← Product Code [GGN](/productcode/GGN) · K113211

# HEMOSIL LA POSITIVE CONTROL & HEMOSIL LA NEGATIVE CONTROL (K113211)

_Instrumentation Laboratory CO · GGN · Jan 4, 2012 · Hematology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113211

## Device Facts

- **Applicant:** Instrumentation Laboratory CO
- **Product Code:** [GGN](/productcode/GGN.md)
- **Decision Date:** Jan 4, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 864.5425
- **Device Class:** Class 2
- **Review Panel:** Hematology

## Indications for Use

HemosIL LA Positive Control For use as an LA Positive Quality Control of Lupus Anticoagulant assays (HemosIL dRVVT Screen/dRVVT Confirm, HemosIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)]. HemosIL LA Negative Control: For use as an LA Negative Quality Control of Lupus Anticoagulant assays (HemosIL dRVVT Screen/dRVVT Confirm, HemsIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)].

## Device Story

HemosIL LA Positive and Negative Controls are lyophilized, assayed quality control materials used to monitor the precision and accuracy of Lupus Anticoagulant (LA) assays. The Positive Control is derived from human donors with anti-phospholipid antibodies; the Negative Control is derived from pooled normal human citrated platelet-poor plasma. These controls are processed on IL Coagulation systems (e.g., ACL TOP, ACL ELITE, ACL Futura, ACL Classic). Healthcare providers use the resulting clot time measurements to verify that the assay system is performing within established normalized ratio (NR) specifications (NR ≥ 1.4 for positive; NR ≤ 1.2 for negative). This ensures the reliability of patient test results for Lupus Anticoagulant, aiding in the clinical assessment of coagulation disorders.

## Clinical Evidence

No clinical data. Bench testing only. Stability data provided to support the 3-week reconstituted frozen storage claim.

## Technological Characteristics

Lyophilized human plasma-based controls. Positive control contains anti-phospholipid antibodies; negative control is pooled normal citrated platelet-poor plasma. Reconstituted stability: 24 hours at 2-8°C, 21 days at -20°C. Compatible with IL Coagulation systems (ACL TOP, ELITE, Futura, Classic).

## Regulatory Identification

A multipurpose system for in vitro coagulation studies is a device consisting of one automated or semiautomated instrument and its associated reagents and controls. The system is used to perform a series of coagulation studies and coagulation factor assays.

## Special Controls

*Classification.* Class II (special controls). A control intended for use with a multipurpose system for in vitro coagulation studies is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 864.9.

## Predicate Devices

- HemosIL LA Positive Control (k102552)
- HemosIL LA Negative Control (k102552)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY

A. 510(k) Number:
k113211

B. Purpose for Submission:
Addition of reconstituted freezing stability claims as well as update to Indications for Use to include use of the controls in the HemosIL dRVVT Screen/Confirm Assays

C. Measurand:
Controls for Lupus Anticoagulant (LA)

D. Type of Test:
Assayed Controls

E. Applicant:
Instrumentation Laboratory Co.

F. Proprietary and Established Names:
HemosIL LA Positive Control
HemosIL LA Negative Control

G. Regulatory Information:
1. Regulation section:
21CFR §864.5425, Multipurpose system for in vitro coagulation studies
2. Classification:
Class II
3. Product code:
GGN (Plasma Coagulation Control)
GGC (Plasma Control, Abnormal)
GIZ (Plasma Control, Normal)
4. Panel:
Hematology (81)

H. Intended Use:
1. Intended use(s):
HemosIL LA Positive Control
For use as an LA Positive Quality Control of Lupus Anticoagulant assays (HemosIL dRVVT Screen/dRVVT Confirm, HemosIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)].
HemosIL LA Negative Control:
For use as an LA Negative Quality Control of Lupus Anticoagulant assays (HemosIL dRVVT Screen/dRVVT Confirm, HemsIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)].
2. Indication(s) for use:
Same as intended use.
3. Special conditions for use statement(s):
For prescription use only

{1}

4. Special instrument requirements:
IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)]

I. Device Description:
HemosIL LA Positive Control is a lyophilized preparation from human donors exhibiting the presence of anti-phospholipid antibodies with added buffer. The device consists of 10 1-mL vials of lyophilized controls per package.

HemosIL LA Negative Control is a lyophilized preparation using human citrated platelet-poor plasma to make pooled normal plasma with added buffer. The device consists of 10 1-mL vials of lyophilized controls per package.

J. Substantial Equivalence Information:
1. Predicate device name(s):
HemosIL LA Positive Control
HemosIL LA Negative Control
2. Predicate 510(k) number(s):
k102552
3. Comparison with predicate:

|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | HemosIL LA Positive Control | Same  |
|  Analyte Tested | Lupus Anticoagulant | Same  |
|  Constituent Material | Lyophilized preparation from human donors exhibiting the presence of anti-phospholipid antibodies with added buffer | Same  |
|  Test System | IL coagulation system: ACL TOP Family, ACL ELITE/ELITE PRO/8/9/10000, ACL Futura/ACL Advance/ACL Advance, ACL Classic (100-7000) | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | HemosIL LA Positive Control | Same  |
|  Intended Use | For use as an LA Positive Quality Control of Lupus Anticoagulant assays (HemosIL dRVVT Screen/dRVVT Confirm, HemosIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)] | For use as an LA Positive Quality Control of Lupus Anticoagulant assays (HemosIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)]  |

