K113190 · Meridian Bioscience, Inc. · MJH · Feb 24, 2012 · Microbiology
Device Facts
Record ID
K113190
Device Name
TRU LEGIONELLA
Applicant
Meridian Bioscience, Inc.
Product Code
MJH · Microbiology
Decision Date
Feb 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 866.3300
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K113190 · Feb 24, 2012
TRU LEGIONELLA
Meridian Bioscience, Inc.
Retrospective human urine specimens from clinical patients
Retrospective clinical specimens were used in a method comparison study to evaluate the performance of the TRU Legionella assay against the predicate device (Binax NOW).
Patients with symptoms of pneumonia or for whom Legionella testing had been ordered; Sample Size: 227 retrospective specimens; Number of Sites: 4 (3 in the US, 1 in The Netherlands)
Binax NOW® Legionella assay
Percent agreement (positive and negative) between TRU Legionella and predicate
Indications for Use
The TRU Legionella assay is an in vitro, rapid, lateral-flow immunoassay for the qualitative detection of Legionella pneumophila serogroup 1 antigen in human urine specimens. It is designed to test specimens from patients with symptoms of pneumonia. Test results are to be used as an aid in diagnosis of Legionella pneumophila serogroup 1 infection. A negative result does not preclude infection with Legionella pneumophila serogroup 1. Test results are to be used in conjunction with information obtained from the patient’s clinical evaluation and other diagnostic procedures.
Device Story
TRU Legionella is a rapid, lateral-flow immunoassay for qualitative detection of Legionella pneumophila serogroup 1 antigen in human urine. The device utilizes test strips containing rabbit polyclonal capture and detection antibodies. The operator mixes urine with sample diluent in a conjugate tube, then applies the mixture to the test strip. Results are interpreted visually after approximately 20 minutes; a pink-red band at the control and test line positions indicates a positive result, while a band at the control line only indicates a negative result. The device is intended for use by healthcare professionals in clinical settings to aid in the diagnosis of Legionnaires' disease. It provides rapid diagnostic information to support clinical decision-making in patients presenting with pneumonia symptoms.
Clinical Evidence
Clinical trials evaluated 428 patient samples (retrospective and prospective) across multiple US sites and one site in the Netherlands. Performance was compared to the Binax NOW Legionella Urinary Antigen test. Results showed an overall positive percent agreement of 96.3% (95% CI: 91.7-98.4%) and negative percent agreement of 100.0% (95% CI: 98.7-100.0%). Reproducibility was assessed at three independent laboratories using blind-coded panels, achieving 100% correlation across all tested conditions.
Technological Characteristics
Lateral-flow immunoassay; utilizes rabbit polyclonal antibodies for capture and detection. Form factor includes test strips, conjugate tubes, and sample diluent. Visual readout method. No electronic components, software, or energy source required. Sterilization not specified.
Indications for Use
Indicated for qualitative detection of Legionella pneumophila serogroup 1 antigen in human urine specimens from patients with symptoms of pneumonia.
Regulatory Classification
Identification
Haemophilus spp. serological reagents are devices that consist of antigens and antisera, including antisera conjugated with a fluorescent dye, that are used in serological tests to identify Haemophilus spp. directly from clinical specimens or tissue culture isolates derived from clinical specimens. The identification aids in the diagnosis of diseases caused by bacteria belonging to the genus Haemophilus and provides epidemiological information on diseases cause by these microorganisms. Diseases most often caused by Haemophilus spp. include pneumonia, pharyngitis, sinusitis, vaginitis, chancroid venereal disease, and a contagious form of conjunctivitis (inflammation of eyelid membranes).
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 866.9.
Predicate Devices
Binax NOW® Legionella Urinary Antigen Test (K982238)
Submission Summary (Full Text)
{1}
1
# 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY
A. 510(k) Number:
K113190
B. Purpose for Submission:
To obtain substantial equivalent determination for this new device
C. Measurand:
Legionella pneumophila serogroup 1 antigen
D. Type of Test:
Rapid lateral flow immunoassay
E. Applicant:
Meridian Bioscience, Inc
F. Proprietary and Established Names:
TRU Legionella assay
G. Regulatory Information:
1. Regulation section:
21CFR 866.3300, Haemophilus spp. Serological Reagents
2. Classification:
Class II
3. Product code:
MJH: Legionella, spp., ELISA
4. Panel:
83 Microbiology
{2}
H. Intended Use:
1. Intended use(s):
The TRU Legionella assay is an in vitro, rapid, lateral-flow immunoassay for the qualitative detection of *Legionella pneumophila* serogroup 1 antigen in human urine specimens. It is designed to test specimens from patients with symptoms of pneumonia. Test results are to be used as an aid in diagnosis of *Legionella pneumophila* serogroup 1 infection. A negative result does not preclude infection with *Legionella pneumophila* serogroup 1. Test results are to be used in conjunction with information obtained from the patient’s clinical evaluation and other diagnostic procedures.
