The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components. General total knee arthroplasty indications include: Painful, disabling joint disease of the knee resulting from: degenerative arthritis, rheumatoid arthritis or post-traumatic arthritis Post-traumatic loss of knee joint configuration and function Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament The Unity Total Knee System is indicated for cemented, single use only.
Device Story
Unity Total Knee System is a fixed-bearing total knee replacement prosthesis. System includes femoral component, tibial insert, tibial tray, and optional all-polyethylene patellar component. Available in cruciate retaining (CR) and posterior stabilized (PS) variants. Used by orthopedic surgeons in clinical settings to replace damaged knee joint articulation. Provides increased mobility and pain reduction for patients with joint disease or trauma. PS variant provides increased anterior-posterior constraint for patients with PCL deficiency. System includes augment components like femoral/tibial wedges and stem extensions. Implanted via surgical procedure; cemented fixation.
Clinical Evidence
No clinical data. Substantial equivalence determined via non-clinical bench testing in accordance with ASTM F2083, including fatigue, endurance, deformation, contact area/pressure, range of motion, constraint, wear testing of UHMWPE inserts, and mechanical integrity of connecting mechanisms.
Technological Characteristics
Fixed-bearing total knee prosthesis. Materials: metal/polymer. Components: femoral, tibial tray, tibial insert, patellar. Variants: CR and PS. Cemented fixation. Testing per ASTM F2083 (fatigue, endurance, deformation, contact mechanics, constraint).
Indications for Use
Indicated for skeletally mature patients requiring total knee arthroplasty due to degenerative, rheumatoid, or post-traumatic arthritis; post-traumatic loss of function; moderate varus, valgus, or flexion deformity; revision of failed knee replacement; or distal femur/proximal tibia fractures. Posterior stabilized variant indicated for PCL instability or absent/non-functioning PCL. Cemented, single-use only.
Regulatory Classification
Identification
A knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis is a device intended to be implanted to replace a knee joint. The device limits translation and rotation in one or more planes via the geometry of its articulating surfaces. It has no linkage across-the-joint. This generic type of device includes prostheses that have a femoral component made of alloys, such as cobalt-chromium-molybdenum, and a tibial component or components and a retropatellar resurfacing component made of ultra-high molecular weight polyethylene. This generic type of device is limited to those prostheses intended for use with bone cement (§ 888.3027).
{1}------------------------------------------------
# K113060
| 3. 510(K) SUMMARY | |
|--------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| 1. Applicant/Sponsor: | Corin USA<br>10500 University Center Drive<br>Suite 190<br>Tampa, Florida 33612<br>Establishment Registration No.: 1056629 |
| 2. Contact Person: | Lucinda Gerber<br>Regulatory Affairs Associate<br>Corin USA<br>813-977-4469<br>lucinda.gerber@coringroup.com |
| 3. Date: | October 13, 2011 |
| 4. Proprietary Name: | Unity™ Total Knee System |
| 5. Common Name: | Knee Prosthesis |
| 6. Classification Name: | Knee joint patella femorotibial polymer/metal/polymer semi- constrained<br>cemented prosthesis (21CFR888.3560) |
| 7. Legally Marketed Devices to which Substantial Equivalence is claimed: | |
| | • Triathlon PS TKA System (K031729) |
| | • Triathlon CR TKA System (K040267) |
| | • Triathlon CR sz 1 Tibial Insert (K042883) |
| | • Triathlon PS sz 1 Tibial Insert (K050539) |
| | • Scorpio CR TKA System (K974556) |
| | • Scorpio PS TKA System (K962152) |
| | • Scorpio NRG System (K042343) |
| | • Scorpio TS (K994128) |
| 8. Device Description: | |
The Corin Unity Knee System is a fixed bearing total knee replacement system that consists of a femoral component, tibial insert, tibial tray and all-polyethylene patellar component for
{2}------------------------------------------------
use in primary total knee arthroplasty. The Unity Knee System is provided in two variants, cruciate retaining (CR) and posterior stabilized (PS). The Unity CR Total Knee System is intended to accommodate the posterior cruciate ligament (PCL) if it is present. The Unity posterior stabilized variant is utilized when total knee replacement is indicated, and the posterior cruciate ligament is non-functioning or absent, resulting in joint instability. The Unity system patellar component is optional and is used in situations where replacement of the articular surface of the patella is required. The system also provides augment components including femoral wedges, tibial wedges, a short keel extension and stem extensions with offset connections.
