← Product Code [FDF](/productcode/FDF) · K113050

# CLEARPATH LOWER GI (K113050)

_Easyglide , Ltd. · FDF · May 16, 2012 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K113050

## Device Facts

- **Applicant:** Easyglide , Ltd.
- **Product Code:** [FDF](/productcode/FDF.md)
- **Decision Date:** May 16, 2012
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The ClearPath Lower GI is intended to connect to standard colonoscopes to help facilitate intraprocedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces, and other bodily fluids and matter, e.g. blood. It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

## Device Story

ClearPath Lower GI is an irrigation/evacuation accessory for standard colonoscopes; comprises suction tube, irrigation tube, and attachment sleeve. Device connects to ClearPath Controller and Tubing (K091305). During colonoscopy, physician operates system to irrigate colon and evacuate debris (feces, blood, irrigation fluid) without removing tools from endoscope working channel. Attachment ring allows compatibility with various endoscope sizes. Used in hospitals, clinics, and doctors' offices by trained medical personnel. Facilitates visualization and cleaning of poorly prepared colons, potentially improving procedure efficiency and diagnostic quality.

## Clinical Evidence

Bench and animal testing only. Bench testing verified dimensional, strength, functional, and mechanical performance, and compatibility with standard endoscopes and ClearPath Controller/Tubing. Animal testing evaluated tip dislodgement, maneuverability, visibility, irrigation quality, suction continuity, and evacuation efficacy. Results demonstrated performance met specifications and was equivalent to predicate.

## Technological Characteristics

Single-use, non-sterile disposable accessory. Materials comply with ISO 10993-1. Distal tip features multi-irrigation holes and a suction hole. Attaches to endoscope via sleeve/ring. Operates in conjunction with external ClearPath Controller (K091305) for fluid management. No software or electronic components.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

## Predicate Devices

- EasyGlide - ClearPath Lower GI ([K091305](/device/K091305.md))
- EasyGlide - ClearPath Upper GI ([K093779](/device/K093779.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Premarket Notification 510(k) Section 5 - Device Description ClearPath Lower GI

| 510(k) Summary |
|----------------|
| Page 1 of 5    |
| 11-Oct-11      |

| EasyGlide Ltd.<br>30 Ha'Ella St.<br>Kfar Truman, 73150<br>Israel | Tel - 011-972-528565644<br>Fax - 011-972-776201003                                   |
|------------------------------------------------------------------|--------------------------------------------------------------------------------------|
| Official Contact:                                                | Izhak Fabian - CEO                                                                   |
| Proprietary or Trade Name:                                       | ClearPath Lower GI                                                                   |
| Common/Usual Name:                                               | Irrigation/evacuation system                                                         |
| Classification Name/Code:                                        | FDF - endoscope and accessories<br>21CFR 876.1500<br>Class 2                         |
| Device:                                                          | ClearPath Lower GI                                                                   |
| Predicate Devices:                                               | K091305 - EasyGlide - ClearPath Lower GI<br>K093779 - EasyGlide - ClearPath Upper GI |

### Device Description:

The proposed modified ClearPath Lower Gl is one component of the predicate ClearPath (K091305) which included a number of components:

- Controller .
- Tubing Set .
- . Irrigation set (now referred to as the ClearPath Lower GI)

We are proposing to only modify the ClearPath Lower GI which attaches to a conventional endoscope.

The ClearPath Lower GI includes one suction tube + tip, one irrigation tube + tip, and a sleeve to attach to the endoscope. The ClearPath Lower GI is connected to the ClearPath Controller and the ClearPath Tubing (K091305) and allows for irrigation and evacuation of debris from the colon during an endoscopic procedure. The device can accommodate several sizes and configurations of endoscopes with a change in the attachment ring.

### Indications for Use:

The ClearPath Lower Gl is intended to connect to standard colonoscopes to help facilitate intraprocedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces, and other bodily fluids and matter, e.g. blood.

{1}------------------------------------------------

### ClearPath Lower Gl

### 510(k) Summary Page 2 of 5

11-Oct-11

It is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

### Patient population:

Individuals undergoing endoscopic procedures.

### Environment of Use:

Hospitals, clinics, and doctors' offices.

