ACE AXCEL CLINICAL CHEMISTRY SYSTEM,ACE ALBUMIN REAGENT,ACE TOTAL PROTEIN REAGENT,ACE BUN/UREANITROGEN REAGENT
K113253 · Alfa Wassermann Diagnostic Technologies, Inc. · JJE · May 17, 2012 · Clinical Chemistry
Device Facts
Record ID
K113253
Device Name
ACE AXCEL CLINICAL CHEMISTRY SYSTEM,ACE ALBUMIN REAGENT,ACE TOTAL PROTEIN REAGENT,ACE BUN/UREANITROGEN REAGENT
Applicant
Alfa Wassermann Diagnostic Technologies, Inc.
Product Code
JJE · Clinical Chemistry
Decision Date
May 17, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 862.2160
Device Class
Class 1
Indications for Use
The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for in vitro diagnostic use in the quantitative measurement of general chemistry assays for clinical use in physician office laboratories or clinical laboratories. The ACE Axcel Clinical System includes an Ion Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only. Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance Potassium measurements are used to monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels. Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. The ACE Glucose Reagent is intended for the quantitative determination of glucose concentration in serum using the ACE Axcel Clinical Chemistry System. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. This test is intended for use in clinical laboratories or physician office laboratories. For in vitro diagnostic use only.
Device Story
Automated, bench-top, random access clinical chemistry analyzer; processes serum samples for glucose, sodium, potassium, and chloride. System includes single pipettor, temperature-controlled reagent compartment, reaction wheel, multi-wavelength holographic diffraction grating spectrophotometer, and integrated Ion Selective Electrode (ISE) module. User operates via touch screen/integrated computer; system prompts for input and displays data. Reagent bottles identified via bottom-applied dot code labels. Glucose measured via enzymatic hexokinase method; Na/K/Cl measured via potentiometric ion-selective electrodes. Healthcare providers use output to diagnose/monitor metabolic and electrolyte disorders. Benefits include rapid, automated quantitative analysis in point-of-care or clinical lab settings.
Clinical Evidence
Bench testing only. Precision/reproducibility studies (n=80/sample) and point-of-care precision studies (n=15/sample) conducted across three physician office laboratories. Linearity confirmed across assay ranges. Interference studies performed for bilirubin, hemoglobin, triglycerides, and ascorbic acid. Method comparison studies (n=111-122 samples) against predicate showed high correlation (R² > 0.98).
Technological Characteristics
Benchtop analyzer; spectrophotometric and potentiometric sensing. Reagents: liquid-stable in plastic bottles with dot code identification. ISE module: crown ether (Na), valinomycin (K), and quaternary ammonium salts (Cl) membranes. Connectivity: integrated computer/touch screen. Calibration: on-demand, automated for ISE. Sterilization: not applicable.
Indications for Use
Indicated for quantitative measurement of glucose, sodium, potassium, and chloride in serum. Used in clinical laboratories or physician office laboratories for diagnosis and treatment of carbohydrate metabolism disorders (e.g., diabetes mellitus, hypoglycemia, pancreatic islet cell carcinoma) and electrolyte/metabolic disorders (e.g., electrolyte imbalance, cystic fibrosis, diabetic acidosis).
Regulatory Classification
Identification
A discrete photometric chemistry analyzer for clinical use is a device intended to duplicate manual analytical procedures by performing automatically various steps such as pipetting, preparing filtrates, heating, and measuring color intensity. This device is intended for use in conjunction with certain materials to measure a variety of analytes. Different models of the device incorporate various instrumentation such as micro analysis apparatus, double beam, single, or dual channel photometers, and bichromatic 2-wavelength photometers. Some models of the device may include reagent-containing components that may also serve as reaction units.
