RAPIDPOINT 500 BLOOD GAS ANALYER

K113216 · Siemens Healthcare Diagnostics · KHP · May 3, 2012 · Clinical Chemistry

Device Facts

Record IDK113216
Device NameRAPIDPOINT 500 BLOOD GAS ANALYER
ApplicantSiemens Healthcare Diagnostics
Product CodeKHP · Clinical Chemistry
Decision DateMay 3, 2012
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1450
Device ClassClass 1
AttributesPediatric

Indications for Use

The RAPIDPoint 500 system is intended for in vitro diagnostic use and is designed to provide the determination in whole blood for the following parameters: Partial pressure of carbon dioxide . Partial pressure of oxygen . pH . Sodium . Potassium . Ionized calcium . Chloride . Glucose . Total hemoglobin and fractions: fO2Hb, fCOHb, fMetHb, fHHb . Neonatal bilirubin . Lactate . This test system is intended for use in point of care or laboratory settings.

Device Story

RAPIDPoint 500 is a point-of-care/laboratory blood gas analyzer. This submission adds a lactate parameter via an integrated lactate sensor chip and modified reagent within a new Measurement Cartridge. Input: whole blood sample. Principle: amperometric enzymatic measurement; lactate oxidase on electrode surface reacts with sample lactate to produce hydrogen peroxide; oxidation of H2O2 generates current proportional to lactate concentration. Output: quantitative lactate concentration (mmol/L). Used in POC/clinical settings by clinicians/technicians. Output informs clinical decision-making regarding acid-base status and lactic acidosis treatment.

Clinical Evidence

Bench testing only. Precision studies (n=128) showed total %CV of 4.5% at 11.38 mmol/L and 6.6% at 0.96 mmol/L. Linearity confirmed across 0.11-30.15 mmol/L range (R²=0.9854). Method comparison against predicate (n=1218) yielded slope 1.042, intercept 0.010, R²=0.9887. Interference testing identified potential bias from isoniazid, acetaminophen, L-ascorbic acid, glycolic acid, and oxalate.

Technological Characteristics

Amperometric biosensor; working electrode contains lactate oxidase enzyme. Electrochemical cell maintains constant polarizing voltage. System is a point-of-care/laboratory blood gas analyzer. Software-enabled parameter (v2.0). No hardware/mechanical changes.

Indications for Use

Indicated for measurement of lactic acid in whole blood to evaluate acid-base status and assist in the diagnosis and treatment of lactic acidosis. For prescription use in point-of-care or clinical laboratory settings.

Regulatory Classification

Identification

A lactic acid test system is a device intended to measure lactic acid in whole blood and plasma. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood).

