The Unistrip1 Test Strips are used with the OneTouch® Ultra®, OneTouch® Ultra2®, OneTouch® UltraMini® and One-Touch® UltraSmart® meters purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The Unistrip1 is meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control. They are for single patient use only and should not be shared. They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm. Testing is done outside the body (in vitro diagnostic use). They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control. Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates. UniStrip1 Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch®Ultra® meter). Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly).
Device Story
UniStrip1 test strips are single-use, in vitro diagnostic consumables for quantitative blood glucose measurement. Used with specific LifeScan OneTouch meters (purchased pre-Oct 2012) via calibration code 49. Input: fresh capillary whole blood (finger, palm, forearm). Principle: amperometric assay using glucose oxidase (Aspergillus niger) and potassium ferricyanide; reaction generates electrical current proportional to glucose concentration. Meter measures current and displays result. Used by patients at home for diabetes management. Output: digital glucose concentration (mg/dL). Benefits: enables self-monitoring of glycemic control. Limitations: not for insulin dose calculation or CGM calibration; AST restricted to steady-state glucose conditions.
Clinical Evidence
No clinical studies were conducted. Evidence consists of non-clinical bench testing including NCCLS EP5-A (precision), NCCLS EP6-A (linearity), ISO 15971 (interference, system accuracy, altitude, control solution), ISO 5725-1 (comparison, user evaluation), and EN 13640 (operational/storage conditions).
Indicated for people with diabetes for self-testing of blood glucose as an aid to monitor diabetes control effectiveness. Used with specific OneTouch meters (purchased before Oct 2012) at calibration code 49. Measures glucose in fresh capillary whole blood from finger, palm, or forearm. Not for diabetes diagnosis/screening or neonatal use.
Regulatory Classification
Identification
A glucose test system is a device intended to measure glucose quantitatively in blood and other body fluids. Glucose measurements are used in the diagnosis and treatment of carbohydrate metabolism disorders including diabetes mellitus, neonatal hypoglycemia, and idiopathic hypoglycemia, and of pancreatic islet cell carcinoma.
Special Controls
*Classification.* Class II (special controls). The device, when it is solely intended for use as a drink to test glucose tolerance, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9.
{0}
1
510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY ONLY TEMPLATE
A. 510(k) Number:
k113135
B. Purpose for Submission:
Alternative blood glucose test strips for use with LifeScan OneTouch Ultra, Ultra2, UltraSmart and UltraMini blood glucose meters purchased before October 2012 with test strip calibration code 49.
C. Measurand:
Capillary whole blood glucose
D. Type of Test:
Quantitative Amperometric assay (Glucose Oxidase)
E. Applicant:
UniStrip Technologies, LLC
F. Proprietary and Established Names:
UniStrip 1
G. Regulatory Information:
1. Regulation section:
21 CFR 862.1345, Glucose test system
2. Classification:
Class II
3. Product code:
NBW, System, Test, Blood Glucose, Over The Counter
CGA, Glucose Oxidase, Glucose
{1}
4. Panel:
Clinical Chemistry (75)
H. Intended Use:
1. Intended use(s):
See indications for use below.
2. Indication(s) for use:
The Unistrip1 Test Strips are used with the OneTouch® Ultra®, OneTouch® Ultra2®, OneTouch® UltraMini® and One-Touch® UltraSmart® meters purchased before October 2012, set at calibration code 49, for measuring glucose (sugar) in whole capillary blood. The Unistrip1 is meant for self-testing of blood glucose as an aid to monitor the effectiveness of diabetes control.
They are for single patient use only and should not be shared.
They are used to quantitatively measure glucose in fresh capillary whole blood samples taken from the finger, palm, or forearm.
Testing is done outside the body (in vitro diagnostic use).
They are indicated for use by people with diabetes in their home as an aid to monitor the effectiveness of diabetes control.
Not intended for the diagnosis of or screening for diabetes mellitus and is not intended for use on neonates.
UniStrip1 Test Strips allow alternate site testing (AST) from the fingertip, palm and/or the forearm. (Use of palm AST is not to be done with OneTouch®Ultra® meter).
Alternative site testing should only be done during steady-state times (when glucose is not changing rapidly).
3. Special conditions for use statement(s):
- For over-the-counter use
- Not for neonatal use
- Not for screening or diagnosis of diabetes mellitus
- Not for use on critically ill patients, patients in shock, dehydrated patients or hyper-osmolar patients
- Alternative site testing (AST) testing should only be done during steady-state times (when glucose is not changing rapidly).
