K113267 · CooperVision, Inc. · LPL · Dec 2, 2011 · Ophthalmic
Device Facts
Record ID
K113267
Device Name
BIOPMEDICS52,52 1DAY,52 TORIC,52 1DAY TORIC
Applicant
CooperVision, Inc.
Product Code
LPL · Ophthalmic
Decision Date
Dec 2, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 886.5925
Device Class
Class 2
Indications for Use
SPHERICAL Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters that does not interfere with visual acuity. TORIC Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 10.00 diopters. The lens may be prescribed for Daily Wear in not-aphakic persons. The eyecare practitioner may prescribe the contact lens for either single-use disposable wear or for frequent replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for single-use disposable wear the lens is to be discarded after each removal. When prescribed for frequent replacement wear, the contact lens may be disinfected using a chemical (not heat) disinfecting system. The Biomedics (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
Device Story
Biomedics 52 (ocufilcon B) is a Group IV, high-water, ionic, silicone hydrogel soft contact lens; hemispherical shell design; incorporates phthalocyanine blue handling tint and UV blocker. Intended for daily wear to correct myopia, hyperopia, and astigmatism. Prescribed by eyecare practitioners for single-use or frequent replacement; frequent replacement requires chemical disinfection. Device provides vision correction and UV radiation protection for the cornea/eye. No changes to material, manufacturing, or lens parameters compared to predicates; submission specifically addresses modifications to shelf-life/stability protocols.
Clinical Evidence
No clinical data required. Technical characteristics, formulation, manufacturing, and sterilization processes remain unchanged from currently marketed predicates. Bench testing only.
Technological Characteristics
Material: ocufilcon B (HEMA-based, Group IV, high water, ionic polymer). Design: Hemispherical shell, spherical and toric configurations. Features: Phthalocyanine blue handling tint, UV blocker. Energy source: N/A. Connectivity: N/A. Sterilization: Unchanged from predicate processes.
Indications for Use
Indicated for correction of visual acuity in non-aphakic, non-diseased eyes with myopia or hyperopia and refractive astigmatism (up to 2.00D for spherical, 10.00D for toric). Prescribed for daily wear, either single-use disposable or frequent replacement.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
Biomedics 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lens for Daily Wear (K020389)
Submission Summary (Full Text)
{1}------------------------------------------------
K113267
.
DEC - 2 2011
# 510(k) Summary
# 1. Applicant Name and Address
CooperVision, Inc. 6150 Stoneridge Mall Drive Suite 370 Pleasanton CA 94588
2. Contact
Gwen Sharp Regulatory Affairs (925) 621-2457 gsharp@coopervision.com
3. Date Prepared
November 3, 2011
4. Device Identification
| Trade Name: | Biomedics 52<br>Biomedics 52 1-Day |
|---------------------------|-------------------------------------------------------------|
| Common Name: | Soft Contact Lens |
| Classification<br>Name: | Soft (hydrophilic) Contact Lens –<br>Daily Wear; Disposable |
| Device<br>Classification: | Class II (21 CFR 886.5925) |
| FDA Material<br>Class: | FDA Group IV, High Water, Ionic<br>Polymer |
| Product Code: | MVN, LPL |
{2}------------------------------------------------
# 5. Device Description
The Biomedics 52 (ocufilcon B) soft contact lens is a Group IV, daily wear silicone hydrogel contact lens that is not surface treated and is characterized by a high oxygen permeability (Dk). The lens material, ocufilcon B. is composed of polymerized material of HEMA and cross linked with other monomers, incorporating phthalocyanine blue as an integrated, handling tint. A UV blocker is added to reduce the amount of ultraviolet light transmitted into the eye. The lenses are manufactured in spherical and toric configurations. The Biomedics 52 (ocufilcon B) Soft (hydrophilic) contact lenses are a hemispherical shell. The physical properties and available dimensions are unchanged from predicate 510(k)s.
# 6. Intended Use
# SPHERICAL
Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters that does not interfere with visual acuity.
### TORIC
Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 10.00 diopters.
The lens may be prescribed for Daily Wear in not-aphakic persons. The eyecare practitioner may prescribe the contact lens for either single-use disposable wear or for frequent replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for single-use disposable wear the lens is to be discarded after each removal. When prescribed for frequent replacement wear, the contact lens may be disinfected using a chemical (not heat) disinfecting system.