### Performance Testing and Summary:

We have performed bench and animal testing to verify that the ClearPath Lower GI performs as expected in conjunction with standard endoscopes and with the cleared ClearPath controller and tubing K0910305. The tests included:

- Dimensional testing o
- o Strength testing
- Functional testing O
- o Mechanical testing
- Compliance with endoscopes o

The animal testing evaluated:

- Any sign of dislodgement of the tip from the endoscope during the o procedure
- Ease of maneuverability and advance in the colon o
- Quality of visibility o
- Quality of irrigation O
- Any sign of occlusion or interruption to continuous suction. 0
- Quality of evacuation of feces and other bodily fluid and matter o

All testing demonstrated that the modified ClearPath Lower GI disposable performed to its specifications and / or was equivalent to the predicate.

### Summarv of substantial equivalence:

We demonstrate that the modified ClearPath Lower GI is substantially equivalent to the predicates in design and performance characteristics:

- Indications Similar to predicate K091305: Cleaning of a poorly prepared colon . by irrigation and evacuation ·

{2}------------------------------------------------

### ClearPath Lower Gl

## 510(k) Summary Page 3 of 5 11-Oct-11

- Technology Identical to predicate K091305: Single use disposable, used with a . standard endoscope, works in conjunction with the ClearPath Controller (K091305) to facilitate cleaning of the colon by irrigation and evacuation.
- Environment of use Identical to predicate K091305: Hospitals, clinics, and . doctors' offices.
- . Materials - Materials were either identical to the predicate or shown to comply with ISO 10993-1.
- . Difference - there are no substantial differences or new features in the proposed device compared to the predicates which raises any new safety or efficacy issue

{3}------------------------------------------------

Premarket Notification 510(k

.

ClearPath Lower Gl

Summary

|                           | Proposed device                                                                                                                                                                                                                                                                                                                                                                                                                                         | Predicate devices                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                         | Proposed device   | Predicate devices                                                                                                                                                             |                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                     |
|---------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Device                    | EasyGlide<br>ClearPath Lower GI                                                                                                                                                                                                                                                                                                                                                                                                                         | K091305<br>EasyGlide<br>ClearPath Lower GI                                                                                                                                                                                                                                                                                                                                                  | K093779<br>EasyGlide<br>ClearPath Upper GI                                                                                                                                                                                                                                                                                                                              | Device            | EasyGlide<br>ClearPath Lower GI                                                                                                                                               | K091305<br>EasyGlide<br>ClearPath Lower GI                                                                                                                                                                                                                                                              | K093779<br>EasyGlide<br>ClearPath Upper GI                                                                                                                                          |
| Design                    | Used as an add-on to standard endoscopes for<br>irrigation and evacuation. Attaches along the<br>endoscope leaving the working channel free.<br>The ClearPath Lower GI is intended to<br>connect to standard colonoscopes to help<br>facilitate intraprocedural cleaning of a poorly<br>prepared colon by irrigating or cleaning the<br>colon and evacuating the irrigation fluid<br>(water), feces, and other bodily fluids and<br>matter, e.g. blood. | Used as an add-on to standard endoscopes for<br>irrigation and evacuation. Attaches along the<br>endoscope leaving the working channel free.<br>The ClearPath is intended to connect to standard<br>colonoscopes to help facilitate intraprocedural<br>cleaning of a poorly prepared colon by irrigating<br>or cleaning the colon and evacuating the irrigation<br>fluid (water) and feces. | Used as an add-on to standard endoscopes for<br>irrigation and evacuation. Attaches along the<br>endoscope leaving the working channel free.<br>The ClearPath-UGI System is intended for<br>irrigating or cleaning the upper digestive tract<br>and evacuating the irrigation fluid, blood and<br>bile in the upper GI tract during endoscopic<br>procedures..          | Accessories       | ClearPath Controller and ClearPath Tubing<br>(K091305)                                                                                                                        | ClearPath Controller and ClearPath Tubing<br>(K091305)                                                                                                                                                                                                                                                  | ClearPath Controller and ClearPath Tubing<br>(K091305)                                                                                                                              |
| Indications for<br>use    | It is for use only by trained medical personnel<br>located in hospitals, clinics, and doctors'<br>offices.                                                                                                                                                                                                                                                                                                                                              | It is for use only by trained medical personnel<br>located in hospitals, clinics, and doctors' offices.                                                                                                                                                                                                                                                                                     | It is for use only by trained medical personnel<br>located in hospitals, clinics, and doctors' offices.                                                                                                                                                                                                                                                                 | Distal tip design | multi irrigation holes<br>Distal suction hole                                                                                                                                 | multi irrigation holes<br>Distal suction hole                                                                                                                                                                                                                                                           | multi irrigation holes<br>Distal suction hole                                                                                                                                       |
| Environment<br>of Use     | Hospitals, clinics, and doctors' offices.                                                                                                                                                                                                                                                                                                                                                                                                               | Hospitals, clinics, and doctors' offices.                                                                                                                                                                                                                                                                                                                                                   | Hospitals, clinics, and doctors' offices.                                                                                                                                                                                                                                                                                                                               | Material          | Comply with ISO 10993                                                                                                                                                         | Comply with ISO 10993                                                                                                                                                                                                                                                                                   | Comply with ISO 10993                                                                                                                                                               |
| Prescriptive              | Yes, only trained medical personnel                                                                                                                                                                                                                                                                                                                                                                                                                     | Yes, only trained medical personnel                                                                                                                                                                                                                                                                                                                                                         | Yes, only trained medical personnel                                                                                                                                                                                                                                                                                                                                     | Performance       | Like the predicates, the modified device was<br>tested for compliance and performance under<br>the working conditions as prescribed by the<br>ClearPath Controller (K091305). | The disposable ClearPath Lower GI was tested<br>with the ClearPath Controller and ClearPath<br>Tubing (which were cleared in the same<br>K091305. Testing demonstrated adequate<br>functionality and performance under the working<br>conditions as prescribed by the ClearPath<br>Controller (K091305) | As detailed in K093779 tests were performed to<br>demonstrate the performance of the device under<br>the working conditions as prescribed by the<br>ClearPath Controller (K091305). |
| Principle of<br>operation | Distal attachment to an endoscope, sleeve<br>ensuring attachment along entire length,<br>suction and irrigation tubes running along the<br>endoscope, suction and irrigations heads at<br>the distal tip. Enables irrigation and suction at<br>any time during the procedure without<br>removing any tools which may be inserted in<br>the endoscope's working channel.                                                                                 | Distal attachment to an endoscope, straps<br>ensuring attachment along entire length, suction<br>and irrigation channels running along the<br>endoscope, suction and irrigations heads at the<br>distal tip. Enables irrigation and suction at any<br>time during the procedure without removing any<br>tools which may be inserted in the endoscope's<br>working channel.                  | Distal attachment to an endoscope, sleeve<br>ensuring attachment along entire length, suction<br>and irrigation tubes running along the<br>endoscope, suction and irrigations heads at the<br>distal tip. Enables irrigation and suction at any<br>time during the procedure without removing any<br>tools which may be inserted in the endoscope's<br>working channel. | Disposable        | Single patient, use, disposable                                                                                                                                               | Single patient, use, disposable                                                                                                                                                                                                                                                                         | Single patient, use, disposable                                                                                                                                                     |
| Packaged                  | Clean, non-sterile                                                                                                                                                                                                                                                                                                                                                                                                                                      | Clean, non-sterile                                                                                                                                                                                                                                                                                                                                                                          | Clean, non-sterile                                                                                                                                                                                                                                                                                                                                                      |                   |                                                                                                                                                                               |                                                                                                                                                                                                                                                                                                         |                                                                                                                                                                                     |