2. Indication(s) for use:
The TRU Legionella assay is an in vitro, rapid, lateral-flow immunoassay for the qualitative detection of *Legionella pneumophila* serogroup 1 antigens in human urine specimens. It is designed to test specimens from patients with symptoms of pneumonia. Test results are to be used as an aid in diagnosis of *Legionella pneumophila* serogroup 1 infection. A negative result does not preclude infection with *Legionella pneumophila* serogroup 1. Test results are to be used in conjunction with information obtained from the patient's clinical evaluation and other diagnostic procedures.
3. Special conditions for use statement(s):
For Prescription Use only
4. Special instrument requirements:
Not applicable
I. Device Description:
TRU Legionella is a single use capture immunoassay to detect *Legionella pneumophila* serogroup 1 in human urine specimens. The test consists of a Conjugate Tube, a Test Strip, and Sample Diluent/Negative Control. The Conjugate Tube contains a lyophilized bead of colloidal gold-linked polyclonal antibody to *Legionella pneumophila* serogroup 1 detector antibody. The Test Strip carries a nitrocellulose membrane with dried capture antibodies placed at a designated Test Line for *Legionella*. The Test Strip holder caps the Conjugate Tube during testing and subsequent disposal to reduce exposure to potential pathogens.
The conjugate bead is first rehydrated in the Conjugate Tube with the Sample Diluent/Negative Control. Patient sample is then added, the contents mixed and the Test Strip added. *Legionella pneumophila* serogroup 1 antigens, if present, bind to the
2
{3}
antibody-colloidal gold conjugate. When the sample migrates up the Test Strip to the Test Line, the antigen-conjugate complex is bound to the capture antibody, yielding a pink-red line. When no antigen is present, no complexes are formed and no pink-red line appears at the Test Line. An internal control line helps determine whether the test has been executed properly, if kit reagents are performing appropriately and that adequate flow has occurred through the Test Strip during a test run. A visible pink-red line at the Control position of the Test Strip should be present each time a specimen or control is tested. If no pink-red control line is seen, the test is considered invalid.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
Binax NOW® Legionella Urinary Antigen Test
2. Predicate 510(k) number(s):
K982238
3. Comparison with predicate:
| Similarities | | |
| --- | --- | --- |
| Item | Device | Predicate |
| | TRU Legionella | Binax NOW® Legionella |
| Qualitative/Quantitative | Qualitative | Qualitative |
| Target Antigen | Legionella pneumophila serogroup 1 | Legionella pneumophila serogroup 1 |
| Specimen Type | Human urine, preserved and unpreserved | Human urine, preserved and unpreserved |
| Detection Antibody | Colloidal gold conjugated Rabbit polyclonal antibody | Colloidal gold conjugated Rabbit polyclonal antibody |
| Reading Method | Visual | Visual |
| Results Interpretation | Negative: A single pink-red band at the Control Line position. Positive: Pink-red bands at both the Control and Legionella Test Line positions. Invalid: No band at the Control Line position, a pink-red band appearing after 21 minutes of incubation, or a band of any color other than pink-red | Negative: Single pink to purple colored Control Line visible in the top half of the window. Positive: Two pink to purple colored lines. Invalid: No line at the Control Line position or no lines at the Control or Sample Line positions. |
{4}
| Differences | | |
| --- | --- | --- |
| Item | Device | Predicate |
| Test Format | Rapid lateral flow immunoassay. Insert test strip into tube containing prepared sample. | Rapid immunochromatographic membrane assay. Insert sample swab into test device well. |
| Reagents/Components | Test Strip, Conjugate Tube, Sample Diluent/Negative Control, Positive Control, Plastic transfer pipettes with 100, 200 and 300 μL marks | Test Device, Reagent A, Positive Control Swab, Negative Control Swab, Swabs |
| Urine Sample Preparation | 1. Add 100 μL of Sample Diluent to Conjugate Tube. Vortex for 10 seconds.