- AST should not be used to calibrate continuous glucose monitors (CGMs).
2
{2}
- AST should not be used for insulin dose calculations.
# 4. Special instrument requirements:
LifeScan OneTouch Ultra, Ultra2, UltraSmart and UltraMini blood glucose meters purchased before October 2012 with test strip calibration code 49.
# I. Device Description:
The UniStrip1 Test Strips are used with the OneTouch Ultra, OneTouch Ultra2, OneTouch UltraMini and OneTouch UltraSmart glucose monitoring meters purchased before October 2012 for qualitatively measuring glucose in fresh capillary whole blood. UniStrip1 Test Strips are used in fingertip, forearm, and palm testing. Each test strip contains glucose oxidase (Aspergillus niger), potassium ferricyanide, buffer and other non-reactive ingredients. UniStrip test strips are only for use with calibration code 49.
The UniStrip test strip package insert instructs the user to check their device using the low and high Prodigy® Control Solutions. Expected control ranges for the low (25-70 mg/dL) and high (200-300 mg/dL) Prodigy Control Solutions are shown on the test strip vial.
# J. Substantial Equivalence Information:
1. Predicate device name(s):
One Touch Test Strips (OneTouch Blood Glucose Test System)
2. Predicate 510(k) number(s):
k923544
3. Comparison with predicate:
| Comparison Table | | |
| --- | --- | --- |
| Item | Device (k113135) | Predicate (k923544) |
| Indications for Use | Quantitative determination of glucose concentrations in blood as an aid in monitoring effectiveness of diabetes control. | Same |
| Alternative Site Testing | Fingertip, forearm, and palm | Fingertip and earlobe |
| Detection method | Amperometry | colorimetric reflectance |
| Enzyme | Glucose Oxidase (Aspergillus niger) | same |
{3}
| Comparison Table | | |
| --- | --- | --- |
| Item | Device (k113135) | Predicate (k923544) |
| Sample volume | 1.0 uL | Not specified |
| Reaction time | 5 seconds | 45 seconds |
| Measurement range | 20-600 mg/dL | 0-600 mg/dL |
| Hematocrit range | 30 - 55% | 25 – 60% |
| Storage conditions | Store at < 30°C until expiration date. | Store at < 30°C until expiration date. Do not refrigerate. |
K. Standard/Guidance Document Referenced (if applicable):
1. ISO 15197. In vitro diagnostic test systems. Requirements for blood-glucose monitoring systems for self-testing in managing diabetes mellitus.
2. ISO 14971:2007. Medical devices-Application of risk management to medical devices.
3. CLSI EP7-A2. Interference Testing in Clinical Chemistry; Approved Guideline (2005).
4. CLSI EP5-A2. Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline (2004).
5. ISO 13640:2002. Stability Testing of In Vitro Diagnostic Reagents. (In Vitro Diagnostics).
6. CLSI EP6-A. Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; Approved Guideline (2003).
7. ISO 5725-1. Accuracy (trueness and precision) of measurement methods and results - Part 1 - General Principles and Definitions.
L. Test Principle:
Glucose in the blood sample mixes with glucose oxidase on the UniStrip1 test strip and a small electrical current is measured by the meter(s) and displayed as your blood glucose results. The strength of this current changes with the amounts of glucose in the blood sample.
M. Performance Characteristics (if/when applicable):
The sponsor states that the following performance evaluation provided for the UniStrip test strips was obtained using the 4 claimed meters (LifeScan OneTouch Ultra, OneTouch Ultra 2, One Touch UltraSmart and OneTouch UltraMini) purchased before October 2012. The sponsor states that UniStrip test strip calibration code 49 was used in each of the performance evaluations.
1. Analytical performance:
a. Precision/Reproducibility:
{4}
Within-run precision was performed in one day using five blood glucose concentrations. Each sample was tested using calibration code 49 with the UniStrip test strips and the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini® and OneTouch® UltraSmart® glucose meters. Three UniStrip test lots were used with each blood glucose meter. Five different meters for each model type above were each used to test each sample in singlicate with all three test strip lots. This resulted in a total of $n = 25$ for each glucose meter model and test strip lot combination for each glucose concentration. The results are summarized below.
