The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician.
Device Story
Arkon Anesthesia Workstation integrates pneumatic circuitry, controls, monitoring, and gas delivery systems; manages oxygen, air, and nitrous oxide distribution; facilitates anesthetic vapor delivery via dismountable vaporizer. Used in hospital operating rooms by qualified physicians. Device provides controlled gas/vapor delivery to patient breathing circuits; supports mechanical ventilation. Healthcare providers monitor system output to manage patient anesthesia and respiratory support during surgical procedures. Benefits include precise control of anesthetic delivery and ventilation, supporting patient stability during surgery.
Clinical Evidence
Bench testing only. No clinical data provided. Performance verified against standards including IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), ASTM F1101-90 (ventilator performance), and IEC 60601-2-13 (anesthetic systems). Usability validated per IEC 60601-1-6 and IEC 62366. Software verified and validated per IEC 62304 and FDA guidance.
Technological Characteristics
Anesthesia workstation with integrated pneumatic circuitry, gas mixing, and mechanical ventilation. Materials verified per ISO 10993-1. Connectivity includes standard medical gas interfaces (CGA V-5). Complies with IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), and IEC 60601-2-13 (anesthetic systems). Software developed per IEC 62304. Includes alarm systems per IEC 60601-1-8.
Indications for Use
Indicated for use in hospital operating rooms by qualified physicians for the delivery of oxygen, air, nitrous oxide, and anesthetic vapors to patients via breathing circuits, with or without mechanical ventilation.
Regulatory Classification
Identification
A continuous ventilator (respirator) is a device intended to mechanically control or assist patient breathing by delivering a predetermined percentage of oxygen in the breathing gas. Adult, pediatric, and neonatal ventilators are included in this generic type of device.
{0}------------------------------------------------
K113051
## MAR 2 2 2012
:
# 510(k) Summary
| Submission Date: | 11 October 2011 |
|-------------------------------|-------------------------------------------------------------------------------------------------------------------|
| Submitter: | Spacelabs Healthcare Ltd.<br>1 Harforde Court, John Tate Road<br>Hertford, SG13 7NW United Kingdom |
| Submitter Contact: | Mr. David J. Geraghty<br>Spacelabs Medical, Inc. (for Spacelabs Healthcare)<br>Phone: +1 (425) 657-7200, ext 5889 |
| | Fax: +1 (425) 657-7210<br>Email: david.geraghty@spacelabs.com |
| Official<br>Correspondent: | Thomas Kroenke<br>Principal Consultant<br>Speed To Market, Inc. |
| | PO Box 3018<br>Nederland, CO 80466 USA<br>tkroenke@speedtomarket.net<br>+1 (303) 956 4232 |
| Manufacturing Site: | Spacelabs Medical, Inc.<br>5150 220th Avenue SE<br>Issaquah, WA 98029 USA |
| Trade Name: | Spacelabs Arkon Anesthesia Workstation |
| Common Name: | Anesthesia workstation with continuous ventilator |
| Classification Name: | Ventilator, continuous, facility use;<br>Gas-machine, anesthesia |
| Classification<br>Regulation: | 21 CFR §868.5895;<br>21 CFR §868.5160 |
| Primary Product<br>Code: | CBK |
| Secondary Product<br>Code: | BSZ |
---
{1}------------------------------------------------
### 510(k) Summary
| Substantially<br>Equivalent Devices: | New Spacelabs Model | Predicate<br>510(k) Number | Predicate<br>Manufacturer / Model |
|--------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|--------------------------------------------------|
| | Spacelabs Arkon<br>Anesthesia Workstation | K101850 | Spacelabs BleaseSirius<br>Anesthesia Workstation |
| Device Description: | The Spacelabs Arkon Anesthesia Workstation (Arkon) is an anesthesia<br>workstation that contains all the pneumatic circuitry, controls,<br>monitoring, ancillaries and storage required to control, distribute and<br>mix medical gases and anesthetic agents in order to deliver them to a<br>patient system. It is capable of delivering oxygen, air and nitrous oxide<br>in a controlled manner to various patient breathing circuits with or<br>without the use of a mechanical ventilator, and may be used for the<br>delivery of anesthetic vapor by use of a dismountable vaporizer. | | |
| Intended Use: | The Spacelabs Arkon Anesthesia Workstation is intended for use in the<br>hospital environment and operating room. It may be used for the<br>delivery of oxygen, air and nitrous oxide in a controlled manner to<br>various patient breathing circuits with or without the use of mechanical<br>ventilator, and may be used for the delivery of anesthetic vapor by use<br>of a dismountable vaporizer. | | |
| | The device is intended for use only by a suitably qualified physician. | | |
| Technology<br>Comparison: | The Spacelabs Arkon Anesthesia Workstation employs the same<br>technological characteristics as the predicate device.fs | | |
| Summary of Performance Testing: | | | |
| Biocompatibility | Patient contact materials within the Arkon and its accessories were | | |
verified in accordance with the following standard: Verification ISO 10993-1: 2009, Biological evaluation of medical devices – Part .