2. Add 100 μL of thoroughly mixed urine sample to the Conjugate Tube.
3. Mix sample and conjugate thoroughly | 1. Dip swab into the urine sample and then insert swab into the bottom hole of the Test Device.
2. Add 2 drops of Reagent A to the bottom hole. |
| Conjugate format | Lyophilized bead of conjugate antibody in the Conjugate tube. | Conjugate antibody dried onto fibrous conjugate pad. |
| General Laboratory Equipment required | Vortex, Disposable latex gloves | Urine collection container, BinaxNOW® Legionella Urinary Antigen Control Swab Pack |
| Testing Time | Approximately 20 minutes | Approximately 15 minutes |
K. Standard/Guidance Documents Referenced:
- User Protocol for Evaluation of Qualitative Test Performance (EP12-A2)
- User Verification of Performance for Precision and Trueness (EP15-A2)
- Guidance on Informed Consent for In Vitro Diagnostic Device Studies Using Leftover Human Specimens that are Not Individually Identifiable - Guidance for Sponsors, Institutional Review Boards, Clinical Investigators and FDA Staff
- Guidance for Industry and FDA Staff - Statistical Guidance on Reporting
{5}
Results from Studies Evaluating Diagnostic Tests
## L. Test Principle:
The conjugate bead is first rehydrated in the Conjugate Tube with the Sample Diluent/Negative Control. Patient sample is then added, the contents mixed and the Test Strip added. *Legionella pneumophila* serogroup 1 antigens, if present, bind to the antibody-colloidal gold conjugate. When the sample migrates up the Test Strip to the Test Line, the antigen-conjugate complex is bound to the capture antibody, yielding a pink-red line. When no antigen is present, no complexes are formed and no pink-red line appears at the Test Line. An internal control line helps determine whether the test has been executed properly, if kit reagents are performing appropriately and that adequate flow has occurred through the Test Strip during a test run. A visible pink-red line at the Control position of the Test Strip should be present each time a specimen or control is tested. If no pink-red control line is seen, the test is considered invalid.
## M. Performance Characteristics (if/when applicable):
### 1. Analytical performance:
a. **Precision/Reproducibility:**
Reproducibility was assessed at three clinical laboratories using a blind coded sample panel. The panel consisted of three contrived moderately positive samples, three low positive samples at a concentration just above the limit of detection, three high negative samples at a concentration just below the limit of blank, and one natural negative sample. The panels were tested in two runs at each site by two operators each day for five non-consecutive days. Reproducibility was 100%.
b. **Linearity/assay reportable range:**
Not applicable
c. **Traceability, Stability, Expected values (controls, calibrators, or methods):**
**Sample storage**
A sample storage study was conducted to verify the sample preservation, handling, and storage criteria included in the TRU Legionella package insert. Aliquots of unpreserved and boric acid preserved positive and negative samples were stored at the following temperatures to determine the worst case storage time for each temperature range: room temperature (20-26 °C) up to
{6}
eight days, refrigerated (2-8 °C) up to eight days, conventional freezer (-16 to -28 °C) up to 31 days, and ultralow freezer (-66 to -84 °C) up to 31 days. Ten contrived samples were tested: four negatives, four positives, a high negative sample, and a limit of detection positive. Samples were tested at various times during storage. All samples produced the expected results. The data supports test performance with preserved and unpreserved samples that are stored for up to seven days at 2-8 °C and 20-26 °C, and/or stored for up to 30 days at ≤ -16 °C.
## Fresh versus frozen samples
A fresh versus frozen specimen study was conducted to verify that frozen samples can be thawed multiple times without affecting the detection of the measurand. Ten contrived samples were tested: four negatives, four positives, a high negative sample, and a limit of detection positive. Unpreserved and boric acid preserved specimens were tested. Baseline tests were run, and then aliquots were tested every cycle for three freeze thaw cycles each at -16 to -28 °C and at -66 to -84 °C. All samples produced the expected results. The data supports test performance with preserved and unpreserved samples that are frozen and thawed up to two times.
