K113163 · Penumbra, Inc. · NRY · Nov 23, 2011 · Cardiovascular
Device Facts
Record ID
K113163
Device Name
PENUMBRA SYSTEM MAX
Applicant
Penumbra, Inc.
Product Code
NRY · Cardiovascular
Decision Date
Nov 23, 2011
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.1250
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
Device Story
Penumbra System MAX components are supplemental catheters and separators designed to provide a larger lumen for efficient thrombus removal from intracranial vessels. The system is used in a clinical setting by physicians to treat acute ischemic stroke. The device is inserted into the vasculature to access the occlusion; it utilizes aspiration to remove thrombus. The system includes reperfusion catheters and separators that track through the vasculature to the site of the clot. By removing the thrombus, the device restores blood flow (revascularization) to the brain, potentially improving patient outcomes. The system is provided sterile and is for single use only.
Clinical Evidence
No human clinical data provided. Evidence consists of biocompatibility testing (ISO 10993), bench-top design verification (tensile strength, burst, friction, flow rate, torsion, simulated use), and an in vivo animal study (swine model). The animal study showed no vessel injury, no abnormal histology, and no significant vascular response.
Technological Characteristics
Catheter-based thrombus removal system. Materials evaluated per ISO 10993. Testing includes ASTM methods for hemolysis. Features include reperfusion catheters (3MAX/4MAX) and separators. Connectivity: None. Energy source: None (mechanical aspiration). Sterilization: 2x sterilized. Software: None.
Indications for Use
Indicated for revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (internal carotid, middle cerebral M1/M2, basilar, and vertebral arteries) within 8 hours of symptom onset.
Regulatory Classification
Identification
A percutaneous catheter is a device that is introduced into a vein or artery through the skin using a dilator and a sheath (introducer) or guide wire.
Penumbra System Separator Flex [026, 032, 041, 054] (K100769)
Submission Summary (Full Text)
{0}------------------------------------------------
NOV 2 3 2011
KIL3163 ps/5
# 10 510(k) Summary of Safety & Effectiveness
(as required by 21 CFR 807.92)
Pursuant to Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, Penumbra Inc. is providing the summary of Substantial Equivalence for the Penumbra System® MAX components.
# 10.1 Sponsor/Applicant Name and Address
Penumbra, Inc. 1351 Harbor Bay Parkway Alameda, CA 94502, USA
### 10.2 Sponsor Contact Information
Seth Schulman Director, Regulatory Affairs Phone: (510) 748-3223 FAX: (510) 217-6414 Email: seth.schulman@penumbrainc.com
### 10.3 Date of Preparation of 510(k) Summary
October 24, 2011
#### 10.4 Device Trade or Proprietary Name
Penumbra System® MAX
#### 10.5 Device Classification
Regulatory Class: II Cardiovascular Classification Panel: Classification Name: Catheter, Thrombus Removal 21 CFR §870.1250 Regulation Number: Product Code: NRY
#### 10.6 Predicate Devices
| 510(k) Number /<br>Clearance Date | Name of Predicate Device | Name of<br>Manufacturer |
|-----------------------------------|--------------------------------------------------------|-------------------------|
| K072718 [28Dec2007] | Penumbra System [026, 032, 041] | Penumbra, Inc. |
| K090752 [21Sep2009] | Penumbra System [054] | Penumbra, Inc. |
| K100769 [21May2010] | Penumbra System Separator Flex<br>[026, 032, 041, 054] | Penumbra, Inc. |
#### 10.7 Device Description
The Penumbra System MAX components are additional components of the currently available Penumbra System. The Penumbra System MAX components provide a larger lumen to assist in the efficient removal of thrombus from the
{1}------------------------------------------------
KL13163
p.2/5
brain. The devices are provided sterile, non-pyrogenic, and intended for single use only.
#### 10.8 Intended Use
The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
#### 10.9 Summary of Non-Clinical Data
As required under Section 12, Part (a)(i)(3A) of the Safe Medical Devices Act of 1990, a summary of any information regarding safety and effectiveness of the device follows.
Included in this section are descriptions of the testing's, which substantiates the safe and effective performance of the Neuron MAX System as well as its substantial equivalence to the predicate devices:
- Biocompatibility .
