K113247 · Ki Works Co., Ltd. · EJI · Mar 27, 2012 · Dental
Device Facts
Record ID
K113247
Device Name
ANESTHESIA INJECTION TOOL SET
Applicant
Ki Works Co., Ltd.
Product Code
EJI · Dental
Decision Date
Mar 27, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 872.6770
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The device is indicated for the injection of local anesthetics for anesthesia administered prior to, or in conjunction with, dental procedures.
Device Story
Anesthesia Injection Tool Set (KIW-1000) is a motor-driven, battery-operated anesthetic delivery device for dental clinics. Operated by dental professionals, it uses a program-controlled motor to regulate injection flow speed, aiming to deliver anesthetics safely and reduce patient pain. The system consists of a main body, hand-piece, battery pack, and adapter. It supports automatic and manual operation with 6-step speed control. The device is used with standard, disposable, one-time-use syringes and needles; the system itself does not contact the patient. The provider selects injection modes (Auto 1 for adults, Auto 2 for children) to manage initial slow injection times, facilitating consistent and comfortable anesthetic delivery.
Clinical Evidence
No clinical data. Safety and effectiveness supported by bench testing, including electrical safety (EN60601-1, EN60601-1-2, EN55011:2007) and performance testing comparing initial slow injection times against the predicate device.
Technological Characteristics
Motor-driven, battery-powered anesthetic delivery system. Features 6-step speed control, automatic/manual operation modes, and internal battery power. Does not contact human body; uses standard disposable syringes/needles. Complies with EN60601-1, EN60601-1-2, and EN55011:2007 standards.
Indications for Use
Indicated for patients requiring local anesthesia prior to or during dental procedures. Contraindications not specified.
Regulatory Classification
Identification
A cartridge syringe is a device intended to inject anesthetic agents subcutaneously or intramuscularly. The device consists of a metal syringe body into which a disposable, previously filled, glass carpule (a cylindrical cartridge) containing anesthetic is placed. After attaching a needle to the syringe body and activating the carpule by partially inserting the plunger on the syringe, the device is used to administer an injection to the patient.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 872.9.
{0}------------------------------------------------
K113247
MAR 2 7 2012
# 510(k) Summary
#### 1. Submitter's Identification:
KI works Co., Ltd.
Gimhae Biomedical Center #113, 155-1, Nongso-ri, Juchon-myeon,
Gyeongnam, Korea
Tel: +82-55-329-707, Fax: +82-55-329-7072
E-mail: woo1819@hananet.net
## 2. 510(k) Preparer
Company name: BSR KOREA Corp.
Address: Migun Technowolrd 2 C-dong 524, #533-1 Yongsan-dong Yuseong-gu, Daejeon, KOREA
Tel: +82-42-476-2977, Fax: +82-42-476-2978
Name of contact person: Eileen Yang
#### 3. Identification of the device
Trade name / proprietary name: Anesthesia Injection Tool Set Model No.: KIW-1000 Common / usual name: Anesthetic delivery device Classification name: Cartridge syringe Regulation number: 872.6770 Product code: EJI
### 4. Equivalent legally marketed device
The KI works' Anesthetic Injection Tool Set (KIW-1000) is substantially equivalent, with respect to questions of safety and effectiveness, to;
Predicate Device 1 - K002387 Comfort Control Syringe System
{1}------------------------------------------------
#### 5. Intended use of the device
The device is indicated for the injection of local anesthesia administered prior to, or in conjunction with, dental procedures.
## 6. Description of the device
KI works' Anesthesia Injection Tool Set (KIW-1000) is a device for injecting local anesthetic for anesthesia prior to, or in conjunction with, dental procedures. This device consists of Main body, Hand-piece, Battery pack and Adapter. All operating displays and the majority of the operator controls are located on front panel of the control unit.
