← Product Code [IYN](/productcode/IYN) · K113184

# LOGICSCAN 64 LOGICSCAN 128 (K113184)

_Telemed · IYN · Jan 26, 2012 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113184

## Device Facts

- **Applicant:** Telemed
- **Product Code:** [IYN](/productcode/IYN.md)
- **Decision Date:** Jan 26, 2012
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1550
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

The LogicScan 64 / LogicScan 128 systems and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body. The applications are: cardiac (adult), fetal, abdominal, pediatric, small organ, transrectal, neonatal cephalic, peripheral vessel and musculo-skeletal (conventional and superficial). It is possible to provide diagnostic information outside of an imaging lab, including at the bedside systems, for navigated medical application, in operating rooms/critical care units.

## Device Story

LogicScan 64/128 are transportable, software-controlled diagnostic ultrasound systems. Hardware consists of a beamformer in a standalone enclosure or PC-integrated module, connected via USB 2.0 to a standard Windows-based PC. Probes (2-12 MHz) connect to the beamformer; raw data is transmitted to the PC for processing. Echo Wave II software on the PC handles image reconstruction, display, measurement, and archiving. Used in clinics, ORs, and bedside by clinicians. Provides real-time B-mode, M-mode, Color/Power Doppler, and PWD imaging. Benefits include portable diagnostic imaging and needle guidance for various clinical applications.

## Clinical Evidence

Bench testing only. Compliance with safety and performance standards (IEC 60601-1, 60601-1-2, 60601-1-4, 60601-2-37, NEMA UD 2, NEMA UD 3) and biocompatibility (ISO 10993) was demonstrated. No clinical data required as indications are not novel.

## Technological Characteristics

Modular ultrasound system; 2-12 MHz transducers; electronic linear/convex scanning. Materials: Silicone rubber, ABS, PPSU. Connectivity: USB 2.0 to host PC. Power: Internal medical-grade supply. Software: Echo Wave II (Windows-based). Standards: IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-2-37, NEMA UD 2, NEMA UD 3, ISO 10993, ISO 14971, IEC 62304.

## Regulatory Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Predicate Devices

- MySono U5 Diagnostic Ultrasound System ([K100186](/device/K100186.md))
- SonoAce R7 Diagnostic Ultrasound System ([K112646](/device/K112646.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/1 description: The image shows a sequence of handwritten digits and letters. The sequence appears to be "K113184". The characters are written in a simple, somewhat uneven style, with varying stroke thicknesses. The image has a white background.

Pg. 1.2-3

# 510(k) Summary of Safety and Effectiveness

#### ldentifying information 1.

JAN 2 6 2012

| Manufacturer   | TELEMED                                              |
|----------------|------------------------------------------------------|
| Address        | Dariaus ir Gireno str. 42 Vilnius LT-02189 Lithuania |
| Telephone      | +370-5 2106272 +370-5 2106273                        |
| Fax            | +370-5 2306733                                       |
| E-mail         | info@telemed.lt yury@telemed.lt                      |
| Contact        | Yury Sokolov / Engineering Manager                   |
| Name of Device | LogicScan 64 / LogicScan 128                         |

### 2. Class and Predicate Information

| Classification Name                      | FR Number                               | Product Cod |
|------------------------------------------|-----------------------------------------|-------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550                                | PIN         |
| Ultrasonic Pulsed Echo Imaging System    | 892.1560                                | IYO         |
| Diagnostic Ultrasonic Transducer         | 892.1570                                | ITX         |
| Common Name                              | Ultrasound imaging system               |             |
| Propietary Name                          | LogicScan                               |             |
| Class                                    | Regulatory Class II                     |             |
| Predicate Device                         | SAMSUNG MEDISON CO., LTD                |             |
|                                          | MySono U5 Diagnostic Ultrasound System  | K100186     |
|                                          | SonoAce R7 Diagnostic Ultrasound System | K112646     |

### 3. Performance Standards

The LogicScan family has been designed to meet the following:

Safety and EMC Requirements for Medical Equipment:

IEC 60601-1: 2000, Part 1: General requirements for safety.