- Design Verification (Bench-Top Testing) .
- Animal Study .
The subject Penumbra System MAX components met all established requirements.
#### 10.9.1 Biocompatibility Testing
The Penumbra System MAX components were classified in accordance with ISO 10993 -1 guidelines (Biological Evaluation of Medical Devices) for limited duration (<24 hours), external-communicating devices, contacting circulating blood. All studies were conducted pursuant to 21 CFR, Part 58, Good Laboratory Practices. In summary, non-clinical testing found the Penumbra System MAX components to be biocompatible according to the requirements of ISO 10993 requirements, including:
{2}------------------------------------------------
K113163
p. 3/5
| Test | Method | Results |
|-------------------------------------------------------|-------------------------------------------------------|--------------------------------------------------------------|
| In Vitro<br>Cytotoxicity | ISO Elution Test (MEM<br>Extract) | No evidence of cell lysis or<br>toxicity |
| Acute<br>Intracutaneous<br>Reactivity<br>(Irritation) | ISO Intracutaneous<br>(Intradermal) Injection Test | No evidence of irritation |
| Acute Systemic<br>Toxicity | ISO Acute Systemic Injection<br>Test | No evidence of systemic<br>toxicity |
| Rabbit Pyrogen<br>Study | USP Material-Mediated<br>Rabbit Pyrogen Test | No evidence of material-<br>mediated pyrogenicity |
| Sensitization | ISO Maximization Test for<br>Delayed Hypersensitivity | Non-Sensitizing |
| Hemo-compatibility | | |
| -In Vitro<br>Hemolysis | ASTM Methode (Extraction &<br>Direct Contact) | Non-hemolytic |
| -In Vitro<br>Coagulation (PT,<br>PTT) | Prothrombin Time (PT) Assay | Coagulation times are within<br>the normal rang |
| | Partial Thromboplastin Time<br>(PTT) Assay | Non-Thrombogenic |
| -Complement<br>Activation | C3a and SC5b-9 through<br>Enzyme Assay | No greater biological response<br>than corresponding control |
# 10.9.2 Bench-top Testing
The physical, mechanical and performance testing of the Penumbra System MAX components demonstrates that the devices are substantially equivalent to the currently marketed predicate devices:
Design Verification testing was conducted to evaluate the physical and mechanical properties of the Penumbra System MAX components. All studies were conducted using good scientific practices and statistical sampling methods as required by the Penumbra Design Control procedures. All testing was performed using units which were 2x sterilized and met finished goods release requirements. The tests performed on the Penumbra System MAX components included:
{3}------------------------------------------------
113163 P. 4/5
| Test / Test Subject | Attribute | Sample Size | Result |
|--------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|--------|
| Pouch Seal | Pouch Seal Strength | 30 | Pass |
| Dimensional / Visual<br>Inspection | These evaluations confirm that the units used in this Design Verification testing meet<br>all inspection criteria for release of finished goods (clinically acceptable) product. | | Pass |
| Simulated Use<br>[Intracranial Access &<br>Thrombus Removal] | These evaluations confirm that the units used in this Design Verification testing meet<br>all inspection criteria for release of finished goods (clinically acceptable) product. | | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Hub /Shaft & Hub / HypotubeTensile Strength | 30 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Joint Tensile Strength | 30 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Hub Air Aspiration | 30 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Burst Test | 30 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Particulate Testing (Hydrophilic Coating) | 10 | Pass |
| Reperfusion Catheter4MAX/<br>Sheath or 8F Guide<br>compatibility<br>(Friction Force) | | | Pass |
| Reperfusion Catheter 4MAX/<br>0.014" Guidewire compatibility<br>(Friction Force) | | | Pass |
| Reperfusion Catheter 3MAX/<br>Reperfusion Catheter 054<br>(Friction Force) | Friction Force | 30 | Pass |
| Reperfusion Catheter 3MAX/<br>0.014" Guide wire<br>compatibility<br>(Friction Force) | | | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Flow Rate | 3 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Elongation to Failure | 30 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Corrosion | 30 | Pass |
| Reperfusion Catheter 3MAX /<br>Reperfusion Catheter 4MAX | Torsion | 30 | Pass |
