K113212 · China Daheng Group, Inc. · GEX · Nov 16, 2011 · General, Plastic Surgery
Device Facts
Record ID
K113212
Device Name
DENTAL LASER THERAPY SYSTEM
Applicant
China Daheng Group, Inc.
Product Code
GEX · General, Plastic Surgery
Decision Date
Nov 16, 2011
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
The PenLase is indicated for incision, excision, hemostasis, coagulation and vaporization of soft tissue including the following indications: Excision and Incision Biopsies Hemostatic assistance Treatment of Apthous Ulcers Frenectomy Frenotomy Gingival Incision and Excision Gingivectomy Gingivoplasty Incising and Draining of Abscesses Operculectomy Oral Papillectomy Removal of Fibromas Soft Tissue Crown Lengthening Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) Tissue retraction for Impression Vestibuloplasty
Device Story
PenLase is a portable, battery-operated semiconductor diode laser system for oral soft tissue surgery. Device uses AlGaAs semiconductor chips to generate 810±10 nm infrared laser radiation. Laser power (0.7W or 1.7W) is delivered to the treatment site via disposable, bendable fiber tips. Operated by clinicians in a dental or clinical setting; no external cords or foot switches required. Charging station recharges batteries. Laser energy facilitates incision, excision, hemostasis, coagulation, and vaporization of soft tissue. Benefits include improved access via bendable tips and cordless portability for multiple procedures per charge.
Clinical Evidence
Bench testing only. Compliance verified with IEC 60825-1 (laser safety), IEC 60601-2-22 (therapeutic laser safety), IEC 60601-1 (general safety), IEC 60601-1-2 (EMC), ISO 10993-5 (cytotoxicity), and ISO 10993-10 (irritation/hypersensitivity). No clinical data provided.
Technological Characteristics
Semiconductor diode laser (AlGaAs); 810±10 nm wavelength; 0.7W/1.7W output power. Battery-powered, cordless, portable handpiece. Disposable fiber tips (400um). Standards: IEC 60825-1, IEC 60601-2-22, IEC 60601-1, IEC 60601-1-2, ISO 10993-5, ISO 10993-10. Biocompatible materials per ISO 10993.
Indications for Use
Indicated for oral soft tissue surgery including biopsies, hemostasis, ulcer treatment, frenectomy, frenotomy, gingival procedures, abscess drainage, operculectomy, papillectomy, fibroma removal, crown lengthening, sulcular debridement, tissue retraction, and vestibuloplasty.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
NOV 16 2011
13212
page ① of ⑤
Premarket Notification 5 10(k) Submission
Section III 510(k) Summary
Project #: M0222011Bd
Section III 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) Number:
- l. Date of Submission: 8/24/2011
- 2. Sponsor
China Daheng Group, Inc. PO Box 8618, #A9 Shangdi Xinxilu Haidian District, Beijing 100085, China Contact Person: Jianli Song Position: Vice general manager Establishment number: 3008705657 Tel: 010-82782668 - 300 Fax: 010-72782669 Email: songjnli@cdhcorp.com.cn
- 3. Submission Correspondent Mr. Tarzan Wang Mid-Link Consulting Co., Ltd P.O. Box 237-023 Shanghai, 200237, China Tel: +86-21-22815850 Fax: 240-238-7587 Email: tarzan.wang@mid-link.net
4. Proposed Device Identification
Proposed Device Name: Dental Laser Therapy System Proposed Device Model: penlase Classification: 2 Product Code: GEX Regulation Number: 21 CFR 878.4810 Review Panel: 878 General and Plastic Surgery
Indication for use Statement:
The PenLase is indicated for incision, excision, hemostasis; coagulation and vaporization of soft
III-1
{1}------------------------------------------------
K1132/2
page (2) of (5)
## Premarket Notification 510(k) Submission
Project #: M0222011Bd
tissue including the following indications:
- . Excision and Incision Biopsies
- 0 Hemostatic assistance
- 0 Treatment of Apthous Ulcers
- o Frenectomy
- . Frenotomy
- . Gingival Incision and Excision
- 0 Gingivectomy
- . Gingivoplasty
- 0 Incising and Draining of Abscesses
- � Operculectomy
- o Oral Papillectomy
- � Removal of Fibromas
- o Soft Tissue Crown Lengthening
- . Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
- . Tissue retraction for Impression
- 0 Vestibuloplasty
#### 5. Predicate Device Identification
510(k) Number: k081214 Product Name: Styla microlaser / Styla Ortho Manufacturer: Zap Lasers, LLC
510(k) Number: k093852 Product Name: iLase Manufacturer: Biolase Technology, Inc.
