ACUSON SC2000 DIAGNOSTIC ULTRASOUND SYSTEM

K113179 · Siemens Medical Solutions USA, Inc., Ultrasound DI · IYN · Dec 7, 2011 · Radiology

Device Facts

Record IDK113179
Device NameACUSON SC2000 DIAGNOSTIC ULTRASOUND SYSTEM
ApplicantSiemens Medical Solutions USA, Inc., Ultrasound DI
Product CodeIYN · Radiology
Decision DateDec 7, 2011
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1550
Device ClassClass 2
AttributesPediatric, 3rd-Party Reviewed

Indications for Use

The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Intraoperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes.

Device Story

Mobile, software-controlled diagnostic ultrasound system; acquires harmonic echo data via transducers/catheters. Operates in B-Mode, M-Mode, PW/CW Doppler, Color Doppler, Color Power Doppler, 3D, and 4D imaging. Inputs include ultrasound signals and patient ECG for cardiac synchronization. System processes data to display anatomical structures and blood flow velocity on a flat panel display. Used by clinicians in clinical settings for diagnostic assessment. Output provides visual images and quantitative measurements; used adjunctively with other medical data to evaluate pathology and guide procedures. Benefits include non-invasive visualization of anatomy and physiology, supporting clinical diagnosis and intraoperative guidance.

Clinical Evidence

Bench testing only. The device underwent extensive safety and performance verification to ensure compliance with industry standards, including electrical safety (UL 60601-1, CSA C22.2 No. 601-1), diagnostic ultrasound safety (IEC 60601-2-37), acoustic output measurement (AIUM/NEMA UD-2, UD-3), and biocompatibility (ISO 10993-1). No clinical trial data was required for this modification.

Technological Characteristics

Mobile ultrasound system with adjustable flat panel display. Supports various transducers and catheters. Connectivity includes on-screen display of thermal/mechanical indices. Complies with IEC 60601-1, IEC 60601-2-37, and ISO 10993-1. Software-controlled signal processing for B-Mode, Doppler, and 3D/4D imaging. Biocompatible materials per ISO 10993-1.

Indications for Use

Indicated for cardiac (adult, pediatric, neonatal, fetal), vascular (carotid, jugular, peripheral), superficial/musculoskeletal, intraoperative neurological, and transcranial imaging. Supports intra-cardiac/intraluminal visualization via compatible catheters. Used for anatomical measurement and blood flow velocity assessment to evaluate pathology.