#### 9. Intended Use / Indications:
The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components.
General total knee arthroplasty indications include:
- Painful, disabling joint disease of the knee resulting from: degenerative arthritis, . rheumatoid arthritis or post-traumatic arthritis
- Post-traumatic loss of knee joint configuration and function .
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be ● returned to adequate function
- . Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate
- Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard ● fracture management techniques
- The posterior stabilized variant is also indicated for PCL instability requiring implant . bearing surface geometries with increased anterior-posterior constraint and absent or non-functioning posterior cruciate ligament
The Unity Total Knee System is indicated for cemented, single use only.
#### 10. Summary of Technologies/Substantial Equivalence:
The Unity Total Knee System has the same intended use and indications and is manufactured from the same materials as the respective cruciate retaining and posterior stabilized predicates Triathlon (CR: K040267, K042883; PS: K031729, K050539) and Scorpio (CR: K974556; PS: K962152, K042343) Total Knee Systems. The range of sizes
{3}------------------------------------------------
#### K113060
available for the Unity Total Knee System falls within the range cleared for the Triathlon and Scorpio devices and the Unity Total Knee System is similar to the predicate devices in regards to design. Based on these similarities, Corin believes that the Unity Total Knee System is substantially equivalent to the predicate devices.
### 11. Non-Clinical Testing:
Non-clinical testing and analysis are provided, including bench testing for both the cruciate retaining and the posterior stabilized designs. Testing was completed in accordance with ASTM F2083 Standard Specification for Total Knee Prosthesis and included component function for femoral, tibial tray and tibial insert (fatigue, endurance and deformation, contact area and contact pressure, range of motion, constraint). Testing also included wear testing of the UHMWPE tibial insert and integrity of connecting mechanisms for static and dynamic shear tests, bending tests, mechanical stop and disassociation. The results of this testing show that the Unity Total Knee System is expected to be safe and effective for the proposed indications and is substantially equivalent to the predicate devices.
## 12. Clinical Testing:
Clinical testing was not necessary to determine substantial equivalence between the Unity Total Knee System and the predicate devices.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the symbol in a circular fashion. The logo is presented in black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 24, 2014
Corin, USA Ms. Lucinda Gerber, BA Regulatory Affairs Associate 10500 University Center Drive, Suite 190 Tampa, Florida 33612
Re: K113060 Trade/Device Name: Unity™ Total Knee System Regulation Number: 21 CFR 888.3560 Regulation Name: Knee joint patellofemorotibial polymer/metal/polymer semi-constrained cemented prosthesis Regulatory Class: Class II Product Code: JWH Dated: August 28, 2012 Received: August 29, 2012
Dear Ms. Gerber:
This letter corrects our substantially equivalent letter of September 10, 2012.
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must
{5}------------------------------------------------
# Page 2 - Ms. Lucinda Gerber
comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set ‹forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely vours.
Lori A. Wiggins
for Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### 2. INDICATIONS FOR USE
# 510(k) Number (if known): K113060
Device Name: Unity™ Total Knee System
# Indications for Use:
The Unity Total Knee System is intended for use in total knee arthroplasty in skeletally mature patients, to provide increased mobility and reduce pain by replacing the damaged knee joint articulation where there is evidence of sufficient sound bone to seat and support the components.
General total knee arthroplasty indications include:
- . Painful, disabling joint disease of the knee resulting from: degenerative arthritis, theurnatoid arthritis or post-traumatic arthritis
- Post-traumatic loss of knee joint configuration and function ●
- Moderate varus, valgus, or flexion deformity in which the ligamentous structures can be returned to adequate function
- . Revision of previous unsuccessful knee replacement or other procedure, where soft tissue stability is adequate
- . Fracture of the distal femur and/or proximal tibia that cannot be stabilized by standard fracture management techniques
- . The posterior stabilized variant is also indicated for PCL instability requiring implant bearing surface geometries with increased anterior-posterior and absent or non-functioning posterior cruciate ligament
The Unity Total Knee System is indicated for cemented, single use only.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
# (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Elizabeth L. Frank -S
Page _ 1 _ of _ 1
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.