SyntheCel® Dura Replacement Devices are intended for use as a dura replacement for the repair of dura mater. SyntheCel® Dura Onlay is indicated for use in adults for the repair of dural defects and it can be placed without sutures. SyntheCel® Dura Substitute is indicated for use in adults for the repair of dural defects and it can be sutured into place.
Device Story
SyntheCel® Dura Replacement Devices consist of biosynthesized cellulose and water; non-woven, interconnected fiber construction. Functions as mechanical layer to protect/repair dura mater; prevents CSF leakage. Non-resorbable; immunologically inert. Used in surgical settings by neurosurgeons for dural defect repair. Onlay version placed without sutures; Substitute version sutured into place. Provides physical barrier to support healing and reduce risk of CSF fistula or pseudomeningocele.
Clinical Evidence
Prospective clinical study of 99 patients (62 investigational, 37 control) across 8 sites. Primary endpoint: success defined as absence of CSF fistula and pseudomeningocele within 6 months post-op via radiographic/physical exam. Results showed SyntheCel® non-inferior to control group.
Technological Characteristics
Biosynthesized cellulose and water; non-woven, interconnected fiber structure. Non-resorbable. Biocompatibility per ISO 10993. Endotoxin levels < 0.06 EU/ml per FDA guidance. Mechanical properties include tensile strength, burst strength, and suture pull-out strength comparable to predicate devices.
Indications for Use
Indicated for adults requiring repair of dural defects. SyntheCel® Dura Onlay is for sutureless placement; SyntheCel® Dura Substitute is for sutured placement.
Regulatory Classification
Identification
A dura substitute is a sheet or material that is used to repair the dura mater (the membrane surrounding the brain).
{0}------------------------------------------------
# K113071
JAN - 9 2012
### 1.0 SUMMARY OF INFORMATION REGARDING SAFETY AND EFFECTIVENESS [510(k) Summary] per the requirements 21 CFR 807.92
Page I of 3 Date Prepared January 03, 2012 Submitter SYNTHES 1301 Goshen Parkway West Chester, PA 19380 United States of America Contact Andrea M. Tasker tasker.andrea@synthes.com phone: (610) 719-6920 Trade Name SyntheCel® Dura Replacement Devices, SyntheCel® Dura Onlay, SyntheCel® Dura Substitute Common Name Dura Replacement Device Classification Dura Substitute Name Classification 21 CFR 882.5910 Device Class I I Review Panel Neurology Product Code(s) GXQ Predicate Devices Bio-Vascular's Dura-Guard (K950956, K973706,K982282) Integra Lifescience's DuraGen (K982180) Gore - Preclude MVP (K021477) SyntheCel® Dura Replacement Devices (Onlay and Substitute) are Device Description composed of biosynthesized cellulose and water with a unique construction of non-woven, interconnected cellulose fibers. SyntheCel® Dura Replacement Devices function as a mechanical layer which protects and repairs the dural defect while preventing further CSF leakage. SyntheCel® Dura Replacement Devices are immunologically inert and have demonstrated minimal foreign body response. They are non-resorbable. Intended Use SyntheCel® Dura Replacement Devices are intended for use as a dura replacement for the repair of dura mater. SyntheCel® Dura Onlay is indicated for use in adults for the repair of dural defects and it can be placed without sutures. SyntheCel® Dura Substitute is indicated for use in adults for the repair of dural defects and it can be sutured into place.
{1}------------------------------------------------
# K113071
## 510(k) Summary (continued)
| Technological<br>Characteristics | The SyntheCel® Dura Replacement Devices are similar to the three<br>predicate devices (Bio-Vascular-Dura-Guard, Integra Lifesciences-<br>DuraGen, and Gore-Preclude MVP) in terms of intended use. All three<br>are intended for dura replacement for the repair of dura mater.<br>The SyntheCel® Dura Replacement Devices are similar to the two<br>predicate devices (Bio-Vascular-Dura-Guard, Integra Lifesciences-<br>DuraGen) in terms of physical properties (form, size, physical integrity,<br>thickness, suturability, drapability, conformability).<br>The SyntheCel® Dura Replacement Devices are similar to the predicate<br>device (Gore-Preclude MVP) in terms of resorbability. Both are non-<br>resorbable. |
|----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Clinical Testing<br>Data | Clinical data were collected to evaluate the safety and effectiveness of<br>the SyntheCel® Dura Replacement Devices as compared to the Control.<br>The objective of this study was to demonstrate that the SyntheCel® Dura<br>Replacement Devices are substantially equivalent to other dura<br>replacement products that were previously cleared for marketing by<br>FDA. A total of 99 patients were enrolled, randomized and treated (62<br>in the investigational SyntheCel® treatment group and 37 in the control<br>group) at 8 clinical sites.<br>Patients were evaluated intra-operatively and immediately post-<br>operatively (1 - 10 days) followed by evaluations at one (1) month, three<br>(3) months and six (6) months. Complications and adverse events,<br>device-related or not, were evaluated over the course of the clinical trial.<br>At each evaluation time-point, the primary and secondary clinical<br>outcome parameters were evaluated. Safety and effectiveness was assessed in all randomized subjects.<br>Primary Efficacy Endpoint: The primary endpoint was based on the<br>findings up to and including the 6 month follow-up visit. An individual<br>patient's treatment was considered successful if and only if the following<br>criterion was met:<br>Success was defined as absence of CSF fistula (drainage from wound or<br>sinus) and pseudomeningocele within 6 months post-operatively<br>confirmed by radiographic evaluation and physical examination of<br>surgical site.<br>The results of overall success analyses indicate that the SyntheCel® is<br>non-inferior to the dura replacement control group. |