Regulatory Classification

Identification

An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ### 510(k) Summary Prepared October18, 2011 | Sponsor: | Siemens Medical Solutions, Inc.,<br>Ultrasound Division<br>1230 Shorebird Way<br>Mountain View, California 94043 | | |------------------|------------------------------------------------------------------------------------------------------------------|----------------------------------| | Contact Person: | Robert F. Lawrence<br>Telephone:<br>Fax: | (650) 694-5984<br>(650) 694-5580 | | Submission Date: | October 18, 2011 | | | Device Name: | Acuson SC2000™ Diagnostic Ultrasound System | | | Common Name: | Diagnostic Ultrasound System | | ## Classification: Regulatory Class: ll Tier II Review Category: Classification Panel: Radiology | Ultrasonic Pulsed Doppler Imaging System | FR # 892.1550 | Product Code 90-IY | |------------------------------------------|---------------|--------------------| | Ultrasonic Pulsed Echo Imaging System | FR # 892.1560 | Product Code 90-IY | | Diagnostic Ultrasound Transducer | FR # 892.1570 | Product Code 90-IT | # A. Legally Marketed Predicate Devices / . Bogany The Acuson SC2000™ Ultrasound System in this 510k is a modification to SC2000™ The Acuson SC2000 - Oltrasound System in this of K072365 and K102017. Diagnostic Ultrasound System previously cleared in K072365 and K102017. ## B. Device Description: The SC2000™ Diagnostic Ultrasound System is a multi-purpose mobile, software controlled The SC2000™ Diagnostic Unrasund System be porporal and mechanical indices diagnostic ultrasound system with an on-screen display for themical indices diagnostic ultrasound system with an on-scieer alspire primary of secondary related to potential bio-effect mechanisms. Its function is to acquire primary of Secondary related to potential bio-enect mechanisms. It in B-Mode, Pulsed (PW) Dopler Mode, harmonic ultrasound echo data and display it in B-Mode, Pulsed (PW) Dopler Mode, a combinati harmonic ultrasound echo data and ulished on inistration of online of Oppler Mode, a combination Continuous (CW) Doppler Mode, Color Dopen Mode, Ampling on a France, Nissalay Continuous (CVV) Dopper Mode, Oolor Doppler Model, Mipmont on a Flat Panel Display. of modes, 3D Imaging, or Harmonic Imaging and 4D imaging on a Flat Panel Display. of mouse, of the SC2000" The SC2000™Ultrasound System has been optimized for user ergonomics with adjustable The SC2000™ Diffasound System nas been oplimized for assessmently adjustable Flat Panel Display. There is an keyboard height and rotation and independently adjustable Flat P available off-line workstation (SC2000WP) {1}------------------------------------------------ #### C. Intended Use The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Intraoperative Petal Gallanc, Pedianie, Transespondged, Aland Musculo-skeletal Superficial applications. The system recorogiou, massure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: ### Cardiac Imaging Applications and Analysis The system transmits ultrasound energy into adult, pediatric, neonatal, and fetal cardiac patients creating The System tantonial List acor Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac and continuous were , and surrounding anatomical structures to evaluate the presence of vathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of r re system aloo bupports can belysiology as well as visualization of other devices in the heart of adult and pediatric patients. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. ### Vascular Imaging Applications and Analysis The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), The system cranillis alsocure energy ... (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Ooks Doppor (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the Doppier (Orro) to other magos and arteries in the arms and legs; and surrounding analomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be The System not Vabodian medical data obtained by a physician for clinical diagnosis purposes. #### Superficial Imaging Applications The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), The System Canadinia and abouter (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Oopler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal Dopping (OTTB) to outdir incatomical structures to evaluate the presence or absence of pathology. The system may be used to aquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. #### Intraoperative imaging Applications The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during neurological intraoperative procedures. ### Transcranial Imaging Applications The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler The System transmile andescribed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) (OD), Outlin images and blood flow velocity of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. {2}------------------------------------------------ ### D. Substantial Equivalence The submission device is a modification is a modification to SC2000™ Diagnostic Ultrasound System previously cleared in K072365 and K102017 with regard to both intended use and technological characteristics. ### E. Performance Data The SC2000™is designed, verified, and validated according to the company's design control process and has been subjected to extensive safety and performance testing before release. Final testing of the SC2000 included various safety and performance testing designed to ensure the device meets all of its specifications. Safety tests have been performed to ensure the device complies with applicable industry and safety standards including: The Acuson SC2000™ has been designed to meet the following product safety standards: - UL 60601-1. Safety Requirements for Medical Equipment l - . IEC 60601-2-37 Diagnostic