## Kit stability
Reagent stability was provisionally established using accelerated (stressed) testing methods. TRU Legionella kits were stored at 3-7 °C, 21-25 °C, 40-44 °C, and 48-52 °C. Test performance was evaluated at various time points for up to 65 or 104 days using a sample panel containing a negative sample, low positive sample, reference standard, and both positive and negative controls. The 2-8 °C temperature range was used as T1 (initial or baseline temperature) when calculating stability using the Q rule, with Q₁₀ = 3. The 40-44 °C and 48-52 °C ranges were used as T2. Real-time stability studies are currently ongoing with testing performed at quarterly intervals and will be completed at the end of 27 months.
All samples produced the expected result with kits stored at the different temperatures and times. The data supports final product storage at 2-8 °C for 18 months.
## High-dose hook effect
L. pneumophila Philadelphia and Bellingham strains were spiked into negative donor urine and then serially diluted. A false negative result obtained with a very high level of target antigen would indicate a high-dose hook effect.
Determination of a high-dose hook effect with L. pneumophila Philadelphia strain ATCC 33152
{7}
| Sample | Replicate | Result |
| --- | --- | --- |
| Positive Control | Positive | Pass |
| Negative Control | Negative | Pass |
| Negative Sample Control | Negative | Pass |
| | | |
| Dilution 1: 2.0 x 108CFU/mL | Positive | Pass |
| Dilution 2: 5.0 x 107CFU/mL | Positive | Pass |
| Dilution 3: 1.25 x 107CFU/mL | Positive | Pass |
| Dilution 4: 3.1 x 106CFU/mL | Positive | Pass |
| Dilution 5: 7.8 x 105CFU/mL | Positive | Pass |
| Dilution 6: 2.0 x 105CFU/mL | Positive | Pass |
Determination of a high-dose hook effect with $L$ pneumophila Bellingham strain NCTC 11404
| Sample | Replicate | Result |
| --- | --- | --- |
| Positive Control | Positive | Pass |
| Negative Control | Negative | Pass |
| Negative Sample Control | Negative | Pass |
| | | |
| Dilution 1: 4.2 x 108CFU/mL | Positive | Pass |
| Dilution 2: 1.0 x 108CFU/mL | Positive | Pass |
| Dilution 3: 2.6 x 107CFU/mL | Positive | Pass |
| Dilution 4: 6.5 x 106CFU/mL | Positive | Pass |
| Dilution 5: 1.6 x 106CFU/mL | Positive | Pass |
| Dilution 6: 4.1 x 105CFU/mL | Negative | Pass* |
*A negative result is not unexpected as Dilution 6 was at a concentration below the claimed limit of detection.
$L$ pneumophila Philadelphia and Bellingham strains did not produce a high-dose hook effect when tested at high concentrations with the TRU Legionella assay.
# d. Detection limit:
The analytical limit of detection (LoD) was determined for $L$ pneumophila Philadelphia and Bellingham strains diluted in a negative human urine matrix. The $L$ pneumophila antigen was serially diluted two-fold and 45 replicates of each dilution were tested to determine the LoD in the TRU Legionella assay.
{8}
| Strain ID | Serogroup | Subgroup | Limit of Detection (LoD) CFU/mL |
| --- | --- | --- | --- |
| L. pneumophila
Philadelphia strain | 1 | Pontiac | 3.76 x 10^{5} |
| L. pneumophila
Bellingham strain | 1 | Non-Pontiac | 5.2 x 10^{5} |
e. Analytical specificity:
Analytical specificity was demonstrated by testing the following potentially cross-reactive bacteria, fungi, and viruses: *Acaligenes faecalis*, *Bacillus cereus*, *Bacillus subtilis*, *Candida albicans*, *Candida glabrata*, *Candida parapsilosis*, *Citrobacter freundii*, *Enterobacter aerogenes*, *Enterobacter cloacae*, *Enterococcus faecalis* (Group D Streptococcus), *Enterococcus faecium*, *Escherichia coli*, *Haemophilus influenza*, *Klebsiella pneumoniae*, *Legionella bozemanii*, *Legionella dumoffii*, *Legionella feelii*, *Legionella gormanii*, *Legionella longbeachae*, *Legionella micdadei*, *Legionella pneumophila* serogroup 2, *Legionella pneumophila* serogroup 3, *Legionella pneumophila* serogroup 4, *Legionella pneumophila* serogroup 5, *Legionella pneumophila* serogroup 6, *Morganella morganii*, *Moraxella osloensis*, *Mycoplasma pneumoniae*, *Nocardia asteroids*, *Proteus mirabilis*, *Proteus vulgaris*, *Pseudomonas aeruginosa*, *Serratia liquefaciens*, *Serratia marcescens*, *Staphylococcus aureus*, *Staphylococcus epidermidis*, *Staphylococcus saprophyticus*, *Streptococcus pyogenes* (Group A), *Streptococcus agalactiae* (Group B), *Streptococcus anginosis* (Group F), *Streptococcus dysgalactiae equisimilis* (Group G), *Streptococcus pneumoniae*, *Adenovirus*, *Coxsackievirus*, *Influenza A*, *Influenza B*, *Parainfluenza virus*, *Respiratory Syncytial Virus A*, and *Respiratory Syncytial Virus B*.