One Touch Ultra
| Lot 1 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| --- | --- | --- | --- | --- | --- |
| Mean (mg/dL) | 47 | 87 | 127 | 213 | 318 |
| SD (mg/dL) | 2.6 | 2.4 | 4.8 | 7.4 | 11.2 |
| %CV | 5.5 | 2.8 | 3.8 | 3.5 | 3.5 |
| Lot 2 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 49 | 87 | 125 | 213 | 316 |
| SD (mg/dL) | 2.3 | 3.1 | 3.8 | 6.3 | 10.9 |
| %CV | 4.7 | 3.5 | 3.0 | 3.0 | 3.4 |
| Lot 3 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 48 | 88 | 126 | 212 | 319 |
| SD (mg/dL) | 2.1 | 3.5 | 5.2 | 6.2 | 9.6 |
| %CV | 4.4 | 4.0 | 4.2 | 2.9 | 3.0 |
One Touch Ultra2
| Lot 1 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| --- | --- | --- | --- | --- | --- |
| Mean (mg/dL) | 48 | 87 | 124 | 213 | 316 |
| SD (mg/dL) | 2.8 | 2.6 | 4.8 | 6.3 | 11.9 |
| %CV | 5.8 | 3.0 | 3.9 | 3.0 | 3.8 |
| Lot 2 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 48 | 88 | 127 | 214 | 320 |
| SD (mg/dL) | 2.5 | 3.1 | 4.4 | 6.0 | 11.4 |
| %CV | 5.2 | 3.5 | 3.5 | 2.8 | 3.6 |
| Lot 3 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 48 | 89 | 125 | 210 | 320 |
| SD | 2.8 | 3.0 | 4.5 | 6.4 | 11.9 |
{5}
One Touch UltraMini
| Lot 1 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| --- | --- | --- | --- | --- | --- |
| Mean (mg/dL) | 47 | 87 | 125 | 212 | 317 |
| SD (mg/dL) | 2.4 | 2.5 | 4.2 | 6.9 | 9.6 |
| %CV | 5.1 | 2.9 | 3.3 | 3.3 | 3.0 |
| Lot 2 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 47 | 88 | 125 | 212 | 321 |
| SD (mg/dL) | 2.7 | 3.0 | 4.6 | 5.9 | 10.0 |
| %CV | 5.7 | 3.4 | 3.7 | 2.8 | 3.1 |
| Lot 3 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 49 | 88 | 125 | 211 | 315 |
| SD (mg/dL) | 2.9 | 2.8 | 5.1 | 6.4 | 11.3 |
| %CV | 5.9 | 3.2 | 4.1 | 3.1 | 3.6 |
One Touch UltraSmart
| Lot 1 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| --- | --- | --- | --- | --- | --- |
| Mean (mg/dL) | 47 | 89 | 126 | 213 | 318 |
| SD (mg/dL) | 2.6 | 3.1 | 4.4 | 6.5 | 11.7 |
| %CV | 5.5 | 3.4 | 3.4 | 3.1 | 3.7 |
| Lot 2 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 48 | 88 | 128 | 213 | 318 |
| SD (mg/dL) | 2.5 | 2.9 | 4.4 | 6.4 | 11.6 |
| %CV | 5.2 | 3.2 | 3.5 | 3.0 | 3.6 |
| Lot 3 | Level 1 | Level 2 | Level 3 | Level 4 | Level 5 |
| Mean (mg/dL) | 49 | 88 | 126 | 214 | 320 |
| SD (mg/dL) | 2.4 | 3.0 | 4.8 | 7.0 | 10.5 |
| %CV | 4.9 | 3.5 | 3.8 | 3.3 | 3.3 |
Between-run precision was performed over ten days using three blood glucose control
{6}
concentrations. Each sample was tested using calibration code 49 with the UniStrip test strips and the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, and OneTouch® UltraSmart® glucose meters. Three UniStrip test lots were used with each blood glucose meter. Five different meters for each model type above were each used to test each sample in singlicate with all three test strip lots. This resulted in $n = 5$ for each glucose meter model and test strip lot combination each day for each glucose concentration tested over ten days for a total of $n = 50$ . Results are summarized below.