Ultrasound Safety Standards - CSA C22.2 No. 601-1, Safety Requirements for Medical Equipment . - AIUM/NEMA UD-3, Standard for Real Time Display of Thermal and Mechanical ■ Acoustic Output Indices on Diagnostic Ultrasound Equipment - AIUMINEMA UD-2. Acoustic Output Measurement Standard for Diagnostic Ultrasound . - Safety and EMC Requirements for Medical Equipment . - = IEC 60601-1 - · IEC 60601-1-1 - IEC 60601-1-2 . - l IEC 1157 Declaration of Acoustic Power - l ISO 10993-1 Biocompatibility {3}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular, with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the circle is an abstract symbol consisting of four curved lines that resemble human profiles facing to the right. Food and Orug Administration 10903 New Hampshire Avenue Silver Spring, MD 20993 Siemens Medical Solutions USA. Inc. - Ultrasound Group % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313 Re: K113179 Trade/Device Name: SC2000TM Diagnostic Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: Class II Product Code: IYN, IYO, and ITX Dated: November 29, 2011 Received: November 30, 2011 Dear Mr. Job: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and we have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. This determination of substantial equivalence applies to the following transducers intended for use with the SC2000™ Diagnostic Ultrasound System, as described in your premarket notification: Transducer Model Number # 9L4 VSM TEE 4V1c 8V3c AUX CW2 4Zlc (Apollo) AcuNav 8F and 10F Ultrasound Catheter ACUSON AcuNav™ V 10F Ultrasound Catheter SoundStar 10F Ultrasound Catheter DEC - 7 2011 {4}------------------------------------------------ If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus permits your device to proceed to market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.html for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. If you have any questions regarding the content of this letter, please contact Jeffrey Ballyns at (301) 796-6105. Sincerely Yours. Mukal D. O'Hare fin Mary S. Pastel, Sc.D. Director Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health {5}------------------------------------------------ #### 1.3 Indications for Use 510(k) Number (if known): #### Device Name: SC2000 ™ Diagnostic Ultrasound System #### Indications for Use: The SC2000 ultrasound imaging system is intended for the following applications: Cardiac, Neo-natal and Fetal Cardiac, Pediatric, Transespohageal, Adult Cephalic, Peripheral Vessel, Intracperative Neurological, Musculo-skeletal Conventional, and Musculo-skeletal Superficial applications. The system also provides the ability to measure anatomical structures and calculation packages that provide information to the clinician that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. The typical examinations performed using the SC2000 Ultrasound System are: #### Cardiac Imaging Applications and Analysis The system transmits ultrasound eneroy into adult, pediatic, neonatal, and fetal cardiac creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Dopter (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound Images. The system also supports catheters which are intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients. The system transmits ultrasound energy from either a transthoracic or transesophageal approach in adult and pediatric patients; and from a transthoracic approach in neonatal and fetal cardiac patients creating 2D (B), 3D, M-Mode (M), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave (PW) Doppler, and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the heart, cardiac valves, great vessels, and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. The system has Cardiac Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. #### Vascular Imaging Applications and Analysis The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of the carotid arteries or juggler veins in the neck; superficial and deep veins and arteries in the arms and legs; and surrounding anatomical structures to evaluate the presence or absence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic capture of ultrasound images. The system has Vascular Measurements and Calculation Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. #### Superficial Imaging Applications The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), Pulsed Wave Doppler (PWD), and Continuous Wave Doppler (CWD) to obtain images and blood flow velocity of conventional or superficial musculoskeletal structures and surrounding anatomical structures to evaluate the presence of pathology. The system may be used to acquire patient electrocardiogram for synchronizing the diastolic and systolic capture of ultrasound images. Page 13 of 1653 {6}------------------------------------------------ #### Intraoperative Imaging Applications The system transmits ultrasound energy into various parts of the body of adult patients creating 2D (B), Color Doppler (CD), Color Power Doppler (CPD), and Pulsed Wave Doppler (PWD) to obtain images and blood flow velocity that provide guidance during neurological intraoperative procedures. #### Transcranial Imaging Applications The system transmits ultrasound energy into the cranium of adult patients creating 2D (B), Color Doppler r no System framerike okled Wave Doppier (PWD), and Continuous Wave Doppler (CWD) (OD), Odlor Fower Dopper (Or D), of the brain and surrounding anatomical structures to evaluate the presence or absence of pathology. The system provides Measurement Packages that provide information that may be used adjunctively with other medical data obtained by a physician for clinical diagnosis purposes. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONȚINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Muhal D.O.R. (Division Sign-Off) 510K k: 113199 Page 14 of 39 Page 14 of 1653 <sup>(</sup>Division Sign-Of Division of Radiological Devices Office of In Vitro Diagnostic Device Evaluation and Safety {7}------------------------------------------------ ### 1.3 Indications for Use Forms ### Diagnostic Ultrasound Indications for Use Form 510(k) Number (if known): #### Device Name: SC2000 Diagnostic Ultrasound System Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color ·<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | Other:<br>Real Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|--------------------------------|-----------------------|-------------------------------|--------------|---------------------------| | Ophthalmic | | | | | | | | | | | | | | Fetal | | P | P | P | P | P | P | | P* | P | | P | | Abdominal | | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | | Intraoperative<br>Neurological | | P | P | P | | P | P | P | P* | P | | | | Pediatric | | P | P | P | P | P | P | | P* | P | P | P | | Small Organ<br>(specify) | | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | | Adult Cephalic | | P | P | P | P | P | P | | P* | P | | | | Cardiac | | P | P | P | P | P | P | | P* | P | P | P | | Trans-esophageal | | P | P | P | P | P | | | P* | | P | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | | Intra-Luminal | | N | N | N | N | N | N | | N* | | | N | | Peripheral Vessel | | P | P | P | P | P | P | P | P* | P | | | | Laparoscopic | | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | P | P | P | | P | P | P | P* | P | | | | Musculo-skeletal<br>Superficial | | P | P | P | | P | P | P | P* | P | | | | Other (Neonatal<br>Cardiac) | | P | P | P | P | P | P | | P* | P | | | | Other (Intra-<br>Cardiac) | | N | N | N | N | N | N | | N* | | | N | N=new indication. P = Previously Cleared in 510(k) K072365, K102017 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF N Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) Page 15 of 39 {8}------------------------------------------------ 510(k) Number (if known): Device Name: 914 Indications for Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | P | P | P | | P | P | P | P* | P | | Pediatric | | | | | | | | | | | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic<br>Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | P | P | P | | P | P | P | P* | P | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | P | P | P | | P | P | P | P* | P | | Musculo-skeletal<br>Superficial | | P | P | P | | P | P | P | P* | P | | Other (specify) | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K072365, K102017 Additional Comments: *Combinations include: B+M, B+PWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE 1000 (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) Michael Dóhm SC2000 510(k) Submission K113179 Page 16 of 39 {9}------------------------------------------------ 510(k) Number (if known): Device Name: ### V5M TEE Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|--------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | | | P* | | P | | Small Organ<br>(specify) | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | | | P* | | P | | Trans-esophageal | | P | P | P | P | P | | | P* | | P | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (specify) | | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K072365, K102017 #### Additional Comments: * Combinations include: B+M, B+PWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color D Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, __ B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) Murhul DÖKhm SC2000 510(k) Submission K113179 Page 17 of 39 . {10}------------------------------------------------ 510(k) Number (if known): Device Name: 4V I c Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | | P | P | P | P | P | P | | P * | P | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | P | P | P | P | P | P | | P * | P | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | P | P | P | P | P | P | | P * | P | | Cardiac | | P | P | P | P | P | P | | P * | P | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (Neonatal<br>Cardiac) | | P | P | P | P | P | P | | P * | P | N=new indication. Previously Cleared in 510(k) K072365, K102017 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color, Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE ............................................................................................................................................ (PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (O1VD) Prescription Use (Per 21 CFR 801.109) Michael D. Ahn SC2000 510(k) Submission Image /page/10/Picture/14 description: The image shows a sequence of handwritten digits and letters. The sequence starts with the letter 'K', followed by the number '113179'. The handwriting style appears casual and slightly uneven, with varying stroke thicknesses. Page 18 of 39 {11}------------------------------------------------ 510(k) Number (if known): Device Name: 8V3c Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | | |----------------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|---| | Ophthalmic | | | | | | | | | | | | | Fetal | | P | P | P | P | P | P | | P * | P | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | P | | P * | P | | | Small Organ<br>(specify) | | | | | | | | | | | | | Neonatal<br>Cephalic<br>Adult Cephalic | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | P | | P * | P | | | Trans-esophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (Neonatal<br>Cardiac) | | P | P | P | P | P | P | P | | P * | P | N=new indication. Previously Cleared in 510(k) K102017 #### Additional Comments: * Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) escription SC2000 510(k) Submission K113179 Page 19 of 39 {12}------------------------------------------------ 