# Design Verification (Bench-Top Testing) Summary – Reperfusion Catheters
# Design Verification (Bench-Top Testing) Summary - Separator 3MAX
| Test / Test Subject | Attribute | Sample<br>Size | Result |
|-------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------|--------|
| Dimensional / Visual Inspection<br>(all sizes) | These evaluations confirm that the units used in this Design<br>Verification testing meet all inspection criteria for release of<br>finished goods (clinically acceptable) product. | | Pass |
| Simulated Use: | Hub Transition | N=30 | Pass |
| Reperfusion Catheter 3MAX /<br>Separator 3MAX / Aspiration<br>Tubing Assembly Performance | Tracking | N=30 | Pass |
| | Separator 3MAX & Reperfusion Catheter<br>Compatible | N=30 | Pass |
| | Separator 3MAX / Reperfusion Catheter / Pump<br>and accessories / Aspiration Tubing (Aspiration<br>Remove Clot) | N=30 | Pass |
| | Separator 3MAX and Reperfusion Catheter | N=30 | Pass |
{4}------------------------------------------------
| Test / Test Subject . | Attribute | Sample<br>Size | Result |
|---------------------------|---------------------------------------------|----------------|--------|
| | Compatibility (Separator Advance / Retract) | | |
| Separator Bond Joint Test | Separator 054 and Wire Joint Break Force | N=30 | Pass |
| Separator Bond Joint Test | Separator 041 and Wire Joint Break Force | N=30 | Pass |
| Separator Bond Joint Test | Separator 032 and Wire Joint Break Force | N=30 | Pass |
| Separator Bond Joint Test | Separator 026 and Wire Joint Break Force | N=30 | Pass |
| Separator Bond Joint Test | Separator 3MAX and Wire Joint Break Force | N=30 | Pass |
The results of the tests appropriately address the physical and mechanical performance expectations of the device. This is further supported by the surgical handling and performance results reported in the in vivo study. Based on these overall results, the physical and mechanical properties of the Penumbra System MAX components are acceptable for the intended use and substantially equivalent to the predicate devices.
# 10.9.3 Animal Study
An animal study was conducted to evaluate the safe use of the Penumbra System MAX in a swine model. The study concluded that:
- No vessel injury was noted on the final angiograms following the vessel . response procedure.
- . No abnormal gross or histology findings were noted in test vessel segments.
- . The use of the Penumbra System MAX components resulted in no significant vascular response in these experimental conditions.
# 10.9.4 Summary of Substantial Equivalence
The Penumbra System MAX components are substantially equivalent to the predicate devices with regard to intended use, operating principle, design concept, materials, shelf-life, packaging and sterilization processes.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" arranged in a circular fashion around the eagle. The logo is black and white and appears to be a simple, vectorized design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
Penumbra, Inc. c/o Mr. Seth A. Schulman Director, Regulatory Affairs 1351 Harbor Bay Parkway Alameda, CA 94502
NOV 2 3 2011
Re: K113163
Trade/Device Name: Penumbra System® MAX Regulation Number: 21 CFR 870.1250 Regulation Name: Percutaneous Catheter Regulatory Class: Class II Product Code: NR Y Dated: October 25, 2011 Received: October 26, 2011
Dear Mr. Schulman:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls: Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Seth A. Schulman
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerety Vours.
.
Melvin R. Felder, MD
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear. Nose and Throat Devices Office of Device Evaluation · Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### Statement of Indication for Use 11
| Indications for Use |
|---------------------|
|---------------------|
| 510(k) Number (if known): | K113163 |
|---------------------------|----------------------|
| Device Name: | Penumbra System® MAX |
Indications for Use:
The Penumbra System is intended for use in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (in the internal carotid, middle cerebral - M1 and M2 segments, basilar, and vertebral arteries) within 8 hours of symptom onset.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR Over The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
> Concurrence of CDRH, Office of Device Evaluation (ODE) JEFFREY TOY
> > (Division Sign-Off) Division of Ophthalmic, Neurological and Ear, Nose and Throat Devices
K113163
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.