- 6. Device Description
PenLase is a surgical equipment, simple tool, portable, and easy to use. It is a tool that can be used for the treatment of oral soft tissue surgery, plastic surgery.
The PenLase utilizes a semiconductor diode which consists of semiconductor "chips" made from Aluminum, Gallium and Arsenide, together commonly referred to as AlGaAs , with invisible infrared radiation as a laser source. The laser power is delivered to the treatment area via a fiber tip. The disposable fiber tips are bendable in a variety of angle for maximum access to all areas of the mouth.
PenLase delivers 0.7 Watts and 1.7 Watts with wavelength 810±10 nm of continue laser power without cords, outlets or foot switches, which makes its battery life last longer.
. ·
{2}------------------------------------------------
ge of 5
Project #: M0222011Bd
The PenLase charging station recharges up to two PenLase batteries simultaneously, a fully charged battery can last for several patients or procedures.
The PenLase is generally indicated for incision, excision, hemostasis, coagulation and vaporization of soft tissue including the following indications:
- Excision and Incision Biopsies �
- . Hemostatic assistance
- � Treatment of Apthous Ulcers
- o Frenectomy
- 0 Frenotomy
- . Gingival Incision and Excision
- � Gingivectomy
- 0 Gingivoplasty
- Incising and Draining of Abscesses
- � Operculectomy
- . Oral Papillectomy
- Removal of Fibromas
- . Soft Tissue Crown Lengthening
- Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
- . Tissue retraction for Impression
- Vestibuloplasty
## Clinical Application:
| Excision and Incision Biopsies | 1.7W |
|----------------------------------------------------------------------------------------------|------|
| Hemostatic assistance | 0.7W |
| Treatment of Apthous Ulcers | 0.7W |
| Frenectomy | 1.7W |
| Frenotomy | 1.7W |
| Gingival Incision and Excision | 1.7W |
| Gingivectomy | 1.7W |
| Gingivoplasty | 1.7W |
| Incising and Draining of Abscesses | 1.7W |
| Operculectomy | 1.7W |
| Oral Papillectomy | 1.7W |
| Removal of Fibromas | 1.7W |
| Soft Tissue Crown Lengthening | 1.7W |
| Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket) | 0.7W |
| Tissue retraction for Impression | 1.7W |
| Vestibuloplasty | 1.7W |
{3}------------------------------------------------
K113212
Page 4 of 5
Premarket Notification 510(k) Submission
Accessories list:
Project #: M0222011Bd
| Accessories | Quantity |
|---------------------------------|----------|
| PenLase Main Handpiece | 1 |
| Charging Station, Power adapter | 1 |
| Laser Protective glasses | 1 |
| Patient goggle | 1 |
| Batteries | 2pcs |
| Disposable Fiber tips(400um) | 6pcs |
| OPERATION MANUAL | 1 |
| Laser warning sign | 1 |
| Quick Start Guide | 1 |
#### 7. Non-Clinical Test Conclusion
Bench tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
IEC 60825-1: 2007, Safety of laser products - Part 1: Equipment classification, requirements.
IEC 60601-2-22: 1995, Medical Electrical Equipment – Part 2: Particular requirements for the safety of diagnostic and therapeutic laser equipment.