IEC 60601-1-2: 2001, Part 1: General requirements for basic safety and essential performance, 2. Collateral standard: Electromagnetic compatibility - Requirements and tests

IEC 60601-1-2: 2001, Part 1: General requirements for basic safety and essential performance, 2.Collateral standard: Electromagnetic compatibility - Requirements and tests

IEC 60601-1-2: 2001, Part 1: General requirements for basic safety and essential performance, 2. Collateral standard: Electromagnetic compatibility - Requirements and tests

IEC 60601-1-2: 2001, Part 1: General requirements for basic safety and essential performance, 2.Collateral standard: Electromagnetic compatibility - Requirements and tests

IEC 60601-1-4: 1996, Part 1: General Requirements for Safety, 4.Collateral Standard: Programmable Electrical Medical Systems

IEC 60601-2-37: 2007-08 Particular requirements for the basic safety and essential

{1}------------------------------------------------

performance of ultrasonic medical diagnostic and monitoring equipment

NEMA UD 2-2004: 2003. Acoustic Output Measurement Standard for Diagnostic Ultrasound Fauipment

NEMA UD 3-2004: 2004, Standard for Real Time Display of Thermal and Mechanical Acoustic Output Indices on Diagnostic Ultrasound Equipment

AIUM MUS: 2002, Medical Ultrasound Safety

ISO-10993-1. Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, 2003

ISO-10993-5. Biological Evaluation of Medical Devices Part 5: Tests for in vitro cytotoxicity , 1999

ISO-10993-10, Biological Evaluation of Medical Devices Part 10: Tests for irritation and delayed-type hypersensitivity, 2002

ISO-10993-11, Biological Evaluation of Medical Devices Part 11: Tests for systemic toxicity, 2006

IEC 62304: 2006 Medical device software -- Software life cycle processes

ISO 14971:2007 Medical devices -- Application of risk management to medical devices

Essential Requirements of Council directive 93/42/EEC (Medical Device Directive)

The system's acoustic output is in accordance with ALARA principle (as low as reasonably achievable)

#### 4. Indication for Use

LogicScan 128 / LogicScan 64 ultrasound imaging systems are intended to be used for applications in cardiac (adult), fetal, abdominal, pediatric, small organ, transrectal, neonatal cephalic, peripheral vessel and musculo-skeletal (conventional and superficial). It is possible to provide diagnostic information outside of an imaging lab, including at the bedside systems, for navigated medical application, in operating rooms/critical care units.

#### 5. Device Description

LogicScan 128 / LogicScan 64 color diagnostic systems are intended for the multipurpose ultrasound examinations, based on electronic linear and convex scanning.

LogicScan 128 / LogicScan 64 is a combination of proprietary hardware and software that has been designed for real-time imaging and is intended to be a basic diagnostic tool. The system is based on a modular and flexible architecture allowing for both mobile and stationary (installed) configurations. The system is designed for imaging with transducer ranges of 2 to 12 MHz.

The devices referenced in this submission represent a transportable, software-controlled, diagnostic ultrasound system with accessories.

USB 2.0 connection between the beamformer and PC is a novel feature for ultrasound systems.

The LogicScan 128 / LogicScan 64 only contains the hardware and firmware, everything else (e.g. ultrasound software, database) is located on a standard PC that is connected to the LogicScan 128 / LogicScan 64 via USB 2.0. Minimum requirements are given for the PC. The probes are connected to the LogicScan 128 / LogicScan 64. All sonograms are saved on the PC and can there be evaluated, printed and archived. The Echo Wave II software was especially designed for the LogicScan 128 / LogicScan 64. Software able to reside in a Windows-based PC. The LogicScan 128 / LogicScan 64 can be used together with the appropriate probes for the entire 

{2}------------------------------------------------

ultrasound diagnostic (2MHz to 12MHz probes). Two probes can work simultaneously for LogicScan 128 2Z modifications.