K11 3050

Page 5.4

DF Page 21 of 1

.

{4}------------------------------------------------

Premarket Notification 510(k)
Section 5 — Device Descriptio

ClearPath Lower Gl

# 510(k) Summary

Page 5.5

PDF Page 22 of 173

T

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Room -WO66-G609 Silver Spring, MD 20993-0002

EasyGlide Ltd. % Mr. Paul Dryden President ProMedic, Inc. 24301 Woodsage Drive BONITA SPRINGS FL 34134

MAY 1 6 2012

Re: K113050

> Trade/Device Name: ClearPath Lower GI Regulation Number: 21 CFR§ 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FDF Dated: April 15, 2012 Received: April 17, 2012

Dear Mr. Dryden:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical

{6}------------------------------------------------

device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerély yours,

Benjamin K. Hicks

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

K113050

### Indications for Use Statement

Page 1 of 1

### 510(k) Number:

# KI | 3050 (To be assigned)

Device Name:

ClearPath Lower GI

:

Indications for Use:

The ClearPath Lower GI is intended to connect to standard colonoscopes to help facilitate intraprocedural cleaning of a poorly prepared colon by irrigating or cleaning the colon and evacuating the irrigation fluid (water), feces, and other bodily fluids and matter, e.g. blood.

lt is for use only by trained medical personnel located in hospitals, clinics, and doctors' offices.

Prescription Use XX (Part 21 CFR 801 Subpart D) or

Over-the-counter use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Signature

ive, Gastro-Renal, and

---

**Source:** [https://fda.innolitics.com/device/K113050](https://fda.innolitics.com/device/K113050)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