Bacteria, fungi (at a final concentrations of $1.2 \times 10^{8}$ CFU/ml), or viruses (at concentrations greater than $1 \times 10^{5}$ TCID$_{50}$/ml) were added to negative urine samples or to urine samples spiked with $L.$ pneumophila serogroup 1 at its limit of detection. None of the tested microorganisms cross-reacted with the TRU Legionella assay.
Analytical reactivity
TRU Legionella was tested for its reactivity with Pontiac and non-Pontiac strains of $L.$ pneumophila. Aliquots of five Pontiac and four non-Pontiac strains of $L.$ pneumophila were adjusted to visually match a MacFarland Standard 4.0 (approximately $1.2 \times 10^{9}$ CFU/mL) and then spiked into a natural negative urine matrix to a final concentration of $1.2 \times 10^{6}$ CFU/mL. Three non-Pontiac strains did not react at the concentration of $1.2 \times 10^{6}$ CFU/mL and were retested at a higher concentration of $4.8 \times 10^{6}$ CFU/mL.
{9}
Analytical reactivity of *L. pneumophila* strains at 1.2 x 10⁶ CFU/mL
| L. pneumophila strain | Source | Designation | CFU/mL | Test Sample | Result |
| --- | --- | --- | --- | --- | --- |
| Positive Control | N/A | N/A | N/A | Positive | Pass |
| Negative Control | N/A | N/A | N/A | Negative | Pass |
| Dilution Control | N/A | N/A | N/A | Negative | Pass |
| | | | | | |
| 13395 | CCUG | Pontiac | 1.2 x 10⁶ | Positive | Pass |
| 12024, Allentown 1 | NCTC | Pontiac | 1.2 x 10⁶ | Positive | Pass |
| 12006, Benidorm | NCTC | Pontiac | 1.2 x 10⁶ | Positive | Pass |
| 12008, OLDA | NCTC | Non-Pontiac | 1.2 x 10⁶ | Positive | Pass |
| 33058, Knoxville | CCUG | Pontiac | 1.2 x 10⁶ | Positive | Pass |
| 12007, France | NCTC | Pontiac | 1.2 x 10⁶ | Positive | Pass |
| 12098, Camperdown | NCTC | Non-Pontiac | 1.2 x 10⁶ | Negative | Fail* |
| 12025, Heysham | NCTC | Non-Pontiac | 1.2 x 10⁶ | Negative | Fail* |
| 12009, Oxford | NCTC | Non-Pontiac | 1.2 x 10⁶ | Negative | Fail* |
*Analytical reactivity of *L. pneumophila* strains at 4.8 x 10⁶ CFU/mL
| L. pneumophila strain | Source | Designation | CFU/mL | Test Sample | Result |
| --- | --- | --- | --- | --- | --- |
| Positive Control | N/A | N/A | N/A | Positive | Pass |
| Negative Control | N/A | N/A | N/A | Negative | Pass |
| Dilution Control | N/A | N/A | N/A | Negative | Pass |
| | | | | | |
| 12098, Camperdown | NCTC | Non-Pontiac | 4.8 x 10⁶ | Positive | Pass |
| 12025, Heysham | NCTC | Non-Pontiac | 4.8 x 10⁶ | Positive | Pass |
| 12009, Oxford | NCTC | Non-Pontiac | 4.8 x 10⁶ | Positive | Pass |
Results indicate that TRU Legionella will detect Pontiac and non-Pontiac strains.