.

.

.

Imaging Modes

- B
- B+B
- 4B
- B+M
- M
- 
- 
- B+CFM/PDI/DPD1+PWD (Triplex) . HPRF(LogicScan 128)

B+PWD (Duplex)

Pulse Wave Doppler (PWD)

Directional Power Doppler (DPDI)

- Color Doppler (CFM)
- Power Doppler (PDI)
- .
- Tissue Harmonic Imaging (THI) (LogicScan 128) .
- The devices included in this submission are as follows:

LogicScan 128 EXT-1Z / EXT-2Z ultrasound system utilizing as hardware and firmware an ultrasound engine contained in a small stand alone enclosure for connection to a host PC via a USB port with internal power supply:

LogicScan 128 CEXT-1Z is a compact version of LogicScan 128 EXT-1Z with internal medical grade power supply;

LogicScan 128 INT-1Z / INT-2Z ultrasound systems utilizing as hardware and firmware an ultrasound engine contained in a small enclosure for insertion to a host PC (to a drive bays);

A probe. 128 element convex array, at a central ultrasonic frequency of approximately 3.5MHz, model C3.5/60/128Z.

A probe, 128 element linear array at a central ultrasonic frequency of approximately 9 MHz, model HL9.0/40/128Z.

A probe, 128 element convex array at a central ultrasonic frequency of approximately 6.5 MHz, model PV6.5/10/128Z.

A biplane probe BiopSee, combination of:

- 128 element convex array at a central ultrasonic frequency of approximately . 6.5 MHz, model BIPC6.5/10/128Z
- . 128 element linear array at a central ultrasonic frequency of approximately 7.5 MHz, model BIPL7.5/70/128Z

LogicScan 64 FLT-1T ultrasound system utilizing as hardware and firmware an ultrasound engine contained in a small stand alone enclosure for connection to a host PC via a USB port;

A probe, 64 element convex array, at a central ultrasonic frequency of approximately 3.5 MHz, model C3.5/60/64.

A probe, 64 element linear array at a central ultrasonic frequency of approximately 8.0 MHz, model HL9.0/40/64.

A probe, 64 element convex array at a central ultrasonic frequency of approximately 6.5 MHz, model PV6.5/10/64.

#### General Safety and Effectiveness 6.

{3}------------------------------------------------

The LogicScan 128 / LogicScan 64 Ultrasound Systems are similar to currently distributed ultrasonic pulsed echo imaging systems.

There are no technological characteristics or features or indications for use in this Submission that are not previously evaluated and approved in the predicate devices, nor are there such technologies, features and indications for use not commonly used in the practice of diagnostic ultrasound.

The LogicScan Ultrasound Systems and its accessories are designed for compliance to all applicable medical devices safety standards, as referenced in DECLARATION OF CONFORMITY (Appendix 05). Prior release for manufacturing, all such devices, so designed, are tested and determined to be in full compliance with acoustic output, biocompatibility, cleaning and disinfection effectiveness. No additional clinical testing is required, as the indications for use are not a novel indication as shown by the predicate devices in Section 1.5 Predicate Device Comparison. The LogicScan has real time acoustic output display with two basic indices, a mechanical index and a thermal index, which are both automatically displayed.

#### 7. Patient Contact Materials

The materials of probes, coming in contact with patient are:

- Silicone Rubber .
- Acrylonitrile Butadien Styrene (ABS)
- Polyphenylsulfone (PPSU)

Standard for the biological evaluation:

ISO-10993-1, Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, 2003 ISO-10993-5, Biological Evaluation of Medical Devices Part 5: Tests for in vitro cytotoxicity , 1999 ISO-10993-10, Biological Evaluation of Medical Devices Part 10: Tests for irritation and delayed-type hypersensitivity, 2002

ISO-10993-11, Biological Evaluation of Medical Devices Part 11: Tests for systemic toxicity, 2006

#### 8. Software

The LogicScan 128 / LogicScan 64 systems contain the hardware and software which collect and pro-processes 'rough' data and send it via USB 2.0 connection to a Windows® based PC. The main application software is Echo Wave II software running on the PC, it is receiving data, processing and showing image/data on the screen. The main user interface shows an ultrasound image, controls and drop-out menus. The ultrasound images and calculated/measured data can be stored in memory.