Predicate Devices
Alfa Wassermann ACE plus ISE/Clinical Chemistry System ACE Reagents (K931786)
Submission Summary (Full Text)
{0}
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY AND INSTRUMENT COMBINATION TEMPLATE
A. 510(k) Number:
k113253
B. Purpose for Submission:
New device
C. Measurand:
Glucose, Sodium, Potassium and Chloride
D. Type of Test:
Quantitative, photometric and ion selective electrodes
E. Applicant:
Alfa Wassermann
F. Proprietary and Established Names:
ACE Axcel Clinical Chemistry System; ACE Ion Selective Electrode (ISE) Module; ACE Glucose Reagent
G. Regulatory Information:
| Product Code | Classification | Regulation Section | Panel |
| --- | --- | --- | --- |
| JJE | I | 862.2160 Discrete photometric chemistry analyzer for clinical use | 75-Chemistry |
| CFR | II | 862.1345, Glucose test system | 75-Chemistry |
| JGS | II | 21 CFR 862.1665 Sodium test system | 75-Chemistry |
| CEM | II | 21 CFR 862.1600 Potassium test system | 75-Chemistry |
| CGZ | II | 21 CFR 862.1170 Chloride test system | 75-Chemistry |
{1}
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
Device Name: ACE Axcel Clinical Chemistry System, ACE Ion Selective Electrode (ISE) Module, ACE Glucose Reagent
The ACE Axcel Clinical Chemistry System is an automated, discrete, bench-top, random access analyzer that is intended for *in vitro* diagnostic use in the quantitative measurement of general chemistry assays for clinical use in physician office laboratories or clinical laboratories.
The ACE Axcel Clinical System includes an Ion Selective Electrode (ISE) module for the measurement of sodium, potassium and chloride in serum. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
- Sodium measurements are used in the diagnosis and treatment of diseases involving electrolyte imbalance
- Potassium measurements are used to monitor electrolyte balance and in the diagnosis and treatment of diseases conditions characterized by low or high blood potassium levels.
- Chloride measurements are used in the diagnosis and treatment of electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis.
The ACE Glucose Reagent is intended for the quantitative determination of glucose concentration in serum using the ACE Axcel Clinical Chemistry System. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma. This test is intended for use in clinical laboratories or physician office laboratories. For *in vitro* diagnostic use only.
3. Special conditions for use statement(s):
For *in vitro* diagnostic use only. For prescription and point-of-care use.
4. Special instrument requirements:
ACE Axcel Clinical Chemistry System
{2}
I. Device Description:
The ACE Axcel Clinical Chemistry System is an automatic discrete, random access wet chemistry system intended for use in clinical laboratories or physician office laboratories that consists of a benchtop analyzer, an operation touch screen/integrated computer that prompts the user for operation input, displays data and reagent kits. The benchtop analyzer includes a single pipettor, temperature controlled reagent compartment, reaction wheel and a multi-wavelength holographic diffraction grating spectrophotometer. The ACE Axcel analyzer also includes an integrated ISE Module.
The ACE reagent kits used with the ACE Axcel Clinical Chemistry System consist of natural or brown plastic bottles containing liquid-stable reagents. The reagents have a dot code label applied to the bottom of the bottle to identify each bottle to the ACE Axcel system. Reagent kits typically have either one reagent (R1) or sometimes a second reagent (R2) and an Evap-Cap.
Glucose reagent kit contains Nicotinamide adenine dinucleotide (NAD) 2 mmol/L, Adenosine 5'-triphosphate (ATP) 4 mmol/L, Magnesium 2 mmol/L, Hexokinase (Yeast) >2000 U/L, Glucose-6-phosphate dehy7drogenase (G-6-PD) (Leuconostoc mesenteroides) >4000 U/L and buffer, stabilizers and preservatives.
The Ion-Selective Electrode (ISE) module includes a sodium electrode, a potassium electrode, a chloride electrode, a reference electrode, a cleaning solution and two calibrators (level 1 and 2). The sodium electrode membrane is a crown ether liquid-membrane. The potassium electrode membrane is a valinomycin liquid-membrane. The chloride electrode membrane is a quaternary ammonium salts polymer membrane. The ISE calibrator 1 and 2 which is included contains the following chemicals: sodium, potassium, chloride.
ISE calibrators have been previously cleared in k933862. Glucose calibrator has been previously cleared in k930104.
