CLEARPATH UPPER GI

K113166 · Easyglide , Ltd. · FDS · May 16, 2012 · Gastroenterology, Urology

Device Facts

Record IDK113166
Device NameCLEARPATH UPPER GI
ApplicantEasyglide , Ltd.
Product CodeFDS · Gastroenterology, Urology
Decision DateMay 16, 2012
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

The ClearPath Upper Gl is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures. It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

Device Story

ClearPath Upper GI is a single-patient-use, non-sterile disposable accessory for standard endoscopes; facilitates irrigation and evacuation of the upper gastrointestinal tract. System comprises a suction tube/tip, irrigation tube/tip, and a sleeve for endoscope attachment. Operates in conjunction with the ClearPath Controller and ClearPath Tubing (K091305). Distal tip design features multi-irrigation holes and a distal suction hole. Device attaches along the endoscope length, leaving the working channel free for other tools. Used by trained medical personnel in hospitals, clinics, and doctors' offices. Enables simultaneous irrigation and suction during endoscopic procedures without removing tools from the working channel. Improves visibility and clears debris (blood, bile, irrigation fluid) to assist clinical decision-making during endoscopy.

Clinical Evidence

Bench and animal testing performed. Bench testing included dimensional, strength, functional, and mechanical testing, and compliance with endoscopes. Animal testing evaluated tip dislodgement, maneuverability, visibility, irrigation quality, suction continuity, and evacuation of bodily fluids. Results demonstrated the modified device performed to specifications and was equivalent to the predicate.

Technological Characteristics

Single-use, non-sterile disposable accessory. Attaches to standard endoscopes via a sleeve. Distal tip with multi-irrigation holes and suction hole. Materials comply with ISO 10993-1. Operates with external ClearPath Controller and Tubing (K091305). Mechanical principle of operation; no electronic or software components.