| Non-Clinical<br>Testing Data | Mechanical testing was performed for tensile strength, burst strength,<br>and suture pull-out strength on samples of SyntheCel® Dura<br>Replacement Devices and the predicates, DuraGen and Dura-Guard. All<br>samples were prepared per respective instructions. SyntheCel® Dura<br>Replacement Device performance was found to be comparable in tensile<br>strength, burst strength, and suture pull-out strength.<br><br>An animal study was performed comparing the SyntheCel® Dura<br>Replacement Devices and predicate devices in a Rabbit Durotomy<br>Implant Study. The results noted at necropsy show no meningitis, CSF<br>leakage, evidence of infection, hydrocephalus, device vascularization,<br>nor significant hemorrhage in any of the animals regardless of the<br>implant type. Overall, the SyntheCel® Dura Substitute performed no<br>different than the other predicate dura substitute materials.<br><br>An additional Rabbit Durotomy Implant Study Model was performed<br>using SyntheCel® Dura Replacement Devices. This study was<br>conducted by NAMSA following GLPs. According to the conclusion,<br>the SyntheCel® Dura Replacement Devices showed no evidence of a<br>toxic leachable or local adverse effect.<br><br>Biocompatibility testing according to standards set forth in ISO 10993<br>demonstrated that the material is non-irritating, non-sensitizing, non-<br>mutagenic, non-cytotoxic, non-hemolytic, and non-pyrogenic.<br><br>Pyrogenicity was evaluated using the Limulus Amebocyte Lysate (LAL)<br>test on the final sterilized SyntheCel® device and found to be less than<br>0.06 EU/ml (Endotoxin Units / ml) per FDA's Guidance Document for<br>Dura Substitute Devices. All SyntheCel® Dura Replacement Device<br>lots will be tested to ensure they are less than 0.06 EU/ml and will be<br>labeled non-pyrogenic. |
| Substantial<br>Equivalence to<br>Predicate Devices | Based on the information presented in this submission, which includes<br>mechanical testing, pre-clinical testing, animal testing, and clinical<br>testing, the SyntheCel® Dura Replacement Devices do not raise new<br>questions of safety and effectiveness. Therefore, it can be concluded<br>that the SyntheCel® Dura Replacement Devices are substantially<br>equivalent to the predicate devices for dura replacement devices in<br>commercial distribution for use as a dura replacement for the repair of<br>dura mater. |
·
{2}------------------------------------------------
510(k) Summary (continued)
### 3 Page 3
(end of summary)
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002
JAN - 9 2012
Synthes c/o Ms. Andrea M. Tasker CMF Regulatory Affairs Manager, Synthes 1301 Goshen Parkway West Chester, PA 19380
Re: K113071
Trade/Device Name: SyntheCel® Dura Replacement Devices Regulation Number: 21 CFR 882.5910 Regulation Name: Dura Substitute Regulatory Class: Class II Product Code: GXO Dated: October 14, 2011 Received: October 18, 2011
Dear Ms. Tasker:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical Of K Part 607), accimig (21 CFR 803); good manufacturing practice requirements as set do records and the overse oversion (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please n you http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part note the regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Kesia Alexander
Image /page/4/Picture/6 description: The image shows a handwritten signature. The signature is in cursive and appears to be a single word. The first letter is a tall loop, and the rest of the letters are connected in a flowing manner. The signature is in black ink on a white background.
Malvina B. Eydelman, M.D. Director Division of Ophthalmic, Neurological, and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "SYNTHES" in bold, black letters. To the left of the word is a circular logo with a design inside. A horizontal line runs beneath the word "SYNTHES" and the logo. A registered trademark symbol is located to the right of the word.
## Indications for Use
K113071 510(k) Number (if known):_
SyntheCel® Dura Replacement Devices Device Name:
Indications For Use:
SyntheCel® Dura Replacement Devices are intended for use as a dura replacement for the repair of dura mater. SyntheCel® Dura Onlay is indicated for use in adults for the repair of dural defects and it can be placed without sutures. SyntheCel® Dura Substitute is indicated for use in adults for the repair of dural defects and it can be sutured into place.
Prescription Use X (Per 21 CFR 801.109)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Sign-Off)
ision of Ophthalmic, Neurological and Ear,
Nose and Throat Devices
K11307 510(k) Number
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.