J. Substantial Equivalence Information:
1. Predicate device name(s):
ACE plus ISE/Clinical Chemistry System, Alfa Wassermann
2. Predicate 510(k) number(s):
k930104
{3}
3. Comparison to predicate
| Similarities and Difference for Glucose | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | Intended for the quantitative determination of glucose concentrations in serum. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia and idiopathic hypoglycemia and of pancreatic islet cell carcinoma. | Same. |
| Calibration | Calibrated by referencing the change in absorbance of the unknown samples to the change in absorbance of the calibrator; the use of GEMCAL reference serum is recommended. | Same |
| Calibration Stability | 30 days | Same |
| On-Board Stability | 30days | Same |
| Basic Principle | Enzymatic method for glucose | Same |
| Analysis Temperature | 37°C | Same |
| Reaction Type | Endpoint | Same |
| Sample Type | Serum | Same |
| Sample Volume | 3 μL | Same |
| Reaction Volume | 318 μL | Same |
| Detection Limit | 3.1 mg/dL | 1 mg/dL |
| Linearity Upper Limit | 757 mg/dL | 750 mg/dL |
| Reportable Range | 3-750 mg/dL | 1-750 mg/dL |
| Similarities and Difference for Na/K/Cl | | |
| --- | --- | --- |
| Item | Candidate Device | Predicate Device |
| Intended Use | intended to measure concentrations of sodium, potassium and chloride in undiluted serum. Sodium measurements are used in the diagnosis and treatment of aldosteronism (excessive secretion of | Same |
| | calibrator). | |
| Calibration Stability | 30 days | Same |
| On-Board Stability | 30 days | Same |
| Basic Principle | Enzymatic method for glucose | Same |
| Analysis Temperature | 37°C | Same |
| Reaction Type | Endpoint | Same |
| Sample Type | Serum | Same |
| Sample Volume | 3 μL | Same |
| Reaction Volume | 318 μL | Same |
| Detection Limit | 3.1 mg/dL | 1 mg/dL |
| Linearity Upper Limit | 757 mg/dL | 750 mg/dL |
| Reportable Range | 3-750 mg/dL | 1-750 mg/dL |
{4}
| Similarities and Difference for Na/K/Cl | | |
| --- | --- | --- |
| | the hormone aldosterone). Diabetes insipidus (chronic excretion of large amounts of dilute urine, accompanied by extreme thirst), adrenal disease, Addison’s disease (caused by destruction of the adrenal glands), dehydration, inappropriate antidiuretic hormone secretion, or other diseases involving electrolyte imbalance. Potassium measurements are used to monitor electrolyte and metabolic disorders such as cystic fibrosis and diabetic acidosis. | |
| Calibration | Calibration is performed automatically by the ISE module. | Same |
| Calibration Stability | Calibration must be performed prior to the initial run. Calibration is stable for 3 hours. | Same |
| Basic Principle | Ion selective electrode | Same |
| Analysis Temperature | Ambient | Same |
| Reaction Type | Electrochemical potential | Same |
| Sample Type | Serum | Same |
| Sample Volume | 156 μL | Same |
| ISE Type | Direct (undiluted) | Same |
| Measuring Range | Na 40-205 mmol/L
K 1.5-15 mmol/L
Cl 50-200 mmol/L | Same |
| Expected Range | Na 136-145 mmol/L
K 3.5-5.1 mmol/L
Cl 98-107 mmol/L | Same |