#### 9. Conclusion

In accordance with the FDA and based on the information provided in this Premarket notification, TELEMED concludes that the LogicScan 128 / LogicScan 64 are safe and effective and substantially equivalent to predicate devices described herein.

{4}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized graphic of what appears to be an abstract caduceus or a symbol representing health and human services.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993

JAN 2 6 2012

Mr. Yury Sokolov Engineering Manager TELEMED Dariaus ir Gireno str. 42 LT-02189 VILNIUS LITHUANIA

Re: K113184

Trade/Device Name: LogicScan 64/LogiScan 128 Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYO, IYN, and ITX Dated: October 28, 2011 Received: October 28, 2011

Dear Mr. Sokolov:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

This determination of substantial equivalence applies to the following transducers intended for use with the DUS 6000 Digital Ultrasonic Imaging System, as described in your premarket notification:

## Transducer Model Number

Probes for LogicScan 64 Convex array C3.5/60/64 Linear array HL9.0/40/64 Convex array PV6.5/10/64

Probes for LogicScan 128 Convex array C3.5/60/128Z Linear array HL9.0/40/128Z Convex array PV6.5/10/128Z BiopSee BIPC6.5/10/128 BIPL7.5/70/128

{5}------------------------------------------------

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801): good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

If you have any questions regarding the content of this letter, please contact Lauren Hefner, Ph.D. at (301) 796-6881.

Sincerely Yours.

Mary Sbotif

Mary S. Pastel. Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health

{6}------------------------------------------------

# Indications for Use

510(k) Number (if known): K113184

Device Name: LogicScan 64 / LogicScan 128

Indications For Use:

The LogicScan 64 / LogicScan 128 systems and transducers are intended for diagnostic ultrasound imaging and fluid analysis of the human body.

The applications are:

cardiac (adult), fetal, abdominal, pediatric, small organ, transrectal, neonatal cephalic, peripheral vessel and musculo-skeletal (conventional and superficial).

It is possible to provide diagnostic information outside of an imaging lab, including at the bedside systems, for navigated medical application, in operating rooms/critical care units.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)

Mary S Pastel
(Division Sign-Off)

Office of In Vitro Diagnostic Device

Page 1 of

{7}------------------------------------------------

# 1.3 Indications for Use

All indications for use of subject devices and probes are identified in the table forms:

| System<br>LogicScan 64  |                                    | Fetal*, Abdominal*, Pediatric*, Small Organ*<br>(Breast, Thyroid, Testicles), Neonatal Cephalic*,<br>Musculo-skeletal* (Conventional), Musculo-<br>skeletal* (Superficial), Cardiac Adult*, Peripheral<br>vessel* | Table 1.3-1 |
|-------------------------|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
|                         | C3.5/60/64                         | Fetal*, Abdominal*, Cardiac Adult*                                                                                                                                                                                | Table 1.3-2 |
| Transducers             | HL9.0/40/64                        | Pediatric*,Musculo-skeletal* (Conventional),<br>Musculo-skeletal* (Superficial), Peripheral vessel*<br>Small Organ* (Breast, Thyroid, Testicles)*                                                                 | Table 1.3-3 |
|                         | PV6.5/10/64                        | Pediatric *, Neonatal Cephalic*, Peripheral vessel*                                                                                                                                                               | Table 1.3-4 |
|                         |                                    |                                                                                                                                                                                                                   |             |
| System<br>LogicScan 128 |                                    | Fetal*, Abdominal*, Pediatric*, Small Organ*<br>(Breast, Thyroid, Testicles), Neonatal Cephalic*,<br>Musculo-skeletal* (Conventional), Musculo-<br>skeletal* (Superficial), Cardiac Adult*, Peripheral<br>vessel* | Table 1.3-5 |
|                         | C3.5/60/128Z                       | Fetal*, Abdominal*, Cardiac Adult*                                                                                                                                                                                | Table 1.3-6 |
| Transducers             | HL9.0/40/128Z                      | Pediatric*,Musculo-skeletal* (Conventional),<br>Musculo-skeletal* (Superficial), Peripheral<br>vessel*, Small Organ* (Breast, Thyroid, Testicles)                                                                 | Table 1.3-7 |
|                         | PV6.5/10/128Z                      | Pediatric *, Neonatal Cephalic*, Peripheral vessel*                                                                                                                                                               | Table 1.3-8 |
|                         | BIPC6.5/10/128Z<br>BIPL7.5/70/128Z | Transrectal*, Intraoperative* (include Needle<br>Guidance)                                                                                                                                                        | Table 1.3-9 |

*Including Imaging for needle guidance

Mary Pastel

---

(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K K113/84

{8}------------------------------------------------

.

### Diagnostic Ultrasound Indications for Use Form

LogicScan 64 System:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Table 1.3-1 ·

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
|                           | Fetal                              | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Abdominal                          | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Intraoperative (specify)           |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Small Organ (specify)              | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Neonatal Cephalic                  | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
| Fetal Imaging<br>& Other  | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       |                   |   |     |     |                  |                       |                    |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Adult                      | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  | N                 | N | N   | N   | N                | Note 1                | Note 2,3           |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary Spostel

(Division Sign-Off Division of Radiological Devic Office of In Vitro Evaluation and Safety

510K. K113.184

{9}------------------------------------------------

Diagnostic Ultrasound Indications for Use Form

.

LogicScan 64 System:

Convex array C3.5/60/64

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Transducer:

| Table 1.3-6               |                                    |                   |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
|                           | Fetal                              | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Abdominal                          | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Intraoperative (specify)           |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          |                   |   |     |     |                  |                       |                    |
|                           | Small Organ (specify)              |                   |   |     |     |                  |                       |                    |
|                           | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                    |
| Fetal Imaging<br>& Other  | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       |                   |   |     |     |                  |                       |                    |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal.<br>(Superficial) |                   |   |     |     |                  |                       |                    |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Adult                      | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary S Patil
(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K.

**Table 1.3-2**

{10}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Form

LogicScan 64 System:

Linear array HL9.0/40/64 Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Table 1.3-3

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
|                           | Fetal                              |                   |   |     |     |                  |                       |                    |
|                           | Abdominal                          |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (specify)           |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Small Organ (specify)              | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                    |
| Fetal Imaging<br>& Other  | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       |                   |   |     |     |                  |                       |                    |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Adult                      |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary S Patel
(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evanuance and Satery

{11}------------------------------------------------

# Diagnostic Ultrasound Indications for Use Form

LogicScan 64 System:

Convex array PV6.5/10/64 Transducer:

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Table 1.3-4               |                                    |                   |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
|                           | Fetal                              |                   |   |     |     |                  |                       |                    |
|                           | Abdominal                          |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (specify)           |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Small Organ (specify)              |                   |   |     |     |                  |                       |                    |
| Fetal Imaging<br>& Other  | Neonatal Cephalic                  | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       |                   |   |     |     |                  |                       |                    |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                    |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Adult                      |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  | N                 | N | N   |     | N                | Note 1                | Note 2,3           |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |

. N= new indication; P= previously cleared by FDA; E= added under Appendix E

- Note 1 Combined Modes: Color+B, Power+B,
- Note 2 Includes Imaging for Needle Guidance
- Note 3 Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

