K113059 · B.Braun Medical, Inc. · CAZ · Jan 24, 2012 · Anesthesiology
Device Facts
Record ID
K113059
Device Name
CONTIPLEX FX
Applicant
B.Braun Medical, Inc.
Product Code
CAZ · Anesthesiology
Decision Date
Jan 24, 2012
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 868.5140
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The B. Braun Contiplex® FX Continuous Nerve Block Set is intended to provide continuous and/or intermittent infusion of local anesthetics and analgesics for peripheral plexus anesthesia and pain management. The Contiplex FX catheter may remain indwelling for up to 72 hours.
Device Story
Contiplex FX Continuous Nerve Block Set facilitates continuous/intermittent delivery of local anesthetics/analgesics for peripheral plexus anesthesia and pain management. Components include 17G Tuohy needle (uncoated, non-stimulating), 19G springwound polyamide catheter with threading assist guide, sideport valve assembly, and clamp-style catheter connector. Used in clinical settings by healthcare professionals. Catheter indwelling time up to 72 hours. Device provides access for anesthetic infusion to manage patient pain; performance verified via biocompatibility and functional testing.
Clinical Evidence
Bench testing only. Performance evaluated against ISO 9626, ISO 7864, ISO 594-1, ISO 594-2, and EN 1618. Biocompatibility testing performed per ISO 10993-1. Results demonstrate device meets applicable standards and is safe/effective for intended use.
Technological Characteristics
Components: 17G Tuohy needle (stainless steel, ISO 9626/7864), 19G springwound polyamide catheter (EN 1618), sideport valve, clamp connector. Luer taper fittings (ISO 594-1/2). Non-stimulating, uncoated needle. No software or electronic components.
Indications for Use
Indicated for patients requiring continuous or intermittent infusion of local anesthetics and analgesics for peripheral plexus anesthesia and pain management. Catheter may remain indwelling for up to 72 hours.
Regulatory Classification
Identification
An anesthesia conduction kit is a device used to administer to a patient conduction, regional, or local anesthesia. The device may contain syringes, needles, and drugs.
{0}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Contiplex FX Continuous Nerve Block Set December 19, 2011
#### ડ. 510(k) SUMMARY
SUBMITTER:
B. Braun Medical Inc. 901 Marcon Boulevard Allentown, PA 18109-9341 610-266-0500
Contact: Lisa Giaquinto, Specialist, Regulatory Affairs Phone: (610) 596-2354 Fax: (610) 266-4962 E-mail: lisa.giaquinto@bbraun.com
DEVICE NAME:
Contiplex® FX Continuous Nerve Block Set
COMMON OR USUAL NAME:
SUBSTANTIAL
EQUIVALENCE:
Anesthesia Conduction Kit
# DEVICE CLASSIFICATION:
Contiplex® Continuous Nerve Block Set, Braun Medical Inc., PREDICATE DEVICES: K090995, Class II, CAZ, 868.5140.
Class II, Product Code CAZ, 868.5140
DESCRIPTION: The B. Braun Continuous Nerve Block Set consists of one 17 gauge Tuohy needle, one 19 gauge springwound catheter with threading assist guide, one sideport valve assembly, and one clamp style catheter connector. The set is used to facilitate the continuous delivery of anesthetics or analgesics to the patient for pain management during regional anesthesia procedures.
INTENDED USE: The B. Braun Continlex® FX Continuous Nerve Block Set is intended to provide continuous and/or intermittent infusion of local anesthetics and analgesics for peripheral plexus anesthesia and pain management. The Contiplex FX catheter may remain indwelling for up to 72 hours.
> The Contiplex® FX Continuous Nerve Block Set has the same intended use and contains similar components as the Contiplex Continuous Nerve Block Set cleared under premarket notification # K090995. Both sets include a Tuohy needle, catheter with threading assist guide, sideport valve assembly, and clamp style catheter connector. The primary differences between the proposed Contiplex FX Continuous Nerve Block Set and the predicate Contiplex
# JAN 2 4 2012
K 113059
{1}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Contiplex FX Continuous Nerve Block Set December 19, 2011
> Continuous Nerve Block Set relate to the Tuohy needle and catheter. The predicate device contains a coated, stimulating Tuohy needle, while the proposed device contains an uncoated, nonstimulating Tuohy needle. Additionally, the predicate device also includes a polyamide catheter, while the proposed device includes a polyamide catheter with inner springwound coil.
> Biocompatibility and functional performance testing have been completed with the proposed Tuohy needle and springwound catheter to verify there are no differences between the proposed and predicate device, which would raise new issues of safety or effectiveness.
#### Performance Testing
The following performance standards have been utilized in the evaluation of the Tuohy needle included in the proposed Contiplex FX Continuous Nerve Block Set:
ISO 9626: "Stainless steel needle tubing for the manufacture of medical devices. "
ISO 7864: "Sterile hypodermic needles for single use."
ISO 594-1: "Conical Fittings with a 6 % Luer taper for syringes, needles and certain other medical equipment - Part I: General Requirements. "
ISO 594-2: "Conical fittings with 6% (Luer) taper for syringes, needles and certain other medical equipment- Part 2: Lock fittings. "
The following performance standards has been utilized in the evaluation of the springwound catheter included in the proposed Contiplex FX Continuous Nerve Block Set:
EN 1618: "Catheters Other Than Intravascular Catheters - Test Methods for Common Properties."
Results of performance testing indicate that the needle and catheter in the Contiplex FX Continuous Nerve Block Set meet applicable sections of the standards referenced and are safe and effective for their intended use.
{2}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Contiplex FX Continuous Nerve Block Set December 19, 2011
#### Biocompatibility
Biocompatibility testing was performed based on the nature and duration of patient contact outlined in ISO 10993-1: "Biological evaluation of medical devices -- Part I: Evaluation and testing within a risk management process." Results of testing demonstrates that the materials used in the construction of the needle and catheter in the proposed Contiplex FX Continuous Nerve Block Set are safe for their intended use.
#### Conclusion
Based on the results of performance and biocompatibility testing, the proposed nerve block set is considered similar to the predicate device identified above, and is safe and effective for it's intended use.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image is a circular seal for the Department of Health & Human Services USA. The seal features the department's name in a circular arrangement around the edge. In the center of the seal is an abstract image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room-WO66-G609 Silver Spring, MD 20993-0002
Ms. Lisa Giaquinto Regulatory Affairs Specialist B. Braun Medical, Incorporated 901 Marcon Boulevard Allentown, Pennsylvania 18109
JAN 2 4 2012
Re: K113059
Trade/Device Name: Contiplex FX Continuous Nerve Block Set Regulation Number: 21 CFR 868.5140 Regulation Name: Anesthesia Conduction Kit Regulatory Class: II Product Code: CAZ Dated: December 20, 2011 Received: December 21, 2011
Dear Ms. Giaquinto:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Giaquinto
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Wh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{5}------------------------------------------------
B. Braun Medical Inc. 510(k) Premarket Notification Contiplex FX Continuous Nerve Block Set
### 4. INDICATIONS FOR USE STATEMENT
Page _________________________________________________________________________________________________________________________________________________________________________
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
#### Device Names:
Contiplex FX Continuous Nerve Block Set
## Indications For Use:
The B. Braun Contiplex® FX Continuous Nerve Block Set is intended to provide continuous and/or intermittent infusion of local anesthetics and analgesics for peripheral plexus anesthesia and pain management. The Contiplex FX catheter may remain indwelling for up to 72 hours.
Prescription Use X (Per 21 CFR 801.109)
OR
Over-The-Counter Use - - -
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
L. Shullins
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
K113059 510(k) Number:
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.