One Touch Ultra
| Lot 1 | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 47 | 93 | 291 |
| SD (mg/dL) | 3.4 | 3.2 | 9.3 |
| %CV | 7.2 | 3.5 | 3.2 |
| Lot 2 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 48 | 93 | 288 |
| SD (mg/dL) | 3.4 | 3.1 | 9.5 |
| %CV | 7.1 | 3.4 | 3.3 |
| Lot 3 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 48 | 94 | 289 |
| SD (mg/dL) | 3.2 | 2.9 | 10.3 |
| %CV | 6.7 | 3.1 | 3.6 |
One Touch Ultra2
| Lot 1 | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 48 | 94 | 291 |
| SD (mg/dL) | 3.5 | 3.5 | 11.2 |
| %CV | 7.3 | 3.8 | 3.8 |
| Lot 2 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 48 | 93 | 290 |
| SD (mg/dL) | 3.5 | 3.5 | 9.4 |
| %CV | 7.3 | 3.8 | 3.2 |
| Lot 3 | Level 1 | Level 2 | Level 3 |
| Mean | 48 | 93 | 289 |
{7}
8
| (mg/dL) | | | |
| --- | --- | --- | --- |
| SD (mg/dL) | 3.4 | 3.3 | 9.7 |
| %CV | 7.1 | 3.6 | 3.4 |
One Touch UltraMini
| Lot 1 | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 48 | 93 | 289 |
| SD (mg/dL) | 3.6 | 3.4 | 8.8 |
| %CV | 7.5 | 3.7 | 3.1 |
| Lot 2 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 48 | 94 | 289 |
| SD (mg/dL) | 4.1 | 3.0 | 8.7 |
| %CV | 8.5 | 3.2 | 3.0 |
| Lot 3 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 47 | 93 | 289 |
| SD (mg/dL) | 3.3 | 3.5 | 9.1 |
| %CV | 7.0 | 3.7 | 3.2 |
One Touch UltraSmart
| Lot 1 | Level 1 | Level 2 | Level 3 |
| --- | --- | --- | --- |
| Mean (mg/dL) | 47 | 94 | 290 |
| SD (mg/dL) | 3.5 | 3.6 | 9.5 |
| %CV | 7.4 | 3.8 | 3.3 |
| Lot 2 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 48 | 93 | 291 |
| SD (mg/dL) | 3.8 | 3.2 | 10.3 |
| %CV | 7.9 | 3.5 | 3.5 |
| Lot 3 | Level 1 | Level 2 | Level 3 |
| Mean (mg/dL) | 47 | 94 | 291 |
| SD (mg/dL) | 3.4 | 3.1 | 9.7 |
| %CV | 7.2 | 3.3 | 3.3 |
{8}
# b. Linearity/assay reportable range:
Linearity was evaluated using 11 venous blood samples with glucose concentrations adjusted to 15, 20, 80, 150, 250, 300, 350, 400, 510, 600 and $620\mathrm{mg / dL}$ and measured by YSI. Each sample was measured 10 times with 3 UniStrip test strip lots using calibration code 49 with the UniStrip test strips and the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, and OneTouch® UltraSmart® glucose meters. Mean values for each sample and glucose meter were calculated and compared to YSI values and used to perform linear regression. All of the individual sample results were within $\pm 10\%$ of the YSI sample results. The linear regression results are shown below.
| | Slope | y-Int. | R2 | r |
| --- | --- | --- | --- | --- |
| ONETOUCH® Ultra Blood Glucose Meters: | 0.97 | 2.0 | 0.998 | 0.999 |
| ONETOUCH® Ultra2 Blood Glucose Meters: | 0.97 | 2.0 | 0.997 | 0.999 |
| ONETOUCH® UltraMini Blood Glucose Meters: | 0.97 | 2.8 | 0.997 | 0.999 |
| ONETOUCH® UltraSmart Blood Glucose Meters: | 0.98 | 0.5 | 0.998 | 0.999 |
The results of the study support the sponsor's claimed glucose measurement range of $20 - 600\mathrm{mg / dL}$ .
# c. Traceability, Stability, Expected values (controls, calibrators, or methods):
The UniStrip test strips are traceable to the NIST SRM #917b D-Glucose (Dextrose). The reference instrument used was the YSI 2300 Glucose Analyzer, which is calibrated by YSI 2747 Glucose Standard. The results of the capillary whole blood samples tested by ONETOUCH series Blood Glucose Meters Ultra, Ultra2, UltraMini and UltraSmart with UniStrip Blood Glucose Test Strips are compared with the results of the corresponded plasma samples tested by the calibrated YSI 2300 Glucose Analyzer.
The UniStrip test strip package insert instructs the user to check their device using the Prodigy® Control Solution. Expected control ranges for the low and high Prodigy Control Solutions are shown on the test strip vial. The controls were previously cleared in k060467.
Stability testing protocols and acceptance criteria for the UniStrip test strips shelf-life were provided. The manufacturer claims shelf life stability of 24 months and opened vial stability of 3 months when stored unopened at $4 - 30^{\circ}\mathrm{C}$ (relative humidity 10-90%) The real time and accelerated stability study protocols and acceptance criteria for these studies were reviewed and found to be adequate.