510(k) Number (if known): Device Name: AUX CW2 Intended Use: Ultrasound imaging or fluid flow analysis of the human body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------| | Ophthalmic | | | | | | | | | | | | Fetal | | | | | | | | | | | | Abdominal | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | Pediatric | | | | P | | | | | | | | Small Organ<br>(specify) | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | Cardiac | | | | | P | | | | | | | Trans-esophageal | | | | | | | | | | | | Transrectal | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | Transurethral | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | Peripheral Vessel | | | | | P | | | | | | | Laparoscopic | | | | - | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | Other (specify) | | | | | | | | | | | N=new indication. Previously Cleared in 510(k) K072365, K102017 Additional Comments: (PLESE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) hul D Othn SC2000 510(k) Submission K113179 Page 20 of 39 {13}------------------------------------------------ 510(k) Number (if known): Device Name: 4Z1c (Apollo) Intended Use: Ultrasound imaging or fluid flow analysis of the buman body as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | P | P | P | P | P | | | P* | P | P | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | | | P* | P | P | | Small Organ<br>(specify) ** | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | | | P* | P | P | | Trans-esophageal | | | | | - | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | | | | | | | | | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (specify) | | | | | | | | | | | | N=new indication. P = Previously Cleared in 510(k) K072365, K102017 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) ul D SAmm KII317C Page 21 of 39 {14}------------------------------------------------ 510(k) Number (if known): Device Name: ### AcuNav 8F and 10F Uitrasound Catheter Intended Use: Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------|--| | Ophthalmic | | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | | Pediatric | | P | P | P | P | P | P | | P* | | | | | Small Organ<br>(specify) ** | | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | P | | P* | | | | | Trans-esophageal | | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | | Intra-Luminal | | P | P | P | P | P | P | | P* | | | | | Peripheral Vessel | | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | | Other (Intra-<br>Cardiac) | | P | P | P | P | P | P | | P* | | | | N=new indication. P = Previously Cleared in 510(k) K071234, K093812 #### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) Presc K113179 {15}------------------------------------------------ 510(k) Number (if known): Device Name: ### ACUSON AcuNav™ V 10F Ultrasound Catheter Intended Use: Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart of adult and pediatric patients as follows: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | N | N | N | N | N | N | N | | N* | | N | | Small Organ<br>(specify) ** | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | P | P | P | P | P | P | P | | P* | | N | | Trans-esophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | P | P | P | P | P | P | P | | P * | | N | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (Intra-<br>Cardiac) | P | P | P | P | P | P | P | | P* | | N | N=new indication. P = Previously Cleared in S10(k) K081808 #### Additional Comments: * Combinations include: B+M, B+PWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler, B+M+Power Doppler, B+PWD+Power Doppler, B+CWD+Power Doppler, B+Clarify VE (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) Prescrip muchal ODA SC2000 510(k) Submission Page 23 of 39 {16}------------------------------------------------ 510(k) Number (if known): Device Name: ### SoundStar 10F Ultrasound Catheter Intended Use: Catheter is intended for intra-cardiac and intraluminal visualization of cardiac and great vessel anatomy and physiology as well as visualization of other devices in the heart: | Clinical<br>Application | A | B | M | PWD | CWD | Color<br>Doppler | Power<br>(Amplitude)<br>Doppler | Color<br>Velocity<br>Imaging | Combined<br>(Specify) | Other:<br>Harmonic<br>Imaging | Other:<br>Real<br>Time<br>3D | |----------------------------------|---|---|---|-----|-----|------------------|---------------------------------|------------------------------|-----------------------|-------------------------------|------------------------------| | Ophthalmic | | | | | | | | | | | | | Fetal | | | | | | | | | | | | | Abdominal | | | | | | | | | | | | | Intraoperative<br>Abdominal | | | | | | | | | | | | | Intraoperative<br>Neurological | | | | | | | | | | | | | Pediatric | | | | | | | | | | | | | Small Organ<br>(specify) ** | | | | | | | | | | | | | Neonatal<br>Cephalic | | | | | | | | | | | | | Adult Cephalic | | | | | | | | | | | | | Cardiac | | P | P | P | P | P | P | | P* | | | | Trans-esophageal | | | | | | | | | | | | | Transrectal | | | | | | | | | | | | | Transvaginal | | | | | | | | | | | | | Transurethral | | | | | | | | | | | | | Intra-Luminal | | P | P | P | P | P | P | | P* | | | | Peripheral Vessel | | | | | | | | | | | | | Laparoscopic | | | | | | | | | | | | | Musculo-skeletal<br>Conventional | | | | | | | | | | | | | Musculo-skeletal<br>Superficial | | | | | | | | | | | | | Other (Intra- | | P | P | P | P | P | P | | P* | | | N=new indication. P = Previously Cleared in 510(k) K070242 ### Additional Comments: *Combinations include: B+M, B+PWD, B+CWD, B+Color Doppler, B+M+ Color Doppler, B+PWD+Color Doppler, B+CWD+Color Doppler, B+Power Doppler (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In-Vitro Diagnostics (OIVD) Prescription Use (Per 21 CFR 801.109) Prescript K113179
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...