## 7. Safety and Effectiveness, comparison to predicate device
This subject device, KI works' Anesthesia Injection Tool Set (KIW-1000), and predicate device are substantially equivalent and have the same intended use. Technological and performance difference do not raise any new questions of safety or effectiveness. Comparison analysis, including comparison tables, of the subject device versus the predicate device is provided in section 7.Comparison chart._____
- 1 + 1 = 2
| Section | Predicate Device<br>DENTSPLY Professional(U.S.A.) | KIWORKS Co., Ltd. |
|---------------------|---------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Product<br>name | MIDWEST<br>Comfort Control Syringe | KIW-1000 (e-joa) |
| Web<br>Address | www.dentsply.com | www.kiworks.co.kr |
| Intended<br>Use | For the injection of local<br>anesthetics for anesthesia prior<br>to, or in conjunction with, dental<br>procedures. | For the injection of local<br>anesthetics for anesthesia<br>administered prior to, or in<br>conjunction with, dental<br>procedures |
| Indications for use | To administer consistent and<br>comfortable injections. | To deliver anesthetics in dental<br>procedure safely and effectively<br>without pain for patient. |
#### 8. Comparison chart
{2}------------------------------------------------
| Target | Anesthesia prior to, or in<br>conjunction with, dental<br>procedures | Anesthesia prior to, or in<br>conjunction with, dental<br>procedures |
|--------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| Point of use | Dental clinic | Dental clinic |
| Performance | To control motor by program to<br>inject anesthetic safely and<br>effectively. It will alleviate the<br>patient's pain. | To control motor by program to<br>inject anesthetic safely and<br>effectively. It will alleviate the<br>patient's pain. |
| FDA<br>clearance | K002387<br>Class II | |
| Sterility | All the system is not contact with<br>human body.<br>Using in conjunction with<br>disposable, one time use,<br>standard syringe and needle. | All the system is not contact with<br>human body.<br>Using in conjunction with<br>disposable, one time use,<br>standard syringe and needle. |
| Control of<br>anesthesia- time | Free control<br>(High/Low speed) | Free control<br>(High/Low speed) |
| anesthesia time | Small doses<br>Short time | Small doses<br>Short time |
| Source of<br>Electric power | Internal | Internal Battery |
| operation | Automatic/Manual<br>(5step speed control) | Automatic/Manual<br>(6step speed control) |
## 9. Statement of substantial equivalence
. .
をする。 …
Please refer to the electrical test report and performance test report. This device, Anesthesia Injection Tool Set (KIW-1000) is very similar with predicate device, DENTSPLY Professional.
·
| Section | Predicate Device<br>DENTSPLY Professional (U.S.A.) | KIWORKS Co., Ltd. |
|-----------------|----------------------------------------------------|-------------------|
| Product<br>name | MIDWEST<br>Comfort Control Syringe | KIW-1000 (e-joa) |
| Web<br>Address | www.dentsply.com | www.kiworks.co.kr |
{3}------------------------------------------------
| Injection<br>mode | 5mode button<br>-Block<br>-Infiltration<br>-Palate<br>-PDL<br>-intraosseous | 3 Mode<br>Auto1<br>Auto2<br>Manual 6 phases |
|-------------------|--------------------------------------------------------------------------------------------------------------|---------------------------------------------|
| Similarity | -An electronic pre-programmed anesthetic delivery device<br>-Use a motor to control the injection flow speed | |
10. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as follows: Testing information demonstrating safety and effectiveness of the Anesthesia Injection Tool Set (KIW-1000) in the intended environment of use is supported by testing that was conducted in accordance with EN60601-1, EN60601-1-2, EN55011:2007 and performance test (Bench). Both KIWORKS device and predicate device have first slow rate injection time in which the patient gets some degree of anesthesia. So for the performance test report, initial slow injection time was performed with predicate device. First slow injection time of Auto 1, KIW-1000 is similar to Predicate device's first 10 seconds. Auto 1 is designed for adult and Auto 2 is for children. For the children the first slow injection time is 22 second. "2013" ------------------------------------------------------------------------------------------------------------------------------------------ーーーーーーーーーーーーーーーーーーーーーーーー
None of the testing demonstrated any design characteristics that violated the requirements of the standards or resulted in any safety hazards.
#### 11. Discussion of Clinical Tests Performed:
Clinical testing was not conducted.
#### 12. Conclusions:
Based on the information provided in this submission we concluded that the Auto controlled injection system is substantially equivalent to the predicate and is safe and effective for its intended use. Technological and performance difference do not raise any new questions of safety or effectiveness.
{4}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized caduceus, a symbol often associated with medicine and healthcare, with three intertwined strands. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
KI Works Company, LTD C/O Ms. Eileen Yang BSR Korea Corporation Migun Technoworld 2 C-Dong 524 #533-1 Yongsan-Dong, Daejeon Republic of Korea KS
MAR 2 7 2012
Re: K113247
Trade/Device Name: Anesthesia Injection Tool Set Model No. KIW-1000 Regulation Number: 21 CFR 872.6770 Regulation Name: Cartridge Syringe Regulatory Class: II Product Code: EJI Dated: March 16, 2012 Received: March 16, 2012
Dear Ms. Yang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Ms. Yang
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K113247.
# Indications for Use
510(k) Number (if known):
Model No. KIW-1000 Device Name: Anesthesia Injection Tool Set
Indications for Use:
The device is indicated for the injection of local anesthetics for anesthesia administered prior to, or in conjunction with, dental procedures.
Prescription Use x (Part 21 CFR 801 Subpart D)
Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
··············································································································································································
Concurrence of CDRH, Office of Device Evaluation (ODE)
Susan Ruaro
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number:
Page 1 of 1
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.