{2}

|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|  Reconstituted Freezing Stability | Stable for 3 weeks at -20°C in the closed original vial | No reconstituted frozen stability claim  |
|  Similarities  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | HemosIL LA Negative Control | Same  |
|  Analyte Tested | Lupus Anticoagulant | Same  |
|  Constituent Material | Lyophilized preparation from human citrated platelet-poor plasma to make pooled normal plasma with added buffer | Same  |
|  Test System | IL coagulation system: ACL TOP Family, ACL ELITE/ELITE PRO/8/9/10000, ACL Futura/ACL Advance/ACL Advance, ACL Classic (100-7000) | Same  |
|  Differences  |   |   |
| --- | --- | --- |
|  Item | Device | Predicate  |
|   | HemosIL LA Negative Control | Same  |
|  Intended Use | For use as an LA Negative Quality Control of Lupus Anticoagulant assays (HemosIL dRVVT Screen/dRVVT Confirm, HemosIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)] | For use as an LA Positive Quality Control of Lupus Anticoagulant assays (HemosIL LAC Screen/LAC Confirm; HemosIL Silica Clotting Time) on IL Coagulation systems [ACL TOP Family; ACL ELITE/ELITE PRO/8/9/10000; ACL Futura/ACL Advance; ACL Classic (100-7000)]  |
|  Reconstituted Freezing Stability | Stable for 3 weeks at -20°C in the closed original vial | No reconstituted frozen stability claim  |

# K. Standard/Guidance Document Referenced (if applicable):

Guidance for Industry and FDA Staff - Assayed and Unassayed Quality Control Material, June 7, 2007
- CLSI EP25-A: Evaluation of Stability of In Vitro Diagnostic Reagents; Approved Guideline (2009)

# L. Test Principle:

LA Positive control is a lyophilized preparation from human donors with anti-phospholipid antibodies with added buffer. LA Negative control is a lyophilized preparation of a pool of normal human citrated platelet-poor plasma. The controls are used to assess the precision and accuracy of Lupus Anticoagulant (LA) assays

{3}

performed on IL Coagulation instrument platforms using HemosIL LA Reagents (LA dRVVT Screen/dRVVT Confirm, LAC Screen/LAC Confirm; Silica Clotting Time).

## M. Performance Characteristics (if/when applicable):

### 1. Analytical performance:

#### a. Precision/Reproducibility:

To evaluate precision 9 vials were run over 1 day from 3 lots (27 samples total) for each control and were analyzed on the ACL TOP base analyzer with both the HemosIL dRVVT Screen and dRVVT Confirm assays. Normalized Ratios (NR) were calculated as described below. Acceptance criteria were set at a NR of $\geq 1.4$ for the Positive Control, and a NR of $\leq 1.2$ for the Negative Control. Both LA Controls met precision specifications.

To evaluate reproducibility, 2 lots of each Control were tested with a single lot of HemosIL dRVVT Screen and dRVVT Confirm assay on representative members of the instrument families for 20 days with 2 runs per day and 2 replicates for each run for each sample level (N=80 per level/instrument/lot).

Testing was done on the following instruments: ACL TOP/500CTS, ACL 10000, ACL Advance, and ACL Classic. Acceptance criteria for both within-run and total reproducibility were set at a CV $\leq 6\%$. Both LA Controls met reproducibility specifications.

#### b. Linearity/assay reportable range:

Not applicable

#### c. Traceability, Stability, Expected values (controls, calibrators, or methods):

**Stability:**

The lyophilized shelf life for the LA Controls is 24 months at $2 - 8^{\circ}\mathrm{C}$.

After reconstitution the LA Controls are stable for 24 hours at $2 - 8^{\circ}\mathrm{C}$ and 21 days at $-20^{\circ}\mathrm{C}$ in the closed original vials.

The on-board stability of the LA Controls after reconstitution are:

- 24 hours on the ACL TOP Family
- 4 hours on the ACL ELITE/ELITE PRO 8/9/10000
- 4 hours on the ACL Futura/ACL Advance
- 4 hours on the ACL Classic (100-7000) System

**Freezing stability claim:**

Fresh vials were reconstituted and assayed at time 0 at the same time that the frozen vials were thawed in a $37^{\circ}\mathrm{C}$ water bath. Two vials per time point per lot were pooled and assayed and the results were compared to the mean of the results from fresh vials at time 0. Each lot was assayed with the LAC Screen, LAC Confirm, SCT Screen, SCT Confirm, dRVVT Screen, and dRVVT Confirm assays.

Acceptance criteria are $\pm 10\%$ of the zero point.

$\% \text{Variation} = (\text{Test clot time (sec)}) - (\text{Zero point clot time (sec)}) / \text{Zero point clot time (sec)} \times 100\%$

Because the LA Negative Controls were not stable at the 32 day time point, the 22 day time point was assayed and the acceptance criteria were met.

The LA Positive Controls was stable at the 32 day time point.

#### d. Detection limit:

Not applicable

{4}

e. Analytical specificity:
Not applicable

f. Assay cut-off:
Not applicable

2. Comparison studies:
a. Method comparison with predicate device:
Not applicable
b. Matrix comparison:
Not applicable

3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
Not applicable

4. Clinical cut-off:
Not applicable

5. Expected values/Reference range:
The expected values were based on determined normalized ratios (NR). The NR acceptance criterion for the HemosIL LA Positive Control was set at $\geq 1.40$, which is the same as that for the HemosIL LA Positive Control and LA Negative Control with the LAC and Silica Clotting Time assays. The Negative Control NR acceptance criterion was established based on the normal cut-off for the dRVVT assay, which is 1.2. Therefore the acceptance criterion for the negative control is set at $\leq 1.2$.

N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

---

**Source:** [https://fda.innolitics.com/device/K113211](https://fda.innolitics.com/device/K113211)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