K. Standard/Guidance Document Referenced (if applicable):
CLSI EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition
CLSI EP6-A: Evaluation of Linearity of Quantitative Measurement Procedures, A Statistical Approach; Approved Guideline
CLSI EP7-A2: Interference Testing in Clinical Chemistry; Approved Guideline-Second Edition
CLSI EP9-A2-IR: Method Comparison and Bias Estimation Using Patient Samples; Approved Guideline-Second Edition
CLSI EP10-A3: Preliminary Evaluation of Quantitative Clinical Laboratory Measurement Procedures; Approved Guideline-Third Edition
{5}
CLSI EP17-A: Protocols for Determination of Limits of Detection and Limits of Quantitation; Approved Guideline
## L. Test Principle:
Potentiometric Chemistries
The sodium, potassium and chloride are measured using ion-selective electrodes. Each electrode uses ion-specific membrane to measure the difference in ionic concentration between an inner electrolyte solution and the sample. The difference causes an electrochemical potential to form on the membrane of the active electrode. The measured voltage difference of the sample and CAL A are used to determine the ion concentration in the sample. Two solutions contained in the ISE reagent pack CAL A and CAL B are used to perform a two-point calibration of the ion selective electrodes (ISE)
Photometric Chemistries
ACE Glucose Reagent - Glucose in serum reacts with adenosine triphosphate (ATP) in the presence of hexokinase (HK) and magnesium with the formation of glucose-6-phosphate (G-6-P) and adenosine diphosphate (ADP). Glucose-6-phosphate dehydrogenase (G-6-PD) catalyzes the oxidation of glucose-6-phosphate with $\mathrm{NAD^{+}}$ to form 6-phosphogluconate and NADH. NADH absorbs strongly at $340~\mathrm{nm}$, whereas $\mathrm{NAD^{+}}$ does not. The total amount of NADH formed is proportional to the initial amount of glucose present. The rate of increase in absorbance, monitored bichromatically at $340~\mathrm{nm}/450~\mathrm{nm}$, is directly proportional to the glucose concentration in the sample.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. Precision/Reproducibility:
In-house precision
Precision studies were conducted by testing human serum pools at four levels. The samples were run 2 times per run, 2 runs per day, for a total of 22 days using one instrument (n = 80/sample). Results are summarized below.
{6}
| Glucose | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (mg/dL) | 91.4 | 360.9 | 628 | 71.7 |
| Within Run | SD | 1.3 | 4.5 | 6.0 | 0.7 |
| | %CV | 1.4 | 1.2 | 1.0 | 1.0 |
| Between Run | SD | 1.2 | 0.0 | 0.0 | 0.3 |
| | %CV | 1.3 | 0.0 | 0.0 | 0.4 |
| Between Day | SD | 0.0 | 1.5 | 3.9 | 0.1 |
| | %CV | 0.0 | 0.4 | 0.6 | 0.1 |
| Total | SD | 1.7 | 4.7 | 7.1 | 0.7 |
| | %CV | 1.9 | 1.3 | 1.1 | 1.0 |
| Sodium | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (mmol/L) | 95.19 | 143.9 | 170.88 | 137.64 |
| Within Run | SD | 0.36 | 1.12 | 1.73 | 0.96 |
| | %CV | 0.4 | 0.8 | 1.0 | 0.7 |
| Between Run | SD | 0.44 | 0.97 | 0.0 | 1.7 |
| | %CV | 0.5 | 0.7 | 0.0 | 1.2 |
| Between Day | SD | 0.51 | 0.61 | 0.57 | 0.0 |
| | %CV | 0.5 | 0.4 | 0.3 | 0.0 |