The Biomedics (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
{3}------------------------------------------------
# 7. Predicate Device(s)
- Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocker Daily Wear Contact Lens . (K003136) January 16, 2001
- Biomedics 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lens for . Daily Wear (K020389) June 4, 2002
# 8. Characteristics of Substantial Equivalence
The soft contact lenses have the following similarities to the predicate lens that previously received 510(k) clearance:
- . have the same indicated use,
- incorporate the same design, .
- incorporate the same materials, .
- have the same shelf life, and .
- . are packaged and sterilized using the same materials and processes.
The modifications to the stability/shelf life protocol include an alternate test method for package integrity and adjusted storage temperatures for the packaged products that will be tested using alternate package integrity test method.
In summary, the ocufilcon B soft contact lenses described in this submission are substantially equivalent to the predicate device.
### 9. Physiochemical Studies
Results from physical, optical and chemical properties were not required as support for this modification to shelf life protocol. Change will not affect physicochemical properties of the lenses.
{4}------------------------------------------------
page 4
K 11 i 320
#### 10. Toxicology Studies
Results from in-vivo and in-vitro studies were not required as support for this modification to shelf life protocol. Change will not affect lenses ability to remain non-toxic and biocompatible with the ocular environment.
#### 11. Conclusions of Non-Clinical Tests Performed:
#### l Physiochemical:
The physical, optical and chemical properties of this lens remain unchanged from the unmodified device, and are within established specifications for the lenses.
#### 트 Toxicology:
Results from in-vivo and in-vitro studies originally conducted remain valid and verify that the lenses remain non-toxic and are biocompatible with the ocular environment.
#### 12. Clinical Studies
The technical characteristics, formulation, manufacturing, and sterilization processes of this lens are not changing and therefore are equivalent to ocufilcon B soft contact lenses currently marketed by Cooper Vision, therefore no clinical data is required.
#### 13. Conclusions
Based on no change to material, no change to manufacturing methods, no change to lens parameters and no change to indicated use, the ocufilcon B soft contact lens described in this document are substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines representing its body and wings. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
CooperVision, Inc. c/o Ms. Gwen Sharp Global Regulatory Affairs 6150 Stoneridge Mall Road Suite 370 Pleasanton, CA 94588
DEC - 2 2011
Re: K113267
Trade/Device Name: Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocker Daily
Wear Contact Lens
Biomedics 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lens for Daily Wear
Regulation Number: 21 CFR 886.5925
Regulation Name: Soft (hydrophilic) contact lenses
Regulatory Class: Class II
Product Code: LPL and MVN
Dated: November 3, 2011
Received: November 4, 2011
Dear Ms. Sharp:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 - Ms. Gwen Sharp
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/5/Picture/7 description: The image shows the name and title of Malvina B. Eydelman, M.D. She is the Director of the Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices. The text also mentions the Office of Device Evaluation and the Center for Devices and Radiological Health.
Enclosure
{7}------------------------------------------------
#### Indications for Use
510(k) Number (if known): K113267
Device Name: Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocker Daily Wear Contact Lens
> Biomedics 52 1-Day (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lens for Daily Wear
#### Indications for Use:
#### SPHERICAL
Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 2.00 diopters that does not interfere with visual acuity.
#### TORIC
Biomedics 52 (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact lenses are indicated for the correction of visual acuity in persons with non-diseased eyes that are myopic (nearsighted) or hyperopic (farsighted) and may exhibit refractive astigmatism of 10.00 diopters.
The lens may be prescribed for Daily Wear in not-aphakic persons. The cyecare practitioner may prescribe the contact lens for either single-use disposable wear or for frequent replacement wear, with cleaning, disinfection, and scheduled replacement. When prescribed for single-use disposable wear the lens is to be discarded after each removal. When prescribed for frequent replacement wear, the contact lens may be disinfected using a chemical (not heat) disinfecting system.
The Biomedics (ocufilcon B) Soft (Hydrophilic) UV Blocking Contact Lenses help protect against transmission of harmful UV radiation to the cornea and into the eye.
Over-The-Counter Use Prescription Use X AND/OR (21 CFR 801 Subpart C) (Part 21 CFR 801 Subpart D) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Rusanand
Page 1 of 1
(Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
510(k) Number K113267
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.