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k113216 B. Purpose for Submission: Addition of lactate parameter to the previously cleared RAPIDPoint 500 system C. Measurand: Lactic Acid/Lactate D. Type of Test: Quantitative, amperometric E. Applicant: Siemens Healthcare Diagnostics F. Proprietary and Established Names: RAPIDPoint 500® System G. Regulatory Information: 1. Regulation section: 21 CFR § 862.1450; Lactic acid test system 2. Classification: Class I, meets limitations of exemptions per 21 CFR § 862.9 (c)(9) 3. Product code: KHP; Lactic Acid, Enzymatic Method 4. Panel: Clinical Chemistry (75) H. Intended Use: 1. Intended use(s): The RAPIDPoint 500 system is intended for in vitro diagnostic use and is designed to provide the determination in whole blood for the following parameters: - Partial pressure of carbon dioxide - Partial pressure of oxygen - pH - Sodium - Potassium {1} - Ionized calcium - Chloride - Glucose - Total hemoglobin and fractions: fO₂Hb, fCOHb, fMetHb, fHHb - Neonatal bilirubin - Lactate This test system is intended for use in point of care or laboratory settings. 2. Indication(s) for use: Lactate: A lactic acid test system is a device intended to measure lactic acid in whole blood. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood). 3. Special conditions for use statement(s): - For prescription use only - For in vitro diagnostic use only - For point-of-care (POC) or clinical laboratory settings 4. Special instrument requirements: RAPIDPoint 500 system I. Device Description: Lactate is a new parameter offered on the RAPIDPoint 500 instrument (formerly the RAPIDPoint 405 instrument) previously cleared under k002738, k020616, and k110277. The RP500 system is a point-of-care and laboratory testing blood gas analyzer and currently measures pCO₂, pO₂, pH, sodium, potassium, calcium, chloride, glucose, fO₂Hb, fCOHb, fMetHb, fHHb and neonatal bilirubin. The measurement of lactate is accomplished through modification of the existing RP500 software and the addition of an integrated lactate sensor into a new RP500 Measurement Cartridge, which is a RP405 Measurement cartridge modified for lactate. To facilitate the lactate measurement, the RP500 Measurement Cartridge includes a modified reagent and the integrated lactate sensor chip. J. Substantial Equivalence Information: 1. Predicate device name(s): RAPIDLab 1200 System 2. Predicate K number(s): k031560 {2} 3. Comparison with predicate: | Item | Lactate on RAPIDPoint 500 | Lactate on RAPIDLab 1200 SYSTEM | | --- | --- | --- | | Similarities | | | | Intended use and indications for use | In vitro diagnostic test for the determination lactate concentration in the whole blood. A lactic acid test system is a device intended to measure lactic acid in whole blood. Lactic acid measurements that evaluate the acid-base status are used in the diagnosis and treatment of lactic acidosis (abnormally high acidity of the blood). | same | | Principle of operation | Blood Gas Analyzer | same | | Test principle | Amperometric lactate oxidase enzymatic | same | | Measured parameter | Lactate | same | | Technology | Electrochemical cells that employ amperometric technology | same | | Specimen type | Whole blood | same | | Calibration | 2 point calibration using automated on-board reagent | same | | Differences | | | | Settings for use | Point-of-care and clinical laboratories | Clinical laboratories | | Reportable range | 0.18 to 30 mmol/L | 0 to 30 mmol/L | K. Standard/Guidance Document Referenced (if applicable): Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices - Guidance for Industry and FDA Staff Format for Traditional and Abbreviated 510(k)s - Guidance for Industry and FDA Staff L. Test Principle: The lactate measurement is performed by a sensor that incorporates amperometric technology. The lactate biosensor is a complete electrochemical cell. A constant voltage, called a polarizing voltage, is maintained during analysis. In the lactate sensor, lactic acid from the sample interacts with the lactate oxidase on the surface of the measuring electrode to form