Indications for Use

Indicated for individuals undergoing endoscopic procedures requiring irrigation or cleaning of the upper digestive tract and evacuation of irrigation fluid, blood, and bile. For use by trained medical personnel in hospitals, clinics, and doctors' offices.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Special Premarket Notification 510(k) Section 5 - Device Description ClearPath Upper Gl | 510(k) Summary | |----------------| | Page 1 of 3 | | 25-Oct-11 | | EasyGlide Ltd.<br>30 Ha'Ella St.<br>Kfar Truman, 73150<br>Israel. | Tel - 011-972-528565644<br>Fax - 011-972-776201003 | |-------------------------------------------------------------------|----------------------------------------------------------------------------| | Official Contact: | Izhak Fabian - CEO | | Proprietary or Trade Name: | ClearPath Upper Gl | | Common/Usual Name: | Irrigation/evacuation system | | Classification Name/Code: | FDS - Endoscope and accessories, flexible/rigid<br>CFR 876.1500<br>Class 2 | | Device: | ClearPath Upper Gl | | Predicate Devices: | K093779 - EasyGlide - ClearPath Upper GI | ### Device Description: The Clearpath Upper GI was cleared under K093779, we have modified the design. . The ClearPath Upper GI includes one suction tube + tip, one irrigation tube + tip, and a sleeve to attach to the endoscope. The ClearPath Upper GI is connected to the ClearPath Controller and the ClearPath Tubing (K091305) and allows for irrigation and evacuation of debris from the upper gastrointenstinal tract during an endoscopic procedure. The device can accommodate several sizes and configurations of endoscopes with a change in the attachment ring. #### Indications for Use: The ClearPath Upper Gl is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures. It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices. ### Patient population: Individuals undergoing endoscopic procedures. {1}------------------------------------------------ ### 510(k) Summarv Page 2 of 3 . 25-Oct-11 ### Environment of Use: Hospitals, clinics, and doctors' offices. ### Performance testing: We have performed bench and animal testing to verify that the ClearPath Upper G1 performs as expected in conjunction with standard endoscopes and with the cleared ClearPath controller and tubing K0910305. The tests include: - o Dimensional testing - 0 Strength testing - Functional testing o - Mechanical testing 0 - Compliance with endoscopes o The animal testing evaluated: - Any sign of dislodgement of the tip from the endoscope during the o procedure - Ease of maneuverability and advance in the GI tract 0 - Quality of visibility o - Quality of irrigation 0 - Any sign of occlusion or interruption to continuous suction. o - Quality of evacuation of blood, bile, and other bodily fluid and matter o All testing demonstrated that the modified ClearPath Upper GI disposable performed to its specifications and / or was equivalent to the predicate. ### Summary of substantial equivalence: We demonstrate that the modified ClearPath Upper Gl is substantially equivalent to the predicate in design and performance characteristics: - . Indications - Identical to predicate K093779: Cleaning, irrigating, and evacuating the upper GI tract during endoscopic procedures - . Technology - Identical to predicate: Single use disposable, applied over a standard endoscope, works in conjunction with the Clearpath Controller (K091305) to facilitate cleaning of the upper GI tract by irrigation and evacuation. - Environment of use Identical to predicate: Hospitals, clinics, and doctors' ● offices. {2}------------------------------------------------ ### 510(k) Summary . Page 3 of 3 25-Oct-11 - Materials Materials were either identical to the predicate or shown to comply with . ISO 10993-1 - . Difference - there are no substantial differences or new features in the proposed device compared to the predicate which raises any new safety or efficacy issue | Comparative Table - Compares the predicates and the proposed modified device | | | |------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | Proposed device | Predicate device | | | EasyGlide | K093779<br>EasyGlide | | Device | Clearpath Upper GI | Clearpath Upper GI | | Design | Used as an add-on to standard endoscopes for<br>irrigation and evacuation. Attaches along the<br>endoscope leaving the working channel free. | Used as an add-on to standard endoscopes for<br>irrigation and evacuation. Attaches along the<br>endoscope leaving the working channel free. | | Indications for use | The ClearPath Upper GI is intended for<br>irrigating or cleaning the upper digestive tract<br>and evacuating the irrigation fluid, blood and<br>bile in the upper GI tract during endoscopic<br>procedures. | The ClearPath Upper GI is intended for<br>irrigating or cleaning the upper digestive tract<br>and evacuating the irrigation fluid, blood and<br>bile in the upper GI tract during endoscopic<br>procedures. | | | It is for use only by trained medical personnel<br>located in hospitals, clinics, and doctors'<br>offices. | It is for use only by trained medical personnel<br>located in hospitals, clinics, and doctors'<br>offices. | | Environment of use | Hospitals, clinics, and doctors' offices. | Hospitals, clinics, and doctors' offices. | | Prescriptive | Yes, only trained medical personnel | Yes, only trained medical personnel | | Principle of<br>operation | Distal attachment to an endoscope, sleeve<br>ensuring attachment along entire length,<br>suction and irrigation tubes running along the<br>endoscope, suction and irrigations heads at<br>the distal tip. Enables irrigation and suction at<br>any time during the procedure without<br>removing any tools which may be inserted in<br>the endoscope's working channel. | Distal attachment to an endoscope, sleeve<br>ensuring attachment along entire length,<br>suction and irrigation tubes running along the<br>endoscope, suction and irrigations heads at<br>the distal tip. Enables irrigation and suction at<br>any time during the procedure without<br>removing any tools which may be inserted in<br>the endoscope's working channel. | | accessories | ClearPath Controller and ClearPath Tubing<br>(K091305) | ClearPath Controller and ClearPath Tubing<br>(K091305) | | Distal tip design | Multi irrigation holes<br>Distal suction hole | Multi irrigation holes<br>Distal suction hole | | Material | Comply with ISO 10993 | Comply with ISO 10993 | | Performance | Like the predicate, the modified device was<br>tested for compliance and performance under<br>the working conditions as prescribed by the<br>ClearPath Controller (K091305).<br>Comparative animal test compared<br>functionality to that of the predicates | As detailed in K093779 tests were performed<br>to demonstrate the performance of the device<br>under the working conditions as prescribed<br>by the ClearPath Controller (K091305). | | Disposable | Single patient, use, disposable | Single patient, use, disposable | | Packaged | Clean, non-sterile | Clean, non-sterile | ### Comparative Table - Compares the predicates and the proposed modified device {3}------------------------------------------------ ## DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002 EasyGlide Ltd. % Mr. Paul Dryden President ProMedic, Inc. 24301 Woodsage Drive BONITA SPRINGS FL 34134 MAY 1 6 2012 Re: K113166 > Trade/Device Name: ClearPath Upper GI Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II . Product Code: FDS Dated: April 15, 2012 Received: April 17, 2012 Dear Mr. Dryden: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean r lease of acreased a determination that your device complies with other requirements of the Act that + Dirderal statutes and regulations administered by other Federal agencies. You must or any I vathar other or strequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical {4}------------------------------------------------ device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ycm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerély yours, Benjamin K. Twitchell Benjarfun R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### Indications for Use Statement Page 1 of 1 ### 510(k) Number: # KJ13166 (To be assigned) Device Name: ClearPath Upper GI Indications for Use: The ClearPath Upper Gl is intended for irrigating or cleaning the upper digestive tract and evacuating the irrigation fluid, blood and bile in the upper GI tract during endoscopic procedures. It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices. Prescription Use XX (Part 21 CFR 801 Subpart D) or Over-the-counter use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) in Sign-Off Reproductive, Gastro-Renal, and
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...