# d. Detection limit:
{9}
See linearity in section M1b above.
# e. Analytical specificity:
Interference testing was performed using three lots of UniStrip test strips and the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, and OneTouch® UltraSmart® meters by spiking venous blood to three glucose concentrations (40, 120, 340 mg/dL) and the potential interferent concentrations shown below. Each interferent and glucose concentration were tested in replicates of five with each lot and glucose meter then compared to the control sample. The sponsor defined no significant interference as $\leq 10\%$ difference between individual interferent samples and the control sample. Results are shown below:
| Interferant | Max Concentration with no significant interference |
| --- | --- |
| Acetaminophen | 15 mg/dL |
| Ascorbic acid | 3.0 mg/dL |
| Caffeine | 100 mg/dL |
| Dopamine | 6.0 mg/dL |
| Ephedrine | 1.0 mg/dL |
| Gentisic Acid | 10 mg/dL |
| L-dopa | 20 mg/dL |
| Mannitol | 5.0 mg/dL |
| Salicylates | 60 mg/dL |
| Tetracycline | 4.0 mg/dL |
| Cholesterol | 700 mg/dL |
| Triglyceride | 3000 mg/dL |
| Uric acid | 15 mg/dL |
The sponsor has the following limitations in the UniStrip test strip insert:
- Hematocrit is the percentage of red blood cells in the blood. Extremes in hematocrit may affect test results. Hematocrit levels below $30\%$ may cause falsely high readings. Hematocrit levels over $55\%$ may cause falsely low readings. If you do not know your hematocrit level, consult your healthcare professional.
- Acetaminophen, salicylates, uric acid, ascorbic acid (vitamin C), and other reducing substances (when occurring in normal blood or normal therapeutic concentrations) do not affect results by more than $10\%$ . However, more than normally high concentrations in blood may cause inaccurately high results.
- Lipemic samples: Cholesterol levels up to $700\mathrm{mg / dL}$ (18.1 mmol/L) and
{10}
triglycerides up to 3000 mg/dL (33.9 mmol/L) do not affect the results. Grossly lipemic patient samples have not been tested. They are not advised for testing with any OneTouch Ultra Meters.
f. Assay cut-off:
Not applicable
2. Comparison studies:
a. Method comparison with predicate device:
A system accuracy study was performed to evaluate the system accuracy of the blood glucose system of ONETOUCH® Series Meters with UniStrip1 Blood Glucose Test Strips compared to Glucose Analyzer YSI 2300. Two each of the following meters were evaluated with three lots of UniStrip1 Blood Glucose Test Strips: OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, and OneTouch® UltraSmart®. All meters were set to code 49 and 100 subjects with YSI blood glucose concentrations ranging from 47 to 444 mg/dL were evaluated over 10 days. Hematocrit for these samples was between 30% and 55%. To obtain samples with glucose < 50 mg/dL, samples were collected with an appropriate anticoagulant (EDTA) and allowed to incubate at room temperature to allow glucose to glycolyze. To obtain samples with glucose concentrations above 400 mg/dL, samples were collected with an appropriate anticoagulant (EDTA) and spiked with glucose. No more than three samples were adjusted with each of these procedures. For each test subject a fingerstick was performed by a healthcare professional and the sample was tested using one test strip for each glucose meter listed above. The procedure was repeated for palm and forearm test sites for all meters, except for the OneTouch®Ultra, which only tested finger and forearm sites. A healthcare professional drew a venous sample from each subject's arm to test hematocrit. An aliquot of this sample was used to test each subject's glucose concentration on the YSI 2300. The glucose concentration from each meter was compared to the YSI 2300. Results are summarized below.