| Total | SD | 0.76 | 1.61 | 1.82 | 1.95 |
| | %CV | 0.8 | 1.1 | 1.1 | 1.4 |
| Potassium | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (mmol/L) | 2.588 | 7.103 | 11.997 | 4.071 |
| Within Run | SD | 0.021 | 0.106 | 0.069 | 0.143 |
| | %CV | 0.8 | 1.5 | 0.6 | 3.5 |
| Between Run | SD | 0.014 | 0.0 | 0.134 | 0.0 |
| | %CV | 0.5 | 0.0 | 1.1 | 0.0 |
| Between Day | SD | 0.023 | 0.052 | 0.164 | 0.0 |
| | %CV | 0.9 | 0.7 | 1.4 | 0.0 |
| Total | SD | 0.034 | 0.118 | 0.223 | 0.143 |
| | %CV | 1.3 | 1.7 | 1.9 | 3.5 |
| Chloride | | Sample 1 | Sample 2 | Sample 3 | Sample 4 |
| --- | --- | --- | --- | --- | --- |
| | Mean (mmol/L) | 68.26 | 117.52 | 168.65 | 104.47 |
| Within Run | SD | 0.49 | 1.04 | 0.84 | 1.02 |
| | %CV | 0.7 | 0.9 | 0.5 | 1.0 |
| Between Run | SD | 0.62 | 0.0 | 1.57 | 0.93 |
| | %CV | 0.9 | 0.0 | 0.9 | 0.9 |
| Between Day | SD | 0.68 | 0.75 | 1.30 | 0.37 |
| | %CV | 1.0 | 0.6 | 0.8 | 0.4 |
{7}
| Total | SD | 1.04 | 1.28 | 2.20 | 1.43 |
| --- | --- | --- | --- | --- | --- |
| | %CV | 1.5 | 1.1 | 1.3 | 1.4 |
# Point-of-Care precision
Precision studies were also conducted at three Physician Office Laboratories (POL) with four trained operators typically found in these settings. Human serum pools at three concentrations were tested three times a day for five days on three instruments (one at each lab) $(n = 15 / \text{sample})$ . The results are presented below:
| GLUCOSE | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 62.4 | 0.4 | 0.6 | 1.1 | 1.8 |
| POL 2 | 1 | 63.3 | 0.7 | 1.1 | 1.0 | 1.6 |
| POL 3 | 1 | 63.5 | 1.4 | 2.2 | 1.4 | 2.2 |
| | | | | | | |
| POL 1 | 2 | 299.7 | 1.4 | 0.5 | 1.4 | 0.5 |
| POL 2 | 2 | 304.3 | 1.9 | 0.6 | 2.7 | 0.9 |
| POL 3 | 2 | 304.0 | 2.5 | 0.8 | 4.0 | 1.3 |
| | | | | | | |
| POL 1 | 3 | 533.3 | 1.8 | 0.3 | 3.3 | 0.6 |
| POL 2 | 3 | 540.2 | 3.9 | 0.7 | 3.9 | 0.7 |
| POL 3 | 3 | 537.7 | 2.9 | 0.5 | 5.3 | 1.0 |
| SODIUM | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 103.26 | 0.62 | 0.6 | 1.17 | 1.1 |
| POL 2 | 1 | 104.07 | 0.78 | 0.7 | 0.87 | 0.8 |
| POL 3 | 1 | 103.09 | 1.08 | 1.0 | 1.42 | 1.4 |
| | | | | | | |
| POL 1 | 2 | 137.53 | 1.26 | 0.9 | 1.81 | 1.3 |
| POL 2 | 2 | 138.02 | 1.08 | 0.8 | 1.11 | 0.8 |
| POL 3 | 2 | 136.78 | 1.03 | 0.8 | 1.62 | 1.2 |
| | | | | | | |
| POL 1 | 3 | 174.77 | 1.26 | 0.7 | 2.32 | 1.3 |
| POL 2 | 3 | 175.01 | 1.42 | 0.8 | 1.43 | 0.8 |
| POL 3 | 3 | 173 | 1.11 | 0.6 | 1.45 | 0.8 |
{8}
| POTASSIUM | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 4.314 | 0.063 | 1.5 | 0.063 | 1.5 |
| POL 2 | 1 | 4.310 | 0.067 | 1.6 | 0.067 | 1.6 |
| POL 3 | 1 | 4.293 | 0.059 | 1.4 | 0.068 | 1.6 |
| | | | | | | |
| POL 1 | 2 | 7.944 | 0.103 | 1.3 | 0.103 | 1.3 |
| POL 2 | 2 | 7.925 | 0.093 | 1.2 | 0.093 | 1.2 |
| POL 3 | 2 | 7.954 | 0.109 | 1.4 | 0.109 | 1.4 |
| | | | | | | |
| POL 1 | 3 | 11.925 | 0.122 | 1.0 | 0.154 | 1.3 |
| POL 2 | 3 | 11.839 | 0.120 | 1.0 | 0.134 | 1.1 |
| POL 3 | 3 | 12.077 | 0.141 | 1.2 | 0.172 | 1.4 |
| CHLORIDE | | | Within Run | | Total | |
| --- | --- | --- | --- | --- | --- | --- |
| Lab | Sample | Mean | SD | %CV | SD | %CV |
| POL 1 | 1 | 58.83 | 0.72 | 1.2 | 0.79 | 1.3 |