- 1 : Evaluation and testing within a risk management process.
Verification results indicated that the materials comply with the standard.
{2}------------------------------------------------
# 510(k) Summary
| Software Testing | The Arkon software was designed and developed according to a robust<br>software development process, and was rigorously verified and<br>validated. Software information is provided in accordance with internal<br>requirements and the following standards and guidance documents:<br>FDA guidance: The content of premarket submissions for software<br>contained in medical devices, 11 May 05; FDA guidance: Off-the-shelf software use in medical devices, 09<br>Sep 99; and FDA guidance: General principles of software validation; Final<br>guidance for industry and FDA staff, 11 Jan 02. IEC 62304: 2006, Medical device software – Software life cycle<br>processes. ISO 14971: 2007, Medical devices – Application of risk<br>management to medical devices Test results indicate that the Arkon complies with its predetermined<br>specifications and the applicable standards and guidance documents. |
|---------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Electrical Safety<br>Testing | The Arkon was tested for patient safety in accordance with the<br>following standards:<br>IEC 60601-1: 1988, Am1: 1991, Am2: 1995, Medical Electrical<br>Equipment, Part 1: General Requirements for Safety; and UL 60601-1: 2006, Medical Electrical Equipment, Part 1:<br>Particular Requirements for Safety. Test results indicated that the Arkon complies with the applicable<br>Standards. |
| Electromagnetic<br>Compatibility<br>Testing | The Arkon was tested for EMC in accordance with the following<br>standard:<br>IEC 60601-1-2: 2007, Medical Electrical Equipment, Part 1: Part<br>1-2: General Requirements for Safety – Collateral Standard:<br>Electromagnetic Compatibility-Requirements and Tests. Test results indicated that the Arkon complies with the applicable<br>Standard. |
.
・
{3}------------------------------------------------
### 510(k) Summary
#### The Arkon was tested for performance in accordance with internal Performance Testing requirements and the following standards:
- IEC 60601-1-6: 2004, Medical electrical equipment Part 1-6: . General requirements for basic safety and essential performance -Collateral standard: Usability.
- IEC 60601-1-8: 2006, Medical electrical equipment General . requirements, tests and guidance for alarm systems in medical electrical equipment and medical electrical systems.
- ASTM F1101-90 (2003) e1, Standard Specification for ventilators . intended for use during anesthesia.
- CGA C-9: 2004, Standard Color Marking of Compressed Gas . Containers for Medical Use.
- CGA V-5: 2008, Diameter Index Safety System . (Noninterchangeable Low Pressure Connections for Medical Gas Applications).
- IEC 60601-2-13: 2003, Am1: 2006, Medical electrical equipment -. Particular requirements for the safety and performance of anesthetic systems.
- IEC 62366: 2007, Medical devices Application of usability . engineering to medical devices.
- ISO 5356-1: 2004, Anaesthetic and respiratory equipment Conical . connectors - Part I: Cones and sockets.
- ISTA Procedure IB, Non-simulation integrity performance test . procedure - Packaged-products over 150 lb (68 kg).
Test results indicate that the Arkon complies with its predetermined specifications and the applicable standards.
Verification and validation activities were conducted to establish the Conclusion performance and safety characteristics of the Arkon. The results of these activities demonstrate that the Arkon is safe and effective when used in accordance with its intended use and labeling.
> Therefore, the Arkon is considered substantially equivalent to the predicate device.
{4}------------------------------------------------
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle or bird-like figure with three wing-like shapes.
#### Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring. MD 20993-0002
MAR 2 2 2012 .
Spacelabs Healthcare, LTD C/O Mr. Thomas Kroenke Principal Consultant Speed To Market, Inc. P.O. Box 3018 Nederland, Colorado 80466
Re: K113051
Trade/Device Name: Spacelabs Arkon Anesthesia Workstation Regulation Number: 21 CFR 868.5895 Regulation Name: Continuous Ventilator Regulatory Class: II Product Code: CBK, BSZ Dated: February 23, 2012 Received: February 24, 2012
Dear Mr. Kroenke:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 - Mr. Kroënke
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices /ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
Nh for
Anthony D. Watson, B.S., M.S., M.B.A. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Indications for Use
K
510(k) Number (if known):
Device Name:
Indications for Use:
Spacelabs Arkon Anesthesia Workstation
The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer.
The device is intended for use only by a suitably qualified physician.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
for. L. Schutheis
(Division Sign-Off)
Division of Anesthesiology, General Hospital Infection Control, Dental Devices
510(k) Number. K11/3
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.