Fingerstick
OneTouch® Ultra vs YSI-2300
Glucose concentration < 75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 11/18 (61%) | 17/18 (94%) | 18/18 (100%) |
Glucose concentration ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 38/82 (46%) | 71/81 (85%) | 79/82 (97%) | 82/82 (100%) |
{11}
OneTouch® Ultra2 vs YSI-2300
Glucose concentration $< 75\mathrm{mg / dL}$
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 11/18 (61%) | 17/18 (92%) | 18/18 (100%) |
Glucose concentration $\geq 75\mathrm{mg / dL}$
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 40/82 (49%) | 70/82 (85%) | 80/82 (98%) | 82/82 (100%) |
OneTouch® UltraMini vs YSI-2300
Glucose concentration $< 75\mathrm{mg / dL}$
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 11/18 (61%) | 17/18 (94%) | 18/18 (100%) |
Glucose concentration $\geq 75\mathrm{mg / dL}$
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 42/82 (51%) | 66/82 (80%) | 80/82 (98%) | 81/82 (99%) |
OneTouch® UltraSmart vs YSI-2300
Glucose concentration $< 75\mathrm{mg / dL}$
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 10/18 (56%) | 17/18 (94%) | 18/18 (100%) |
Glucose concentration $\geq 75\mathrm{mg / dL}$
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 42/82 (51%) | 71/82 (87%) | 81/82 (99%) | 82/82 (100%) |
Palm
OneTouch® Ultra2 vs YSI-2300
Glucose concentration $< 75\mathrm{mg / dL}$
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 12/18 (67%) | 17/18 (94%) | 18/18 (100%) |
Glucose concentration $\geq 75\mathrm{mg / dL}$
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 37/82 (45%) | 63/82 (77%) | 77/82 (95%) | 81/82 (99%) |
OneTouch® UltraMini vs YSI-2300
Glucose concentration $< 75\mathrm{mg / dL}$
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 12/18 (67%) | 16/18 (89%) | 18/18 (100%) |
Glucose concentration $\geq 75\mathrm{mg / dL}$
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 45/82 (55%) | 69/82 (84%) | 80/82 (98%) | 81/82 (99%) |
OneTouch® UltraSmart vs YSI-2300
{12}
Glucose concentration <75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 12/18 (67%) | 17/18 (94%) | 17/18 (94%) |
Glucose concentration ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 44/82 (54%) | 69/82 (84%) | 80/82 (98%) | 82/82 (100%) |
# Forearm
OneTouch® Ultra vs YSI-2300
Glucose concentration <75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 13/18 (72%) | 17/18 (94%) | 18/18 (100%) |
Glucose concentration ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 37/82 (41%) | 68/82 (80%) | 79/82 (97%) | 82/82 (100%) |
OneTouch® Ultra2 vs YSI-2300
Glucose concentration <75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 9/18 (50%) | 18/18 (100%) | 18/18 (100%) |
Glucose concentration ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 37/82 (45%) | 62/82 (76%) | 81/82 (99%) | 82/82 (100%) |
OneTouch® UltraMini vs YSI-2300
Glucose concentration <75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 11/18 (61%) | 16/18 (89%) | 18/18 (100%) |
Glucose concentration ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 43/82 (52%) | 67/82 (82%) | 79/82 (96%) | 81/81 (99%) |
OneTouch® UltraSmart vs YSI-2300
Glucose concentration <75 mg/dL
| Within ±5 mg/dL | Within ±10 mg/dL | Within ±15 mg/dL |
| --- | --- | --- |
| 8/18 (446%) | 178/18 (94%) | 18/18 (100%) |
Glucose concentration ≥75 mg/dL
| Within ±5% | Within ±10% | Within ±15% | Within ±20% |
| --- | --- | --- | --- |
| 35/82 (43%) | 67/82 (82%) | 81/82 (99%) | 81/82 (99%) |
{13}
Linear regression analysis results are summarized below:
Fingerstick testing vs YSI 2300
| | Slope | y-int. | R2 |
| --- | --- | --- | --- |
| ONETOUCH® Ultra Blood Glucose Meters: | 0.997 | 1.18 | 0.981 |
| ONETOUCH® Ultra2 Blood Glucose Meters: | 0.992 | -0.08 | 0.984 |
| ONETOUCH® UltraMini Blood Glucose Meters: | 0.979 | 1.18 | 0.982 |
| ONETOUCH® UltraSmart Blood Glucose Meters: | 0.992 | 0.38 | 0.985 |
AST Palm testing vs YSI 2300
| | Slope | y-int. | R2 |
| --- | --- | --- | --- |
| ONETOUCH® Ultra2 Blood Glucose Meters: | 0.994 | -1.15 | 0.978 |
| ONETOUCH® UltraMini Blood Glucose Meters: | 0.982 | 1.03 | 0.983 |
| ONETOUCH® UltraSmart Blood Glucose Meters: | 0.987 | 0.33 | 0.984 |
AST Forearm testing vs YSI 2300
| | Slope | y-int. | R2 |
| --- | --- | --- | --- |
| ONETOUCH® Ultra Blood Glucose Meters: | 1.001 | -2.06 | 0.981 |
| ONETOUCH® Ultra2 Blood Glucose Meters: | 1.015 | -3.724 | 0.981 |
| ONETOUCH® UltraMini Blood Glucose Meters: | 0.976 | 1.96 | 0.981 |
| ONETOUCH® UltraSmart Blood Glucose Meters: | 0.986 | 1.22 | 0.981 |
b. Matrix comparison:
Not applicable
3. Clinical studies:
a. Clinical Sensitivity:
Not applicable
b. Clinical specificity:
Not applicable
c. Other clinical supportive data (when a. and b. are not applicable):
{14}
A lay user study was performed with 152 volunteers between the ages of 15-80 years old. There were approximately equal numbers of male and female volunteers and they did not have experience performing blood glucose tests on themselves. Each volunteer read the user's guide before performing the testing. This was the only instruction provided. The users performed a fingerstick and measured their blood glucose. This was followed by the users performing palm and forearm sticks and testing, after which the technician performed a fingerstick and blood glucose measurement. A technician then performed a venous blood draw to measure hematocrit and glucose using the YSI-2300. The testing was performed with the UniStrip test strip and the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, and OneTouch® UltraSmart® meters. Results are summarized below.