f. Assay cut-off:
Not applicable
g. Interfering substances:
The following substances were tested for interference with the TRU Legionella assay:
Amphotericin B (0.22 mg/mL), Antihistamine (0.22 mg/mL), Ascorbic acid (1.0 mg/mL), Beet root (0.01%), Bilirubin (0.2 mg/mL), Boric acid (2.63 mg/mL), Caffeine, purified (0.4%), Chlorophyll (0.81 mg/mL), Ciprofloxacin (0.22 mg/mL), Cold and flu tablets (50 mg/ml), Cough drops (0.22 mg/mL), Cough syrup (0.20 mg/mL), Decongestant (0.22 mg/mL), Erythromycin (0.067 mg/mL), Glucose (20 mg/mL), Itraconazole (0.22 mg/mL), Miconazole (5%), Oxalic acid (0.01%), Prednisone (0.22 mg/mL), Protein (BSA) (5 mg/mL), Rifampicin (0.09 mg/mL), Tobacco, purified (0.4%), Urea (20 mg/mL), Vaginal contraceptive gel with nonoxynol-9, 4% (5%), Water-
{10}
based personal lubricant (KY Jelly) (5%), White blood cells (10%), Whole blood (10%).
The substances were added to three negative samples and three samples spiked with L. pneumophila antigen at the limit of detection. These substances did not interfere with the results of the TRU Legionella assay.
## 2. Comparison studies:
a. Method comparison with predicate device:
Clinical performance of the TRU Legionella assay was demonstrated by comparison to the predicate Binax NOW® Legionella assay. Results were reported as percent agreement between the two assays. The method comparison study was performed independently at three sites in the US and one site in The Netherlands using three lots of TRU Legionella. 123 prospective and 227 retrospective human urine specimens were tested from patients with symptoms of pneumonia or for whom Legionella testing had been ordered. Retrospective samples were randomized and masked prior to testing.
Percent agreement of TRU Legionella to Predicate Assay for Retrospective and Prospective Specimens from Patients >21 years of age.
| | Predicate Assay | | |
| --- | --- | --- | --- |
| TRU Legionella | Positive | Negative | Total |
| Positive | 106 | 0 | 106 |
| Negative | 5 | 239 | 244 |
| Total | 111 | 239 | 350 |
| | | | 95% CI |
| Positive Agreement | 106/111 | 95.5% | 89.9 – 98.1% |
| Negative Agreement | 239/239 | 100.0% | 98.4 – 100.0% |
{11}
Percent agreement of TRU Legionella to Predicate Assay for Retrospective Specimens from Patients >21 years of age
| | Predicate Assay | | |
| --- | --- | --- | --- |
| TRU Legionella | Positive | Negative | Total |
| Positive | 101 | 0 | 101 |
| Negative | 5 | 121 | 126 |
| Total | 106 | 121 | 227 |
| | | | 95% CI |
| Positive Agreement | 101/106 | 95.3% | 89.4 – 98.0% |
| Negative Agreement | 121/121 | 100.0% | 96.9 – 100.0% |
Percent agreement of TRU Legionella to Predicate Assay for Prospective Specimens from Patients >21 years of age
| | Predicate Assay | | |
| --- | --- | --- | --- |
| TRU Legionella | Positive | Negative | Total |
| Positive | 5 | 0 | 5 |
| Negative | 0 | 118 | 118 |
| Total | 5 | 118 | 123 |
| | | | 95% CI |
| Positive Agreement | 5/5 | 100.0% | 56.6 – 100.0% |
| Negative Agreement | 118/118 | 100.0% | 96.8 – 100.0% |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
See method comparison with predicate device.
{12}
4. Clinical cut-off:
Not applicable
5. Expected values/Reference range:
*Legionella pneumophila* serogroup 1 is an important cause of travel, community and hospital acquired pneumonia worldwide. Legionnaires’ disease (LD) is known to occur sporadically and in outbreak settings. Recent data indicates that 0.5-5.0% of adults hospitalized for pneumonia has LD. Approximately 8,000-18,000 cases of LD occur each year in the USA. With this assay, 111 positive samples were tested, 5 were misidentified as negative; and 239 negative samples were tested, all were correctly identified.
N. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
O. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
12
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.