| POL 2 | 1 | 58.15 | 0.79 | 1.4 | 1.13 | 1.9 |
| POL 3 | 1 | 58.67 | 0.89 | 1.5 | 1.52 | 2.6 |
| | | | | | | |
| POL 1 | 2 | 101.53 | 1.13 | 1.1 | 1.13 | 1.1 |
| POL 2 | 2 | 102.45 | 1.24 | 1.2 | 1.24 | 1.2 |
| POL 3 | 2 | 103.21 | 1.38 | 1.3 | 1.38 | 1.3 |
| | | | | | | |
| POL 1 | 3 | 148.36 | 1.29 | 0.9 | 1.59 | 1.1 |
| POL 2 | 3 | 148.73 | 1.61 | 1.1 | 1.61 | 1.1 |
| POL 3 | 3 | 152.40 | 2.04 | 1.3 | 296 | 1.9 |
# b. Linearity/assay reportable range:
Linearity across the assay range was confirmed by spiking serum samples to a high concentration of analyte, then diluting the sample to obtain 1-14 levels to cover the measuring range of each assay. The assigned value of the highest sample was set to its mean value. The assigned values of the other levels were calculated by multiplying the mean value by the ratios obtained from the manufacturer. Each level was tested in replicates of four. Results are summarized below:
| Analyte tested | Linear regression | r2 | Samples range tested |
| --- | --- | --- | --- |
| Gluc (mg/dL) | y = 0.988x - 0.88 | 0.999 | 2 - 757 |
| Na (mmol/L) | y = 0.991x -1.94 | 0.9985 | 40 - 205 |
| K (mmol/L) | y = 1.001x - 0.44 | 0.9969 | 1.5 - 15 |
| Cl (mmol/L) | y = 0.997x - 2.1 | 0.9987 | 50 - 200 |
{9}
Results of the study support the sponsor claims for the following measuring/linearity ranges:
| Sodium | Potassium | Chloride | Glucose |
| --- | --- | --- | --- |
| 40-205 mmol/L | 1.5 – 15 mmol/L | 50 – 200 mmol/L | 3 – 750 mg/dL |
c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The ISE calibrator material was previously cleared under k933862 and glucose calibrator (Gemcal) was previously cleared under k930104.
d. Detection limit:
The sponsor determined that the detection limit was defined by the linear range study. Please refer to the linearity study for Na, K and Cl.
In addition, the sponsor performed a detection limits study for glucose based on a modified protocol according to the CLSI EP17-A. The limit of blank was determined by assaying five low samples (serum samples) and five true blanks (human serum albumin in saline). Testing was carried out over three days on two ACE Axcel Clinical Chemistry Analyzers. Serum samples and true blanks were assayed every day for a total of 60 measurements. LoD and LoQ were determined by running 5 low samples testing 8 replicates/day for 5 days for a total of 40 measurements. The LoB was calculated to be $2.5\mathrm{mg / dL}$ and LoD was calculated to be $3.1\mathrm{mg / dL}$ . The LoQ was determined to be $3\mathrm{mg / dL}$ for glucose with an inter-assay precision of $< 11.6\%$ CV.
e. Analytical specificity:
Interference studies were performed to determine the effects from potential interferents. The various concentration of interferent was spiked into serum pools containing glucose, sodium, potassium and chloride at normal and abnormal concentrations. All samples were tested in triplicate. Seven levels were tested for each interferent. The sponsor states that interferences are considered to be non-significant if the bias between the tested and control samples are within $\pm 10\%$ for all the analytes.