# Fingerstick results
ONETOUCH® Ultra versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 15/21 (71%) | 21/21 (100%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 89/131 (68) | 112/131 (85%) | 128/131 (98%) | 131/131 (100%) |
ONETOUCH® Ultra2 versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 17/21 (81%) | 21/21 (100%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 88/131 (67) | 103/131 (79%) | 126/131 (96%) | 131/131 (100%) |
ONETOUCH® UltraMini versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
{15}
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 19/21 (90%) | 21/21 (100%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 88/131 (67) | 106/131 (81%) | 125/131 (96%) | 131/131 (100%) |
ONETOUCH® UltraSmart versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 16/21 (76%) | 21/21 (100%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 98/131 (75%) | 111/131 (85%) | 131/131 (100%) | 131/131 (100%) |
AST Palm Results:
ONETOUCH® Ultra2 versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 17/21 (81%) | 17/21 (81%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 85/131 (65%) | 105/131 (80%) | 123/131 (94%) | 131/131 (100%) |
ONETOUCH® UltraMini versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
{16}
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 88/131 (67%) | 109/131 (83%) | 124/130 (95%) | 128/131 (98%) |
ONETOUCH® UltraSmart versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 17/21 (81%) | 20/21 (95%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 96/131 (738%) | 111/131 (85%) | 126/131 (96%) | 131/131 (100%) |
AST Forearm results
ONETOUCH® Ultra versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 20/21 (95%) | 21/21 (100%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 98/131 (75%) | 107/131 (82%) | 123/131 (94%) | 129/131 (98%) |
ONETOUCH® Ultra2 versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
{17}
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 89/131 (68%) | 104/131 (79%) | 122/131 (93%) | 128/131 (98%) |
ONETOUCH® UltraMini versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 18/21 (86%) | 20/21 (95%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 85/131 (65%) | 104/131 (79%) | 123/131 (94%) | 139/131 (98%) |
ONETOUCH® UltraSmart versus YSI 2300
For glucose concentration $< 75\mathrm{mg / dL}$
| Within±5 mg/dL (Within±0.28 mmol/L) | Within±10 mg/dL (Within±0.56 mmol/L) | Within±15 mg/dL (Within±0.83 mmol/L) |
| --- | --- | --- |
| 16/21 (76%) | 18/21 (86%) | 21/21 (100%) |
For glucose concentration $\geq 75\mathrm{mg / dL}$
| Within±5 % | Within±10 % | Within±15 % | Within±20 % |
| --- | --- | --- | --- |
| 94/131 (72%) | 118/131 (90%) | 130/131 (99%) | 131/131 (100%) |
Linear regression analysis results are summarized below:
AST Palm testing vs YSI 2300
| | Slope | y-int. | R2 |
| --- | --- | --- | --- |
| ONETOUCH® Ultra2 Blood Glucose Meters: | 1.020 | 0.03 | 0.977 |
| ONETOUCH® UltraMini Blood Glucose Meters: | 1.023 | -1.54 | 0.976 |
| ONETOUCH® UltraSmart Blood Glucose Meters: | 1.047 | -4.28 | 0.979 |
{18}
AST Forearm testing vs YSI 2300
| | Slope | y-int. | R2 |
| --- | --- | --- | --- |
| ONETOUCH® Ultra Blood Glucose Meters: | 1.008 | -0.35 | 0.976 |
| ONETOUCH® Ultra2 Blood Glucose Meters: | 1.019 | -1.06 | 0.980 |
| ONETOUCH® UltraMini Blood Glucose Meters: | 1.009 | -0.57 | 0.971 |
| ONETOUCH® UltraSmart Blood Glucose Meters: | 1.022 | -0.14 | 0.983 |
# 4. Clinical cut-off:
Not applicable
# 5. Expected values/Reference range:
The package insert states the following:
Reference values: Please consult your physician to establish your target ranges. The American Diabetes Association (ADA) suggests the following targets for most nonpregnant adults without diabetes. More or less stringent goals may be appropriate for each individual.
