an ear, nose, and throat surgical drape, a latex sheet drape with self-retaining finger cot, a disposable urological drape, a Kelly pad, an ophthalmic patient drape, an ophthalmic microscope drape, an internal drape retention ring (wound protector), or a surgical drape that does not include an antimicrobial agent, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 878.9.
{1}------------------------------------------------
K113268
## 510(k) SUMMARY
Frans VandenBroek
January 25, 2013
Wound protector/retractor
fvandenbroek@appliedmedical.com
# FEB 0 8 2013
Applied Medical Resources Corporation 510(k) OWNERS NAME: 22872 Avenida Empresa Rancho Santa Margarita, CA-92688 (949) 713-8000 (949) 713-8205 (FAX)
CONTACT PERSON:
DATE OF PREPARATION:
TRADE NAME:
Alexis® O Wound Protector/retractor models C8405 and C8406
COMMON NAME:
CLASSIFICATION NAME:
Surgical drape and drape accessory, General & Plastic Surgery. Class II, 21CFR §878.4370, Product Code KKX
PREDICATE DEVICE:
Applied Medical Alexis Wound Retractor (K031889)
DEVICE DESCRIPTION: Applied Medical's Alexis O wound protector/retractors are disposable single-use surgical devices used to establish access to internal body cavities and spaces. The protector/retractors are designed for soft tissue retraction and provide 360 degrees of a circumferential retraction and 360 degrees of protection of the wound margins. The predicate Alexis devices are made in sizes ranging from Small to XLarge that accommodate incisions ranging from 2.5 to 17cm. This submission request clearance for adding an XXL and XXXL sizes to the product family. The XXL is for incisions ranging from 17 to 25cm; the XXXL is for incisions ranging from 25 to 30cm.
The protector/retractors have a shelf life of three years and are delivered with a template that is used to determine the incision length for the corresponding retractor chosen for the procedure. Protector/retractors are packaged in a Tyvek pouch placed inside a carton.
INTENDED USE: The Applied Medical Alexis wound protector/retractor models C8405 and C8406 have the same intended use as the predicate devices: accessing the abdominal cavity during surgery through an atraumatically retracted incision. The protector/retractors deliver maximum exposure of the abdominal cavity with minimum incision size and protect against wound contamination during laparoscopic and open surgery.
{2}------------------------------------------------
SUMMÀRY OF TECHNOLOGICAL CHARACTERISTICS: The Alexis O Wound protector/retractor has the same technological characteristics as the predicate devices of K031889. Predicate and proposed device are identical in design, principle of action and clinical utility. Both consist of a clear flexible membrane formed into a cylindrical sleeve with two semi-rigid polymer rings at each end. Alexis is placed by advancing one of the rings through an incision in the abdominal wall. The ring establishes the sheath's internal anchor.
The second ring remains outside the body and is placed in traction and repeatedly rolled over itself with the thumbs until the ring contacts the abdominal wall. The tension in the sheath retracts the abdominal tissue and the incision becomes a circular opening. Alexis allows access to the peritoneum while protecting the incised tissue from contamination and injury throughout the procedure.
DISCUSSION OF NONCLINICAL TESTS SUBMITTED: There are no FDA recognized standards that specify performance of wound protectors/retractors. For that reason. Applied Medical created non-clinical test protocols specifically designed to confirm substantial equivalence of the proposed device relative to the predicate device. In addition, the Alexis sheath was exposed to a microbial challenge per ANSI AAMI PN70: 2003/R 2009. "Liquid Barrier Performance and Classification of Protective Apparel and Drapes Intended for Use in Health Care Facilities." This standard determines a barrier classification level for protection against blood borne pathogens.
Testing included:
- Visual inspection to verify mechanical integrity of the device. .
- Reliability (cvcling) testing to verify the device can complete five activations . without losing structural integrity.
- Wound retraction testing in simulated and in porcine tissue to verify the device . retracts straight incisions into circular openings.
- Biocompatibility testing per ISO10993 (cytotoxicity, sensitization and irritation) to ● verify the device materials did not elicit a tissue response.
- Microbial performance testing per ANSI AAMI PN70: 2003/R 2009 to verify the . device met the requirements of a level 4 surgical gown as defined by test method ASTM F1671, "Standard Test Method for Resistance of Materials Used in Protective Clothing to Penetration by Blood Borne Pathogens using Phi-X174 Bacteriophage Penetration as a Test System."
## DISCUSSION OF CLINICAL TESTS SUBMITTED:
No clinical testing was performed on the subject devices. However, the following study, offered as supporting material, was performed using the predicate devices of K031889:
- . Reid K .. et al. Barrier wound protection Decreases Surgical Site Infection in Open Elective Colorectal Surgery: A Randomized Clinical Trial. Dis Colon Rectum. 2010 Oct: 53(10); 1374-1380
{3}------------------------------------------------
This randomized clinical trial compared the use of the Alexis wound retractor with standard retraction in open elective colorectal resectional surgery. One hundred thirty consecutive patients. 18 years and older. were randomly assigned to have either wound retraction/protection with Alexis (n=64) or standard wound retraction (n=66). Randomization was performed after the patient was anesthetized. The primary end point was the incidence of superficial or deep surgical site infection occurring within 30 days of the procedure as defined by the Centers for Disease Control and Prevention. The secondary end point was performance (effectiveness) of the Alexis wound protectors as assessed by participating surgeons.
The results revealed a clinically significant reduction in SSI in the trial population subjected to the Alexis device. SSI rate with Alexis was 4.7%, without, 22.7%. Seventy eight percent of the SSI events were diagnosed after discharge from the hospital and there was no difference in rates of reoperation, readmission or formal wound drainage between the two groups. There were no adverse effects or complications reported as a result of the Alexis device in any surgeries using the Alexis device. Seven of the 8 surgeons (78%) in the trial are adapting Alexis as part of their standard setup.
CONCLUSIONS DRAWN FROM TESTING: Applied Medical's testing demonstrated that the subject device is safe and effective, and performs as well as the predicate device of K031889.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three tail feathers, representing the three levels of government: federal, state, and local. The eagle is surrounded by a circle of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA".
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
#### February 8, 2013
Mr. Frans Vandenbroek Principal Regulatory Specialist Applied Medical Resources Corporation 22872 Avenida Empresa RANCHO SANTA MARGARITA CA 92688
Re: K113268
Trade/Device Name: Alexis O Wound Protector/Retractors, models C8405 and C8406 Regulation Number: 21 CFR 878.4370 Regulation Name: Surgical Drape and Drape Accessories Regulatory Class: II Product Code: KKX Dated: January 28, 2013 Received: January 30, 2013
Dear Mr. Vandenbroek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Ilsting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH0ffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
hori
Theti
Starshfield.
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# INDICATIONS FOR USE
#### 510(k) Number (if known): k113268
Device Name: Alexis O Wound Protector/Retractors, models C8405 and C8406
Indication for use: The Applied Wound Protector/Retractor is indicated for use to:
- . Access the abdominal cavity during surgery through an atraumatically retracted incision.
- Deliver maximum exposure of the abdominal cavity with minimum incision size. ●
- Protect against wound contamination during laparoscopic and open surgery. .
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
| Concurrence of CDRH, Office of Device Evaluation (ODE) |
|--------------------------------------------------------|
|--------------------------------------------------------|
Elizabeth F. Clayerie 2013.02.07 17:05:51 -05'00' (Division Sign-Off) (Division Sign-Off)
Division of Anesthesiology, General Hospital
Division of Anesthesiology, General Devices Division of Antool, Dental Devices 3268 510(k) Number:
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.