- Note 5 Intraoperative applications: Include Needle Guidance
Mary S/astel
(Division Sign-Off)

510

Division Sign-Off)
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation anc Sarety

{12}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

Table 1.3-5

System: LogicScan 128

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
|                           | Fetal                              | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Abdominal                          | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Intraoperative (specify)           | N                 |   |     |     | N                | Note 1                | Note 2,5           |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Small Organ (specify)              | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Neonatal Cephalic                  | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
| Fetal Imaging<br>& Other  | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       | N                 |   |     |     | N                | Note 1                | Note 2,5           |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethrai                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Musculo-skeletal<br>(Superficial)  | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Adult                      | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
|                           | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Peripheral .<br>Vessel    | Peripheral vessel                  | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary S. Patel
(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

510K. K113184

{13}------------------------------------------------

## Diagnostic Ultrasound Indications for Use Form

System: LogicScan 128 Transducer:

Convex array C3.5/60/128Z

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

| Table 1.3-6               |                                    |   |   |     |     |                   |                       |                    |
|---------------------------|------------------------------------|---|---|-----|-----|-------------------|-----------------------|--------------------|
| Clinical Application      |                                    |   |   |     |     | Mode of Operation |                       |                    |
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B | M | PWD | CWD | Color<br>Doppler  | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |   |   |     |     |                   |                       |                    |
|                           | Fetal                              | N | N | N   |     | N                 | Note 1                | Note 2,3,4         |
|                           | Abdominal                          | N | N | N   |     | N                 | Note 1                | Note 2,3,4         |
|                           | Intraoperative (specify)           |   |   |     |     |                   |                       |                    |
|                           | Intraoperative (Neuro)             |   |   |     |     |                   |                       |                    |
|                           | Laparoscopic                       |   |   |     |     |                   |                       |                    |
|                           | Pediatric                          |   |   |     |     |                   |                       |                    |
|                           | Small Organ (specify)№             |   |   |     |     |                   |                       |                    |
|                           | Neonatal Cephalic                  |   |   |     |     |                   |                       |                    |
| Fetal Imaging<br>& Other  | Adult Cephalic                     |   |   |     |     |                   |                       |                    |
|                           | Trans-rectal                       |   |   |     |     |                   |                       |                    |
|                           | Trans-vaginal                      |   |   |     |     |                   |                       |                    |
|                           | Trans-urethral                     |   |   |     |     |                   |                       |                    |
|                           | Trans-esoph. (Non-Card)            |   |   |     |     |                   |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) |   |   |     |     |                   |                       |                    |
|                           | Musculo-skeletal<br>(Superficial)  |   |   |     |     |                   |                       |                    |
|                           | Intravascular                      |   |   |     |     |                   |                       |                    |
|                           | Other (specify)                    |   |   |     |     |                   |                       |                    |
| Cardiac                   | Cardiac Adult                      | N | N | N   |     | N                 | Note 1                | Note 2,3,4         |
|                           | Cardiac Pediatric                  |   |   |     |     |                   |                       |                    |
|                           | Intravascular (Cardiac)            |   |   |     |     |                   |                       |                    |
|                           | Trans-esoph. (Cardiac)             |   |   |     |     |                   |                       |                    |
|                           | Intra-cardiac                      |   |   |     |     |                   |                       |                    |
|                           | Other (specify)                    |   |   |     |     |                   |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |   |   |     |     |                   |                       |                    |
|                           | Other (specify)                    |   |   |     |     |                   |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Man S Patel
(Division Sign-Off)

DIVISION
Division of Radiological Devices
Office of In Vitro Diagnostic Device Evaluation and Safety

510K: K113184

{14}------------------------------------------------

510(k) Premarket Submission

Pg. 1.3-8

## Diagnostic Ultrasound Indications for Use Form

LogicScan 128 System:

Transducer: Linear array HL9.0/40/128Z

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Tahlo

| ויטיו ביתריו במוספו<br>Clinical Application |                                    | Mode of Operation |   |     |     |         |           |            |  |  |
|---------------------------------------------|------------------------------------|-------------------|---|-----|-----|---------|-----------|------------|--|--|
| General                                     | Specific                           |                   |   |     |     | Color   | Combined  | Other      |  |  |
| (Track 1 Only)                              | (Track 1 & 3)                      | B                 | M | PWD | CWD | Doppler | (specify) | (specify)  |  |  |
| Ophthalmic                                  | Ophthalmic                         |                   |   |     |     |         |           |            |  |  |
|                                             | Fetal                              |                   |   |     |     |         |           |            |  |  |
|                                             | Abdominal                          |                   |   |     |     |         |           |            |  |  |
|                                             | Intraoperative (specify)           |                   |   |     |     |         |           |            |  |  |
|                                             | Intraoperative (Neuro)             |                   |   |     |     |         |           |            |  |  |
|                                             | Laparoscopic                       |                   |   |     |     |         |           |            |  |  |
|                                             | Pediatric                          | N                 | N | N   |     | N.      | Note 1    | Note 2,3,4 |  |  |
|                                             | Small Organ (specify)              | N                 | N | N   |     | N       | Note 1    | Note 2,3,4 |  |  |
|                                             | Neonatal Cephalic                  |                   |   |     |     |         |           |            |  |  |
| Fetal Imaging                               | Adult Cephalic                     |                   |   |     |     |         |           |            |  |  |
| & Other                                     | Trans-rectal                       |                   |   |     |     |         |           |            |  |  |
|                                             | Trans-vaginal                      |                   |   |     |     |         |           |            |  |  |
|                                             | Trans-urethral                     |                   |   |     |     |         |           |            |  |  |
|                                             | Trans-esoph. (Non-Card)            |                   |   |     |     |         |           |            |  |  |
|                                             | Musculo-skeletal<br>(Conventional) | N                 | N | N   |     | N       | Note 1    | Note 2,3,4 |  |  |
|                                             | Musculo-skeletai<br>(Superficial)  | N                 | N | N   |     | N       | Note 1    | Note 2,3,4 |  |  |
|                                             | Intravascular                      |                   |   |     |     |         |           |            |  |  |
|                                             | Other (specify)                    |                   |   |     |     |         |           |            |  |  |
|                                             | Cardiac Adult                      |                   |   |     |     |         |           |            |  |  |
|                                             | Cardiac Pediatric                  |                   |   |     |     |         |           |            |  |  |
| Cardiac                                     | Intravascular (Cardiac)            |                   |   |     |     |         |           |            |  |  |
|                                             | Trans-esoph. (Cardiac)             |                   |   |     |     |         |           |            |  |  |
|                                             | Intra-cardiac                      |                   |   |     |     |         |           |            |  |  |
|                                             | Other (specify)                    |                   |   |     |     |         |           |            |  |  |
| Peripheral                                  | Peripheral vessel                  | N                 | N | N   |     | N       | Note 1    | Note 2,3,4 |  |  |
| Vessel                                      | Other (specify)                    |                   |   |     |     |         |           |            |  |  |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary S Patil
(Division Sign-Off)

Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety

૧૮૮૮૪

{15}------------------------------------------------

### Diagnostic Ultrasound Indications for Use Form

System: LogicScan 128 Transducer:

Convex array PV6.5/10/128Z

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Table 1.3-8

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
|                           | Fetal                              |                   |   |     |     |                  |                       |                    |
|                           | Abdominal                          |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (specify)           |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
|                           | Small Organ (specify)              |                   |   |     |     |                  |                       |                    |
|                           | Neonatal Cephalic                  | N                 | N | N   |     | N                | Note 1                | Note 2,3,4         |
| Fetal Imaging<br>& Other  | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       |                   |   |     |     |                  |                       |                    |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                    |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Adult                      |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
|                           | Peripheral<br>Vessel               | Peripheral vessel | N | N   | N   |                  | N                     | Note 1             |
| Other (specify)           |                                    |                   |   |     |     |                  |                       |                    |

· N= new indication; P= previously cleared by FDA; E= added under Appendix E

Note 1 - Combined Modes: Color+B, Power+B,

Note 2 - Includes Imaging for Needle Guidance

Note 3 - Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary S. Patel
Division Sign-Off)

Division of Radiological Devices vice Evaluation and Safety Office of In Vitro Diagn

Office of In Vitro Diagnostic Device Evaluation and Se
810K. K113184

.