Glycemic control Time of Day ADA Suggested Reading
Preprandial plasma glucose $70 - 100\mathrm{mg / dl}$ (before a meal)
Postprandial plasma glucose $< 140\mathrm{mg / dl}$ (after a meal)
Source: American Diabetes Association. Standards of medical care in diabetes - 2012. 2012;35 (supp.1 Diabetes Care):S11-563.
# P. Other Supportive Instrument Performance Characteristics Data Not Covered In The "Performance Characteristics" Section above:
1. Software validation testing was performed demonstrating that the error codes displayed by the meters functioned as intended with the UniStrip test strips.
2. The package insert instructs the use to call 1.888.729.9674 if the user needs any additional assistance, if the test strips vial is open or damaged, or if the instructional
{19}
materials or results seem unclear. The insert states that the firm can be contacted 24 hours a day / 7 days a week.
3. Hematocrit Study: The effect of different hematocrit levels on the performance of the UniStrip test strip with the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, OneTouch® UltraSmart® meters was evaluated using venous whole blood samples with hematocrit levels of 25, 30, 35, 45, 55 and 60% spiked with glucose to 5 concentrations (approximately 40, 80, 130, 200 and 320 mg/dL). Each sample was then tested in replicates of 10 using each meter type with three lots of UniStrip test strips and the values were compared with those obtained from YSI analyzer. The % biases relative to YSI were acceptable within the claimed hematocrit range of 30 to 55%.
4. The UniStrip test strip labeling states that the test strips can be used at a temperature range of 6-44°C and a relative humidity (RH), range of 10-90%. In order to demonstrate this performance, OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, OneTouch® UltraSmart® meters and the UniStrip test strips were placed in environmental chambers with the following temperature/relative humidity combinations: 6°C/10% RH, 6°C/90% RH, 25°C/10% RH, 25°C/90% RH, 44°C/10% RH and 44°C/90% RH. Fresh venous blood samples with glucose concentrations of approximately 50, 150 and 315 mg/dL were tested five times with each glucose meter model at all of the temperature/RH combinations shown above and the results from each measurement were compared to the YSI values for each sample. The results from each measurement were within the sponsor's acceptance criteria. These results demonstrate that the performance of the UniStrip test strips with the claimed meters is unaffected by the temperature and humidity ranges specified in the device labeling.
5. Altitude Study: The effect of altitude on the performance of the UniStrip test strip with the OneTouch®Ultra, OneTouch®Ultra2®, OneTouch UltraMini®, OneTouch® UltraSmart® meters was evaluated using venous whole blood samples spiked with glucose to 5 concentrations (approximately 40, 80, 130, 200 and 320 mg/dL). Each meter model was used to test each sample five times at altitudes of 298, 3,608, 7,575 and 10,745 feet above sea level. The results from each measurement were compared to the YSI values for each sample. The results from each measurement were the sponsor's acceptance criteria. These results demonstrate that the performance of the UniStrip test strips with the claimed meters is unaffected by altitudes up to 10,745 feet.
The sponsor has the following limitations in the UniStrip test strip insert:
- UniStrip1 Test Strips may be used at altitudes up to 10,000 feet (3048 meters) without an effect on test results. Accurate results were shown in clinical studies
20
{20}
performed at heights up to 5,280 feet (1609 meters) and in studies simulating heights up to 10,745 feet (3275 meters).
- Patients going through oxygen therapy may yield falsely low results.
6. The sponsor performed a readability assessment of the test strip package insert and the results showed the insert was readable at an 8th grade level.
7. The sponsor evaluated the minimum blood sample size required by testing sample volumes ranging from 0.5 to 2.0 µl on each meter. 5 glucose concentrations were tested ranging from 40~400 mg/dL, as determined by the YSI. Appropriate sample volume was determined if the meter results matched the YSI results. The data shows that sample volume of ≥0.8 µL is the smallest volume to produce acceptable results.
Q. Proposed Labeling:
The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.
R. Conclusion:
The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
21
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.