The tested ranges and analyte concentrations are listed in the tables below:
Glucose
| Interferent | No Significant Interference at or below |
| --- | --- |
| Unconjugated Bilirubin | 31 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Triglycerides | 327 mg/dL |
| Ascorbic Acid | 6 mg/dL |
{10}
Sodium
| Interferent | No Significant Interference at or below |
| --- | --- |
| Unconjugated Bilirubin | 50 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Lipemia (Intralipid) | 689 mg/dL |
| Ascorbic Acid | 6 mg/dL |
| Potassium Phosphate, Dibasic | 50 mmol/L |
| Glucose | 500 mg/dL |
| Calcium Chloride | 216 mmol/L |
Potassium
| Interferent | No Significant Interference at or below |
| --- | --- |
| Unconjugated Bilirubin | 50 mg/dL |
| Hemoglobin | 125 mg/dL |
| Lipemia (Intralipid) | 659 mg/dL |
| Ascorbic Acid | 6 mg/dL |
| Sodium Chloride | 513 mmol/L |
| Sodium Bicarbonate | 625 mmol/L |
| Glucose | 500 mg/dL |
Chloride
| Interferent | No Significant Interference at or below |
| --- | --- |
| Unconjugated Bilirubin | 50 mg/dL |
| Hemoglobin | 1000 mg/dL |
| Lipemia (Intralipid) | 620 mg/dL |
| Ascorbic Acid | 6 mg/dL |
| Potassium Phosphate, Dibasic | 100 mmol/L |
| Sodium Bicarbonate | 1250 mmol/L |
| Glucose | 500 mg/dL |
Based on the hemolysis interference, the sponsor has the following limitations in their labeling:
"Do not use hemolyzed samples for potassium since significant hemolysis may increase K concentration because of high levels of K in erythrocytes".
"Drugs and other substances may affect sodium, potassium, chloride, and glucose determinations. See Young, D.S.¹ for a compilation of reported interferences"
1. Young, D.S., Effects of Drugs on Clinical Laboratory Tests, 5th edition, AACC Press, Washington D.C., 2000.
{11}
f. Assay cut-off:
Not applicable.
## 2. Comparison studies:
### a. Method comparison with predicate device:
An in-house method comparison study to the predicate device was performed with serum patient samples. A total of 122 glucose (108 native, 5 diluted and 9 spiked) serum samples, 113 Sodium (102 native, 6 diluted and 5 spiked) serum sample, 115 Potassium (103 native, 4 diluted and 8 spiked) serum samples and 111 Chloride (103 native, 3 dilutes and 5 spiked) serum samples covering the assay range were tested. The results are presented in the table below:
| Test | n | Regression Equation | R² | Standard Error | Sample range |
| --- | --- | --- | --- | --- | --- |
| Glucose | 122 | y=1.005x-0.7 | 0.9998 | 3.1 | 6-729 mg/dL |
| Sodium | 113 | y=1.008x-1.34 | 0.9963 | 1.65 | 45.1-194 mmol/L |
| Potassium | 115 | y=1.002x+0.022 | 0.9974 | 0.146 | 1.57-14.20 mmol/L |
| Chloride | 111 | Y=0.970x+2.28 | 0.9855 | 2.05 | 63.4-176 mmol/L |
Additional method comparison studies were performed at three Physician Office Laboratories, with four operators. Operators assayed serum samples a total of 166 glucose (137 native, 22 spiked and 7 diluted), a total of 155 sodium (126 native, 15 spiked and 14 diluted), a total of 166 potassium (132 native, 23 spiked and 11 diluted) and a total of 155 chloride (127 native, 20 spiked and 8 diluted) samples on the Ace Axcel clinical chemistry analyzer and the ACE clinical chemistry System. The results are presented in the tables below:
Glucose
| POL | n | Regression Equation | R² | Standard Error | Sample range |
| --- | --- | --- | --- | --- | --- |
| 1 | 63 | y=0.980x+1.5 | 0.9995 | 5.4 | 24-712 mg/dL |
| 2 | 56 | y=1.004x+1.3 | 0.9992 | 7.0 | 9-693 mg/dL |
| 3 | 47 | y=1.014x-1.5 | 0.9998 | 3.6 | 20-718 mg/dL |
Sodium
| POL | n | Regression Equation | R² | Standard Error | Sample range |
| --- | --- | --- | --- | --- | --- |
{12}