IEC 60601-1:1988+A1:1991+A2:1995, Medical Electrical Equipment - Part1: General requirements for safety.
IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and tests (Edition 3).
ISO 10993-5: 2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity.
ISO 10993-10: 2002/Amd.1:2006, Biological evaluation of medical devices - Part 10: Tests for irritation and delayed-type hypersensitivity.
#### 8. Substantially Equivalent Conclusion
The proposed device, Dental Laser Therapy System, is determined to be Substantially Equivalent (SE) to the predicate device, Styla microlaser / Styla Ortho (k081214) and iLase (k093852), in respect of safety and effectiveness.
Analysis 1: The output power of proposed device is between that of the two predicates devices, and all of them comply with IEC 60825-1 and IEC 60601-2-22. So the difference will not affect the safety and effectiveness of the proposed device.
III-4
{4}------------------------------------------------
Project #: M0222011Bd
III-5
Analysis 2: the class of the aiming beam of proposed device is lower than that of the iLase, which means the proposed device has lower hazards. And all of them comply with IEC 60825-1 and IEC 60601-2-22. So the difference will not affect the safety and effectiveness of the proposed device.
Analysis 3: the Fiber tip diameter, beam diameter, spot size of proposed device are similar to that of the Styla Microlaser / Styla Ortho and iLase, and all of them comply with IEC 60825-1 and IEC 60601-2-22. So the difference will not affect the safety and effectiveness of the proposed device.
Analysis4: the button switch and foot switch are both intended to provide the handpiece on/off capabilities. And all of them comply with IEC 60825-1 and IEC 60601-2-22. So the difference will not affect the safety and effectiveness of the proposed device.
Analysis5: the dimension of the three devices does not affect the effectiveness of laser. And all of them comply with IEC 60825-1 and IEC 60601-2-22. So the difference will not affect the safety and effectiveness of the proposed device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES · USA" arranged around the perimeter. Inside the circle is an abstract symbol that resembles an eagle or a bird in flight, composed of three stylized, curved lines.
## Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
NOV 1 6 2011
China Daheng Group, Inc. % Underwriters Laboratories, Inc. Mr. Jeffrey D. Rongero 12 Laboratory Drive Research Triangle, North Carolina 27709
Re: K113212
Trade/Device Name: Dental Laser Therapy System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: October 31, 2011
Received: November 01, 2011
Dear Mr. Rongero:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Jeffrey D. Rongero
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Mark N. Melkerson
Director Division of Surgical, Orthopedic and Restorative Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
K1/32/2
Premarket Notification 510(k) Submission Section II Indications for Use Project #:M022201IBd
# Section II Indications for Use
510(k) Number: Device Name: Dental Laser Therapy System
Indications for Use:
The PenLase is indicated for incision, excision, hemostasis, coagulation and vaporization of soft tissue including the following indications:
- � Excision and Incision Biopsies
- � Hemostatic assistance
- . Treatment of Apthous Ulcers
- Frenectomy
- Frenotomy
- . Gingival Incision and Excision
- Gingivectomy
- . Gingivoplasty
- Incising and Draining of Abscesses
- Operculectomy
- . Oral Papillectomy
- . Removal of Fibromas
- . Soft Tissue Crown Lengthening
- . Sulcular Debridement (removal of diseased or inflamed soft tissue in the periodontal pocket)
- Tissue retraction for Impression
- Vestibuloplasty
ZPRESCRIPTION USE (Part 21 CFR 801 Subpart D) OOVER-THE-COUNTER USE (21 CFR 801 Subpart C)
NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
| (Division Signoffice of CDRH, Office of Device Evaluation (ODE) |
|-----------------------------------------------------------------|
|-----------------------------------------------------------------|
- Division of Surgical, Orthopedic,
- and Restorative Devices
Page 1 of 1
510(k) Number K113212
II-1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.