pyruvic acid and hydrogen peroxide. The polarizing voltage is sufficient to cause oxidation of the hydrogen peroxide to oxygen. The loss of electrons in the oxidation of $\mathrm{H}_2\mathrm{O}_2$ creates a current flow that is directly proportional to the lactate concentration in the sample. M. Performance Characteristics (if/when applicable): 1. Analytical performance: a. Precision/Reproducibility: {3} The sponsor performed precision studies in-house following the guidelines provided in “Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline—Second Edition” using 4 lithium heparinized whole blood samples (prepared fresh for every run), 3 instruments, at least 2 operators, 3 lots of sensors and 3 lots of reagents. Each blood sample was measured in duplicate, twice a day in both modes (syringe and capillary) for a minimum of 20 days. The results are summarized below: Pooled Capillary | Mean (mmol/L) | Within run SD | Within run %CV | | --- | --- | --- | | 0.409 | 0.05 | 13.3 | | 1.408 | 0.10 | 7.0 | | 2.516 | 0.18 | 7.0 | | 22.05 | 1.65 | 7.5 | Pooled Syringe | Mean (mmol/L) | Within run SD | Within run %CV | | --- | --- | --- | | 0.361 | 0.03 | 8.7 | | 1.337 | 0.09 | 6.5 | | 2.487 | 0.12 | 4.7 | | 22.89 | 1.83 | 8.0 | The sponsor also evaluated 4 levels of control material, 3 instruments, 2 operators, 3 lots of sensors and 3 lots of reagents. Each level was measured in duplicate, twice a day for a minimum of 20 days in syringe mode. The results are summarized below: Pooled control material | Mean (mmol/L) | Within run SD | Within run %CV | Within Lab SD | Within Lab %CV | | --- | --- | --- | --- | --- | | 0.501 | 0.02 | 3.5 | 0.044 | 8.7 | | 0.912 | 0.02 | 2.3 | 0.039 | 4.3 | | 2.939 | 0.09 | 3.0 | 0.128 | 4.4 | | 22.96 | 1.06 | 4.6 | 1.58 | 6.9 | The sponsor also performed a reproducibility study at 3 point of care (POC) sites using a minimum of 3 typical POC operators per site and 3 levels of QC material. Each level was tested in quadruplicate in each run and 2 runs were performed each day using one instrument per site and one lot of reagents using QC mode. The results are summarized below: | Site | N | Mean (mmol/L) | Within run SD | Within run %CV | Total SD | Total %CV | | --- | --- | --- | --- | --- | --- | --- | | 1 | 40 | 11.77 | 0.26 | 2.2 | 0.33 | 2.8 | | 2 | 41 | 11.43 | 0.26 | 2.2 | 0.38 | 3.3 | {4} | 3 | 47 | 11.00 | 0.35 | 3.2 | 0.36 | 3.2 | | --- | --- | --- | --- | --- | --- | --- | | all | 128 | 11.38 | 0.30 | 2.6 | 0.51 | 4.5 | | 1 | 41 | 1.01 | 0.03 | 3.0 | 0.05 | 5.0 | | 2 | 42 | 0.97 | 0.03 | 3.6 | 0.04 | 4.6 | | 3 | 44 | 0.91 | 0.02 | 2.0 | 0.04 | 4.9 | | all | 127 | 0.96 | 0.03 | 3.0 | 0.06 | 6.6 | | 1 | 40 | 3.33 | 0.15 | 4.5 | 0.26 | 7.9 | | 2 | 41 | 3.19 | 0.05 | 1.6 | 0.18 | 5.6 | | 3 | 45 | 3.06 | 0.05 | 1.5 | 0.11 | 3.7 | | all | 126 | 3.19 | 0.10 | 3.2 | 0.22 | 6.8 | # b. Linearity/assay reportable range: A linearity study was conducted with lithium heparinized whole blood samples ranging from 0.11 to $30.15\mathrm{mmol}$ . The samples were run in replicates of six on one analyzer in syringe mode. The result of linear regression analysis is summarized below: | Slope | 0.979 | | --- | --- | | Intercept | 0.087 | | R² | 0.9854 | The execution of first, second and third order linear regression did not indicate statistically significant non linear regression coefficients in either the $2^{\mathrm{nd}}$ order or $3^{\mathrm{rd}}$ order models. The reportable range of this device for lactate measurements is 0.18 to 30 mmol/L. c. Traceability, Stability, Expected values (controls, calibrators, or methods): Calibrators: There is a 2 point calibration using automated on-board reagents. The $2.0\mathrm{mmol / L}$ lactate calibrator value assignment is conducted using the lactate oxidase spectrophotometric method and is traceable to a commercially available standard. The production lots are value assigned against the standard in several test runs using multiple samples in multiple replicates against multiple replicates of the standard on one instrument. Previously released lots are tested in each run. The