| 1 | 57 | y=1.022x-3.43 | 0.9994 | 0.91 | 40.4-185.4 mmol/L |
| --- | --- | --- | --- | --- | --- |
| 2 | 50 | y=0.999x+0.47 | 0.9995 | 0.74 | 56.9-185.9 mmol/L |
| 3 | 48 | y=1.028x-3.27 | 0.9917 | 3.07 | 42.5-204 mmol/L |
Potassium
| POL | n | Regression Equation | R² | Standard Error | Sample range |
| --- | --- | --- | --- | --- | --- |
| 1 | 62 | y=1.017x-0.088 | 0.9978 | 0.182 | 1.99-14.73 mmol/L |
| 2 | 53 | y=0.968x+0.174 | 0.9996 | 0.064 | 1.67-14.07 mmol/L |
| 3 | 51 | y=1.008x-0.058 | 0.9973 | 0.167 | 1.56-14.77 mmol/L |
Chloride
| POL | n | Regression Equation | R² | Standard Error | Sample range |
| --- | --- | --- | --- | --- | --- |
| 1 | 55 | y=1.019x-2.63 | 0.9990 | 0.89 | 59.7-195.1 mmol/L |
| 2 | 52 | y=0.984x+1.36 | 0.9996 | 0.55 | 51.5-180 mmol/L |
| 3 | 48 | y=1.041x-3.05 | 0.9885 | 3.05 | 52.4-189.6 mmol/L |
b. Matrix comparison:
The device is being cleared for serum use only.
3.Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable
4.Clinical cut-off:
Not applicable
{13}
5. Expected values/Reference range:
Reference Values are provided in the labeling according to literature as follows:
| Analyte | Serum |
| --- | --- |
| Glucose | 70-105 mg/dL |
| Sodium | 136-145 mEq/L |
| Potassium | 3.5-5.1 mEq/L |
| Chloride | 98-107 mEq/L |
Tietz:N.W.Clinical Guide to Laboratory tests 3rd Ed., (WB Saunders Co. Philadelphia USA), (1995).
Tietz:N.W.(Ed.), Fundamentals of Clinical Chemistry, (WB Saunders eds. Philadelphia USA), (1976).
N. Instrument Name:
ACE Axcel Clinical Chemistry Analyzer
O. System Descriptions:
1. Modes of Operation:
This instrument is capable of testing several assays via self-contained reagent bottles. The instrument identifies the assay through reading a dot code label on the bottom of each reagent bottle.
2. Software:
FDA has reviewed applicant’s Hazard Analysis and software development processes for this line of product types:
Yes ☐ X or No ☐
3. Specimen Identification:
Barcoding or manual entry
4. Specimen Sampling and Handling:
Samples are manually placed on the instrument either by sample tube or sample cup. The system can run an individual sample or a batch of samples. Once tested the samples are removed.
{14}
5. Calibration:
On demand calibration. It is recommended to perform a calibration measurement after installing a new or fresh bottle of reagent and/or intervals that are defined for a particular test. It is recommended to recalibrate ISEs after installing a new lot of calibration solution. ISE calibration is required every 3 hours or when quality control results fall outside the established range after replacing electrode, and after ISE cleaning maintenance.
6. Quality Control:
Controls are run manually and recommended daily. Results can be stored in instrument memory for future use.
7. Room Temperature:
An ambient temperature study was conducted to show the affects on sample results when room temperature fluctuates (from 59 – 80°F). Five serum samples with analyte concentrations covering the medical decision points were tested. Calibrations were performed at each of these temperatures (59°F, 70°F, and 80°F) and testing was performed at all the three temperatures for each calibration temperature. Room temperatures were set and maintained by thermostats in the facility. Temperatures were recorded during the study. Protocols and acceptance criteria provided are found to be adequate. Based on the ambient temperature study, the sponsor claimed that ambient temperature between 59 – 80°F will not affect the performance of the device.
P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
None
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.