{16}------------------------------------------------

510(k) Premarket Submission

LogicScan Diagnostic Ultrasound System

Pg. 1.3-10

### Diagnostic Ultrasound Indications for Use Form

System:

Logic Scan 128

Transducer:

BiopSee BIPC6.5/10/128 BIPL7.5/70/128

Intended Use: Diagnostic ultrasound imaging or fluid flow analysis of the human body as follows:

Table 1.3-9 ·

| Clinical Application      |                                    | Mode of Operation |   |     |     |                  |                       |                    |
|---------------------------|------------------------------------|-------------------|---|-----|-----|------------------|-----------------------|--------------------|
| General<br>(Track 1 Only) | Specific<br>(Track 1 & 3)          | B                 | M | PWD | CWD | Color<br>Doppler | Combined<br>(specify) | Other<br>(specify) |
| Ophthalmic                | Ophthalmic                         |                   |   |     |     |                  |                       |                    |
| Fetal Imaging<br>& Other  | Fetal                              |                   |   |     |     |                  |                       |                    |
|                           | Abdominal                          |                   |   |     |     |                  |                       |                    |
|                           | Intraoperative (specify)           | N                 |   |     |     | N                | Note 1                | Note 2,5           |
|                           | Intraoperative (Neuro)             |                   |   |     |     |                  |                       |                    |
|                           | Laparoscopic                       |                   |   |     |     |                  |                       |                    |
|                           | Pediatric                          |                   |   |     |     |                  |                       |                    |
|                           | Small Organ (specify)              |                   |   |     |     |                  |                       |                    |
|                           | Neonatal Cephalic                  |                   |   |     |     |                  |                       |                    |
|                           | Adult Cephalic                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-rectal                       | N                 |   |     |     | N                | Note 1                | Note 2,5           |
|                           | Trans-vaginal                      |                   |   |     |     |                  |                       |                    |
|                           | Trans-urethral                     |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Non-Card)            |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Conventional) |                   |   |     |     |                  |                       |                    |
|                           | Musculo-skeletal<br>(Superficial)  |                   |   |     |     |                  |                       |                    |
|                           | Intravascular                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Cardiac                   | Cardiac Adult                      |                   |   |     |     |                  |                       |                    |
|                           | Cardiac Pediatric                  |                   |   |     |     |                  |                       |                    |
|                           | Intravascular (Cardiac)            |                   |   |     |     |                  |                       |                    |
|                           | Trans-esoph. (Cardiac)             |                   |   |     |     |                  |                       |                    |
|                           | Intra-cardiac                      |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |
| Peripheral<br>Vessel      | Peripheral vessel                  |                   |   |     |     |                  |                       |                    |
|                           | Other (specify)                    |                   |   |     |     |                  |                       |                    |

N= new indication; P= previously cleared by FDA; E= added under Appendix E

- Note 1 Combined Modes: Color+B, Power+B,
- Note 2 Includes Imaging for Needle Guidance
- Note 3 Small Organs (specifically Breast, Thyroid, Testicles)

Note 4 - Tissue harmonic Imaging (THI)

Note 5 - Intraoperative applications: Include Needle Guidance

Mary S. Pastel

Division valuation and Safety Office of In Vitro Diagnostic Device

SIOK

---

**Source:** [https://fda.innolitics.com/device/K113184](https://fda.innolitics.com/device/K113184)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