protocols and acceptance criteria for the sponsor's pre-defined acceptance criteria for precision and accuracy were reviewed and found to be adequate. Quality Control Materials: Quality Control materials recommended for use with this system were previously cleared under k970956. The automatic quality control materials (AQC) cartridge was previously cleared under k020616. Reagent and calibrator stability: The sponsor evaluated the device's stability claims following a protocol designed to evaluate the measurement cartridge shelf life under customer use conditions. The sponsor monitors the {5} performance of the device using whole blood samples, quality control samples and calibration verification materials that are evaluated at a day (pre-determined as the designated test point) for each monthly stability interval. Since the calibrators are within the reagent cartridge, the stability of the calibrators is monitored along with the reagent stability. The proposed protocol and acceptance criteria were reviewed and found to be adequate. d. Detection limit: For the Detection Limit studies, 2 lots of sensors and 2 lots of reagents were used. The Limit of Blank (LoB) was determined using the wash reagent (no lactate added). One lot of wash reagent was analyzed in replicates of 11 on 2 instruments over 5 days (for a total of 5 reagent lots). The results were found not to be Gaussian. Therefore, the LOB was determined as the non-parametric $95^{\text{th}}$ percentile. The LoB was calculated to be $0.09 \, \text{mmol/L}$. The Limit of Detection (LoD) was determined using lithium heparinized whole blood collected from 5 donors. Eleven replicates of each sample were analyzed on 2 analyzers over 5 days. The LoD was calculated as $\mathrm{LoD} = \mathrm{LoB} + \mathrm{C}_{\beta} * \mathrm{SD}$ where SD is the pooled standard deviation and $C_{\beta}$ was calculated according to the following formula: $C_{\beta} \div (1 - 1 \div (4 \times \text{degrees of freedom of the estimated SD}))$. The LoD was calculated to be $0.11 \, \text{mmol/L}$. The limit of quantitation (LoQ) was determined using lithium heparinized whole blood collected from 5 donors. Each sample was analyzed in replicates of 5 on 4 instruments over 7 days. The LoQ was defined as the lowest concentration tested at which the $\% \mathrm{CV}$ of the device is less than $20\%$. The LoQ was determined to be $0.18~\mathrm{mmol / L}$. e. Analytical specificity: Interference studies: Lithium heparinized whole blood samples were spiked to either $0.7\mathrm{mmol / L}$ or $2.6\mathrm{mmol / L}$ lactate. The samples were split and either spiked with an interferent or an equivalent volume of solvent and then run on 2 instruments in syringe mode using 2 lots of sensors. The low lactate samples were measured in replicates of 8 and the high lactate samples were measured in replicates of 5. The bias was calculated using the following formula: Bias = (mean test-mean control) The percent interference was calculated using the following formula: $\%$ interference = (bias/mean control)*100. The sponsor defined no interference if the bias was within $\pm 10\%$. | Potential Interferent | Highest Concentration Tested (mg/dL) | | --- | --- | {6} | Acetaminophen | 2 | | --- | --- | | Acetoacetate | 20 | | Chlorpromazine | 0.2 | | Conjugated bilirubin | 20 | | Creatinine | 5 | | Dopamine | 0.1 | | Ethanol | 400 | | Glycolic acid | 5 | | Guaifenesin | 120 | | Hemoglobin | 2 g/L | | Heparin | 15000 U/L | | Hydroxybutyrate (beta) | 10 | | Ibuprofen | 12.5 | | Isoniazide | 1 | | L-Ascorbic acid | 1 | | Oxalate | 1 | | Penicillamine | 2.4 | | Phenobarbital | 9.6 | | Pyruvate | 2.7 | | Salicylate (sodium) | 70 | | Theophylline | 4 | | Thiocyanate | 41 | | Unconjugated bilirubin | 20 | | Urea | 257 | | Uric Acid | 20 | The sponsor states that isoniazid, acetaminophen and L-ascorbic acid demonstrate interference within the therapeutic range. The sponsor states that pralidoxime iodide interferes. The sponsor included the following under "Limitations" in the device's labeling: - Glycolic acid levels above $5\mathrm{mg / dL}$ and oxalate levels above $1\mathrm{mg / dL}$ may interfere. Both are metabolites of ethylene glycol". - Therapeutic levels of isoniazid, acetaminophen and L-ascorbic acid interfere". f. Assay cut-off: Not applicable # 2. Comparison studies: a. Method comparison with predicate device: The sponsor performed a method comparison study in-house. Lithium heparinized whole blood samples were collected and run in replicates of 2 on 6 proposed devices and 2 predicate devices, using 3 lots of reagents in syringe {7} mode and capillary mode using 3 operators. Ordinary least squares regression was used to determine the coefficient of determination $(\mathbf{r}^2)$ and Orthogonal Deming regression analysis was performed to determine the slope and intercept using singlicate values from the proposed device and the predicate device. Summary data (by individual instrument) is shown in the tables below: Capillary mode | n | Range | Slope (95% CI) | Intercept (95% CI) | r² | | --- | --- | --- | --- | --- | | 102 | 0.22 – 26.38 | 1.027 (1.005 – 1.049) | 0.039 (-0.189 – 0.266) | 0.9775 | | 101 | 0.22 – 26.38 | 1.038 (1.015 – 1.061) | -0.038 (-0.279 – 0.204) | 0.9756 | | 102 | 0.22 – 26.38 | 1.056 (1.036 – 1.076) | 0.017 (-0.194 – 0.227) | 0.9817 | | 102 | 0.30 – 29.56 | 1.089 (1.065 – 1.114) | -0.059 (-0.329 – 0.210) | 0.9742 | | 102 | 0.30 – 29.56 | 1.019 (0.994 – 1.043) | -0.128 (-0.395 – 0.139) | 0.9710 | | 101 | 0.30 – 29.56 | 1.019 (0.997 – 1.041) | -0.022 (-0.261 – 0.217) | 0.9767 | Syringe mode | n | Range | Slope (95% CI) | Intercept (95% CI) | r² | | --- | --- | --- | --- | --- | | 101 | 0.23 – 27.63 | 1.031 (1.012 – 1.050) | 0.119 (-0.092 – 0.330) | 0.9821 | | 102 | 0.23 – 27.63 | 1.007 (0.983 – 1.030) | 0.103 (-0.155 – 0.362) | 0.9714 | | 102 | 0.23 – 27.63 | 1.066 (1.046 – 1.085) | 0.132 (-0.079 – 0.343) | 0.9829 | | 101 | 0.29 – 27.37 | 1.067 (1.045 – 1.088) | 0.077 (-0.161 – 0.316) | 0.9799 | | 101 | 0.29 – 27.37 | 1.069 (1.051 – 1.087) | -0.143 (-0.346 – 0.059) | 0.9855 | | 101 | 0.29 – 27.37 | 1.016 (0.999 – 1.034) | 0.047 (-0.150 – 0.244) | 0.9849 | All instruments, all reagents, both modes combined | n | Range | Slope (95% CI) | Intercept (95% CI) | r² | | --- | --- | --- | --- | --- | | 1218 | 0.22 – 29.56 | 1.042 (1.036 – 1.049) | 0.010 (-0.058 – 0.079) | 0.9887 | The sponsor also performed method comparison studies at 3 point of care (POC) sites using at least 3 typical POC operators and one instrument per site. One cartridge lot and one reagent lot was used for this study. Left over clinical specimens were analyzed in duplicate using either syringe $(n = 82)$ or capillary $(n = 67)$ sampling modes, however, the statistical analysis used the first replicate only. Less than $10\%$ of the samples in the study were spiked samples. Ordinary least squares regression was used to determine the coefficient of determination $(r^2)$ and Orthogonal Deming regression analysis was performed to determine the slope and intercept. The results are summarized below: {8} | Site | n | Range (mmol/L) | Slope (95% CI) | Intercept (95% CI) | r2 | | --- | --- | --- | --- | --- | --- | | 1 | 46 | 1.75 – 25.84 | 1.100 (1.062 – 1.137) | -0.215 (-0.602 – 0.172) | 0.988 | | 2 | 60 | 1.41 – 24.04 | 1.006 (0.961 – 1.052) | -0.026 (-0.357 – 0.305) | 0.972 | | 3 | 43 | 1.48 – 21.28 | 1.008 (0.952 – 1.064) | 0.019 (-0.451 – 0.488) | 0.970 | | All | 149 | 1.41 – 25.84 | 1.054 (1.027 – 1.081) | -0.142 (-0.374 – 0.089) | 0.976 | b. Matrix comparison: This device is intended for use on lithium heparinized whole blood only. 3. Clinical studies: a. Clinical Sensitivity: Not applicable. b. Clinical specificity: Not applicable. c. Other clinical supportive data (when a. and b. are not applicable): Not applicable. 4. Clinical cut-off: Not applicable. 5. Expected values/Reference range: The sponsor references lactate values from Tietz NW, Textbook of Clinical Chemistry (2008), pg 852 Heparinized whole blood at bed rest: Venous: $0.56 - 1.39\mathrm{mmol / L}$ Arterial: $0.36 - 0.75\mathrm{mmol / L}$ N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...